Yuqi Shi1, Yubing Hu1, Nan Jiang2, Ali K Yetisen1. 1. Department of Chemical Engineering, Imperial College London, South Kensington, London, SW7 2BU, United Kingdom. 2. West China School of Basic Medical Sciences and Forensic Medicine, Sichuan University, Chengdu 610041, China.
Abstract
Personalized and point-of-care (POC) diagnoses are critical for ocular physiology and disease diagnosis. Real-time monitoring and continuous sampling abilities of tear fluid and user-friendliness have become the key characteristics for the applied ophthalmic techniques. Fluorescence technologies, as one of the most popular methods that can fulfill the requirements of clinical ophthalmic applications for optical sensing, have been raised and applied for tear sensing and diagnostic platforms in recent decades. Wearable sensors in this case have been increasingly developed for ocular diagnosis. Contact lenses, as one of the commercialized and popular tools for ocular dysfunction, have been developed as a platform for fluorescence sensing in tears diagnostics and real-time monitoring. Numbers of biochemical analytes have been examined through developed fluorescent contact lens sensors, including pH values, electrolytes, glucose, and enzymes. These sensors have been proven for monitoring ocular conditions, enhancing and detecting medical treatments, and tracking efficiency of related ophthalmic surgeries at POC settings. This review summarizes the applied ophthalmic fluorescence sensing technologies in tears for ocular diagnosis and monitoring. In addition, the cooperation of fabricated fluorescent sensor with mobile phone readout devices for diagnosing ocular diseases with specific biomarkers continuously is also discussed. Further perspectives for the developments and applications of fluorescent ocular sensing and diagnosing technologies are also provided.
Personalized and point-of-care (POC) diagnoses are critical for ocular physiology and disease diagnosis. Real-time monitoring and continuous sampling abilities of tear fluid and user-friendliness have become the key characteristics for the applied ophthalmic techniques. Fluorescence technologies, as one of the most popular methods that can fulfill the requirements of clinical ophthalmic applications for optical sensing, have been raised and applied for tear sensing and diagnostic platforms in recent decades. Wearable sensors in this case have been increasingly developed for ocular diagnosis. Contact lenses, as one of the commercialized and popular tools for ocular dysfunction, have been developed as a platform for fluorescence sensing in tears diagnostics and real-time monitoring. Numbers of biochemical analytes have been examined through developed fluorescent contact lens sensors, including pH values, electrolytes, glucose, and enzymes. These sensors have been proven for monitoring ocular conditions, enhancing and detecting medical treatments, and tracking efficiency of related ophthalmic surgeries at POC settings. This review summarizes the applied ophthalmic fluorescence sensing technologies in tears for ocular diagnosis and monitoring. In addition, the cooperation of fabricated fluorescent sensor with mobile phone readout devices for diagnosing ocular diseases with specific biomarkers continuously is also discussed. Further perspectives for the developments and applications of fluorescent ocular sensing and diagnosing technologies are also provided.
As the second most complicated
and important part of the human body, the human eye is responsible
for receiving, transforming, transmitting, and assimilating informative
messages.[1] Ocular defects have affected
more than 2.2 billion people around the world with different severities
of visual dysfunction based on the analytical results from the World
Health Organization (WHO).[2−4] Thirty-six million people from
this investigation are blind and over 50; moreover, the majority of
these patients with ocular dysfunctions are from developing countries.[3] Visual impairment is usually caused by different
common ocular diseases, including refractive errors, corneal opacification,
age-related macular degeneration (AMD), trachoma, glaucoma, cataracts,
diabetic retinopathy, and some undetermined diseases (Figure ). Blindness and visual impairment
can be the worst influence of these diseases and are accompanied by
other functional impairments, such as intellectual disability, cerebral
palsy, and epilepsy.[4,5] Moreover, the total expenditure
of ophthalmic health care costs has reached around $3 trillion,[5] and the prescription drug expenditure was reported
to be over $1 billion during 2013.[6] In
addition, the largest direct medical cost is the hospitalization medical
services for diagnosing and treating vision impairment and blindness
at the primary phase.[7] In recent decades,
ocular therapeutic treatments and diagnosis have been highly developed
and applied for personalized treatment, and some therapeutic technologies
have been applied for real-time ophthalmic monitoring. Ophthalmologic
technologies have been improved from both clinical trials and fundamental
laboratory-based research. Current clinical ophthalmic spectroscopic
applications contain both routine assessment photographic technologies
and ocular segment imaging technologies. Routine detection of ocular
disease diagnosis consists of retinal scopes,[8] external eye photography in cooperation with a hand-held smartphone,[9] and gonioscopy technologies.[10,11] Further developments for spectroscopic assessment of ocular diseases
have been established during the recent two decades and utilized for
ophthalmologists clinically for real-time and dynamic monitoring,
such as optical coherence tomography (OCT),[12−19] confocal scanning laser ophthalmoscope (cSLO),[20−22] scanning laser
polarimeter (GDx),[23] and fundus autofluorescence
imaging.[24] Several ophthalmic technologies
have also been applied with the hand-held spectroscopic devices at
point-of-care (POC) settings in recent years, such as digital cameras
and smartphone devices, and are treated as one of the emerging clinical
applications in ophthalmology for ocular imaging diagnosis and monitoring.[9]
Figure 1
Summary of tear capacities for detection of various biomarkers
in ocular diseases (top) and ocular complication disease fluorescent
sensing (bottom). Biomarkers in purple are potential indicators for
specific ocular diseases and ocular-related diseases, such as cancer,
Parkinson’s disease, sclerosis disease, and diabetic retinopathy.[30,62−77] Fluorescent sensing technologies can be applied for tear biomarker
sensing aiming to indicate and monitor related ocular disease to achieve
personal and POC diagnosis.
Summary of tear capacities for detection of various biomarkers
in ocular diseases (top) and ocular complication disease fluorescent
sensing (bottom). Biomarkers in purple are potential indicators for
specific ocular diseases and ocular-related diseases, such as cancer,
Parkinson’s disease, sclerosis disease, and diabetic retinopathy.[30,62−77] Fluorescent sensing technologies can be applied for tear biomarker
sensing aiming to indicate and monitor related ocular disease to achieve
personal and POC diagnosis.Sensing of tears for diagnosing and monitoring ophthalmic diseases
has been emerging in the past decade.[25,26] There are
three layers within tear fluid, including the outer lipid layer, the
aqueous layer, and the mucous layer (Figure ).[27] As tears
obtain similar but simple compositions compared to blood, their potential
for diagnosing various diseases has been evaluated. Mitochondrial
energy metabolism and some other specific metabolic processes occur
during plasma leakage and lead to transference of the components from
the blood through the barrier to tears.[27,28] Hence, tear
fluid has a wide potential detection range for developing innovative
diagnostic platforms of ocular diseases and other developmental ocular
dysfunctions, such as cancer, neurological disorders, and diabetes.[29−34] A large number of analytes within the tear fluid have been examined
and treated as potential biomarkers for diagnosing and monitoring
various ophthalmic diseases, including ocular diseases and ocular
complication diseases (Figure ).[35−46] Therefore, the cooperation of specific technologies for target biomarkers
in tear fluid provides more possibilities to identify the pathophysiology
of ocular diseases. Instead of direct biomarker monitoring and detection
within tears, other evaluation criteria have also been studied for
ocular disease diagnosis such as moisture content and intraocular
pressure.[47,48] These developed sensors for physical signals
commonly collaborate with the contact lens-based detection and aim
especially for dry eye diseases and glaucoma examinations. There have
been various biosensors fabricated for tear analyte detection in ophthalmological
diagnosis on either paper-based or contact lens-based platforms during
the past decade, including colorimetric,[25,49−51] fluorescent,[52−55] electrochemical,[56−60] and photonic crystal sensing.[61]Fluorescence sensing of biomarkers in tears is remarkable,
especially
for glucose,[78] pH, and electrolytes.[52] Performance of different fluorescent tear sensors
(especially contact lens type of sensor) varies depending on different
status, the range of detection for biomarkers, and the response time
of the sensor, for instance. Fluorescence sensors have always been
treated for various scientific applications, such as fluorescent labeling,
biological detections, mineralogy, gemmology, and cosmic-ray detection.[79] Based on the advantages of high sensitivity,
specificity, easy operation, and low cost, fluorescence sensors have
emerged in recent decades. Moreover, according to the WHO health report
of the global population, 75% of them are claimed as subhealth and
over 68.5% of these people are the health care workers during the
COVID-19 pandemic.[80] The demand for POC
diagnostic platforms continues to increase. It is crucial to detect
and monitor the biological and chemical molecules in tears under minimal
concentrations within the physiological conditions through rapid and
accurate methodologies in the modern world.[81] Fluorescence biosensors, therefore, are among the target examples
that can rapidly detect analytes. Reversibility of the fluorescence
biosensors is also an important criterion for POC settings, as the
patients require multisensing and real-time monitoring for ocular
disease diagnosis. The evolution of typical fluorescence sensing technologies
can access the opportunities in real-time monitoring, diagnosing specific
ocular diseases, and understanding the physiological conditions within
the eye systems.[29] Typical fluorescence
sensing techniques, such as Forster or fluorescence resonance energy
transfer (FRET), can also apply one detection system for multianalyte
monitoring in tears. In this circumstance, the ocular disease that
is evaluated by different biomarkers and the accuracy of diagnosis
can be improved for a future clinical study.This review aims
to summarize the applied clinical and experimental
fluorescence sensing technologies for ophthalmic diagnosis and real-time
monitoring in recent decades. The importance of fluorescence sensing
for tear diagnosis will be discussed. The development of portable
smart readout devices for sensing at POC settings will also be included.
