Kandi Zhang1, Jing Zhou1, Tiantian Zhang1, Zongqi Zhang1, Shanliang Jin2, Qing He1, Junfeng Zhang1.
Abstract
BACKGROUND: Left atrial appendage (LAA) closure (LAAC) can safely and effectively prevent stroke events caused by atrial fibrillation. However, the structure of the LAA is highly variable among individuals, and the optimal method for obtaining measurements remains unknown. HYPOTHESIS: We aimed to study the accuracy of left atrial computed tomography angiography (CTA), three-dimensional (3D) reconstruction using CTA, two-dimensional transesophageal echocardiography (2D-TEE), and digital subtraction angiography (DSA) for measuring the diameter of the LAA and compare their value for selecting occluder size.Entities:
Keywords: 3D reconstruction; atrial fibrillation; left atrial appendage occlusion; occluder size
Mesh:
Year: 2022 PMID: 35673993 PMCID: PMC9346965 DOI: 10.1002/clc.23869
Source DB: PubMed Journal: Clin Cardiol ISSN: 0160-9289 Impact factor: 3.287
Baseline characteristics of the study population.
|
| |
|---|---|
| Age, year | 72.70 ± 8.30 |
| Female | 65 (43.92%) |
| BMI, kg/m2 | 25.17 ± 2.19 |
| Paroxysmal AF | 76 (51.35%) |
| CHA2DS2‐VASc score | 4.45 ± 1.51 |
| HAS‐BLED score | 2.92 ± 1.06 |
| Hypertension | 121 (81.76%) |
| Diabetes | 36 (24.32%) |
| Vascular disease | 61 (41.22%) |
| Chronic heart disease | 19 (12.84%) |
| Stroke or TIA | 66 (44.59%) |
Abbreviations: BMI, body mass index; TIA, transient ischemic attacks.
Figure 1LAA orifice diameter as measured by different imaging methods. (A) Digital subtraction angiography at the 30° right anterior oblique position +20° caudal position to measure the LAA orifice diameter. (B, C) LAA orifice diameter was measured by 3D reconstruction of CTA using Mimics software. (D–G) LAA orifice diameter was measured in four sections at 0°, 45°, 90°, and 135°, respectively, by transesophageal echocardiography. (H–J) LAA orifice diameter was measured by multiplanar reconstruction with CT angiography in different positions (coronal, sagittal, and transverse positions). CT, computed tomography; CTA, computed tomography angiography; LAA, left atrial appendage.
LAA orifice diameter as measured by different imaging methods.
| Imaging methods | Mean ± standard deviation (mm) | Median (quartiles) (mm) |
|---|---|---|
| DSA | 23.15 ± 2.79 | 23.08 (21.31, 25.33) |
| 2D‐TEE | 23.48 ± 2.82 | 23.61 (21.82, 25.40) |
| CTA | 24.07 ± 2.57 | 24.23 (22.42, 25.00) |
| CTA 3D reconstruction | 25.20 ± 2.85 | 25.27 (23.30, 27.23) |
| Occluder size | 28.12 ± 3.04 | 27.00 (27.00, 30.00) |
Note: Since the 2D‐TEE, CTA, and occlude size did not meet the normal distribution, the Kruskal–Wallis H test was used and the Bonferroni method was used for pairwise comparison.
Abbreviations: 2D‐TEE, two‐dimensional transesophageal echocardiography; 3D, three‐dimensional; CTA, computed tomography angiography; DSA: digital subtraction angiography; LAA, left atrial appendage.
Indicates that compared with occluder size, p < .05.
Indicates that compared with DSA, p < .05.
Indicates that compared with ultrasound, p < .05.
Indicates that compared with CTA, p < .05.
Correlation between LAA orifice diameter as measured by different imaging methods and occluder size.
| Imaging methods |
|
|
|
|---|---|---|---|
| DSA and occluder size | .57 | 232252.10 | <.001 |
| 2D‐TEE and occluder size | .57 | 232162.30 | <.001 |
| CTA and occluder size | .88 | 66013.57 | <.001 |
| CTA 3D reconstruction and occluder size | .89 | 58381.65 | <.001 |
Note: Since the 2D‐TEE, CTA, and occlude size did not meet the normal distribution, the Spearman rank correlation analysis and Wilcoxon signed‐rank test were used.
Abbreviations: 2D‐TEE, two‐dimensional transesophageal echocardiography; 3D, three‐dimensional; CTA, computed tomography angiography; DSA, digital subtraction angiography; LAA, left atrial appendage.
Figure 2Correlation between LAA orifice diameter as measured by different imaging methods and occluder size. 2D‐TEE, two‐dimensional transesophageal echocardiography; 3D, three‐dimensional; CTA, computed tomography angiography; DSA, digital subtraction angiography; LAA, left atrial appendage.
Figure 3Consistency between LAA orifice diameter as measured by different imaging methods and occluder size. (Bland–Altman) (A) DSA, (B) 2D‐TEE, (C) CTA, (D) CTA 3D reconstruction. 2D‐TEE, two‐dimensional transesophageal echocardiography; 3D, three‐dimensional; CTA, computed tomography angiography; DSA, digital subtraction angiography; LAA, left atrial appendage.
Consistency between LAA orifice diameter as measured by different imaging methods and occluder size.
| Imaging method | ICC |
|
|---|---|---|
| DSA | −0.067 | .793 |
| 2D‐TEE | 0.006 | .469 |
| CTA | 0.241 | .001 |
| CTA 3D reconstruction | 0.519 | <.001 |
Abbreviations: 2D‐TEE, two‐dimensional transesophageal echocardiography; 3D, three‐dimensional; CTA, computed tomography angiography; DSA, digital subtraction angiography; ICC, intraclass correlation coefficient; LAA, left atrial appendage.