Chun-Kai Chen1, Li-Ying Cheng1,2, Shan-Wei Hsu1, Min-Tsun Liao1,3, Po-Wen Ku4, Yen-Bin Liu3,5.
Abstract
Background: Physical activity (PA) has become an important health issue for decades. Cardiovascular implantable electronic devices (CIEDs) have built-in PA-recording functions. We aimed to compare PA measurements using an external accelerometer (ActiGraph GT3X+) and internal accelerometers (Abbott, Biotronik, and Medtronic CIEDs).Entities:
Keywords: ActiGraph GT3X+; accelerometer; cardiac implantable electric device; physical activity; validation
Year: 2022 PMID: 35694655 PMCID: PMC9184442 DOI: 10.3389/fcvm.2022.898086
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
FIGURE 1Study flow diagram.
Baseline characteristics of study participants.
| Clinical characteristics | Abbott | Biotronik | Medtronic | |
|
| ||||
| Age (year) | 77.5 ± 12.7 | 74.6 ± 12 | 70.4 ± 18.5 | 0.312 |
| Male | 9 (45) | 10 (50) | 8 (40) | 0.817 |
| BMI | 25.1 ± 3.3 | 24.7 ± 3.8 | 25 ± 5.9 | 0.955 |
| Indication for PPM | 0.344 | |||
| SSS | 15 (75) | 17 (85) | 13 (65) | |
| AVB | 5 (25) | 3 (15) | 7 (35) | |
| Hypertension | 10 (50) | 14 (70) | 11 (55) | 0.410 |
| Diabetes mellitus | 6 (30) | 6 (30) | 6 (30) | 1.000 |
| Coronary artery disease | 4 (20) | 4 (20) | 1 (5) | 0.308 |
| Congestive heart failure | 3 (15) | 2 (10) | 2 (10) | 0.851 |
| Cerebral vascular accidence | 2 (10) | 2 (10) | 1 (5) | 0.804 |
| Atrial fibrillation | 3 (15) | 4 (20) | 3 (15) | 0.887 |
| Peripheral artery disease | 0 (0) | 0 (0) | 1 (5) | 0.362 |
| CKD | 1 (5) | 2 (10) | 2 (10) | 0.804 |
| ESRD | 1 (5) | 0 (0) | 1 (5) | 0.596 |
| COPD | 2 (10) | 2 (10) | 0 (0) | 0.343 |
| Hyperlipidemia | 8 (40) | 4 (20) | 5 (25) | 0.344 |
| Smoke | 5 (25) | 8 (40) | 4 (20) | 0.344 |
| NYHA status | ||||
| No heart failure | 17 (85) | 18 (90) | 18 (90) | 0.838 |
| Class I | 2 (10) | 1 (5) | 2 (10) | |
| Class II | 1 (5) | 1 (5) | 0 | |
| Class III | 0 | 0 | 0 | |
|
| ||||
| Alpha-blockers | 0 (0) | 3 (15) | 1 (5) | 0.153 |
| Beta-blockers | 7 (35) | 5 (25) | 5 (25) | 0.720 |
| ACEi/ARB | 8 (40) | 6 (30) | 2 (10) | 0.092 |
| Calcium channel blockers | 7 (35) | 7 (35) | 8 (40) | 0.931 |
| Diuretics | 6 (30) | 3 (15) | 5 (25) | 0.521 |
| Antiarrhythmia agents | 1 (5) | 1 (5) | 1 (5) | 1.000 |
BMI, body mass index; NYHA, New York Heart Association; LVEF, left ventricular ejection fraction; ACEi, angiotensin-converting-enzyme inhibitor; ARB, angiotensin receptor blocker.
Continuous values are expressed as mean ± standard deviation (SD) or median (interquartile range); categorical variables are reported as numbers (percentages).