Moreover, the expected aspect for fluorescence tear sensing is overwhelmed
by the possibilities for drug delivery.
Fluorescence Sensing Technologies
In the fluorescence mechanism, normally the excitation of one molecule
from the ground state (S0) to a singlet state (S2) would occur by absorption (eq ); then the relaxation occurs by emitting photon energy to
a lower energy state (S1) (eq ). The absorption can be treated quantitatively using
the Beer–Lambert Law. The ending state (S1) does
not have to be the ground state, and the remaining energy within the
molecule may be emitted through further fluorescence processes or
dissipated by nonradiative relaxation energy such as heat. Therefore,
the fluorescence process is rotative; the same fluorophore can be
excited and detected repeatedly as long as the fluorophore is not
destroyed at the excited state. Various principles have been applied
to define a fluorescence process, including quantum yield, lifetime,
quenching, photobleaching, and energy transfer.where S0 indicates the ground state;
S1 and S2 indicate singlet states; h indicates the Planck’s constant; νex and
νem indicate the frequency of the photon.
Characteristics
for Fluorescence
The fluorescence process
usually can be visualized with a Jablonski diagram (Figure a) which illustrates molecular
electronic states and demonstrates the transitions between them via
a diagram. The states are arranged in two directions, the energy levels
are measured vertically, and the spin multiplicity is grouped horizontally.
Squiggly arrows indicate nonradiative transitions, and radiative transitions
are noted by straight arrows (Figure b). The absorption spectrum and the emission image
are mirror images for some fluorescent molecules and can be explained
by the Frank–Condon principle.[82] As a consequence, the vibrational levels are similar between the
excited state and the ground state, and the nucleus does not move.
Moreover, nonradiative transitions exhibit various mechanisms and
with different labels in the diagram. The vibrational relaxation presents
the relaxation from the molecule’s excited state to its lowest
vibrational level. The isolated molecules do not exhibit this process,
as the energy from the molecule to its surroundings would be dissipated.
Furthermore, internal conversion (IC) and intersystem crossing (ISC)
are also two types of nonradiative transition. An examination of the
Jablonski diagram indicates that emission energy is universally less
than absorption energy.[82] As a consequence,
lower energy level or longer wavelengths can motivate fluorescence
reactions. Stokes shifts usually indicate the energy difference from
the absorbed fluorescent molecule to the emitted fluorescent molecule.
The rapid decay of vibration and the heat caused by extra vibrational
energy are the initial causes for Stokes shifts (Figure ).[83] Additionally, excited state reactions, solvent effects, energy transfer,
and complexation reactions can lead to further Stokes shifts.
Figure 2
Schematic explanation
for fluorescence and current fluorescent
sensing methods. (a) The Jablonski diagram demonstrates one molecule
(A) excited from the ground state (1A) followed by two
procedures: including direct fluorescence and intersystem crossing
to its triplet state (3A) phosphorescence to the ground
state after that. The relative figures for different sections of the
Jablonski diagram.[84−86] Fluorescence image. Reproduced with permission from
ref (84). Copyright
2014, Public Library of Science. Intersystem crossing image. Reproduced
from ref (85). Copyright
2018, American Chemical Society. Phosphorescence image. Reproduced
with permission from ref (86). Copyright 2020, Nature Portfolio. (b) Energy transfer
for Jablonski diagram demonstration of fluorescence reaction. The
system excitation reaction occurs electronically and vibrationally;
then the high-energy photon is absorbed by an electron. The system
relaxes by vibrational reactions, and the fluorescence at a longer
wavelength is triggered eventually. (c) Relative common approaches
for turn-on or radiometric fluorescence sensing detection mechanisms:
(i) bond-cleavage reaction; (ii) organic addition or metal–ligand
substitution reaction; and (iii) tandem reaction cascades for fluorogenic
scaffolds.[87]
Schematic explanation
for fluorescence and current fluorescent
sensing methods. (a) The Jablonski diagram demonstrates one molecule
(A) excited from the ground state (1A) followed by two
procedures: including direct fluorescence and intersystem crossing
to its triplet state (3A) phosphorescence to the ground
state after that. The relative figures for different sections of the
Jablonski diagram.[84−86] Fluorescence image. Reproduced with permission from
ref (84). Copyright
2014, Public Library of Science. Intersystem crossing image. Reproduced
from ref (85). Copyright
2018, American Chemical Society. Phosphorescence image. Reproduced
with permission from ref (86). Copyright 2020, Nature Portfolio. (b) Energy transfer
for Jablonski diagram demonstration of fluorescence reaction. The
system excitation reaction occurs electronically and vibrationally;
then the high-energy photon is absorbed by an electron. The system
relaxes by vibrational reactions, and the fluorescence at a longer
wavelength is triggered eventually. (c) Relative common approaches
for turn-on or radiometric fluorescence sensing detection mechanisms:
(i) bond-cleavage reaction; (ii) organic addition or metal–ligand
substitution reaction; and (iii) tandem reaction cascades for fluorogenic
scaffolds.[87]The efficient rate of one fluorescence process is called the fluorescence
quantum yield (QY) and is usually equal to the amount of emission
photons divided by the amount of absorption photons.[79] An alternative explanation of QY is by the decay rate of
the excited state (eq ), and the nonradiative rates that are caused by mechanisms are included
where kisc indicates ISC, kic indicates IC, kpd and kd indicates predissociation and dissociation
respectively, and kec is the external
coversion. Therefore, the fluorescence QY can be affected by the variation
of the rate in any pathway. However, QY is independent or has less
dependency on the wavelength of exciting radiation according to the
Kasha–Vavilov rule, because the fluorescence emission usually
takes place after the decay of excited molecule transfer to the lowest
vibrational level.[82,88] The excited state lifetime may
also be influenced by the rate changes, and the fluorescence can be
explained by a lifetime using first-order kinetics (eq ). The fluorescence lifetime is
critical for some applied fluorescent technologies, such as Förster
resonance energy transfer (FRET) and fluorescence-lifetime imaging
microscopy. A fluorophore’s fluorescence lifetime (τ)
usually indicates an exponential decay of the radiative (kf) and nonradiative (knr)
process during the depopulation of excited state molecules (eq ). Subsequently, with an
infinitesimally short excitation process, the fluorescence intensity
decay (It) with time can be expressed
(eq ). The fluorochrome
may experience different reactions, for instance, diffusion of molecules,
reaction process of molecule through conformational changes, or molecular
interactions with surrounding molecules during the lifetime. Consequently,
it provides chances for lifetime measurements to probe these actions.
Fluorescent lifetime is an important parameter for many tear sensing
studies.[89,90]where kf indicates
the rate constant of radiation with spontaneous emission; ∑k indicates all rates of excited state decay in total; [S1] indicates excited state molecules’ concentrations
at time t; [S1]0 indicates
the original concentration and Γ indicates the decay rate; τ
indicates the lifetime; knr indicates
the rate constant of nonradiative decay process; I0 indicates the initial intensity, I indicates the fluorescence intensity at
time t, and t indicates time.Additionally, the fluorescence polarization can be measured through
the orientation of the transition moment of fluorochromes at rapid
emission, as the measurement of anisotropy can determine the rotation
of fluorochromes. Fluorescence quenching, a phenomenon in which a
molecule (quencher) interacts with the fluorophore, leads to a reduction
in quantum yield or lifetime. Furthermore, autofluorescence is also
one of the species in fluorescent sensing technology, and it occurs
from cellular components with fluorescence properties instead of being
developed from the fluorochrome of interest. The most typical example
would be flavins and extracellular matrix components such as elastin,
lipofuscin, and collagen.[91,92] Photobleaching is an
important photochemical process that can be applied in several practical
fluorescence sensing technologies. The chemical reactivity of fluorochrome
is high, and the fluorochrome lives longer than singlet states under
its dark triplet excited state; hence, the photochemical process occurs
predominantly.Fluorescence can be characterized by different
parameters and applied
in various scientific areas. Fluorophores for fluorescent detection
are important elements, as they obtain high specificity and satisfy
basic principles. Fluorophores can be applied as diverse characters,
which can be utilized alone as a substrate of enzymes, probes, or
indicators. In the meantime, they can be bonded covalently to a macromolecule
as a marker for bioactive reagents.[93−95] Hence, various fluorescence
technologies and their practical applications on the platform of tear
fluids can be utilized for future personalized treatment and applied
at POC settings.
