Manuela Cesaretti1,2, Nicolas Poté3, Federica Dondero1, François Cauchy1, Anne Sophie Schneck1, Olivier Soubrane1, Valerie Paradis3, Alberto Diaspro2, Andrea Antonini4,5.
Abstract
Light microscopy is an essential tool in histological examination of tissue samples. However, the required equipment for a correct and rapid diagnosis is sometimes unavailable. Smartphones and mobile phone networks are widespread, and could be used for diagnostic imaging and telemedicine. Macrovesicular steatosis (MS) is a major risk factor for liver graft failure, and is only assessable by microscopic examination of a frozen tissue section. The aim of this study was to compare the microscopic assessment of MS in liver allograft biopsies by a smartphone with eyepiece adaptor (BLIPS device) to standard light microscopy. Forty liver graft biopsies were evaluated in transmitted light, using an Iphone 5s and 4 different mini-objective, add-on lenses. A significant correlation was reported between the two different approaches for graft MS assessment (Spearman's correlation coefficient: rs = 0.946; p < .001). Smartphone with eyepiece adaptor had similar discriminatory power to identify MS in liver grafts than standard light microscopy. Based on these findings, a smartphone integrated with a low-cost eyepiece adaptor can achieve adequate accuracy in the assessment of MS in liver graft, and could be used as an alternative to standard light microscope when unavailable.Entities:
Keywords: eyepiece adaptor; liver steatosis; pathology; smartphone