Male infertility affects up to 12% of men. Although manual testing using microscope examination and computer-assisted semen analysis are standard methods of measuring sperm count and motility, these methods are limited by being laboratory based. To investigate the usefulness of a novel semen analysis device using a smartphone camera. This prospective multicenter randomized parallel design trial enrolled 200 men aged ≥19 years of age between August and December 2018. Each subject was advised to use the Smart Sperm Test for OVIEW-M at home after 5 days of abstinence. The accuracy of the OVIEW-M test relative to the in-hospital test was determined. A questionnaire was administered to assess subject likelihood of using the OVIEW-M. Measurements using standard methods and the OVIEW-M showed similar sperm counts and similar motile sperm counts. Correlation analysis showed significant correlations between sperm count and sperm motility when measured by OVIEW-M tests (r = 0.893, p < 0.01) and standard microscope examination (r = 0.883, p < 0.01). Of the subjects who responded to questionnaires, 43% regarded the results of the OVIEW-M tests as reliable and 18% as unreliable. Semen analysis with the smartphone-based application and accessories yielded results not inferior to those of laboratory tests. Men who visit the hospital for evaluation of infertility can easily perform OVIEW-M semen tests at home.

Entities:  

Mesh:

Year:  2022        PMID: 35709172      PMCID: PMC9202868          DOI: 10.1371/journal.pone.0269894

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.752


Introduction

Male infertility is a serious global problem, with around 485 million couples experiencing infertility [1]. The causes of infertility have been identified in 15% of couples, with the incidence found to be equal in men and women [2]. Although 2.5% to 12% of men worldwide are thought to be infertile, estimates of male infertility in many countries are inaccurate. Most surveys are of couples and of women who are trying to have children. In contrast, causes of infertility cannot be easily determined in countries that allow male supremacy or polygamy due to cultural differences. Historically, male infertility was not recognized as a disease until recently, and definitions of male infertility differ among countries [1]. Semen test criteria for male infertility based on the World Health Organization (WHO) classification [3]. Semen tests measure sperm count, motility, and malformation [4, 5]. Causes of male infertility have been classified into those involving the hypothalamus or pituitary gland, the testicles, or sperm delivery by sex glands associated with the penis and genitalia [6]. In the Republic of Korea, the declining fertility rate is becoming a social problem. Sperm counts may be reduced or sperm damaged due to various causes, such as environmental pollution and the use of electronic devices during the process of industrialization. Causes of male infertility include oligozoospermia (low-spermatozoa count), asthenozoospermia (poor sperm motility), and teratozoospermia (abnormal sperm morphology). Abnormal semen measurements are idiopathic in 30–40% of infertile men. Although treatment of some men may enhance the likelihood of pregnancy, other men are resistant to treatment [7]. Screening for male infertility requires measurements of sperm counts and sperm motility. Men may be uncomfortable visiting a professional medical institution for diagnosis of male infertility. The OVIEW-M is a self-testing device that can measure the number and activity of sperm using an optical camera on the front side of an individual’s smartphone. The accuracy of the OVIEW-M relative to standard laboratory methods for measuring sperm count and motility has not been determined. This clinical therefore compared the accuracy of the OVIEW-M with that of standard methods.

Materials and methods

Study design

This prospective multicenter randomized parallel design trial enrolled 200 men aged >19 years of age from August to December 2018. All subjects provided written informed consent using a form developed by one of the investigators. Subjects were administered a questionnaire that evaluated ejaculation and use of contraceptive devices within 5 days; history of vasectomy and other contraceptive procedures; medical history within the previous 6 months, including surgery; and medications being taken. Subjects were included if they were healthy male adults aged ≥19 years, had no specific disease that could affect sperm count or motility, were not taking hormone medications, had not been assessed during the previous 5 days, and provided written informed consent. Females were excluded, as were males aged <19 years and those unable to provided written informed consent. Also excluded were men with congenital or genetic diseases affecting spermatogenesis and nocturnal emission, and subjects who had participated in other clinical trials within the previous 30 days. While in the hospital, subjects were asked to provide a >1 cc semen sample by masturbation in an uncontaminated cup. Subjects were instructed to use the Smart Self-Sperm Test (OVIEW-M) at home after 5 days of abstinence at home. The accuracy of the OVIEW-M test was assessed by comparison with standard in-hospital methods. After the test, a questionnaire was administered to each subject to determine their likelihood of using the OVIEW-M.