Fluorescence Sensors for Ophthalmic Diagnosis
pH and
Electrolyte Sensing
Electrolytes and pH levels
in tears have always been crucial and popular for tear sensing and
diagnosis. The composition of the tear film is based on the ion and
water transported from the ocular surface epithelia and secreted fluid
from lacrimal glands. Hence, major electrolytes within tears (pH,
Na+, K+, Ca2+, Mg2+, Cl–, and Zn2+) are the most popular analytes
for real-time fluorescence monitoring in ophthalmic platforms. Moreover,
variation of these electrolytes would be related with a series of
ocular diseases, such as dry eye disease (DED), ocular infections
caused by parasites, and thyroid eye disease.[96−98] Besides the
quantification of different ions for various ocular disease demonstrations,
DED is one of the most common examples.[52,53] Existing clinical
diagnostic approaches for DED can be varied by symptom identification,[99] ocular surface examination with aid of slit
lamp,[100] quantification tests (e.g., Schrimer’s
test),[101−103] physical tear fluid analysis (e.g., TearLab),[104] and lateral flow assay detections (e.g., InflammaDry
and Tearscan).[105,106] However, most of the detection
methods lack specificity, commercial availability, and understanding
of pathophysiology of DED. The potential of tear fluid for DED diagnosis
and differentiation should be explored. The development of fluorescent
biosensing technologies would enable high selectivity for continuous
detection of typical analytes and major ions within tears as well
as achieving a relatively rapid and cheap detection. One fluorescence
detection method was developed to analyze the concentrations of electrolytes
in tear fluid from anesthetized mice.[107] The dual-wavelength fluorescent indicators were selected accompanying
the application of the ratio imaging fluorescence microscope. Both
red- and green-colored fluorescence images were obtained from the
self-designed dual fluorescent membrane-impermanent indicators for
Na+ (Figure a(i)), K+ (Figure a(ii)), and Cl– (Figure a(iii)), and a bis(carboxyethyl)-carboxyfluorescein
fluorescence-conjugated dextran was utilized for pH detection (Figure a(iv)).[107] The remarkable universal concentrations of
electrolytes within the wild-type mice were reported, and the Na+ ion level was claimed to be significantly higher in AQP5
null mice and declined after the ocular surface was exposed to a humid
environment.[107] The in vivo fluorescence analytic method for tear analyte determination was
conducted using the ratio fluorescence microscopic technology. The
overall experiment provided the idea for specifically synthesized
fluorophores applying for tear sensing and indicated the biocompatibility
for in vivo measurement. However, the experimental
element was anesthetized during the staining process, noninvasive
detection could be considered and developed for commercial applications.
Figure 3
Chemical
structures of the sensing probes for pH, sodium, potassium,
and chloride ions and experimental results. (a) Chemical structure
demonstration of fluorescent probes (i) Na+; (ii) K+; (iii) Cl+; and (iv) pH sensing probe. (b) Chemical
structure of sodium ion detection within silicone hydrogel contact
lens (SiHG). (c) pH sensing probe that was used for tear diagnosis
(left) and the resulting fluorescent contact lens sensors (right);
the blue color fluorescent contact lens was at pH = 4.0, and the yellow
fluorescence image was obtained under pH = 10.0. Reproduced with permission
from ref (89). Copyright
2017, Elsevier. (d) Reversibility test of the detection of Na+ and Cl– ions. Reproduced with permission
from ref (90). Copyright
2020, Elsevier.
Chemical
structures of the sensing probes for pH, sodium, potassium,
and chloride ions and experimental results. (a) Chemical structure
demonstration of fluorescent probes (i) Na+; (ii) K+; (iii) Cl+; and (iv) pH sensing probe. (b) Chemical
structure of sodium ion detection within silicone hydrogel contact
lens (SiHG). (c) pH sensing probe that was used for tear diagnosis
(left) and the resulting fluorescent contact lens sensors (right);
the blue color fluorescent contact lens was at pH = 4.0, and the yellow
fluorescence image was obtained under pH = 10.0. Reproduced with permission
from ref (89). Copyright
2017, Elsevier. (d) Reversibility test of the detection of Na+ and Cl– ions. Reproduced with permission
from ref (90). Copyright
2020, Elsevier.Instead of fluorescence sensing
with the aid of injection through
ocular surfaces, other types of devices have been studied and investigated
during the recent decade. Wearable sensing technologies are therefore
emerging and developed during recent years. One silicone hydrogel
(SiHG) contact lens sensor was fabricated and examined to distinguish
the concentration levels of pH, Na+, and Cl– ion for ocular disease diagnosis.[89] During
the fabrication and evaluation procedure at the beginning stage of
the SiHG sensor, the stabilities and other properties were examined
for three sensing probes. As for the pH sensing probe, three chemicals
including polarity-sensitive probes, 1-anilinonaphthalene-8-sulfonic
acid (1,8-ANS, ANS), and a 4-(1-octylamine)-7-nitrobenzoxadiazole
(NBD-C18) were used. Both hydrophilic and hydrophobic pH sensing probes
were detected and compared to exhibit the sensing property of the
fabricated probe. The chloride ion sensing was fabricated from 6-A
methoxyquinolinium-containing (SPQ) probe.[89] As for the mechanisms of pH and chloride ion sensing, hydrophobic
and hydrophilic fluorescence interactions were detected and evaluated
through the C18 and C3 alky chains. Among different types of contact
lenses, SiHG contact lenses were selected for their high Dk value
(permeability of the material) and silicone content (52%).[89] The hydrophobic fluorophores for ion sensing
were hardly removed from aqueous solutions due to a strong binding
between SiHG lenses and the selected fluorophore. With a similar fabrication
process, another tear sensing experiment was employed for sodium and
chloride ions detection by SiHG lenses. The binding site of the Cl–-sensitive fluorophore between the octadecyl side chain
and SiHG lenses was hydrophobic.[90] As for
the sodium-sensitive fluorophore, the sodium green (SG) and poly(l-lysine) (PL) were combined through covalent conjugation (Figure b). The fabricated
contact lens sensor achieved a wider concentration range of detection
for Na+ and Cl– ions, and the response
of each fluorescent was independent.Moreover, a pH-independent
SiHG was also detected with the selected
labeled fluorophore and showed a significant fluorescence change under
pH = 4 (blue) and pH = 10 (yellow) (Figure d). The reversibility of the fluorescence
sensor was examined for Na+ and Cl– ion
detection as well (Figure d).[90] By testing the fabricated
fluorescent probe, 5 cycles were obtained and the contact lenses were
rinsed various times within 3 mL of buffered solution in each cycle.
The fluorescent lifetime was also detected between 0 and 340 mmol
L–1 of NaCl and claimed no effect on different concentration
of NaCl, where 0 mmol L–1 of NaCl indicated 1.5
ns and 340 mmol L–1 indicated 2.8 ns at fluorescence-reversible
lifetime detection. The interference of tear proteins for the reversibility
of the probe was also evaluated and reported no effect even after
rinsing the lenses for 2 h within the tear proteins buffered solutions.
The interfacial region of SiHG lenses were well-established and explored
through these studies, and the developed silicone hydrogel fluorescent
contact lens sensor can be established and integrated with other tear
analytes for typical ocular disease diagnosis and monitoring of ophthalmic
physiologies. Furthermore, the approved detection of reversibility
of fluorophores for continuous tear monitoring enhanced the advantages
of fluorescence tear monitoring by comparing with other tear diagnostic
methods, such as disposable colorimetric test strips.Quantitative
analysis and real-time monitoring of analytes in tear
fluid are essential for diagnosing eye diseases at the early stage
for a POC platform, especially for tear ion detection. A type of paper-based
fluorescence sensor for tear diagnosis was investigated with a multidetection
mode strip for pH, Na+, K+, and Ca2+ ions using chelation reactions (Figure a(i,ii)).[53] The
fluorescence sensor was fabricated incorporating an optical readout
device and a smartphone diagnostic system (Figure a(iii,iv)). As a continuous study for this
type of sensor, one fluorescent scleral contact lens sensor was then
fabricated to detect pH, Na+, K+, Ca2+, Mg2+, and Zn2+ ions within the tear physiological
range. A smartphone readout was accompanied by this sensor for analyzing
specific ion-sensitive data quantitatively.[52] The constructed sensor is advantageous in the ability to recognize
the severity of DEDs and their different types (Table ). Chelation reactions were utilized for
different fluorophore detection of individual concentrations (Figure b(i–iii)).
