Longitudinal nuclear spin relaxation of 129 Xe in solution and in hollow fiber membranes at low and high magnetic field strengths.

Nicholas Bryden1,2, Christian T McHugh1,2, Michele Kelley1,2, Rosa T Branca1,2.   

Abstract

PURPOSE: To measure dissolved-phase 129 Xe T1 values at high and low magnetic fields and the field dependence of 129 Xe depolarization by hollow fiber membranes used to infuse hyperpolarized xenon in solution.
METHODS: Dissolved-phase T1 measurements were made at 11.7T and 2.1 mT by bubbling xenon in solution and by using a variable delay to allow spins to partially relax back to thermal equilibrium before probing their magnetization. At high field, relaxation values were compared to those obtained by using the small flip angle method. For depolarization studies, we probed the magnetization of the polarized gas diffusing through an exchange membrane module placed at different field strengths.
RESULTS: Total loss of polarization was observed for xenon diffusing through hollow fiber membranes at low field, while significant polarization loss (>20%) was observed at magnetic fields up to 2T. Dissolved-phase 129 Xe T1 values were found consistently shorter at 2.1 mT compared to 11.7T. In addition, both O2 and Xe gas concentrations in solution were found to significantly affect dissolved-phase 129 Xe T1 values.
CONCLUSION: Dissolved-phase 129 Xe measurements are feasible at low field, but to assess the feasibility of in vivo dissolved-phase imaging and spectroscopy the T1 of xenon in blood will need to be measured. Both O2 and Xe concentrations in solution are found to greatly affect  dissolved-phase 129 Xe T1 values and may explain, along with RF miscalibration, the large discrepancy in previously reported results.
© 2022 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  dissolved-phase T1; hollow fiber membranes; hyperpolarized xenon; low field

Mesh:

Substances:

Year:  2022        PMID: 35726363      PMCID: PMC9420755          DOI: 10.1002/mrm.29362

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   3.737


  23 in total

1.  A system for low field imaging of laser-polarized noble gas.

Authors:  G P Wong; C H Tseng; V R Pomeroy; R W Mair; D P Hinton; D Hoffmann; R E Stoner; F W Hersman; D G Cory; R L Walsworth
Journal:  J Magn Reson       Date:  1999-12       Impact factor: 2.229

2.  129Xe-Xe molecular spin relaxation.

Authors:  B Chann; I A Nelson; L W Anderson; B Driehuys; T G Walker
Journal:  Phys Rev Lett       Date:  2002-02-28       Impact factor: 9.161

3.  In vivo MR imaging and spectroscopy using hyperpolarized 129Xe.

Authors:  M E Wagshul; T M Button; H F Li; Z Liang; C S Springer; K Zhong; A Wishnia
Journal:  Magn Reson Med       Date:  1996-08       Impact factor: 4.668

4.  NMR of laser-polarized xenon in human blood.

Authors:  A Bifone; Y Q Song; R Seydoux; R E Taylor; B M Goodson; T Pietrass; T F Budinger; G Navon; A Pines
Journal:  Proc Natl Acad Sci U S A       Date:  1996-11-12       Impact factor: 11.205

5.  Xenon-129 MR imaging and spectroscopy of rat brain using arterial delivery of hyperpolarized xenon in a lipid emulsion.

Authors:  G Duhamel; P Choquet; E Grillon; L Lamalle; J L Leviel; A Ziegler; A Constantinesco
Journal:  Magn Reson Med       Date:  2001-08       Impact factor: 4.668

6.  Spin-lattice relaxation of laser-polarized xenon in human blood.

Authors:  J Wolber; A Cherubini; A S Dzik-Jurasz; M O Leach; A Bifone
Journal:  Proc Natl Acad Sci U S A       Date:  1999-03-30       Impact factor: 11.205

7.  An open-source, low-cost NMR spectrometer operating in the mT field regime.

Authors:  Nicholas Bryden; Michael Antonacci; Michele Kelley; Rosa T Branca
Journal:  J Magn Reson       Date:  2021-09-27       Impact factor: 2.734

Review 8.  Hyperpolarized 129Xe MRI of the human lung.

Authors:  John P Mugler; Talissa A Altes
Journal:  J Magn Reson Imaging       Date:  2013-02       Impact factor: 4.813

9.  Measurement of 129Xe T1 in blood to explore the feasibility of hyperpolarized 129Xe MRI.

Authors:  M S Albert; V D Schepkin; T F Budinger
Journal:  J Comput Assist Tomogr       Date:  1995 Nov-Dec       Impact factor: 1.826

10.  Relaxation and exchange dynamics of hyperpolarized 129Xe in human blood.

Authors:  Graham Norquay; General Leung; Neil J Stewart; Gillian M Tozer; Jan Wolber; Jim M Wild
Journal:  Magn Reson Med       Date:  2014-08-28       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.