Overview of the OVIEW-M device and testing procedure

The OVIEW-M is a self-testing device that measures the number and activity of sperm using an optical camera on the front side of an individual’s smartphone. It consists of a collecting cup and stick and a testing device composed of a cover, chamber, and body, which is connected to the smartphone. Each test was performed using both an Android smartphone and an iPhone, with the OVIEW-M application downloaded from Google Play and the iPhone App Store, respectively. While at the hospital, each subject tested the device and application before returning home to confirm that the application was working properly and that the testing device was visible to the camera.

Standard semen testing procedure

A semen sample in a disposable cup was allowed to separate for 15 minutes, as above, and a drop placed at the center of a 1 cc slide glass using a pipette. The sample was covered with a cover slip and lightly moved to spread the sample evenly, and the slide was viewed on an optical microscope (400×). The numbers of spermatozoa and of active spermatozoa were determined using a grid.

Ethics statement

This study was approved by the clinical study ethics committee (IRB) of the institution performing the clinical trial (MPUC-A-1808-S-01). The trial was conducted in accordance with the standards of the KGCP (Clinical Trial Management of Medical Devices) and the principles of the Declaration of Helsinki.

Statistical analysis

To determine whether the use of standard methods by specialists at different medical institutions yielded similar results, semen samples were randomly sampled and assayed by laboratories at 13 regional medical institutions in the Manpower Urology Network, a representative urology clinic in Korea. To verify the difference between standard and OVIEW-M methods, 210 people were recruited, based on a 5% drop out rate. However, because no subjects were excluded or dropped out, 200 subjects were recruited. Although most hospitals evaluated more than 15 subjects, some did not meet this quota, with data from these subjects being replaced by those from hospitals that evaluated more than 15 subjects used. Results obtained using the OVIEW-M device and standard method were compared by Student’s t-tests. All statistical analyses were performed using SPSS for windows, version 20.0.

Results

Design of optical equipment using low power

Briefly, a semen sample was added to the sampling cup and mixed well with a stick. The sperm and suspended matter were allowed to separate for 15 minutes to reduce the viscosity of the semen. The semen was placed in the square section of the chamber of the testing device, and the chamber was placed between the cover and the body of the device. The camera was connected and the OVIEW-M application started on the smartphone (Fig 1). While the camera was running, the light source was centered using the external stand on the camera screen. The testing device was positioned on the front camera, and the measurement start button was pressed. Results were displayed within 10 seconds.
Fig 1

Installed OviewM and View of Sperm analysis using Oview M.

(a) Consist of Oview M (b) View of Sperm analysis using Oview M.

Installed OviewM and View of Sperm analysis using Oview M.

(a) Consist of Oview M (b) View of Sperm analysis using Oview M.

Validity evaluation (accuracy evaluation)

Mean sperm counts (49.49 ± 31.10 × 103/μl vs. 49.6 ± 31.06 × 103/μl; t = 0.68, p = 0.50) and sperm motility (49.59 ± 24.72 × 103/μl vs. 49.45 ± 24.44 × 103/μl; t = .66, p = 0.51) did not differ significantly when measured by OVIEW-M tests and standard microscope examination (Table 1). Correlation analysis showed significant correlations between sperm counts and sperm mobility when measured by OVIEW-M tests (r = 0.893, p < 0.01) and standard microscope examination (r = 0.883, p < 0.01). These results confirm the reliability of the OVIEW-M measurement tool in assessing sperm count and sperm motility.
Table 1

Repeatability, precision, and correlation coefficients of motile sperm concentration and motility.

