Bassam Lajin1, Joerg Feldmann2, Walter Goessler1. 1. Institute of Chemistry, Analytical Chemistry for the Health and Environment, University of Graz, Universitaetsplatz 1, 8010 Graz, Austria. 2. Institute of Chemistry, TESLA (Trace Element Speciation Laboratory), University of Graz, Universitaetsplatz 1, 8010 Graz, Austria.
Abstract
The inductively coupled plasma mass spectrometry (ICPMS) has been attracting increasing attention for many applications as an element-selective chromatographic detector. A major and fundamental limitation in coupling ICPMS with liquid chromatography is the limited compatibility with organic solvents, which has so far been addressed via a tedious approach, collectively referred to as the "organic ICPMS mode", that can decrease detection sensitivity by up to 100-fold. Herein, we report 1,2-hexanediol as a new eluent in high-performance liquid chromatography-ICPMS which enables avoiding the current limitations. Unlike commonly used eluents, 1,2-hexanediol was remarkably compatible with ICPMS detection at high flow rates of 1.5 mL min-1 and concentrations of at least 30% v/v, respectively, under the standard conditions and instrumental setup normally used with 100% aqueous media. Sensitivity for all tested elements (P, S, Cl, Br, Se, and As) was enhanced with 10% v/v 1,2-hexanediol relative to that of 100% aqueous media by 1.5-7-fold depending on the element. Concentrations of 1,2-hexanediol at ≤30% v/v were superior in elution strength to concentrations at >90% v/v of the common organic phases, which greatly decreases the amount of carbon required to elute highly hydrophobic compounds such as lipids and steroids, enabling detection at ultra-trace levels. The proposed approach was applied to detect arsenic-containing fatty acids in spiked human urine, and detection limits of <0.01 μg As L-1 were achieved, which is >100-fold lower than those previously reported using the organic ICPMS mode. Nontargeted speciation analysis in Allium sativum revealed the presence of a large number of hydrophobic sulfur-containing metabolomic features at trace levels.
The inductively coupled plasma mass spectrometry (ICPMS) has been attracting increasing attention for many applications as an element-selective chromatographic detector. A major and fundamental limitation in coupling ICPMS with liquid chromatography is the limited compatibility with organic solvents, which has so far been addressed via a tedious approach, collectively referred to as the "organic ICPMS mode", that can decrease detection sensitivity by up to 100-fold. Herein, we report 1,2-hexanediol as a new eluent in high-performance liquid chromatography-ICPMS which enables avoiding the current limitations. Unlike commonly used eluents, 1,2-hexanediol was remarkably compatible with ICPMS detection at high flow rates of 1.5 mL min-1 and concentrations of at least 30% v/v, respectively, under the standard conditions and instrumental setup normally used with 100% aqueous media. Sensitivity for all tested elements (P, S, Cl, Br, Se, and As) was enhanced with 10% v/v 1,2-hexanediol relative to that of 100% aqueous media by 1.5-7-fold depending on the element. Concentrations of 1,2-hexanediol at ≤30% v/v were superior in elution strength to concentrations at >90% v/v of the common organic phases, which greatly decreases the amount of carbon required to elute highly hydrophobic compounds such as lipids and steroids, enabling detection at ultra-trace levels. The proposed approach was applied to detect arsenic-containing fatty acids in spiked human urine, and detection limits of <0.01 μg As L-1 were achieved, which is >100-fold lower than those previously reported using the organic ICPMS mode. Nontargeted speciation analysis in Allium sativum revealed the presence of a large number of hydrophobic sulfur-containing metabolomic features at trace levels.
