BACKGROUND: To understand the accuracy of volume calibration syringes used in China and compare the difference between new and old volume calibration syringes, technical testing was performed on volume calibration syringes in clinical lung function instruments.Entities:
Keywords: Leak test; Quality control; Volume accuracy; Volume calibration syringes
Mesh:
Year: 2022 PMID: 35659277 PMCID: PMC9166523 DOI: 10.1186/s12890-022-02010-1
Source DB: PubMed Journal: BMC Pulm Med ISSN: 1471-2466 Impact factor: 3.320
Fig. 1Operation diagram of the volume calibration syringe of the 1180 capacity calibration cylinder verifier from Hans Rudolph Company in the United States. Note: ① optical encoder, ② graphite piston, ③ controller, ④ rotary valve, and ⑤ volume calibration syringe to be verified
Results of the leak test and volume accuracy test for 16 volume calibration syringes
| The evaluation index | NEW-D1 | NEW-D2 | NEW-D3 | NEW-D4 | NEW-D5 | NEW-D6 | NEW-D7 | NEW-D8 | OLD-D1 | OLD-D2 | OLD-D3 | OLD-D4 | OLD-D5 | OLD-D6 | OLD-D7 | OLD-D8 | Overall pass rate/% |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Use fixed number of years | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 2.5 | 2.5 | 1 | 1.5 | 3 | 3 | 2 | 5 | |
| The target volume (mL) | 1000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 100.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 100.0 | |
| Objective to reveal (mL/min) | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | |
| Actual average leakage (mL/min) | 0.1 | 0.2 | 0.9 | 0.5 | 2.5 | 0.0 | 0.8 | 0.3 | 0.3 | 0.1 | 0.4 | 0.2 | 1.0 | 1.9 | — | 0.1 | 75 |
| Reveal the mean error value (mL/min) | 0.1 | 0.2 | 0.9 | 0.5 | 2.5 | 0.0 | 0.8 | 0.3 | 0.3 | 0.1 | 0.4 | 0.2 | 1.0 | 1.9 | — | 0.1 | 75 |
| Whether through (Y/N) | Y | Y | Y | Y | N | Y | Y | Y | Y | Y | Y | Y | N | N | N | Y | 75 |
| The target volume (mL) | 1000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 100.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 3000.0 | 100.0 | |
| Average volume measurement (mL) | 992.2 | 2994.7 | 2996.9 | 3003.0 | 2990.9 | 2995.7 | 3001.2 | 96.9 | 2994.4 | 2998.6 | 2992.5 | 2990.6 | 2995.4 | 2998.9 | 2979.2 | 92.1 | 88 |
| Volume mean error range (mL) | 0.5 | 0.4 | 0.6 | 0.5 | 0.8 | 1.4 | 0.4 | 0.4 | 1.5 | 0.6 | 1.0 | 1.2 | 0.4 | 1.8 | 13.5 | 0.9 | 69 |
| Volume mean error (mL) | − 7.8 | − 5.3 | − 3.1 | 3.0 | − 9.1 | − 4.3 | 1.2 | − 3.1 | − 5.6 | 1.4 | − 7.5 | − 9.4 | − 4.6 | − 1.1 | − 20.8 | − 7.9 | 88 |
| Whether through (Y/N) | N | Y | Y | Y | Y | Y | Y | N | Y | Y | Y | Y | Y | N | N | N | 69 |
| Two tests whether through (Y/N) | N | Y | Y | Y | N | Y | Y | N | Y | Y | Y | Y | N | N | N | N | 56 |
“—” indicates that the leak test could not be completed due to serious gas leakage during the leak test