NMean ± SD, x 103/μlMean ± SD, difference, x 103/μlp-valueCorrelation coefficient (concordance)p-value
Sperm Count
Microscope analysis20049.60±31.080.11±2.290.500.997<0.01
OVIEW-M device20049.49±31.10
Sperm Motility
Microscope analysis20049.45±24.43-0.14±2.910.510.993<0.01
OVIEW-M device20049.59±1.75

Questionnaire analysis

Of all respondents, 34% planned to use the smart self-sperm OVIEW-M test to measure sperm count and sperm motility in the future, with 76% of the respondents being optimistic about the results obtained with this device (Fig 2). Although 43% of respondents regarded the results obtained with OVIEW-M test as reliable and 18% did not, most respondents did not have sufficient confidence in the results of the smartphone test (Fig 3).
Fig 2

Desire to buy O`view M (smart phone based self- semen analysis tester).

Fig 3

Reliability of using O`view M to subjects.

Discussion

This study showed that an application using a smartphone and accessory parts could accurately determine sperm counts and sperm motility in semen samples, when compared with results obtained by cell counting in hospital laboratories test. These results were reliable and may be used to assess confirm sperm counts and sperm motility not in infertile men, but also in men who have undergone vasectomy and vasovasostomy. Smartphones were reported to be used by up to 96.8% of subjects in developed countries in 2015 [8]. Smartphones have powerful computing systems and internet access, making them useful in many applications, including the facility to attach low-cost microscopic devices [9]. The development of micro-technology and the rapid growth of consumer electronics have laid a solid foundation for the development of mobile health technology that can change the paradigm of world health today. In daily life, we can exercise with using smartphone with measuring vital signs, total calories that we consumed, and distance of we jogged. Now, many research showed application of smartphone in medical field with various kind of ways. We can use smartphone application using fractal image analysis of moles [10]. Also, Add on lenses on smartphone is feasible for checking up macrovesicular steatosis in liver allograft biopsies [11]. Smartphone applications in urology include the University of Washington Peyronie’s Examination Network smartphone (UWPEN), which was developed to assess penile angulation in a home environment [12]. Other smartphone applications include a voiding diary, the tracking of "forgotten" ureteral stents, and applications to manage enuresis and ureteral stones [Patient Saf Surg. 2017 ">13-15]. Smartphone applications for diagnostic purposes include a guide for surgical treatment using three-dimensional and virtual reality technology [16]. In addition, smartphone applications used in combination with various technologies combined with are being developed to assist diagnosis and treatment [17, 18]. To date, almost all semen samples have been tested in clinical laboratories by a specialist. Standard tests include measurements of the number of spermatozoa, their mobility and their deformity. The results of these tests have been used to diagnose causes of male infertility and to analyze the effects of treatment. Couples who wish to get pregnant before going to a hospital can search the medical literature and the internet to find a pregnancy tester that assesses the causes of infertility. To date, however, no tool or method outside a hospital is available to test male infertility. Couples who do not go for testing must wait for the next and subsequent cycles, increasing psychological problems such as anxiety and depression. Antonie van Leeuwenhoek in the 17th century was the first to visualize spermatozoa using single-ball handheld microscopes with high quality lenses that magnified objects 270-fold [19]. Using a smartphone camera with OVIEW-M magnified objects 230~250 fold. Smartphone cameras can magnify objects more than five-fold, with the camera and the OVIEW-M device magnifying objects 230~250 fold. This easy-to-use Smartphone-based semen analyzer could accurately measure sperm concentration, sperm motility, total sperm count, total motile sperm count, and changes in these parameters using small amounts of non-washed semen samples loaded onto disposable microchips. Use of these diagnostic assays can make male fertility tests as accessible, easy, fast, and private as pregnancy tests. Test results can be stored and imaged on smartphones and can enable tele-diagnosis [20]. This smartphone-based semen analyzer can also be used for home-based monitoring of men undergoing vasectomy and vasovasostomy. More than 33 million married couples worldwide prefer vasectomy for contraception, believing it to be safer, cheaper, and simpler than female sterilization [21]. In the US, between 175,000 and 550,000 vasectomy procedures are performed per year [22]. Although the vasectomy failure is <1%, semen should be analyzed subsequently. Sperm concentrations in semen 8–16 weeks after surgery should be below 100,000/ml [23]. However, rates of postvasectomy follow-up semen analysis have been extremely poor [24]. The smartphone based semen analyzer may improve compliance with semen analysis after vasectomy by providing a fast, easy to use, inexpensive, and personalized method that accurately measures sperm concentration and motility. Similarly, the semen of men who have undergone vasovasostomy and varicocelectomy for infertility should be evaluated postoperatively to analyze sperm quality. The smartphone-based semen analyzer would also be applicable to these individuals. Application to animals of breeding can be another way to using smartphone-based semen analyzer at home, laboratory, or zoo. Collected semen sample by animal breeders, they need to carry semen in certain condition with any harm, or trauma to check exact semen condition. However, using smartphone-based semen analyzer can give comfort to user without giving damage to semen. Animal breeder can make quick decision whether to go hospital or use the sample on research. Given the difference between human semen and animal semen in terms of sperm concentration, motility and sample volume, the current version of the smartphone-based semen analyzer should be extended for application in animals. Considering that many men are only clinically tested for fertility problems at a relatively late stage, the use of such smartphone-based devices can improve patient care. In other medical fields such as obstetrics, pediatrics, hematology, or ophthalmology, smartphone technology is already being used to improve efficiency, patient satisfaction and empowerment with respect to smartphone [25]. There is a debate over the case where the quality of sperm is deteriorated by COVID-19 [26], and a tool to measure this is needed. In the era of COVID-19, any equipment which can easily measure for semen at home is needed, and for this, it must have an equal effect compared to existing measurements. This study was conducted with a structure similar to that of RCT, but there is a limitation in that it was conducted without prior registration such as clinical trial. Although there are some limitations in this article, this tool can be use as the standard of the measurement about semen analysis in the era of COVID-19.