The
employment of the inductively coupled plasma mass spectrometry
(ICPMS) as an element-selective detector for liquid chromatography
has gained increasing popularity in previous years,[1,2] particularly
since the introduction of the tandem mass spectrometry (MS/MS) technology
to the technique,[3] which effectively resolved
polyatomic interferences.[4,5] Coupling chromatography
with an element-selective detector enables a unique approach to chemical
speciation analysis, involving comparative and simultaneous detection
and quantification of chemical forms of multiple elements[6] as well as serving as a tool for the discovery
of novel compounds of environmental, biological, or industrial origin,[7−10] particularly through simplifying molecular metabolomic data in nontargeted
analysis.[11,12]Reversed-phase liquid chromatography
is by far the most common
separation mode but involves the employment of mobile phases containing
high percentages of an organic solvent. Apart from the general environmental
aspects, the use of high quantities of organic solvents presents a
specific challenge for chromatographic detection with ICPMS due to
the known low tolerability of carbon by the inductively coupled plasma,
resulting in plasma shutdown at high carbon load. The general intolerability
of carbon by the inductively coupled plasma has been a well-recognized
and thoroughly investigated subject.[13,14] For chromatographic
speciation analysis using an ICPMS detector, this obstacle has so
far been partly overcome by applying modifications in terms of instrumental
setup and experimental conditions,[15−17] which can be collectively
referred to as the “organic ICPMS mode”.In current
practice, the organic ICPMS mode involves combinations
of the following: (1) using oxygen as an optional gas to help volatilize
carbon and prevent its buildup on the cones, which is associated with
decreased interface and analyzer pressure, signal drift, and clogging
of the sampler and skimmer cones; (2) replacing the Cu/Ni cones with
the more expensive Pt cones, which is recommended under continuous
operation with oxygen as an optional gas to prevent corrosion; (3)
mobile phase flow rate splitting (usually 1:5–1:10) and/or
postcolumn dilution; (4) employing sub-zero temperature for the spray
chamber; and (5) using a plasma torch with a 1.5 mm injector rather
than the standard 2.5 mm injector.Many of the above components
of the organic ICPMS mode can negatively
impact sensitivity, not only through flow splitting/postcolumn dilution
but also through the addition of oxygen to the plasma which may decrease
sensitivity, for example, by increasing the formation of competing
oxides and polyatomic species. Using a narrow injector plasma torch
increases carrier gas velocity and therefore shortens sample residence
time in the plasma, which can negatively impact drying, decomposition,
and ionization efficiency.One case exemplifying the consequences
of the organic ICPMS mode
is the speciation analysis of arsenolipids, where detection limits
of 1.0–10 μg As L–1 are typically reported,[18−20] which is >100-fold higher than typical detection limits reported
for low-molecular-weight hydrophilic arsenic species not requiring
the ICPMS organic mode (0.005–0.03 μg As L–1).[21−23] The latter low detection limits are otherwise achievable
for the arsenolipids with molecular mass spectrometric detection,[24] which is often used in combination with ICPMS
detection for identifying novel species in this class of actively
explored arsenic compounds.[17,25] However, the current
gap in detectability between the two techniques renders the powerful
combination of elemental and molecular MS incapable of identifying
new species in this class of compounds at such low concentration levels,
where ICPMS is the limiting component.Furthermore, the increased
complexity and reduced convenience associated
with the organic ICPMS mode has apparently been a deterring factor
in many nontargeted analysis studies where low organic content mobile
phases were employed,[26,27] and for some elements, this has
likely lead to a gap (i.e., low column recovery) between the total
elemental content and the sum of the individual species detected.[28−30] It is plausible that the inclusion of mobile phases with higher
elution strength in nontargeted screening speciation analyses using
high-performance liquid chromatography (HPLC)-ICPMS on a routine basis
would enable the identification of more novel compounds.Organic
solvents are known to show a wide variation in their tolerability
by the inductively coupled plasma depending on a set of key physicochemical
properties including boiling point, vapor pressure, viscosity, surface
tension, and density.[13] As a proof of concept,
we previously introduced dimethylcarbonate as a new solvent in HPLC-ICPMS,[31] which was found to offer superior elution strength
to commonly used organic solvents such as acetonitrile and methanol,
but the utility of this solvent was limited to concentrations ≤10%
v/v, and these concentrations were insufficient to elute highly hydrophobic
compounds (e.g., lipids).[31] The aim of
the present work was to find alternative organic solvents that can
show high compatibility with ICPMS detection and provide exceptionally
strong chromatographic elution at lower eluent concentrations in order
to enable coupling reversed-phase chromatography with ICPMS detection
under default conditions and standard experimental setup without the
need to employ any of the components of the organic ICPMS mode, which
would have the advantage of increasing the detection capability of
the technique while rendering analysis more convenient.We herein
introduce 1,2-hexanediol as a new eluent to speciation
analysis via liquid chromatography coupled with ICPMS detection, and
we examine its properties and highlight its advantages, remarkable
tolerability by the plasma, and potential to eliminate the need for
the organic ICPMS mode and its associated disadvantages.
Experimental Section
Chromatographic Separation
All chromatographic
investigations were performed using the reversed-phase column Zorbax
Eclipse Plus C18, 50 mm × 2.1 mm i.d., 1.8 μm (Agilent
Technologies, Waldbronn, Germany). A short column length was chosen
to enable the observation of the chromatographic behavior and the
calculation of retention factors up to k = 50 within
reasonable retention times (<30 min) in all experiments. It is
important to keep in mind that chromatographic retention is correctly
measured by the retention factor, which is independent of the column
length, rather than the retention time. The retention factors (capacity
factors) were calculated based on the compound retention time (tR) and column void time (t0) using the formula k = (tR – t0)/t0 and used as a measure of retention throughout the study.