ActiGraph GT3X+ and cardiovascular implantable electronic device-measured physical activity of study participants.
| GT3X+ parameters | Abbott | Biotronik | Medtronic | |
| Device-measured PA (hours per day) | 3.0 ± 1.5 | 2.6 ± 1.8 | 3.8 ± 2.5 | 0.126 |
| Wearing days of ActiGraph | 6.7 ± 0.56 | 6.55 ± 0.75 | 6.96 ± 0.08 | 0.065 |
| 7-day average activity (hours per day) | 6.9 ± 2.8 | 6.0 ± 2.4 | 6.4 ± 2.5 | 0.531 |
| Light activity (hours per day) | 4.3 ± 1.6 | 3.9 ± 1.2 | 3.9 ± 1.1 | 0.487 |
| Moderate activity (hours per day) | 2.4 ± 1.4 | 2 ± 1.3 | 2.2 ± 1.4 | 0.678 |
| Vigorous activity (hours per day) | 0.2 ± 0.3 | 0.2 ± 0.2 | 0.3 ± 0.4 | 0.459 |
| 7-day average steps per day | 8,276 ± 4,769 | 6,949 ± 4,486 | 7,537 ± 4,515 | 0.660 |
| 7-day average VMU per day | 1,487,187 ± 816,746 | 1,269,637 ± 765,616 | 1,449,361 ± 902,263 | 0.677 |
| y (vertical) axis per hour | 33,732 ± 18,858 | 27,438 ± 16,623 | 32,742 ± 20,955 | 0.530 |
| x (anterior-posterior) axis per hour | 33,530 ± 19,305 | 28,036 ± 16,357 | 33,577 ± 21,172 | 0.575 |
| z (medial-lateral) axis per hour | 39,368 ± 21,426 | 35,285 ± 22,119 | 37,853 ± 23,268 | 0.843 |
| Average kcal per day | 68.9 ± 39 | 55.9 ± 36.9 | 61.5 ± 39.8 | 0.565 |
| MET | 1.4 ± 0.3 | 1.3 ± 0.2 | 1.4 ± 0.3 | 0.467 |
MET, metabolic equivalent; PA, physical activity; VMU, vector magnitude units.
Correlations between cardiovascular implantable electronic device-measured physical activity and Actigraph GT3X+ parameters.
| GT3X+ parameters | Abbott | Biotronik | Medtronic | |||
| 7-day average activity (hours per day) | 0.534 | 0.015 | 0.465 | 0.039 | 0.677 | 0.001 |
| Light activity (hours per day) | 0.559 | 0.010 | 0.336 | 0.147 | 0.457 | 0.043 |
| Moderate activity (hours per day) | 0.413 | 0.070 | 0.445 | 0.049 | 0.696 | 0.001 |
| Vigorous activity (hours per day) | 0.248 | 0.292 | 0.475 | 0.035 | 0.655 | 0.002 |
| 7-day average steps per day | 0.398 | 0.082 | 0.352 | 0.128 | 0.666 | 0.001 |
| 7-day average VMU per day | 0.419 | 0.066 | 0.583 | 0.007 | 0.693 | 0.001 |
| x(anterior-posterior) axis | 0.401 | 0.080 | 0.699 | 0.001 | 0.651 | 0.002 |
| y(vertical) axis | 0.422 | 0.064 | 0.666 | 0.001 | 0.730 | < 0.001 |
| z(medial-lateral) axis | 0.416 | 0.068 | 0.726 | < 0.001 | 0.687 | 0.001 |
| Average kcal per day | 0.385 | 0.093 | 0.662 | 0.001 | 0.651 | 0.002 |
| MET | 0.364 | 0.115 | 0.614 | 0.004 | 0.694 | 0.001 |
VMU, vector magnitude units.
FIGURE 2Correlation between the 7-day average physical activity (hours per day) of the pacemakers and ActiGraph GT3X+. (A) All patients, (B) Abbott, (C) Biotronik, (D) Medtronic.
FIGURE 3Bland-Altman difference plot of device- and ActiGraph-measured 7-day average physical activity (hours per day) with 95% limits of agreement (± 1.96 SD). (A) All patients, (B) Abbott, (C) Biotronik, (D) Medtronic.
Intraclass correlation coefficient between average daily physical activity measured by the Actigraph GT3X+ and cardiovascular implantable electronic devices (hours per day).
| ICC (95% CI) | F test with true value 0 | ||||
| Value | df1 | df2 | |||
| Abbott | 0.614 (0.026, 0.847) | 2.593 | 19 | 19 | 0.022 |
| Biotronik | 0.618 (0.034, 0.849) | 2.615 | 19 | 19 | 0.021 |
| Medtronic | 0.807 (0.513, 0.924) | 5.184 | 19 | 19 | 0.000 |