Conclusion

This study showed that the smartphone-based semen test was as accurate and reliable as laboratory-based tests. These smartphone-based tests can be performed easily at home by men who visit the hospital for evaluation of infertility. This equipment can be used for measurement of semen in the era of COVID-19.

Validity evaluation (accuracy evaluation).

(DOCX) Click here for additional data file. 2 Jan 2022
PONE-D-21-36708
Ability and Accuracy of the Smartphone-based O`VIEW-M® Sperm Test: Useful tool in the era of Covid-19.
PLOS ONE Dear Dr. Jung Ki Jo, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Feb 16 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:
A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. We look forward to receiving your revised manuscript. Kind regards, Joël R Drevet, Ph.D. Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. Please provide additional details regarding participant consent. In the ethics statement in the Methods and online submission information, please ensure that you have specified (1) whether consent was informed and (2) what type you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed). If your study included minors, state whether you obtained consent from parents or guardians. If the need for consent was waived by the ethics committee, please include this information. If you are reporting a retrospective study of medical records or archived samples, please ensure that you have discussed whether all data were fully anonymized before you accessed them and/or whether the IRB or ethics committee waived the requirement for informed consent. If patients provided informed written consent to have data from their medical records used in research, please include this information. 3. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match. When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section. 4. Thank you for stating the following financial disclosure: “Yes. This article was funded by the Bio & Medical Technology Development Program of the National Research Foundation (NRF) and funded by the Korean government (MSIT) [grant number 2019M3E5D1A01069353].” Please state what role the funders took in the study.  If the funders had no role, please state: "The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript." If this statement is not correct you must amend it as needed. Please include this amended Role of Funder statement in your cover letter; we will change the online submission form on your behalf. 5. Thank you for stating the following in the Competing Interests section: “Ji Hoon Kim is the Chief Executive Officer of INTIN Corporation.” Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf. 6. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability. Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized. Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access. We will update your Data Availability statement to reflect the information you provide in your cover letter. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: This is the multicenter study which is showing good correlation between the new smart-phone-based semen analyzing kit and manual semen analysis. The tool OVIEW-M seems to have good potential as a clinical tool and this article is showing interesting result. I have some suggestions and questions below. - Why did the author use only manual semen analysis in hospital? Don’t they use CASA? - Why is the abstinence 5 days? According to WHO guideline, it should be 2-7 days. If there is specific reason, the author should mention. - From this result, readers can only know the mean and SD. The graph such as simple correlation etc. which is showing the actual plots should be added. - This study cohort only includes healthy donors. It is questionable that OVIEW-M is also reliable as actual semen analysis in patient cohort. Author should mention about this issue in discussion. - In page 9, there are already some home test kits available for semen test in the world (see example at: https://seem.life/en/), although they are not always so accurate but aimed to just notify the people to go to hospital. Author should discuss about those kits and its merit and demerit. - According to the result, the tool is very accurate compared to manual semen analysis. Do the author think this can replace going to the hospital, or they just want to notify the result roughly and indicate to go to the hospital? Author should discuss more about future expectation for clinical use. Reviewer #2: To evaluate the accuracy of smartphone semen analysis using O`VIEW-M, the authors investigated the relationship between smartphone semen analysis results and laboratory semen analysis of 200 men. Although we think this article is very interesting, there are fix points, comments and some of the questions in this paper. 