The column void time (t0) was estimated
to be 0.55 min based on the retention time of the unretained sulfate
anion.The following general chromatographic conditions were
employed for all experiments unless otherwise stated: mobile phase
flow rate: 0.25 mL min–1; column temperature: 50
°C; injection volume: 1.0–3.0 μL; and mobile phase
composition: formic acid 0.1% v/v (ACS grade, purity >98%) with
variable
contents of the different organic solvents (1,2-hexanediol, methanol,
acetonitrile, or isopropanol). Elution was performed isocratically,
and the mobile phases were prepared via online mixing of a 2.0% v/v
solution of formic acid, purified water produced in-house using a
Milli-Q water purification system (18.2 MΩ cm, Merck Millipore
GmbH, Vienna, Austria), and pure acetonitrile, methanol, or isopropanol.
A solution of 1,2-hexanediol was prepared offline at a concentration
of 30% v/v in purified water, mixed, and sonicated for 10 min. The
resulting aqueous solution of 1,2-hexanediol was then treated similarly
to the other pure organic solvents, as described above. Chromatographic
reagents and solvents, including 1,2-hexanediol (purity 98%, CAS-Number
6920-22-5), were purchased from Sigma-Aldrich (Steinheim, Germany).A group of hydrophobic compounds with a LogP within the range of
2.4–8.2 (computed using XLogP3 3.0[32]) were selected as model compounds (Figure ) in order to investigate the elution properties
of 1,2-hexanediol in comparison with those of the commonly used organic
solvents methanol, acetonitrile, and isopropanol. The selected compounds
are of medicinal/environmental interest and detectable via ICPMS through
a heteroatom. Standard solutions were prepared in pure methanol at
concentrations of 20–50 mg element L–1, unless
otherwise stated, and injected onto the column using the conditions
described above.
Figure 1
Chemical structures of the hydrophobic model compounds
chosen in
the current study (Log P 2.4–8.2). ICPMS/MS
detection was based on the highlighted heteroatoms.
Chemical structures of the hydrophobic model compounds
chosen in
the current study (Log P 2.4–8.2). ICPMS/MS
detection was based on the highlighted heteroatoms.
Chromatographic Detection
ICPMS/MS
detection was performed using an Agilent 8900 ICPQQQ system coupled
with an Agilent 1100 HPLC system. The ICPMS/MS system consisted of
an AriMist polyether ether ketone (PEEK) nebulizer (maximum nebulizer
gas flow rate: 0.8 L min–1), a glass double pass
spray chamber (cooled at 2 °C), a nickel/copper sampler and skimmer
cones, and a quartz plasma torch with an injector inner diameter of
2.5 mm. The use of oxygen as an optional gas or flow splitting/postcolumn
dilution was deliberately avoided in all experiments with 1,2-hexanediol.The ICPMS/MS detector was operated using the following parameters:
RF power: 1550 W; plasma gas: 15.0 L min–1; auxiliary
gas: 0.9 L min–1; RF matching: 1.3–1.8 V
[depending on the concentration of 1,2-hexanediol in the mobile phase
(0–30% v/v) at a 0.25 mL min–1 flow rate];
sampling position (sampling depth): 5.0 mm; nebulizer gas flow rate:
0.65 L min–1; makeup gas (argon) flow rate: 0.25–0.45
L min–1 (used to yield a total carrier gas flow
of 0.9–1.1 L min–1 depending on the organic
content; lower organic content required a higher total carrier gas
flow rate for optimum sensitivity); optional gas: 0.0%; nebulizer
pump speed (for drainage): 0.50 rps (ca. 2.0 mL min–1); and S/C (spray chamber) temperature: 2 °C. Chlorine was detected
in the hydrogen mode (H2 flow rate: 3.5 mL min–1) by monitoring the transition m/z 35 → 37 (as 35Cl1H2+), and all other elements in the oxygen mode (O2 flow rate: 0.3 mL min–1) by monitoring the transitions
corresponding to m/z M+ → M+ + 16.Chromatographic experiments involving
mobile phases containing
>10–25% of methanol, acetonitrile, or isopropanol could
not
be performed with ICPMS detection under the standard conditions described
above due to plasma instability. For simplicity, an Agilent 1260 spectrophotometric
detector was employed to investigate these solvents, using chromatographic
conditions identical to those described above with ICPMS detection.
Furthermore, confirmation of the elution patterns displayed by the
most hydrophobic compounds included in the study (cholesterol sulfate
and arsenic-containing fatty acids) was performed using a molecule-selective
detector (Agilent triple quadrupole Ultivo ESIMS/MS system) by monitoring m/z 363 and 419 for the arsenic-containing
fatty acids AsFA 362 and AsFA 418, respectively, in the positive mode,
and m/z 465 for cholesterol sulfate
in the negative mode using the following source settings: nebulizer
gas temperature and flow rate: 350 °C and 10 L min–1, respectively; sheath gas temperature and flow rate: 400 °C
and 12 L min–1, respectively; nebulizer pressure:
35 psi; and capillary voltage: 3000 V.
