| Literature DB >> 35708142 |
Xin Tan1, Bei Zhang1, Lisheng Zheng2, Hongbin Shi3, Dandan Liu1, Yuanmin Sun1, Xue Li1, Huiqiang Li1.
Abstract
BACKGROUND: Specific IgE (sIgE) testing has become one of the most important tools for diagnosing IgE-mediated food allergy. Enzyme-linked immunosorbent assay (ELISA) and dot-enzyme-linked immunosorbent assay (Dot-ELISA) have been used to measure sIgE in clinical widely. Light-initiated chemiluminescence assay (LICA) is a new method for measuring allergen-sIgE. We aimed to establish a LICA method for quantitative detection of egg white-sIgE and evaluate its performances.Entities:
Keywords: egg white-specific IgE; evaluation; light-initiated chemiluminescent assay; performance; quantification
Mesh:
Substances:
Year: 2022 PMID: 35708142 PMCID: PMC9279973 DOI: 10.1002/jcla.24544
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 3.124
FIGURE 1Comparison of coupling method 1 and method 2. Nine serum pools which clinical values ranged from 0.11 to 0.34 kUA/L were measured with chemibeads of two coupling methods, respectively, by indirect light‐initiated chemiluminescence assay (LICA). (A) A summary of two methods S/N ratios. Two methods showed no statistical difference analyzed by paired Wilcoxon signed‐rank test. (B) Each serum pools S/N ratios of two methods
FIGURE 2Reaction conditions optimization. (A) A comparison of three mass ratios chemibeads which ranged from 80:1 to 10:1. Differences between groups were analyzed using Friedman test. p < 0.05(*) and p < 0.01(**) indicated a statistically difference. (B) The effect of different dilution ratios of chemibeads. (C) The effect of different dilution ratios of anti‐Human IgE antibody
FIGURE 3Buffer system optimization. (A) The S/N ratios of four serum pools in seven buffer systems. (B) The chemiluminescence (CL) signals of four serum pools in seven buffer systems. (C) The S/N ratios of four serum pools in phosphate buffer (PB) with three pH values. (D) The CL signals of four serum pools in PB with three pH values. (E) The S/N ratios of four serum pools in PB (pH 7.4) with three concentrations. (F) The CL signals of four serum pools in PB (pH 7.4) with three concentrations. (G) The S/N ratios of four serum pools in 0.01 M PB (pH 7.4) with three surfactants. (H) The CL signals of four serum pools in 0.01 M PB (pH 7.4) with three surfactants
FIGURE 4Calibration curve of light‐initiated chemiluminescence assay (LICA) for detecting egg white‐sIgE in human sera based on four‐parameter logistic equation. The corresponded concentration range is 0–45.5 kUA/L.
Repeatability and intermediate precision of LICA
| Serum pool | Repeatability | Within‐run precision | ||||
|---|---|---|---|---|---|---|
| Average measured value (kUA/L) | SD (kUA/L) | CV (%) | Average measured value (kUA/L) | SD (kUA/L) | CV (%) | |
| Low | 0.275 | 0.020 | 7.29 | 0.267 | 0.023 | 8.64 |
| Middle | 1.015 | 0.028 | 2.72 | 1.007 | 0.050 | 4.93 |
| High | 2.608 | 0.113 | 4.34 | 2.586 | 0.151 | 5.85 |
Abbreviations: CV, coefficient of variation; LICA, light‐initiated chemiluminescence assay; SD, standard deviation.
FIGURE 5Linearity of the assay
Method comparison of LICA and ELISA (HOB, China) for detection of egg white‐sIgE
| LICA | ELISA | Total | |
|---|---|---|---|
| Positive | Negative | ||
| Positive | 64 | 4 | 68 |
| Negative | 4 | 31 | 35 |
| Total | 68 | 35 | 103 |
Abbreviations: ELISA, enzyme‐linked immunosorbent assay; LICA, light‐initiated chemiluminescence assay.