1. Can O`VIEW-M devices and apps be used on all smartphones? Because smartphones use different optical cameras and software. 2. Do O`VIEW-M devices and apps automatically display semen test results? Or does the user need to count the sperm count on the screen? 3. Please indicate whether each user performed the inspection in a different place. This is because microscopes using a single ball lense are strongly influenced by the light source. 4. Was each test using O`VIEW-M done by an individual? Or did the representative perform? ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 13 Apr 2022 Revision Letter Dear Editor, Edrian Nim Tolentino We are pleased to submit to you a revised version of our manuscript #: PONE-D-21-36708, Title: "Ability and Accuracy of the Smartphone-based O`VIEW-M® Sperm Test: Useful tool in the era of Covid-19.” with hope of it being published in PLOS ONE. Followings are comments on our manuscript along with our replies. Sincerely, Corresponding author �  The followings are our responses to the queries given by Reviewers. 1. We note that you provided the following response to our query regarding any patents or products in development related to this research: "Yes. The products in development or marketed products associated with this research are as follows. A measuring device of salive for smart phone (Reg. No. 10-1533343), The test device for Body fluid analysis using natural light (Reg. No. 10-1813866), CHAMBER FOR INSPECTING BODY-FLUID AND BODY-FLUID INSPECTING APPARATUS USING THE SAME (Reg. No. 10-0029588) and so on." Please clarify what is meant by "and so on", and whether there are any additional patents, products in development or marketed products associated with this research to declare that were not included in your response above. “Yes. As you mentioned, we have confirmed that there are no additional patents, products in development or products on the market related to this research. Therefore, there are A measuring device of salive for smart phone (Reg. No. 10-1533343), The test device for Body fluid analysis using natural light (Reg. No. 10-1813866), CHAMBER FOR INSPECTING BODY-FLUID AND BODY-FLUID INSPECTING APPARATUS USING THE SAME (Reg. No. 10-0029588) products related to this research.” 2. Please review our data sharing policy: https://journals.plos.org/plosone/s/data-availability PLOS journals require authors to make all data necessary to replicate their study’s findings publicly available without restriction at the time of publication. When specific legal or ethical restrictions prohibit public sharing of a data set, authors must indicate how others may obtain access to the data. “Yes. We submit the raw data used in this study as a supplement to indicate how others may obtain access to the data.” Submitted filename: Revisionl letter_ver.4.doc Click here for additional data file. 30 May 2022 Ability and Accuracy of the Smartphone-based O`VIEW-M® Sperm Test: Useful tool in the era of Covid-19. PONE-D-21-36708R1 Dear Jung Ki Jo, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Joël R Drevet, Ph.D. Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: All my comments included in my revision were now all addressed in the manuscript. I think this manuscript is ready for publication. Reviewer #2: (No Response) ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: No 6 Jun 2022 PONE-D-21-36708R1 Ability and Accuracy of the Smartphone-based O`VIEW-M® Sperm Test: Useful tool in the era of Covid-19 Dear Dr. Jo: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Prof. Joël R Drevet Academic Editor PLOS ONE
  22 in total