Results
and Discussion
Selection of 1,2-Hexanediol
The degree
of plasma tolerability for organic matrices depends on various physicochemical
properties, most notably, boiling point, vapor pressure, and viscosity,[13] and the selection of 1,2-hexanediol was based
on these properties. In particular, an extremely low vapor pressure
of 2.7 Pa at 20 °C (0.02 mmHg) and a high boiling point of 224
°C would be expected to result in low vapor transport and carbon
load on the plasma. On the other hand, a high LogP of 0.7[32] for 1,2-hexanediol would enable higher chromatographic
elution strength at lower organic content. The superiority of 1,2-hexanediol
becomes clear when comparing the above-mentioned properties with those
for the commonly employed solvents methanol, acetonitrile, and isopropanol
(see Supporting Information Table S1).This is not the first report describing the incorporation of 1,2-hexanediol
in a mobile phase for reversed-phase liquid chromatography. Li and
Fritz reported the addition of 1% v/v 1,2 hexanediol as a chromatographic
modifier to improve the separation of hydrophilic organic acids (e.g.,
formic acid and acetic acid) under spectrophotometric and conductometric
detection.[33] However, the behavior of 1,2-hexanediol
as a general chromatographic eluent (i.e., at concentrations >1%)
rather than a chromatographic modifier was not previously reported,
and its employment with ICPMS detection has not been previously described.
Chromatographic Elution Behavior
A group
including highly hydrophobic compounds (up to a LogP of 8.2,
computed using XLOGP3 3.0[32]) was selected
(Figure ). These compounds
are amenable to detection via ICPMS through the presence of a heteroatom
and are relevant to pharmaceutical, biological, and/or environmental
applications.1,2-Hexanediol showed superior elution strength
relative to commonly used solvents even when these were employed at
much higher concentrations (Figure ). Overall, direct experimental data as well as calculations
based on the linear regression of the Log C (eluent
concentration) versus Log k (retention factor) relationship
revealed that 1,2-hexanediol at concentrations within the range of
1.0–25% v/v can replace methanol, acetonitrile, and isopropanol
within the concentration ranges of 20–95, 10–85, and
5.0–60%, respectively (Figure ). For highly hydrophobic compounds such as cholesterol
sulfate, the elution strength of 1,2 hexanediol at >25% v/v could
not be matched by that of any concentration of acetonitrile or methanol
(Supporting Information Figure S1). It
is noteworthy that 1,2-hexanediol was also found to be applicable
for compounds with low hydrophobicity (Log P <1.0)
when used at concentrations as low as 0.5–1.0% v/v, matching
10–20% v/v methanol (Supporting Information Figure S2).
Figure 2
Comparing the elution strength of 1,2-hexanediol (A) with
solvents
commonly used as eluents in reversed-phase chromatography, namely,
isopropanol (B), acetonitrile (C), and methanol (D). Only 1,2-hexanediol
was compatible with direct ICPMS/MS detection at the investigated
concentration range (for conditions, see the Experimental
Section), and therefore, detection with the other eluents was
undertaken using a spectrophotometric detector at 254 nm. The column
void time is 0.55 min. Note that the selectivity (and peak order)
for 1,2-hexanediol is similar to that of isopropanol and differs from
that of acetonitrile and methanol. Peak a: cloxacillin; peak b: mometasone
furoate; peak c: diclofenac; peak d: pantoprazole; and peak e: ethoxysulfuron.
Figure 3
Concentrations of eluents showing comparable elution strength
to
that of 10% (A) and 25% v/v (B) of 1,2-hexanediol. The values were
calculated based on linear regression lines (r2 = 0.9990–0.9999) constructed by plotting log k (retention factor) values against log C % (percentage concentration), based on experiments where retention
times were recorded under varying organic solvent proportions. Note
that for highly hydrophobic compounds (e.g., cholesterol sulfate investigated
in the present study), the elution strength of >20% v/v 1,2-hexanediol
may not be matched by any concentration of methanol or acetonitrile
(see Supporting Information Figure S1).
Pa, pantoprazole; Clox, cloxacillin; Eth, ethoxysulfuron; Mom, mometasone
furoate; and Dicl, diclofenac.
