Thomas Graf1, Gitanas Kancerevycius2, Linas Jonušauskas2, Patric Eberle1.
Abstract
Glaucoma is a common, irreparable eye disease associated with high intraocular pressure. One treatment option is implantation of a stent to lower the intraocular pressure. A systematic approach to develop a microchannel stent meshwork that drains aqueous humor from the anterior chamber of the eye into the subconjunctival space is presented. The stent has a large number of outlets within its mesh structure that open into the subconjunctiva. The development approach includes a flow resistance model of the stent. Local adaption of the stent's tubular dimensions allows for adjustment of the flow resistance. In this way, an evenly distributed outflow into the subconjunctiva is achieved. We anticipate that microblebs will form at the stent outlets. Their size is crucial for drainage and control of intraocular pressure. An analytical model for bleb drainage is developed based on the porous properties of the subconjunctival tissue. Both models-the stent flow resistance model and the bleb drainage model-are verified by numerical simulation. The models and numerical simulation are used to predict intraocular pressure after surgery. They allow for a systematic and personalized design of microchannel stents. Stents designed in this way can stabilize the intraocular pressure between an upper and lower limit.Entities:
Keywords: IOP stabilization; bleb drainage; flow resistance model; glaucoma; microfluidic stent
Year: 2022 PMID: 35744591 PMCID: PMC9229318 DOI: 10.3390/mi13060978
Source DB: PubMed Journal: Micromachines (Basel) ISSN: 2072-666X Impact factor: 3.523
Figure 1(a) Sectional anatomy of the eye with glaucoma stent. Part A is placed in the anterior chamber to collect AH. Part B conducts the AH into the meshwork of part C. Part C drains the fluid into the subconjunctival tissue. (b) Drainage pathways and flow balance in the eye after surgery. AH is produced in the ciliary body and drained by the trabecular meshwork (TM) via Schlemm’s Canal, the uveoscleral pathway (UP) and the stent. At the stent outlets, blebs form in the subconjunctiva on the scleral barrier. The indicated flow rates were used in numerical simulation.
Figure 2The meshwork consists of honeycomb cells and is wide and high. Each hexagonal segment is a microtube (microchannel) with a square internal cross-section wide. Each honeycomb cell contains an outlet tube with specific dimensions and a corresponding flow resistance (see enlarged detail). The black numbers are the dimensions; the red numbers denote the columns and rows of the meshwork. The liquid flows evenly from part B above into the meshwork.
Figure 3Tubular structure of the stent and fluid resistance model. (a) The honeycomb meshwork geometry is the same everywhere; only the outlet tubes have different dimensions. The star-shaped part, colored in red, can be expressed as a fluidic resistance between two subsequent outlet tubes. (b) Equivalent circuit diagram of a single column of the stent, from the fluid inlet (top) to the individual outlets. The numbered resistors to correspond to the various outlet tubes along the stent column. Due to the different dimensions, the outlet tubes have distinct flow resistances. The dimensions are chosen so that the same rate of liquid flows out of each tube. The diagram in (c) illustrates how the resistance can be calculated using the flow resistance of a straight channel segment (marked in blue). Drawing (d) depicts the relevant geometric entities for the calculation of the flow resistance.
Equations to calculate the outlet tube resistances as a function of the connecting resistance and the bottom outlet resistance . Only the equations for the lowest outlet tubes are shown in the table. The general equation for the outlet resistance of row is presented in the text (Equation (4)). is the flow resistance of the circuit below row .
|
|
|
|
|---|---|---|
| 1 |
| |
| 2 |
|
|
| 3 |
|
|
| 4 |
|
|
| 5 |
|
|
Resistances, flow rates and pressures of the meshwork of Figure 1 and Figure 2. The resistances and are obtained from the channel geometry. The stent flow rate is an example value after surgery. The pressure difference across the stent is then calculated with the stent fluid model using the mentioned quantities or with COMSOL simulations using the specified geometry.
