Katarzyna Ewa Nowik1, Kamil Nowik1, Piotr Kanclerz2,3, Jacek Paweł Szaflik1.
Abstract
Purpose: Extended depth of focus intraocular (EDOF) IOLs form a bridge between single- and multifocal IOL design. This study aimed to compare clinical outcomes obtained after implanting two different optical designs of EDOF IOLs: the Mini Well Ready (SIFI Medtech, Catania, Italy) and Tecnis Symfony (Abbott Laboratories, Illinois, USA).Entities:
Keywords: cataract; defocus curve; extended depth of focus; intraocular lens
Year: 2022 PMID: 35651536 PMCID: PMC9150758 DOI: 10.2147/OPTH.S341698
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Patients’ Occupation in Both Groups
| Occupation | Group 1 (Mini Well Ready) | Group 2 (Symfony) |
|---|---|---|
| Clerk | 8 patients | 8 patients |
| Medical staff | 5 patients | 5 patients |
| Lawyer | 3 patients | 3 patients |
| IT specialist | 7 patients | 6 patients |
| Clergyman | 1 patient | 0 patients |
| Mechanic | 0 patients | 1 patient |
| Pensioner | 8 patients | 6 patients |
Demographic Data of Patients Before Surgery
| Parameters | Group 1 (Mini Well Ready IOL) | Group 2 (IOL Symfony) |
|---|---|---|
| Age (years) | 64.59 ± 10.63 | 68.34 ± 10.59 |
| Gender | Women- 20, men-12 | Women-19, men-10 |
| Mean keratometry K1, K2 (D) | K1: 43.21 ± 1.56, K2: 43.74 ± 1.67 | K1: 42.82 ± 1.45, K2: 43.34 ± 1.41 |
| Mean anterior astigmatism (D) | 0.46 ± 0.19 | 0.01 ± 0.45 |
| Mean anterior chamber depth (mm) | 3.23 ± 0.44 | 3.12 ± 0.33 |
| Mean axial length (mm) | 23.68 ± 1.56 | 23.56 ± 0.82 |
| Mean spherical equivalent (D) | −0.20 ± 2.27 | 0.24 ± 2.24 |
| Mean target refraction error (D) | −0.09 ± 0.21 | −0.33 ± 0.20 |
| Mean IOL dioptric power (D) | 21.29 ± 3.03 | 21.9 ± 2.22 |
| Mean WTW (white-to-white diameter- mm) | 12.12 ± 0.46 | 12.22 ± 0.5 |
| Mean lens thickness (mm) | 4.44 ± 0.42 | 4.58 ± 0.32 |
Note: All results are expressed as a number or mean absolute average ± standard deviation (SD).
Pre- and Post-Operative Visual Acuity Results in Patients with Mini Well and Symfony Implanted Lenses
| Mean VA (logMAR) | Group 1 (Mini Well) Preoperative | Group 1 (Mini Well) Postoperative | P-value | Group 2 (Symfony) Preoperative | Group 2 (Symfony) Postoperative | P-value |
|---|---|---|---|---|---|---|
| UDVA monocular (4m) | 0.77±0.48 | 0.04±0.03 | 0.0005 | 0.66±0.34 | 0.00±0.06 | 0.0042 |
| Monocular CDVA (4m) | 0.43±0.38 | −0.01±0.05 | 0.0021 | 0.44±0.20 | −0.02±0.07 | 0.0002 |
| Binocular UDVA (4m) | 0.60±0.15 | 0.02±0.04 | 0.005 | 0.65±0.16 | −0.02±0.04 | 0.0002 |
| Binocular CDVA (4m) | 0.38±0.14 | −0.01±0.07 | 0.0000 | 0.39±0.18 | −0.05±0.05 | 0.0000 |
| Monocular UIVA (60 cm) | Not measured | −0.01±0.04 | N/A | Not measured | 0.03±0.06 | N/A |
| Monocular DCIVA (60cm) | Not measured | −0.01±0.04 | N/A | Not measured | 0.00±0.02 | N/A |
| Binocular UIVA (60 cm) | Not measured | −0.01±0.05 | N/A | Not measured | 0.01±0.03 | N/A |
| Binocular DCIVA (60 cm) | Not measured | −0.02±0.05 | N/A | Not measured | −0.01±0.05 | N/A |
| Monocular UNVA (40 cm) | Not measured | 0.00±0.05 | N/A | Not measured | 0.11±0.09 | N/A |
| Monocular DCNVA (40cm) | Not measured | 0.00±0.04 | N/A | Not measured | 0.02±0.12 | N/A |
| Binocular UNVA (40 cm) | Not measured | −0.01±0.05 | N/A | Not measured | 0.05±0.06 | N/A |
Abbreviations: VA, visual acuity; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; UIVA, uncorrected intermediate distance visual acuity; DCIVA, distance-corrected intermediate visual acuity; UNVA, uncorrected near visual acuity; DCNVA, distance-corrected near visual acuity.
Figure 1Binocular defocus curve in patients implanted with the Mini Well and Symfony EDOF-IOL. The vertical bars represent the standard deviation of defocus.