The pH probe, for instance, was examined and evaluated in the concavity
of the contact lens. Moreover, the Na+ sensing probe (15.6
mmol L–1) was selected by the crown ether derivatives
within 0–100 mmol L–1 of the detection range,
and the K+ sensing probe could be examined from 0 to 50
mmol L–1 with a LOD at 8.1 mmol L–1.[52] Acid-based probes were utilized for
Ca2+ and Zn2+ monitoring ranging from 0.50 to
1.25 mmol L–1 and from 0.5 to 0.8 mmol L–1 respectively. In addition, a sensitivity of 1 μmol L–1 was claimed for the Zn2+ ion sensor and the range of
detection was 10–20 μmol L–1.[52] Moreover, the microfluidic channel of the contact
lens was examined for the final fabricated contact lens sensor (Figure c).
Figure 4
Multifluorescence sensing
devices and fluorescence chelation reaction.
(a) Paper-based microfluidic device for multianalyte sensing in tears:
(i) fabricated paper-based sensing probes for pH, Na+,
K+, and Ca2+ ion detection. (scale bar: 2 mm);
(ii) finalized sensing device for tear diagnosis (scale bar: 1 cm);
(iii) guidance of the smartphone readout device; scale bar = 1 cm;
(iv) photograph for reading the paper-based sensing device within
the black box before excitation of the fluorophores (scale bar: 4
mm). Reproduced with permission from ref (53). Copyright 2017, The Royal Society of Chemistry.
(b) Chelation fluorescent reaction mechanisms for (i) Ca2+, (ii) Zn2+, and (iii) Mg2+. (c) Fabrication
of microfluidic contact lens sensor: (i) closed microfluidic channel
for 2 μL of fluid flowing in the microchannel; (ii) fabricated
contact lens sensor for multianalytes detection. (d) Photograph of
readout system: (i) readout box for multidetection of fluorescence
sensors, consisting of three excitation switches of the filter and
three readout emission filters with different wavelengths of light;
(ii) smartphone readout demonstration for detection the ions within
contact lens sensor. Reproduced with permission from ref (52). Copyright 2020, Wiley-VCH.
Table 1
Ion Detection within Tear for the
Diagnosis of Dry Eye Disease and Its Subtypes[52,53]
Abnormal
Level within Tears (mmol L–1)
Dry
Eye Disease Type
Electrolytes
Detection Range within Tears (mmol L–1)
MGD
LGD
MGD and LGD
Sensitivity (mmol L–1)
pH
∼7.4
∼7.9
0.12
Na+
120–165
133.2–136.1
133.2–142.2
133.2–145.1
15.6
K+
20–42 (ave. 24)
24.6
24.9
25.4
0.8
Ca2+
0.4–1.1 (ave. 0.8)
0.82
0.84
0.86
0.02–0.05
Mg2+
0.5–0.9 (ave. 0.61)
0.61
0.63
0.65
0.01–0.03
Multifluorescence sensing
devices and fluorescence chelation reaction.
(a) Paper-based microfluidic device for multianalyte sensing in tears:
(i) fabricated paper-based sensing probes for pH, Na+,
K+, and Ca2+ ion detection. (scale bar: 2 mm);
(ii) finalized sensing device for tear diagnosis (scale bar: 1 cm);
(iii) guidance of the smartphone readout device; scale bar = 1 cm;
(iv) photograph for reading the paper-based sensing device within
the black box before excitation of the fluorophores (scale bar: 4
mm). Reproduced with permission from ref (53). Copyright 2017, The Royal Society of Chemistry.
(b) Chelation fluorescent reaction mechanisms for (i) Ca2+, (ii) Zn2+, and (iii) Mg2+. (c) Fabrication
of microfluidic contact lens sensor: (i) closed microfluidic channel
for 2 μL of fluid flowing in the microchannel; (ii) fabricated
contact lens sensor for multianalytes detection. (d) Photograph of
readout system: (i) readout box for multidetection of fluorescence
sensors, consisting of three excitation switches of the filter and
three readout emission filters with different wavelengths of light;
(ii) smartphone readout demonstration for detection the ions within
contact lens sensor. Reproduced with permission from ref (52). Copyright 2020, Wiley-VCH.The development of a portable readout device for
data collection
and process is another advantage of this series of research. This
device was constructed with light-emitting diodes (LEDs) and bandpass
optical filters for the sensor excitation. Data collection was another
innovative point for the research (Figure d(i)). One smartphone camera was applied
to assist with the fabricated readout device to deliver measurements
quantitatively (Figure d(ii)). Then, the finalized fluorescent biosensor based on scleral
lenses was explored for diagnosing and detecting the severity stage
and distinguishing the subtypes of DEDs, meibomian gland dysfunction
(MGD), and lacrimal gland dysfunction (LGD), for instance. The conversion
of fluorescence into readable output data using a smartphone is advantageous
in a personalized POC platform by providing the possibility for patients
to collect data corresponding with their ocular conditions at any
moment necessitated. Future development of the build-up application
of smartphones can also be enriched with different sections of a variety
of ocular diseases, and it would be user-friendly for real-time POC
monitoring. Ascorbic acid, as an important antioxidant biomarker for
ocular inflammations, has also been studied for fluorescence detections.[108,109] However, the detection of ascorbic acid at POC platforms was conducted
in aqueous humor instead of tear fluid with a high accuracy over than
80%.[108] The sensor was mainly developed
for diagnostics of ocular globe injuries and glaucoma care. Further
fluorescence detection of ascorbic acid should be developed for tear
fluid examination.
Glucose Sensing
Glucose has been
considered one of
the critical biomarkers for diabetes diagnosis. There have been extensive
studies conducted for glucose monitoring within various biological
fluids, including blood,[110,111] interstitial fluid,[112,113] urine,[110,114,115] sweat,[116−118] saliva,[114,119] and tear
fluid in the recent decade.[120−124] Tear glucose monitoring has been recently established for diagnosing
diabetes and diabetic retinopathy. A normal level of tear glucose
at 0.16 ± 0.03 mmol L–1 was claimed as compared
to the diabetic patients 0.35 ± 0.04 mmol L–1, and the glucose level within tears is 0–5 mmol L–1.[125,126] Different technologies have been applied
for tear glucose sensing, especially fluorescence sensing technologies.
Fluorescence resonance energy transfer (FRET) is one of the popular
methods for low volume glucose detection. One highly sensitive nanostructured
fluorescent biosensor was fabricated by utilizing FRET to monitor
tear glucose levels (Figure a).[127] In this work, the interaction
between the selected nanoparticles FITC-dextran-silica and tetramethyl
rhodamine isothiocyanate-labeled Concanavalin A (TRITC-Con A) occurred
through binding Con A to dextran molecules. The FRET pairs were then
formed by fluorescein isothiocyanate and tetramethyl rhodamine isothiocyanate
(FITC-TRITC). The fluorescent chip device for tear glucose sensing
was then formatted by the deposition of these fabricated fluorescent
nanomaterials on the poly(dimethylsiloxane) (PDMS) surface. The insertion
of glucose could replace the TRITC-Con A from PDMS. The different
levels of glucose concentration would be detected through the fluorescence
resonance energy transfer (FITC-TRITC) ratio. The morphology of assembled
FITC-dextran-Con A-TRITC mesoporous silica nanoparticles (MSN) on
PDMS was characterized using a scanning electron microscope (SEM)
(Figure b(i)), and
the diameter of the nanoparticles were indicated as 60 ± 5 nm
on average. Further fluorescence images of FRET sensors from laser
confocal scanning microscopy (LCSM) were also shown to demonstrate
the glucose concentration (Figure b(ii,iii)). As a result, the finalized FRET biosensor
could reach a detection range within 0.04–4 mmol L–1, and the data of the sensor could be obtained within 2 min (Figure c). Moreover, the
chip can still be functionalized within 5 days.[127] The obtained fluorescence image was finally converted into
readable data with the aid of MATLAB coding. Different concentrations
of glucose were evaluated, and the image and data of glucose under
0.05 mmol L–1 in Figure d(i) and 0.10 mmol L–1 in Figure d(ii) were the corresponding
results, respectively. The binding performance of the fabricated sensor
was further investigated within the hydrogel contact lens material;
the fluorescence property and biocompatibility were proven. With the
readable developed MATLAB data, it could be applied to smart readout
devices for the future perspective.