|
|
|
| Equation (1) |
|
|
|
|
| Equation (2) | |
|
|
|
| Equation (3) |
|
|
|
|
| Equation (5) | Resistance of a whole column composed of 20 rows |
|
|
|
|
| Resistance of a whole stent meshwork composed of 20 rows and 42 columns |
|
| 1.7 |
| specified | typical stent flow rate |
|
|
|
|
| Flow rate of a whole column of 20 outlet tubes |
|
|
|
|
| Flow rate of a single outlet tube |
|
|
|
|
| |
|
|
|
| COMSOL simulations |
Figure 4Pressure and flow velocity field in the microchannel mesh computed with the program COMSOL. The boundary conditions were the inlet flow rate of and the outlet pressure of at each orifice. (a) Pressure distribution in the stent of the leftmost and rightmost part of the mesh. The maximum value is at the inlet, above row 20. (b) Flow velocities in the midplane of the meshwork. Shown are flow details at top and bottom of the stent, along the centerline of the honeycomb structure.
Figure 5(a) COMSOL simulation of the outflow from a bleb with radius . The inlet boundary condition is the mass flow rate at the hemispherical bleb surface. This corresponds to , given in Table 2. The outer boundary condition is the pressure at the domain boundary at . is calculated by means of Equations (8) and (11) and used to mimic an infinite domain. The white lines and arrows indicate the flow direction given by the simulation. The axis of rotational symmetry is shown as a vertical, dashed red line. (b) Pressure as a function of distance from the origin. The blue line is the COMSOL result; the red circles are obtained using Equations (8) and (11).
Figure 6(a) Microbleb array of of the current stent meshwork. The microblebs lie on the scleral barrier and are spaced 300 µm apart. All blebs have a radius of . AH flows from the bleb surfaces into the thick subconjunctiva. (b,c) Drainage pressure field in the subconjunctival tissue determined by COMSOL simulation. The total flow rate of the bleb array is . Figure 6b shows a quarter of the simulation domain. Figure 6c is an enlarged view of 6b. The pressure inside the blebs is and reaches at the surface of the subconjunctiva. The pressure decreases rapidly in the plane of the array within the characteristic length .
Figure 7Drainage resistance obtained by COMSOL simulations of hexagonal arrays as in Figure 6. (a) as a function of bleb spacing . The bleb radius is constant . The curve reaches the value of for large separations, which is consistent with Equation (13) for free blebs. (b) as a function of bleb radius . The bleb spacing is constant at . For bleb radii , the curve terminates in a constant value corresponding to the drainage resistance of a shallow macrobleb of surface area.
Figure 8as a function of microbleb radius . The bleb separation is corresponding to the geometric period in the current honeycomb design. The numbers in the boxes are the IOP values before surgery in units of . The curves were calculated using Equation (16) and ranges from and for bleb radii greater than . The green shaded area corresponds to the for AH production rate varying between and and an IOP before surgery of . The green shaded area demonstrates that the IOP is within a healthy range even with daily fluctuating production rates [4].
The step-by-step procedure for personalized stent design.
| Steps | Input | Input | Output | Equation |
|---|---|---|---|---|
| 1 | Serial outflow resistance | |||
| 2 | Hemispherical bleb radius for COMSOL simulation, e.g., | Bleb spacing in array for COMSOL simulation, e.g., | Bleb array drainage resistance | |
| 3 | Stent flow resistance |
| ||
| 4 | Number of stent columns, e.g., | Number of stent rows, e.g., | Number of stent outlets and micro blebs |
|
| 5 | Stent column resistance | Fix lowest stent outlet resistance, | Stent flow resistance | |
| 6 | Hexagonal segment flow resistance | Bleb spacing | Channel length | |
| 7 | Lowest stent outlet resistance | Flow resistance | ||
| 8 | Length | Row number | Cross section width of outlet tube |
Governing parameters for calculations.
| Parameter | Value | SI Units | Description |
|---|---|---|---|
|
|
|
| Density of fluid, of AH |
|
|
|
| Dynamic viscosity of liquid |
|
|
|
| Hydraulic conductivity in subconjunctival tissue [ |
|
|
|
| Fluid permeability in subconjunctival tissue, used in Equation (6) |
|
|
|
| Used in Equation (6) |
|
|
|
| Hydraulic permeability of blood vessel wall [ |
|
|
|
| Vessel wall area per tissue volume [ |
|
|
|
| Characteristic drainage length, used in Equation (8) |
|
|
|
| Typical AH production rate |
|
|
|
| Constant outflow rate through uveoscleral pathway |
|
|
|
| Episcleral venous pressure, ranging from |