Validation of spirometer calibration syringes.

Flemming Madsen1.   

Abstract

The volume calibration syringe is probably the single most important instrument in pulmonary function laboratories, yet no validation results have been published. In this study a sample of volume calibration syringes was validated. We weighed a 1-L and two 3-L calibration syringes before and after emptying them of water and determined the corresponding volume of gas by using a modified rolling seal spirometer. In this way we established an unbroken calibration chain between a certified weight of water and the corresponding volume of gas. The volume of a spirometer calibration syringe could be verified with an accuracy of ± 15 ml. The modified rolling seal spirometer had an accuracy of ± 11 ml at a volume of 1 L and ± 13 ml at a volume of 3 L. A sample of spirometer calibration syringes was validated and all syringes except for two small 1-liter syringes all had volumes within the label claimed volume ± 0.5%. Spirometer calibration syringes have a stable stroke volume even after many years of use and storage but have to be calibrated yearly to comply with international standard.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22998085     DOI: 10.3109/00365513.2012.723739

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  4 in total

1.  Quality inspection and result analysis of the spirometer calibration cylinder.

Authors:  Zhongping Wu; Yongyi Peng; Kuiqing Lin; Ruibo Huang; Jinping Zheng; Yi Gao
Journal:  BMC Pulm Med       Date:  2022-06-04       Impact factor: 3.320

2.  Characterization of optical-surface-imaging-based spirometry for respiratory surrogating in radiotherapy.

Authors:  Guang Li; Jie Wei; Hailiang Huang; Qing Chen; Carl P Gaebler; Tiffany Lin; Amy Yuan; Andreas Rimner; James Mechalakos
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

3.  Standardization of Spirometry 2019 Update. An Official American Thoracic Society and European Respiratory Society Technical Statement.

Authors:  Brian L Graham; Irene Steenbruggen; Martin R Miller; Igor Z Barjaktarevic; Brendan G Cooper; Graham L Hall; Teal S Hallstrand; David A Kaminsky; Kevin McCarthy; Meredith C McCormack; Cristine E Oropez; Margaret Rosenfeld; Sanja Stanojevic; Maureen P Swanney; Bruce R Thompson
Journal:  Am J Respir Crit Care Med       Date:  2019-10-15       Impact factor: 21.405

4.  A pilot study of hot-wire, ultrasonic and wedge-bellows spirometer inter- and intra-variability.

Authors:  Marit E Aardal; Lene L Svendsen; Sverre Lehmann; Tomas M Eagan; Ingvild Haaland
Journal:  BMC Res Notes       Date:  2017-10-10
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.