Comparing the elution strength of 1,2-hexanediol (A) with
solvents
commonly used as eluents in reversed-phase chromatography, namely,
isopropanol (B), acetonitrile (C), and methanol (D). Only 1,2-hexanediol
was compatible with direct ICPMS/MS detection at the investigated
concentration range (for conditions, see the Experimental
Section), and therefore, detection with the other eluents was
undertaken using a spectrophotometric detector at 254 nm. The column
void time is 0.55 min. Note that the selectivity (and peak order)
for 1,2-hexanediol is similar to that of isopropanol and differs from
that of acetonitrile and methanol. Peak a: cloxacillin; peak b: mometasone
furoate; peak c: diclofenac; peak d: pantoprazole; and peak e: ethoxysulfuron.Concentrations of eluents showing comparable elution strength
to
that of 10% (A) and 25% v/v (B) of 1,2-hexanediol. The values were
calculated based on linear regression lines (r2 = 0.9990–0.9999) constructed by plotting log k (retention factor) values against log C % (percentage concentration), based on experiments where retention
times were recorded under varying organic solvent proportions. Note
that for highly hydrophobic compounds (e.g., cholesterol sulfate investigated
in the present study), the elution strength of >20% v/v 1,2-hexanediol
may not be matched by any concentration of methanol or acetonitrile
(see Supporting Information Figure S1).
Pa, pantoprazole; Clox, cloxacillin; Eth, ethoxysulfuron; Mom, mometasone
furoate; and Dicl, diclofenac.The ability of 1,2-hexanediol to elute highly hydrophobic compounds
at low concentrations can be explained by its high hydrophobicity
as it has a Log P value of 0.7, which is remarkably
higher than that of commonly employed solvents (Supporting Information Table S1). Furthermore, 1,2-hexanediol has an
amphiphilic structure with a long hydrocarbon chain and two adjacent
polar hydroxyl groups. This has two consequences. First, 1,2-hexanediol
can efficiently compete with the hydrophobic analytes and strongly
adsorb onto the hydrophobic C18 stationary phase, which in turn results
in a decrease in the overall hydrophobicity of the latter due to coating
with the adjacent hydroxyl groups in 1,2-hexanediol. This adsorption
of 1,2-hexanediol and the resulting availability of hydrogen bonding
with the coated reversed phase would be expected to influence chromatographic
selectivity. Indeed, the peak order with 1,2-hexanediol was different
from that with methanol and acetonitrile and similar to that with
the more structurally related isopropanol (compare peaks c and b in Figure ). This influence
of 1,2-hexanediol on chromatographic selectivity was also observed
in a previous report where 1,2-hexanediol was used as a modifier to
alter selectivity in electrokinetic chromatography.[34] Second, the amphiphilic structure of 1,2-hexanediol enables
micelle formation, which can enhance solubilization of the hydrophobic
analytes in the mobile phase and the elution strength. Indeed, we
observed a change in the slope for the curve depicting the relationship
between Log C and Log k around a
concentration of 8–9% v/v of 1,2-hexanediol (Supporting Information Figure S3), which was found to be commensurate
with the previously reported critical micelle concentration (cmc)
of 1,2-hexanediol (0.7 M).[35]It is
notable that 1,2-hexanediol has a markedly higher viscosity
(87 mPa s at 20 °C) relative to that of the other commonly employed
eluents (Supporting Information Table S1). However, the viscosity of its aqueous mixtures drops sharply with
temperature.[36] We therefore recommend operating
at a column temperature of ≥45 °C to ensure compatibility
with standard 4.00 × 107 Pa (400 bar) pumps at a chromatographic
column length of 250 mm. At 30% v/v 1,2-hexanediol and a mobile phase
flow rate of 1.0 mL min–1, the resulting backpressure
using a reversed-phase (C18) column with a length of 250 mm, 4.6 mm
i.d., and 5 μm particle size (Phenomenex Synergi Fusion-RP)
was 2.80 × 107 Pa (280 bar) at 50 °C column temperature.
Supporting Information Figure S4 illustrates
the backpressure profile of aqueous mixtures of 1,2-hexanediol in
comparison with those of methanol.