Figure 5
Highly sensitive glucose sensor based
on FRET within tears. (a)
Detection mechanism for tear glucose sensing within MSNs. (b) Microphotographs
for the nanoparticles and aqueous glucose sensing: (i) micrograph
(SEM) of assembled FITC-dextran-ConA-FRITX@MSNs; LCSM photograph of
fabricated FRET sensor before (ii) and after (iii) the addition of
glucose (20 μL) within 0.10 mmol L–1 of aqueous
solution. (c) Fluorescent spectrum with various glucose concentrations.
(d) Image of the FRET sensor and its conversion of readable MATLAB
data under the glucose concentration of (i) 0.05 and (ii) 0.10 mmol
L–1, respectively. Reproduced with permission from
ref (127). Copyright
2013, SAGE. (e) Mechanism for tear glucose sensing: (i) illustrated
FRET glucose detection included ConA conjugation quantum dots as the
donor and MG as the acceptor; (ii) schematic mechanism of sensor immobilization
on silicone hydrogel. (f) Images for fabricated nanoparticles and
glucose sensing: (i) micrograph of QDs coated with ZnO nanorods using
SEM; (ii) TEM image of QDs coated with ZnO nanorods; the resulting
fluorescence images for the patterned FRET sensors on silicone hydrogel
under aqueous glucose at 0.04 mmol L–1 (iii), and
0.4 mmol L–1 (iv). Reproduced from ref (128). Copyright 2017, Elsevier.
Highly sensitive glucose sensor based
on FRET within tears. (a)
Detection mechanism for tear glucose sensing within MSNs. (b) Microphotographs
for the nanoparticles and aqueous glucose sensing: (i) micrograph
(SEM) of assembled FITC-dextran-ConA-FRITX@MSNs; LCSM photograph of
fabricated FRET sensor before (ii) and after (iii) the addition of
glucose (20 μL) within 0.10 mmol L–1 of aqueous
solution. (c) Fluorescent spectrum with various glucose concentrations.
(d) Image of the FRET sensor and its conversion of readable MATLAB
data under the glucose concentration of (i) 0.05 and (ii) 0.10 mmol
L–1, respectively. Reproduced with permission from
ref (127). Copyright
2013, SAGE. (e) Mechanism for tear glucose sensing: (i) illustrated
FRET glucose detection included ConA conjugation quantum dots as the
donor and MG as the acceptor; (ii) schematic mechanism of sensor immobilization
on silicone hydrogel. (f) Images for fabricated nanoparticles and
glucose sensing: (i) micrograph of QDs coated with ZnO nanorods using
SEM; (ii) TEM image of QDs coated with ZnO nanorods; the resulting
fluorescence images for the patterned FRET sensors on silicone hydrogel
under aqueous glucose at 0.04 mmol L–1 (iii), and
0.4 mmol L–1 (iv). Reproduced from ref (128). Copyright 2017, Elsevier.Another similar fluorescence tear glucose sensor
has been further
developed by using fluorescent patterned arrays. The illustrated mechanism
(Figure e(i)) introduced
a FRET pair consisting of Con-A-conjugation quantum dots as a donor
and MG as an acceptor, and the quenched fluorescence was restored
by a competitive affinity of glucose over MG. The nanostructured FRET
quenching sensor was immobilized onto ZnO nanorod arrays which were
attached to the silicon hydrogel (Figure e(ii)).[128] Both
the fluorescence camera and the fluorometer could be utilized for
analyzing glucose concentrations based on the uniquely designed procedure.
During this fabrication procedure, the patterned ZnO nanorod arrays
on the hydrogel were treated as a substrate, and ConA was conjugated
onto the hybrid nanorods after conjugating CdSe/ZnS quantum dots (QDs)
with ZnO nanorods. The fluorescence quenching molecule was claimed
to be malachite green modified dextran and was bound onto Con A. After
introducing the glucose, the dextran molecule was replaced by glucose
competitively and the QDs fluorescence recovered. The QDs coated ZnO
nanorod arrays were further analyzed by SEM (Figure f(i)) and transmission electron microscope
(TEM) (Figure f(ii)).
The FRET sensor on SiHG was evaluated with different concentrations
of glucose (Figure f(iii–iv)), and a detection range of glucose was stated within
0.03–3 mmol L–1. The possibility of conducting
a FRET sensor onto an SiHG was claimed. Hence, it would be promising
to develop this series of FRET technologies onto an SiHG biosensor
attached with contact lenses for real-time tear glucose monitor at
POC settings. However, the biocompatibility for applying QDs to in vivo studies is needed to be considered.Instead
of FRET imaging for glucose detection within tears, other
contact lens sensors have been developed. For example, a typical SiHG
contact lens sensor has been examined and fabricated for tear glucose
monitoring continuously in the last five years. In this study, the
monitoring mechanism was simplified, and a glucose-sensitive fluorophore
(Glu-SFs) named Quin-C18 was utilized for examination.[126,129] The interpenetrating polymer network of the SiHG lenses was evaluated
for the characterization of the sensor during fabrication. One polarity-sensitive
probe (Prodan) was applied for the lenses used within water and pure
silicone regions. In order to confine the glucose-sensitive fluorescent
probe within the interfacial area of the contact lens, Quin-C18 was
formed and attached with one hydrophobic chain (Figure a).[129] The applied
glucose sensor would also suitable for in vitro examination
and can be verified with different glucose concentrations (Figure b,c). The fabricated
contact lens sensor was claimed to be consistent for glucose detection
and proved that Quin-C18 can bound strongly to the lenses. Other properties
such as leaching rate was also examined during detection. The leaching
rate was extremely low after several rinsing processes, and a continuous
detection result of glucose sensing was reported as being similar
after a three-month storage of lenses within water. Hence, the developed
glucose-sensitive sensor is advantageous in continuous POC detecting
glucose within tears.
Figure 6
Fluorescence sensing of glucose in tears. (a) Synthetic
mechanism
for Quin-C18, the fluorophore that was used for glucose sensing. (b)
Photographs for characterization of the contact lens glucose sensor:
(i) micrograph for indicating the influence of dwell time over the
increased intensity and phase angle, and the absence of the signal
at the outside surrounding circles claimed the fluorophore could not
be detected; photographs of Quin-C18-doped contact lens (Comfilcon
A) under room light (ii) and UV light (iii) without an emission filter.
(c) Glucose detection for optimization of contact lens sensor: (i)
glucose concentration change versus the emission intensity in the
phosphate-buffered solution (pH = 7.2); (ii) emission spectra for
Quin-C18-coated Comfilcon A contact lens with multiple washing steps;
(iii) finalized glucose-dependent emission results of Quin-18-coated
Comfilcon A SiHG contact lens. Reproduced with permission from ref (129). Copyright 2018, Society
of Photo-Optical Instrumentation Engineers Digital Library. (d) Mechanism
for fluorescent copolymer formation for tear glucose sensing. (e)
Paper-based microfluidic system for artificial tear glucose detection:
(i) illustrated structure of a Schirmer test for tear glucose sensing;
(ii) resulting emission spectrum for the variation of glucose detection
under gel-encapsulated B5 sensing system; the inset chart indicated
the calibration curve of glucose sensing within artificial tears (n = 3); here, the percentage emission change was evaluated
with different glucose concentrations. Reproduced with permission
from ref (130). Copyright
2018, Multidisciplinary Digital Publishing Institute. (f) Glucose-sensitive
fluorescent contact lens sensor: (i) reversible detection mechanism
on contact lens; (ii) photos for artificial eyes with contact lens
and the fluorescent signal change with the variation of glucose; (iii)
fluorescent image captured by smartphone from the rabbit with 1 mmol
L–1 (upper) and 5 mmol L–1 (bottom)
of glucose injection. Reproduced with permission from ref (131). Copyright 2022, Elsevier.
Fluorescence sensing of glucose in tears. (a) Synthetic
mechanism
for Quin-C18, the fluorophore that was used for glucose sensing. (b)
Photographs for characterization of the contact lens glucose sensor:
(i) micrograph for indicating the influence of dwell time over the
increased intensity and phase angle, and the absence of the signal
at the outside surrounding circles claimed the fluorophore could not
be detected; photographs of Quin-C18-doped contact lens (Comfilcon
A) under room light (ii) and UV light (iii) without an emission filter.
