Assessment of Heating on Titanium Alloy Cerebral Aneurysm Clips during 7T MRI.

S Tsutsui1, T Matsuda2, K Takeda2, M Sasaki2, Y Kubo1, K Setta1, S Fujiwara1, K Chida1, K Ogasawara3.   

Abstract

BACKGROUND AND
PURPOSE: Patients with cerebral aneurysms often undergo MR imaging after microsurgical clipping. Ultra-high-field MR imaging at 7T may provide high diagnostic capability in such clinical situations. However, titanium alloy clips have safety issues such as adverse interactions with static magnetic fields and radiofrequency-induced heating during 7T MR imaging. The purpose of this study was to quantitatively assess temperature increases on various types of titanium alloy aneurysm clips during 7T MR imaging.
MATERIALS AND METHODS: Five types of titanium alloy aneurysm clips were tested, including combinations of short, long, straight, angled, and fenestrated types. Each clip was set in a phantom filled with gelled saline mixed with polyacrylic acid and underwent 7T MR imaging with 3D T1WI with a spoiled gradient recalled acquisition in the steady-state technique. Temperature was chronologically measured at the tips of the clip blade and head, angled part of the clip, and 5 mm from the tip of the clip head using MR imaging-compatible fiber-optic thermometers.
RESULTS: Temperature increases at all locations for right-angled and short straight clips were <1°C. Temperature increases at the angled part for the 45° angled clip and the tip of the clip head for the straight fenestrated clip were >1°C. Temperature increases at all locations for the long straight clip were >2°C.
CONCLUSIONS: Temperature increases on the right-angled and short straight clips remained below the regulatory limit during 7T MR imaging, but temperature increases on the 45° angled, straight fenestrated, and long straight clips exceeded this limit.
© 2022 by American Journal of Neuroradiology.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35738672      PMCID: PMC9262056          DOI: 10.3174/ajnr.A7561

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   4.966


  25 in total

Review 1.  Radiofrequency energy-induced heating during MR procedures: a review.

Authors:  F G Shellock
Journal:  J Magn Reson Imaging       Date:  2000-07       Impact factor: 4.813

2.  Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field.

Authors:  Vasily L Yarnykh
Journal:  Magn Reson Med       Date:  2007-01       Impact factor: 4.668

3.  Intracranial time-of-flight MR angiography at 7T with comparison to 3T.

Authors:  Cornelius von Morze; Duan Xu; Derk D Purcell; Christopher P Hess; Pratik Mukherjee; David Saloner; Douglas A C Kelley; Daniel B Vigneron
Journal:  J Magn Reson Imaging       Date:  2007-10       Impact factor: 4.813

4.  Small vessel disease in patients with subarachnoid hemorrhage: Prevalence and associations with vasospasm occurrence, severity and clinical outcomes.

Authors:  Adrien Villain; Grégoire Boulouis; Wagih Ben Hassen; Christine Rodriguez-Regent; Denis Trystram; Myriam Edjlali; François Nataf; Xavier Sauvageon; Tarek Sharshar; Jean F Meder; Catherine Oppenheim; Olivier Naggara
Journal:  Neuroradiol J       Date:  2019-09-30

5.  Safety assessment of spine MRI in deep brain stimulation patients.

Authors:  Alexandre Boutet; Gavin J B Elias; Robert Gramer; Clemens Neudorfer; Jürgen Germann; Asma Naheed; Nicole Bennett; Bryan Li; Dave Gwun; Clement T Chow; Ricardo Maciel; Alejandro Valencia; Alfonso Fasano; Renato P Munhoz; Warren Foltz; David Mikulis; Ileana Hancu; Suneil K Kalia; Mojgan Hodaie; Walter Kucharczyk; Andres M Lozano
Journal:  J Neurosurg Spine       Date:  2020-02-14

6.  First clinical study on ultra-high-field MR imaging in patients with multiple sclerosis: comparison of 1.5T and 7T.

Authors:  K Kollia; S Maderwald; N Putzki; M Schlamann; J M Theysohn; O Kraff; M E Ladd; M Forsting; I Wanke
Journal:  AJNR Am J Neuroradiol       Date:  2009-01-15       Impact factor: 3.825

7.  Magnetic resonance angiography with ultrashort echo times reduces the artefact of aneurysm clips.

Authors:  F Gönner; K O Lövblad; O Heid; L Remonda; R Guzman; A Barth; G Schroth
Journal:  Neuroradiology       Date:  2002-08-14       Impact factor: 2.804

8.  Symptomatic vasospasm after subarachnoid haemorrhage: assessment of brain damage by diffusion and perfusion-weighted MRI and single-photon emission computed tomography.

Authors:  X Leclerc; A Fichten; J Y Gauvrit; B Riegel; M Steinling; J P Lejeune; J P Pruvo
Journal:  Neuroradiology       Date:  2002-05-28       Impact factor: 2.804

9.  Non-enhanced magnetic resonance imaging of unruptured intracranial aneurysms at 7 Tesla: Comparison with digital subtraction angiography.

Authors:  Karsten H Wrede; Toshinori Matsushige; Sophia L Goericke; Bixia Chen; Lale Umutlu; Harald H Quick; Mark E Ladd; Sören Johst; Michael Forsting; Ulrich Sure; Marc Schlamann
Journal:  Eur Radiol       Date:  2016-03-18       Impact factor: 5.315

10.  Perforating arteries originating from the posterior communicating artery: a 7.0-Tesla MRI study.

Authors:  Mandy M A Conijn; Jeroen Hendrikse; Jaco J M Zwanenburg; Taro Takahara; Mirjam I Geerlings; Willem P Th M Mali; Peter R Luijten
Journal:  Eur Radiol       Date:  2009-12       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.