Tolerability
and Plasma Stability
Under standard conditions (i.e., without
oxygen as the optional gas,
platinum cones, or flow splitting/postcolumn dilution) and the standard
instrumental setup (including a standard 2.5 mm i.d. plasma torch
and an AriMist nebulizer) and using a 2.1 mm i.d. chromatographic
column operated at its conventional flow rate of 0.25 mL min–1, we observed no plasma instability throughout the study with mobile
phases containing concentrations of up to 30% v/v of 1,2-hexanediol
(higher concentrations were not tested), and reflected power remained
constant at 0 W with an applied RF matching value of 1.3–1.8
V (depending on the concentration of 1,2-hexanediol). Prolonged operation
under the above conditions did not produce a significant change in
analyzer pressure or visible carbon buildup on the interface (Supporting
Information Figure S5). Additionally, a
plasma torch with a 1.5 mm i.d. was also tested with similar results
except that no adjustment for the RF matching value was required (default
value of 1.3 V was used).Even though narrow-bore chromatographic
columns are generally preferable due to reduced solvent consumption
and increased sensitivity, we tested higher mobile phase flow rates
frequently employed with 4.6 mm i.d. columns. At 30% v/v of 1,2-hexanediol
(which provides superior elution strength to that of >90% v/v methanol
or acetonitrile, as described above) and mobile phase flow rates of
up to 1.5 mL min–1, the plasma was stable (<3
W reflected power), and the sensitivity was the highest at a total
carrier gas flow rate (nebulizer gas + argon make-up gas) of 0.90
L min–1 including a 0.65 L min–1 nebulizer gas flow rate (AriMist nebulizer) and a 0.25 L min–1 make-up (argon) gas mobile phase flow rate (Supporting
Information Figure S6). Lower organic contents
were found to require a higher total carrier gas flow rate for maximum
sensitivity (0.90–1.0 L min–1 for 5.0–30%
v/v 1,2-hexanediol).For high mobile phase flow rates (>0.75
mL min–1) combined with a high 1,2-hexanediol content
(>25% v/v), the RF
matching had to be increased gradually up to 2.2 V to maintain the
reflected power at <3 W and plasma stability. It is also noteworthy
that for these combinations of high mobile phase flow rates and organic
content, we observed plasma instability when using certain combinations
of a low nebulizer gas flow rate (<0.60 L min–1) and a low total carrier gas flow rate (<0.9 L min–1). These conditions were however practically irrelevant as they were
associated with a decrease or no significant change (within ±20%)
in sensitivity (e.g., see Supporting Information Figure S6). Similar patterns were observed using a micromist
nebulizer, and maximum sensitivity and plasma stability were achieved
at 0.90–0.95 L min–1 nebulizer/carrier gas
flow rates (no argon make-up gas required) at 30% v/v 1,2-hexanediol
and a mobile phase flow rate of up to 1.5 mL min–1 (higher concentrations of 1,2-hexanediol and mobile phase flow rates
were not tested).The exceptionally high tolerability of 1,2-hexanediol
becomes most
evident when comparing it with the other commonly used solvents in
reversed-phase chromatography. At conditions comparable to the above
described (including a 2.5 mm i.d. plasma torch), it was not possible
to sustain a stable plasma even at mobile phase flow rates of <0.25
mL min–1 at concentrations >10–25% of
acetonitrile,
methanol, or isopropanol, which is in sharp contrast with the observed
stability of 1,2-hexanediol at 30% v/v concentration and up to 1.5
mL min–1 mobile phase flow rate. The use of the
1.5 mm torch, which is known to confer a much higher tolerability
for organic solvents, was deliberately avoided in order to demonstrate
the tolerability of 1,2-hexanediol in comparison with that of the
commonly used organic solvents.It is generally known that solvents
with high boiling points and
low vapor pressure are better tolerated by the plasma because of the
reduced carbon load due to vapor transfer. Indeed, when comparing
with methanol, the current data showed that at v/v % concentrations
corresponding to equal carbon molarity, 1,2-hexanediol results in
roughly half the carbon load, as estimated by monitoring the 40Ar 12C signal (Supporting Information Figure S7). A thorough discussion of the criteria
contributing to a higher plasma tolerance for organic solvents can
be found elsewhere.[13]
Influence on Sensitivity
The impact
of 1,2-hexanediol and methanol on the sensitivity of detection for
six commonly involved elements in speciation analysis using HPLC-ICPMS/MS
was compared within the concentration range of 5.0–20% v/v
(note that the carbon molarity values in pure 1,2-hexanediol and pure
methanol are 48 and 25 M, respectively). It was found that 1,2-hexanediol
resulted in higher sensitivity than methanol when normalized to 100%
aqueous media (Figure ). Overall, 1,2-hexanediol at concentrations of up to 20% v/v was
found to enhance the signal for all elements tested (relative to 100%
aqueous solution) including a slight enhancement for chlorine, which
among the tested elements has the highest ionization potential at
13.0 eV (Figure ).
By contrast, in order to match the elution strength of the above concentrations
of 1,2-hexanediol, up to 90% v/v methanol would be needed, which would
require employing the organic ICPMS mode and therefore significantly
compromise sensitivity. The limits of detection achievable when using
a mobile phase containing 10% v/v 1,2-hexanediol were estimated based
on the S/N = 3 method to be 0.1 μg P L–1,
0.3 μg S L–1, 0.01 μg Se L–1, 0.003 μg As L–1, 4.2 μg Cl L–1, and 2.0 μg Br L–1 (injection
volume: 50 μL, based on the chromatographic peak for the inorganic
form with a peak width of 0.3 min).