(c) Glucose detection for optimization of contact lens sensor: (i)
glucose concentration change versus the emission intensity in the
phosphate-buffered solution (pH = 7.2); (ii) emission spectra for
Quin-C18-coated Comfilcon A contact lens with multiple washing steps;
(iii) finalized glucose-dependent emission results of Quin-18-coated
Comfilcon A SiHG contact lens. Reproduced with permission from ref (129). Copyright 2018, Society
of Photo-Optical Instrumentation Engineers Digital Library. (d) Mechanism
for fluorescent copolymer formation for tear glucose sensing. (e)
Paper-based microfluidic system for artificial tear glucose detection:
(i) illustrated structure of a Schirmer test for tear glucose sensing;
(ii) resulting emission spectrum for the variation of glucose detection
under gel-encapsulated B5 sensing system; the inset chart indicated
the calibration curve of glucose sensing within artificial tears (n = 3); here, the percentage emission change was evaluated
with different glucose concentrations. Reproduced with permission
from ref (130). Copyright
2018, Multidisciplinary Digital Publishing Institute. (f) Glucose-sensitive
fluorescent contact lens sensor: (i) reversible detection mechanism
on contact lens; (ii) photos for artificial eyes with contact lens
and the fluorescent signal change with the variation of glucose; (iii)
fluorescent image captured by smartphone from the rabbit with 1 mmol
L–1 (upper) and 5 mmol L–1 (bottom)
of glucose injection. Reproduced with permission from ref (131). Copyright 2022, Elsevier.Therefore, the fabricated fluorescent contact lens
sensor enables
people to obtain the ophthalmic conditions and properties, and to
explore the typical ophthalmological pathology through the examined
results of specific tear biomarkers. Corresponding with the data-processing
portable smart readout devices, it would be a solid benefit for clinical
applications, personal diagnosis, and treatments for ophthalmic diseases.
Another tear glucose fluorescent microfluidic paper-based analytical
device was also developed based on lateral flow assay detection (Figure d,e), the fluorescent
copolymer formation mechanism was developed and applied for the tear
glucose sensing process. As a result, a detection range for the level
of glucose in tears was achieved within the concentration between
0.1 and 4.0 mmol L–1.[130] This sensor provided a faster response time and a wider range of
glucose sensing. Meanwhile, it was claimed for practical diagnosis
for diabetes. The development of the lateral flow assay detection
provided an alternative measurement for glucose and diagnostic methodology
to monitor the diabetic patients at POC platform.One recent
wearable glucose-sensitive fluorescent contact lens
sensor was developed by integrating and immobilizing one glucose fluorescent
probe and a reference fluorescent reference probe for calibration
within the hydrogel network to achieve highly sensitive glucose detection
on contact lens (Figure f).[131] The fabricated glucose fluorescent
contact lens sensor was able to recognize a concentration variation
from 23 μmol L–1 to 1.0 mmol L–1 with a fluorescent color change from pink to blue (Figure f(ii).[131] One smartphone RGB signal region was developed for collecting
and transferring the fluorescent color of this experiment. Additionally,
further in vivo rabbit experiments were conducted
to indicate the biocompatibility of the fabricated contact lens sensor
(Figure f(iii)). It
is advantageous in providing a rapid and noninvasive method for real-time
tear glucose examination, and the constructed glucose-sensitive contact
lens sensor was able to detect as low as 9.3 μmol L–1 through the fluorescence spectrophotometer.[131] The idea for implementing and stabilizing specific fluorescent
probe for a target biomarker within the hydrogel network and attach
it with a contact lens broadens the possibility for the detection
of other tear analytes, especially for proteins within tear fluid.
Most of the developed glucose fluorescent sensors can cover the detection
range within the tear fluid and the achieved sensitivity of the sensor
enable a differentiation ability between nondiabetic and diabetic
individuals. However, few comparison experiments were conducted for in vivo identities before and after taking food. Because
the tear glucose level is significantly lower than the glucose in
blood (3.9–30 mmol L–1),[125] it would be harder to achieve the variation examination
within tears. Further developments could be made to reach a lower
sensitivity of the fabricated fluorescent tear glucose sensor. Moreover,
the clinical data collection and analysis for tears between healthy
and diabetic groups of people would also benefit a future evaluation
of the fluorescent tear glucose sensor.
Proteins Sensing
Other than the integration of the
fluorescent biosensor onto the contact lens, the microfluidic paper-based
analytical device (μPAD) is also emerging for tear diagnosis.
LF is one of the most important and abundant ion-binding proteins
within the human body fluid, especially in tears.[132] It is responsible for antibacterial and anti-inflammatory
activities. Abnormal increase of LF levels within tears can lead to
serious ocular diseases, for example, xerophthalmia and early-stage
inflammatory bowel diseases.[132,133] One fluorescent μPAD
was developed for tear lactoferrin (LF) detection without antibodies
(Figure ).The inkjet
printer and UV-curable ink were utilized for fabricating microfluidic
patterns on μPADs.[55] The fabricated
fluorescent sensor exhibited a limit of detection (LOD) of LF at 0.3
mg mL–1 from 0.5 to 3 mg mL–1 of
LF in the tear level.[134] During the fabrication
of μPAD, filter papers, A4 copy papers, and the EPSON inkjet
printer were used. It was claimed to be important to attach the filter
paper onto a sheet of copy paper, because the inkjet printer used
for this work was not suitable to handle round shapes. The filter
paper faces were then fitted with the round-shaped cut-out of the
copy paper. During the inking process, both octadecyl acrylate and
1,10-decanediol diacrylate were applied for UV-curable ink. After
the ejection of the paper from the printer, it was placed to cool
at 10 °C.[134] The filter paper with
a circular cut area of 81 cm2 was utilized for microfluidic
channel patterns. The designed sensor was then fabricated by a straight
channel including two square areas (Figure a). During the detection, the sensing areas
consisted of TbCl3 solution (1 mmol L–1) mixing with ethylene glycol (15 vol %); then the pattern was soaked
in poly(vinyl alcohol) for 5 min.[134] The
paper was then dried to prevent LF adsorption on the paper surface.
After that, the 25 mmol L–1 of the NaHCO3 solution was pipetted onto the sampling areas. The buffered solutions
(HEPES pH 7.4, 50 mmol L–1) were used during all
procedures. Finally, the single μPADs were cut from the papers
after the soaking process of substrates. As a result, the LF was first
detected within the buffered solution from 0 to 1 mg mL–1 (Figure b) and the
paper-based device was examined with the obtained range of 0.63–2.9
mg mL–1 (Figure c).[134] Instead of conducting
the measurement with a signal readout instrument, distance-based LF
was also developed for similar research (Figure d). Further treatment on the filter paper
used was claimed (Figure e). With the fabricated ι-Cg (ι-carrageenan)-coated
filter paper (Figure e(iii)), an even lower LOD at 0.1 mg mL–1 of LF
was obtained.[135] The LF within tears was
eligible for detection with 0–4 mg mL–1 of
LF with this μPAD (Figure f(i)). Instead of correlation from the fabricated device
(Figure f(ii)), further
evaluation was conducted between the standard ELISA test and μPADs
(Figure f(iii)). After
the success of detection in tears using μPADs, the device can
be explored further by integrating paper-based sensors onto contact
lenses or other portable devices to fulfill the aim at the POC diagnostic
platform for ocular diseases.[26] In addition,
with the integration and investigation of fluorescent paper-based
detection, more protein biomarkers within tear fluid including interleukin
6 (IL-6) and immunoglobin G (IgG) can obtain the potential for fluorescence
detection with relative technologies, such as immunofluorescent assay
and encapsulation of nanocluster fluorescence detection.[136,137]
Figure 7
Developed
fluorescent μPAD for tear LF detection. (a) Schematic
(upper) and photograph (lower) of μPAD for tear LF analysis.
(b) Emission spectrum of LF (0–1 mg mL–1)
within a pH = 7.4 level of HEPES (50 mmol L–1) buffered
solution. (c) Calibration curve of human LF on μPAD; the dashed
line was LOD and the dashed square area represented regular the physiological
detection range of the tear LF (n = 3). Reproduced
with permission from ref (134). Copyright 2014, The Royal Society of Chemistry. (d) Distance
μPAD for human LF detection: (i) illustrated outline structure
of μPAD; (ii) UV illumination (λmax = 254 nm)
photographs of filter paper test for LF mobility within water; 100
mmol L–1 of NaCl aqueous solution; solution with
lysozyme (3 mg mL–1); pseudotear fluid and water
with treated lysozyme (3 mg mL–1), from left to
right, respectively; (iii) UV illumination (λmax =
254 nm) photographs for visualizing Tb3+-LF (1.5 mg mL–1, 0.5 μL) reaction after elution with different
fluids: pure water, pseudotear fluid, LF in water and in the pseudotear
fluid. Photograph on the right was the control results of LF buffered
solution (HEPES 50 mmol L–1, pH = 7.4) without elution.