1.  Use of three-dimensional virtual images for planning surgery of complex anal fistulas: a new technology available via smartphone.

Authors:  E C Sousa Júnior; W M N Eulálio Filho; A T Nogueira; B A Rocha; A D Meneses
Journal:  Tech Coloproctol       Date:  2019-07-24       Impact factor: 3.781

Review 2.  Microfluidics for sperm analysis and selection.

Authors:  Reza Nosrati; Percival J Graham; Biao Zhang; Jason Riordon; Alexander Lagunov; Thomas G Hannam; Carlos Escobedo; Keith Jarvi; David Sinton
Journal:  Nat Rev Urol       Date:  2017-10-31       Impact factor: 14.432

3.  Significance of sperm characteristics in the evaluation of male infertility.

Authors:  Kiran P Nallella; Rakesh K Sharma; Nabil Aziz; Ashok Agarwal
Journal:  Fertil Steril       Date:  2006-03       Impact factor: 7.329

4.  An automated smartphone-based diagnostic assay for point-of-care semen analysis.

Authors:  Manoj Kumar Kanakasabapathy; Magesh Sadasivam; Anupriya Singh; Collin Preston; Prudhvi Thirumalaraju; Maanasa Venkataraman; Charles L Bormann; Mohamed Shehata Draz; John C Petrozza; Hadi Shafiee
Journal:  Sci Transl Med       Date:  2017-03-22       Impact factor: 17.956

5.  Testing feasibility of an accurate microscopic assessment of macrovesicular steatosis in liver allograft biopsies by smartphone add-on lenses.

Authors:  Manuela Cesaretti; Nicolas Poté; Federica Dondero; François Cauchy; Anne Sophie Schneck; Olivier Soubrane; Valerie Paradis; Alberto Diaspro; Andrea Antonini
Journal:  Microsc Res Tech       Date:  2017-10-17       Impact factor: 2.769

6.  Compliance with semen analysis.

Authors:  Jon Duplisea; Thomas Whelan
Journal:  J Urol       Date:  2013-01-25       Impact factor: 7.450

7.  Comparison of clinician and patient users of a mobile phone application to assess penile curvature in Peyronie's disease.

Authors:  Wayne G Brisbane; Marc J Rogers; Ryan S Hsi; Saneal Rajanahally; George R Schade; Lauren Trew; Kevin A Ostrowski; Hunter Wessells; Thomas J Walsh
Journal:  Int J Impot Res       Date:  2019-05-08       Impact factor: 2.896

Review 8.  Trends of male factor infertility, an important cause of infertility: A review of literature.

Authors:  Naina Kumar; Amit Kant Singh
Journal:  J Hum Reprod Sci       Date:  2015 Oct-Dec

9.  A new patient safety smartphone application for prevention of "forgotten" ureteral stents: results from a clinical pilot study in 194 patients.

Authors:  Wilson R Molina; Rodrigo Pessoa; Rodrigo Donalisio da Silva; McCabe C Kenny; Diedra Gustafson; Leticia Nogueira; Mark E Leo; Michael K Yu; Fernando J Kim
Journal:  Patient Saf Surg       Date:  2017-04-04
View more
  1 in total

Review 1.  Increase in Direct-to-Consumer Telemedicine in Urology.

Authors:  Erin Jesse; Nannan Thirumavalavan; Aram Loeb
Journal:  Curr Sex Health Rep       Date:  2022-08-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.