Figure 4
Influence of 1,2-hexanediol and similar
concentrations of methanol
on the sensitivity for the detection of multiple elements (in inorganic
forms) relative to pure water. The y-axis shows the
signal enhancement/suppression factors calculated based on the signal
response ratio between the investigated composition of the organic
solvent and 100% aqueous solution (all with 0.1% formic acid) for
inorganic forms of the investigated elements injected onto the reversed-phase
chromatographic column. The error bars represent the standard deviation
(n = 3). Attention has to be paid to the difference
in carbon molarity between pure methanol and pure 1,2-hexanediol (25
and 48 M for methanol and 1,2-hexanediol, respectively). It is noteworthy
that sensitivity in ICPMS/(MS) detection depends on not only carbon
concentration but also the nebulization properties of the eluent,
which is a product of several factors including viscosity, surface
tension, and droplet size distribution. Therefore, while general conditions
employed for these experiments were identical (see the Experimental Section), each composition of the organic solvent
required applying slightly different optimum carrier gas (higher organic
compositions were found to necessitate lower carrier gas flow rates).
Influence of 1,2-hexanediol and similar
concentrations of methanol
on the sensitivity for the detection of multiple elements (in inorganic
forms) relative to pure water. The y-axis shows the
signal enhancement/suppression factors calculated based on the signal
response ratio between the investigated composition of the organic
solvent and 100% aqueous solution (all with 0.1% formic acid) for
inorganic forms of the investigated elements injected onto the reversed-phase
chromatographic column. The error bars represent the standard deviation
(n = 3). Attention has to be paid to the difference
in carbon molarity between pure methanol and pure 1,2-hexanediol (25
and 48 M for methanol and 1,2-hexanediol, respectively). It is noteworthy
that sensitivity in ICPMS/(MS) detection depends on not only carbon
concentration but also the nebulization properties of the eluent,
which is a product of several factors including viscosity, surface
tension, and droplet size distribution. Therefore, while general conditions
employed for these experiments were identical (see the Experimental Section), each composition of the organic solvent
required applying slightly different optimum carrier gas (higher organic
compositions were found to necessitate lower carrier gas flow rates).A high concentration of carbon generally suppresses
the ICPMS signal,
particularly for elements with high ionization energy such as the
halogens.[37] However, a few elements, notably
arsenic, selenium, and phosphorous, are known to show an increase
in the signal at moderate carbon concentrations through the so-called
“carbon enhancement effect”,[38] which has been previously investigated.[39] The outcome of organic media on detection via ICPMS is however not
straightforward since the final signal response is governed not only
by carbon concentration but also by the impact of the presence of
an organic solvent on physical properties such as viscosity and surface
tension, which affect the nebulization process.The basis of
the superior sensitivity with 1,2-hexanediol compared
with that of methanol is not clear. It is worth noting however that
due to its amphiphilic structure, 1,2-hexanediol can act as a surfactant
and greatly lower the surface tension of water. Indeed, the surface
tension of 5.0–20% v/v of 1,2 hexanediol was reported to be
within the range of 25–26 mN m–1 at 20 °C,[40] which is considerably lower than that for corresponding
concentrations of methanol (50–65 mN m–1).[41] It can be assumed that lower surface tension
might result in a more efficient nebulization process by producing
a finer spray, which can result in not only more efficient droplet
transfer to the plasma but also more rapid desolvation of the fine
droplets within the plasma. The effects of surfactants on the nebulization
efficiency in atomic spectrometric techniques have been previously
discussed.[42] Additionally, it is plausible
that more rapid desolvation of finer droplets may have contributed
to the observed high plasma stability with 1,2-hexanediol.
Proof-Of-Concept Applications
Sulfur
Speciation in Allium
sativum (Garlic)
A methanolic extract of
freshly minced garlic (ca. 0.5 g mL–1) was analyzed
following direct injection into the HPLC-ICPMS/MS system, and the
sulfur metabolomic profiles were compared between those resulting
with 5.0–20% v/v 1,2-hexanediol and that with 20% v/v methanol
(higher methanol concentrations extinguished the plasma under the
standard conditions and instrumental setup employed). The major sulfur
species detected was allicin at 190 mg S L–1 (Figure ), which eluted at k = 12 with 20% v/v methanol and k = 3.0
with as little as 5% v/v 1,2-hexanediol. Identification was confirmed
using molecular MS at m/z 163 →
41.[43] Moreover, 1,2-hexanediol enabled
the detection of a larger number of hydrophobic sulfur compounds,
including at least five major compounds (20–200 mg L–1) and >10 minor compounds (0.1–1.0 mg S L–1), see Figure a–c.
Note that the lower sensitivity under an organic ICPMS mode, which
would otherwise be necessary to elute these minor compounds, may render
these undetectable (Figure a,c). Furthermore, the simple profile under the high elution
strength of 20% v/v 1,2-hexanediol over a prolonged elution time (Figure c) suggests that
the elution of the sulfur metabolome is likely complete and renders
missing yet-to-be identified compounds less likely.