(e) SEM of utilized filter paper (i) and results of the final fabricated
μPAD; (ii) SEM for ι-Cg coated filter paper; (iii) images
for analyzing different levels of LF (0.1, 0.6, 1, 2, 3, and 4 mg
mL–1 from left to right) onto the ι-Cg coated
μPAD. Scale bar: 10 μm. (f) Graphs for LF evaluations:
(i) calibration curve between LF concentration and the length of emitted
fluorescence line (0.5 mm was related to 1 scale of increment); (ii)
further correlation on observed results for four sets of batches of
fabricated μPAD; (iii) correlation curve between ELISA (n = 4) and μPAD (n = 6). Reproduced
from ref (135). Copyright
2015, American Chemical Society.
Developed
fluorescent μPAD for tear LF detection. (a) Schematic
(upper) and photograph (lower) of μPAD for tear LF analysis.
(b) Emission spectrum of LF (0–1 mg mL–1)
within a pH = 7.4 level of HEPES (50 mmol L–1) buffered
solution. (c) Calibration curve of human LF on μPAD; the dashed
line was LOD and the dashed square area represented regular the physiological
detection range of the tear LF (n = 3). Reproduced
with permission from ref (134). Copyright 2014, The Royal Society of Chemistry. (d) Distance
μPAD for human LF detection: (i) illustrated outline structure
of μPAD; (ii) UV illumination (λmax = 254 nm)
photographs of filter paper test for LF mobility within water; 100
mmol L–1 of NaCl aqueous solution; solution with
lysozyme (3 mg mL–1); pseudotear fluid and water
with treated lysozyme (3 mg mL–1), from left to
right, respectively; (iii) UV illumination (λmax =
254 nm) photographs for visualizing Tb3+-LF (1.5 mg mL–1, 0.5 μL) reaction after elution with different
fluids: pure water, pseudotear fluid, LF in water and in the pseudotear
fluid. Photograph on the right was the control results of LF buffered
solution (HEPES 50 mmol L–1, pH = 7.4) without elution.
(e) SEM of utilized filter paper (i) and results of the final fabricated
μPAD; (ii) SEM for ι-Cg coated filter paper; (iii) images
for analyzing different levels of LF (0.1, 0.6, 1, 2, 3, and 4 mg
mL–1 from left to right) onto the ι-Cg coated
μPAD. Scale bar: 10 μm. (f) Graphs for LF evaluations:
(i) calibration curve between LF concentration and the length of emitted
fluorescence line (0.5 mm was related to 1 scale of increment); (ii)
further correlation on observed results for four sets of batches of
fabricated μPAD; (iii) correlation curve between ELISA (n = 4) and μPAD (n = 6). Reproduced
from ref (135). Copyright
2015, American Chemical Society.
Enzymes Sensing
Instead of proteins detected within
the tear, enzymes have also been studied for fluorescence detection
within tears. One contact lens sensing method was developed for sensing
analytes within tears. The contact lenses were treated as collectors
of samples, and the subsequent analysis was accompanied by one field-portable
and cost-effective reader (Figure a(i)–(iii)).[138] Lysozyme,
as one of the most prevalent and important naturally occurring enzymes
within tears,[139] was therefore selected
to be quantified. Moreover, the time-lapse imaging technology was
utilized with the mobile reader to observe the increase of fluorescence
signal within a standard well-plate. The obtained data was indirectly
inferred to the change of lysozyme concentration through a standard
curve. The best-suited contact lens and the assay were chosen empirically
for tear collection and detection. The variation of lysozyme concentrations
were then monitored within nine healthy human objectives over 2 weeks.
The results were used for a comparison with the objectives with DED.
A time dependency experiment was conducted with a mobile-based microplate
readout that exhibited the data of each ELISA well with three fibers
to get the green channels (Figure b). The fluorescence data increased over 10 min, and
a calibration curve was delivered from the constructed smartphone
reader with the variation of lysozyme concentration. The experimental
data were then compared with the in vivo human objectives
with and without DEDs (Figure c). As a result, the concentration of lysozyme increased from
6.89 ± 2.02 μg mL–1 to 10.72 ± 3.22
μg mL–1 (mean ± SD) was observed for
six participants of nine who wear contact lenses regularly, and these
objectives were detected with the induction of a digital ocular strain
model during the period of contact lens wear. Moreover, a lower mean
lysozyme level of a patient with DED was claimed compared to the healthy
participants, the mean levels of concentrations were 2.43 ± 1.66
μg mL–1 and 6.89 ± 2.02 μg mL–1, respectively. The main advantages of this study
would be introducing a simple and noninvasive sampling method for
detection as well as the measurement system was considered to be relatively
rapid, user-friendly, and cost-effective for indicating the physiological
change within human ocular objectives. Future tear-fluid studies could
be conducted with the application of this methodology, and it would
be significant for tear biomarker multiplex measuring on a POC platform.
Figure 8
Detection
of lysozyme within tears using developed fluorescence
mobile phone-based microplate reader. (a) Detection method and schematic
demonstration of well-plate reader developed by smartphone, (i) the
final step of mixing the wash solution and the Micrococcus
lysodeikticus cell solution within the ELISA well for fluorescent
detection; (ii) a schematic illustration of the smartphone-based well-plate
reader; (iii) photograph of the fabricated product. (b) The recorded
results for fluorescent observation with time dependence, where the
green channel indicated the fiber bundled image that was taken from
the smartphone (three fibers per well) (upper) and the lower graph
indicated the standard curve for fluorescence well-plate assay over
time (10 min). (c) Experimental results for lysozyme detection: (i)
Standard curve for fluorescence well-plate assay: the vertical and
horizontal lines state the LOD (1.99 μg mL–1); (ii) overall resulting data of lysozyme concentration detection
for healthy participants (n = 30) without DED for
over 5-day monitoring period of contact lens wear and one-time measurement
for patients (n = 6) with DED. The red horizontal
line claimed the median concentration of lysozyme in each measurement
group and the top and bottom line of the blue rectangular represented
75% and 25% of the data, respectively. Reproduced with permission
from ref (138). Copyright
2020, The Royal Society of Chemistry.
Detection
of lysozyme within tears using developed fluorescence
mobile phone-based microplate reader. (a) Detection method and schematic
demonstration of well-plate reader developed by smartphone, (i) the
final step of mixing the wash solution and the Micrococcus
lysodeikticus cell solution within the ELISA well for fluorescent
detection; (ii) a schematic illustration of the smartphone-based well-plate
reader; (iii) photograph of the fabricated product. (b) The recorded
results for fluorescent observation with time dependence, where the
green channel indicated the fiber bundled image that was taken from
the smartphone (three fibers per well) (upper) and the lower graph
indicated the standard curve for fluorescence well-plate assay over
time (10 min). (c) Experimental results for lysozyme detection: (i)
Standard curve for fluorescence well-plate assay: the vertical and
horizontal lines state the LOD (1.99 μg mL–1); (ii) overall resulting data of lysozyme concentration detection
for healthy participants (n = 30) without DED for
over 5-day monitoring period of contact lens wear and one-time measurement
for patients (n = 6) with DED. The red horizontal
line claimed the median concentration of lysozyme in each measurement
group and the top and bottom line of the blue rectangular represented
75% and 25% of the data, respectively. Reproduced with permission
from ref (138). Copyright
2020, The Royal Society of Chemistry.Another type of fluorescent detection of lysozyme within tears
was developed through the inner filter effect of gold nanoparticles
on CdTe quantum dots.[140] This sensor was
fabricated by utilizing lysozyme reacting with the lysozyme binding
aptamer to avoid the reaction between gold nanoparticles and the lysozyme
binding aptamer. The gold nanoparticles in this case would aggregate
and a strong blue fluorescence could be observed under this circumstance.
The finalized sensor was examined and evaluated within real tear and
saliva samples, and the detection range was claimed between 1.0 nmol
L–1 and 20 nmol L–1.[140] The fabricated sensor is advantageous in high
sensitivity and label-free detection of lysozyme. With the application
of this method, further integration of this sensor onto paper-based
microfluidic detection could be applied for POC detection. However,
further biocompatibility tests should be qualified if the sensor could
direct contact with human eyes.
