| Literature DB >> 35632216 |
Shuang Liao1, Peng Ye1, Cheng Chen1, Jie Zhang1, Lin Xu1, Feng Tan1.
Abstract
Blood viscosity measurements are crucial for the diagnosis of cardiovascular and hematological diseases. Traditional blood viscosity measurements have obvious limitations because of their expensive equipment usage and large sample consumption. In this study, blood viscosity was measured by the oscillating circuit method and impedance analysis method based on single QCM. In addition, the effectiveness of two methods with high precision and less sample is proved by the experiments. Moreover, compared to the result from a standard rotational viscometer, the maximum relative errors of the proposed oscillating circuit method and impedance analysis method are ±5.2% and ±1.8%, respectively. A reliability test is performed by repeated measurement (N = 5), and the result shows that the standard deviation about 0.9% of impedance analysis is smaller than that of oscillating circuit method. Therefore, the impedance analysis method is superior. Further, the repeatability of impedance analysis method was evaluated by regression analysis method, and the correlation coefficient R2 > 0.965 demonstrated that it had excellent reproducibility.Entities:
Keywords: BVD circuit; blood viscosity; high precision; linear regression analysis; quartz crystal microbalance (QCM)
Mesh:
Year: 2022 PMID: 35632216 PMCID: PMC9147212 DOI: 10.3390/s22103804
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.847
Figure 1Test schematic for the measurement of viscosity by QCM.
Figure 2Equivalent BVD circuit with viscoelastic film loading.
Characteristic Parameters of QCM.
| Parameters | Values | Units |
|---|---|---|
|
| 2651 | Kg·m−3 |
|
| 9.27 × 10−3 | Pa·s |
|
| 2.92 × 10−5 | m2 |
|
| 1.66 × 10−4 | m |
|
| 7.44 × 10−3 | |
|
| 2.947 × 1010 | N·m−2 |
|
| 3.32 × 103 | N1/2·Kg−1/2·m5/2 |
|
| 9.657 × 10−2 | A·s·m−2 |
|
| 3.982 × 10−11 | A2·s4·Kg−1·m−3 |
Measured pressure-frequency coefficient and stress-frequency coefficient.
|
|
|
|---|---|
| 1.39 × 105 | −5.704 × 106 |
Figure 3Response of the QCM sensor based on oscillating circuit analysis method.
Measured the blood viscosity by oscillating circuit analysis method.
| Concentration | Viscosity (mPa·s) | ||||||
|---|---|---|---|---|---|---|---|
| Reference Value | Method 1 Measured Value | Standard Deviation | |||||
| 1 | 2 | 3 | 4 | 5 | |||
| 10 | 1.14 | 1.13 | 1.127 | 1.12 | 1.20 | 1.135 | 5.5% |
| 20 | 1.48 | 1.44 | 1.455 | 1.456 | 1.458 | 1.41 | 2.1% |
Measured the blood viscosity by impedance analysis method.
| Testing Times | Concentration (wt%) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| 10% | 20% | |||||||||
| R (Ω) |
| Measured Viscosity | Reference | Standard Deviation | R (Ω) |
| Measured Viscosity | Reference | Standard Deviation | |
| 1 | 83.56 | 1167.8 | 1.119 | 1.14 | 0.9% | 102.05 | 1195.5 | 1.459 | 1.48 | 0.5% |
| 2 | 84.03 | 1193.2 | 1.14 | 101.79 | 1189.9 | 1.465 | ||||
| 3 | 83.69 | 1186.7 | 1.12 | 101.9 | 1182.9 | 1.47 | ||||
| 4 | 83.72 | 1185.4 | 1.121 | 101.79 | 1200.1 | 1.458 | ||||
| 5 | 83.98 | 1189.1 | 1.13 | 101.59 | 1171.7 | 1.46 | ||||
Figure 4Repeatability experiment (N = 8) of blood viscosity QCM sensor based on impedance analysis method.