Figure 5
Applications involving
the use of 1,2-hexanediol as an eluent for
speciation analysis using HPLC-ICPMS. Chromatograms A–D show
the sulfur metabolomic profile in a methanolic extract of freshly
minced garlic (Allium Sativum) at ca.
0.5 g mL–1. Different concentrations of 1,2-hexanediol
and methanol (indicated on the chromatograms) as eluents were used,
and the resulting profiles were compared. A larger number of features
was observed under 1,2-hexanediol, including at least five major compounds
(20–200 mg S L–1), such as allicin, which
is known as a dominant sulfur compound in garlic, along with >10
minor
and trace compounds (0.1–1 mg S L–1). Chromatogram
E shows the detection of two arsenic fatty acids in spiked urine at
a concentration of 0.05 μg As L–1 (injection
volume: 50 μL). Inorganic arsenic (iAs), dimethylarsinate (DMA),
and arsenobetaine (AB) elute in the front.
Applications involving
the use of 1,2-hexanediol as an eluent for
speciation analysis using HPLC-ICPMS. Chromatograms A–D show
the sulfur metabolomic profile in a methanolic extract of freshly
minced garlic (Allium Sativum) at ca.
0.5 g mL–1. Different concentrations of 1,2-hexanediol
and methanol (indicated on the chromatograms) as eluents were used,
and the resulting profiles were compared. A larger number of features
was observed under 1,2-hexanediol, including at least five major compounds
(20–200 mg S L–1), such as allicin, which
is known as a dominant sulfur compound in garlic, along with >10
minor
and trace compounds (0.1–1 mg S L–1). Chromatogram
E shows the detection of two arsenic fatty acids in spiked urine at
a concentration of 0.05 μg As L–1 (injection
volume: 50 μL). Inorganic arsenic (iAs), dimethylarsinate (DMA),
and arsenobetaine (AB) elute in the front.
Detection of Arsenic-Containing Fatty Acids
in Spiked Human Urine
The elution of arsenolipids in HPLC-ICPMS
has been performed using mobile phases containing 70–100% organic
solvent with detection limits typically reported in the range of 1.0–10
μg As L–1,[18−20] which are considerably
higher than those typically reported for low-molecular-weight hydrophilic
arsenic species analyzed under standard conditions (0.005–0.030
μg As L–1).[21−23]Figure e shows the detection of arsenic-containing
fatty acids with C14 and C18 carbon chains (AsFA 362 and AsFA 418)
at a concentration of 0.05 μg As L–1 in spiked
urine using as little as 10% v/v 1,2-hexanediol. Morning first-pass
urine was collected from a healthy volunteer and directly injected
without sample preparation other than filtration using a 0.22 μm
pore size Nylon syringe filter. The calculated limit of detection
for the arsenic-containing fatty acids in the spiked urine based on
the S/N = 3 definition was 0.003 μg As L–1 (injection volume 50 μL), see Supporting Information Figure S8.
Safety
and Environmental Aspects
1,2-Hexanediol is widely used in
the cosmetic industry as a preservative
with antibacterial activity as well as an emulsifying and moisturizing
agent at concentrations of >2%, and its safety for human use has
been
investigated.[44−46] According to current data from the European Chemicals
Agency (ECHA),[47] the toxicities of 1,2-hexanediol
was tested in daphnia and microorganisms with EC10 (48 h) > 110
mg
L–1 and EC50 (3 h) > 1000 mg L–1, respectively. The oral LD50 for 1,2-hexanediol in rats
was reported to be >5000 mg kg–1, compared with
values of 1187 mg kg–1 for methanol and 617 mg kg–1 for acetonitrile. Furthermore, 1,2-hexanediol was
categorized by the European chemicals agency (ECHA) as “readily
biodegradable” with 83% degradation in 28 days in a biodegradability
test performed according to the OECD 301B guideline. Overall, 1,2-hexanediol
appears to be less toxic than acetonitrile and methanol, and in light
of the far lower concentrations of 1,2-hexanediol needed for elution,
it might be considered as a greener alternative as a general eluent
in reversed-phase liquid chromatography with and without ICPMS detection.
Conclusions
1,2-Hexanediol is shown to be
well-tolerated by the plasma, does
not negatively impact the detection sensitivity of ICPMS, and provides
strong chromatographic elution of highly hydrophobic compounds at
low carbon concentrations. The employment of 1,2-hexanediol in mobile
phases for HPLC-ICPMS at concentrations of <30% v/v can be a replacement
for >90% v/v of common organic eluents, eliminating the inconvenience
and the negative impact of the organic ICPMS mode on detection sensitivity.
This can increase the likelihood of detecting low levels of novel
hydrophobic compounds in nontargeted analysis and enables quantification
in targeted analysis at trace levels.