Future Prospective
During the past decade, ophthalmology-related fluorescence technologies
have been commonly studied and have been explored for ophthalmic diagnosis
from spectroscopic technologies to portable biosensors. The possibility
of integration of smart readout devices and cooperation with a hand-held
camera for fluorescence tear sensing has also been proposed that would
be helpful for monitoring and diagnosis on POC platforms. Ophthalmic
fluorescent monitoring has been well-established in lab-based studies
(Table ). Nevertheless,
the detected analytes within tears are still limited in glucose and
typical ions.[26,61,141,142] More proteins and enzymes need
to be explored for further ocular diagnosis, such as reactive oxygen
species (ROS), immunoglobulins, and interleukins.[143] Nanoscale carriers in fluorescent detection on targeted
biomarkers in ophthalmologic diagnosis and drug delivery have received
much attention recently, especially for relatively large-sized proteins
and cytokines.[144−146] Relative fluorescent technologies can also
be employed for tear analyte detection, such as quantum dots, carbon
dots, and FRET. As discussed, different means of tear sensing platforms
are also one of the future directions for bringing biosensors to the
commercial industry. For instance, paper-based detection, lateral
flow assays (LFA), capillary tube detection, and three-dimensional
(3D) printing could be established and combined with fluorescence
sensing technologies to achieve personalized treatments on POC platforms.[147] Biocompatibility should be one of the criteria
for evaluation of the sensing platforms developments. Therefore, the
development for biologically modified nanoparticles could also be
one of the directions for ophthalmic therapeutic monitor and examinations,
especially for diagnosing and monitoring cancer-derived ocular diseases.[148,149]
Table 2
Conclusion of the Developed Fluorescent
Sensors for Tear Biomarker Detection
Tear Biomarker
Detection Platform
Sensitivity
Linearity
Animal Test
Response/Reaction
Time
ref
pH
Silicone hydrogel contact lens
-
4.2–10.0
No
-
(89)
Paper-based microfluidic channel
-
7.0–8.0
No
-
(53)
Scleral contact lens
0.12
7.0–8.0
No
-
(52)
Na+
Silicone hydrogel contact lens
0.2–0.3 mmol L–1
0–150 mmol L–1
No
-
(90)
Paper-based microfluidic channel
1.5 mmol L–1
0–200 mmol L–1
No
-
(53)
Scleral contact lens
15.6 mmol L–1
0–100 mmol L–1
No
-
(52)
K+
Silicone hydrogel contact
lens
-
0–200 mmol L–1
No
-
(90)
Paper-based microfluidic
channel
0.9 mmol L–1
0–50 mmol L–1
No
-
(53)
Scleral contact lens
8.1 mmol L–1
0–50 mmol L–1
No
-
(52)
Ca2+
Paper-based microfluidic channel
0.03 mmol L–1
0–2 mmol L–1
No
-
(53)
Scleral contact lens
0.02–0.05 mmol L–1
0.50–1.25 mmol L–1
No
-
(52)
Mg2+
Scleral contact lens
0.01–0.44 mmol L–1
0.5–0.8 mmol L–1
No
-
Zn2+
0.001 mmol L–1
0.01–0.02 mmol L–1
No
-
Glucose
Matlab imaging
-
0.04–4 mmol L–1
No
<2 min
(127)
Fluorescence spectrum
-
0.03–3 mmol L–1
No
30 s
(128)
Silicone hydrogel contact lens
-
0–100 mmol L–1
No
-
(129)
Paper-based microfluidic channel
0.08 mmol L–1
0.1–4.0 mmol L–1
No
0.3 s
(130)
Contact lens
0.0093 mmol L–1
0.023–1.0 mmol L–1
Yes
3–5 s
(131)
Lactoferrin
Paper-based microfluidic channel
0.3 mg mL–1
0.5–3 mg mL–1
Yes
15 min
(134)
0.1 mg mL–1
0–4 mg mL–1
Yes
<10 min
(135)
Lysozyme
Contact lens
1.99
μg mL–1
0–25 μg mL–1
Yes
10 min
(138)
Fluorescence spectrum
0.55 nmol L–1
1.0–20 nmol
L–1
Yes
100 s
(140)
It was predicted that the global contact lens market
will increase
to over $19 billion in 2024.[150] Instead
of converting the contact lens for existing therapeutic and cosmetic
uses, various countries have begun treating contact lenses as potential
medical tools for broad applications in ocular-related disease diagnosis
and drug delivery. Cooperation with a novel series of hand-held readout
smart devices to enhance the facilities of minimally or noninvasive
fluorescence contact lens sensors even brought a higher value for
these novel generated contact lens sensors. The market for these kinds
of smart contact lens sensors is claimed to possess over $24.12 billion
by 2029.[150,151] Therefore, both manufacturers
and patients worldwide could obtain a promising potential market with
the development of contact lens sensors. However, the low volume and
concentration of sampled proteins and enzymes in tears are still negligible.[152,153] Principles of fabricating fluorescence biosensors for in situ contact
lens sensing to achieve POC in indication and real-time detection
of ocular diseases should focus on the mechanism of sensing technologies,
types of cross-linkers, and fluorophores. Another crucial ultimate
prospective of fluorescence tear sensors is drug dosage and delivery.
Microneedle arrays, as a popular developing technology in delivering
drug, biosensing analytes, and neural interfaces,[154−156] can be utilized to combine with fluorescence sensing technologies
for drug delivery in tears. One of the advantages of the microneedle
fabrication process is an easy operation by using low-cost 3D printing
techniques.[157,158] Subsequently, the fabricated
fluorescent microneedle arrays can reach not only the tear fluid interface
but other intraocular positions, such as aqueous humor. Then, one
suitable design could be selecting each one-phase monitoring and diagnosis
for ocular disease and combining the single test channel into the
multidetection phase, simulating and optimizing the interferences
for the final diverse sensing mode within the fluorescent sensor,
and finally producing one microneedle channel for target therapeutic
aim. Moreover, the experimental data can be compared and optimized
with the obtained clinical ophthalmic technologies, such as ophthalmic
spectroscopies and clinical chemistry analyzer, for instance. With
the golden standard comparisons, the permitted tolerance of errors
would be optimized to be as minor as possible.
Conclusion
In
conclusion, tear fluid detection has attracted continuous attention
in scientific, technological, and clinical studies of healthcare diagnosis.
Fluorescence sensing materials have been well-established for wide
applications in the past decade. Fluorescent sensing in tear fluid
offers a sensitive, cost-effective, and noninvasive platform for early
diagnosis of ocular-related diseases, including various cancers, neurological
disorders, sclerosis diseases, and Parkinson’s disease. One
immediate future area could be directly working on immediate diagnosis
and data collection for ocular disease at the POC platform in which
the patients would not need to rely heavily on hospitality and could
achieve self-detection and data collection. The development of biological
assays would also be one of the intermediate future directions on
ocular disease monitoring. The integration of the portable readout
devices and wearable sensing tools (contact lens) can be helpful at
the POC platform to improve the user experience and convenience. The
fluorescence sensors then should be sufficient in sensitivity, selectivity,
accuracy, and reproductivity. Hence, potential principles for fluorescence
sensing and sufficient fluorophores need to be developed and investigated.
Moreover, the biocompatibility of the fluorescent detection within
tears should be considered for future commercialized trials. As for
the intermediate future of fluorescent tear sensing, more complicate
biomarkers within tears and even aqueous human fluid such as BDNF,
IL-6, and other proteins need to be explored. Some neuro-related diseases
could be achieved at a POC diagnosis and monitor. Moreover, a multichannel
diagnosing fluorescent sensing could be developed to achieve multidetection
on one ocular-related disease and that would lead to a more accurate
data outcome for patients.It is also necessary to achieve minimally
or noninvasive tear sensing
and real-time monitoring for ophthalmic diagnosis and ocular physiological
index analysis. The achievement of drug delivery at the POC platform
is one of the most critical longer future directions to tear diagnosis.
Fluorescence contact lens sensors, as one of the most effective techniques
for biosensing and real-time monitoring, therefore, should be used
to conduct more investigations in intermediate ophthalmological diagnosis
and long-term database collection for hospitalization. Sufficient
fluorescence sensing technologies as well as the cooperation with
portable hand-held smart devices should be evaluated and explored
to enhance the functions of fluorescence sensing and diagnosis within
tears at the POC platforms. These fabricated fluorescent sensors could
be applied clinically for real-time monitoring, one-time detection,
drug delivery, etc. In some cases, the continuous monitoring would
not be necessary, but the annual or regular detection and monitoring
would be required. For example, some glaucoma patients would need
to do the examination every 6 months, where some of the ocular diseases
such as inflammation would require a more frequent detection (once
per week). Furthermore, real-time monitoring would be beneficial to
the surgical monitoring before and afterward.