| Literature DB >> 35630698 |
Maximilian Weiss1, Thomas Glechner2, Victor U Weiss1, Helmut Riedl2,3, Andreas Limbeck1.
Abstract
The increased demand for sustainability requires, among others, the development of new materials with enhanced corrosion resistance. Transition metal diborides are exceptional candidates, as they exhibit fascinating mechanical and thermal properties. However, at elevated temperatures and oxidizing atmospheres, their use is limited due to the fact of their inadequate oxidation resistance. Recently, it was found that chromium diboride doped with silicon can overcome this limitation. Further improvement of this protective coating requires detailed knowledge regarding the composition of the forming oxide layer and the change in the composition of the remaining thin film. In this work, an analytical method for the quantitative measurement of depth profiles without using matrix-matched reference materials was developed. Using this approach, based on the recently introduced online-LASIL technique, it was possible to achieve a depth resolution of 240 nm. A further decrease in the ablation rate is possible but demands a more sensitive detection of silicon. Two chromium diboride samples with different Si contents suffering an oxidation treatment were used to demonstrate the capabilities of this technique. The concentration profiles resembled the pathway of the formed oxidation layers as monitored with transmission electron microscopy. The stoichiometry of the oxidation layers differed strongly between the samples, suggesting different processes were taking place. The validity of the LASIL results was cross-checked with several other analytical techniques.Entities:
Keywords: boride thin films; high-temperature oxidation; online-LASIL; quantitative depth profiling
Year: 2022 PMID: 35630698 PMCID: PMC9147019 DOI: 10.3390/molecules27103221
Source DB: PubMed Journal: Molecules ISSN: 1420-3049 Impact factor: 4.927
Figure 1Demonstration of the optimization process for the make-up and carrier solutions: If the analyte was not stabilized enough, it was absorbed onto the walls, and it appeared as a peak if it was desorbed by concentrated acid.
Figure 2Illustration of the signal during a depth profile measurement with standard addition.
Figure 3Dependence of the ablation rate and the ICP-MS signal from the applied laser energy.
Comparison of the results of the stoichiometry determination between the liquid ICP-OES (n = 3) and the online-LASIL analysis performed on the native samples with high laser energy (0.51 mJ). The cations were normalized to 100%.
| Sample | Measurement | Cr at% | Si at% | B at% |
|---|---|---|---|---|
| Sample A | ICP-OES | 27.1 ± 0.1 | 8.9 ± 0.3 | 64.0 ± 0.3 |
| LASIL bulk analysis | 25.0 ± 1.8 | 7.7 ± 3.2 | 67.3 ± 1.5 | |
| Sample B | ICP-OES | 25.6 ± 0.2 | 15.8 ± 0.1 | 58.6 ± 0.1 |
| LASIL bulk analysis | 22.8 ± 1.0 | 18.8 ± 1.5 | 58.4 ± 0.6 |
Figure 4LASIL depth profiles with corresponding bright-field TEM cross-sections of the samples: (a) sample A (native stoichiometry Cr0.27Si0.9B0.64); (b) sample B (native stoichiometry Cr0.26Si0.16B0.58). The scale bar represents 1000 nm, and the images were rotated to correspond to the depth profiles. The thin red line in the TEM image of sample B indicates the boundary of the oxide and was determined with high-angle annular dark-field imaging (HAAFD) STEM (scanning transition electron microscopy).
Figure 5(Left): CAD exploded drawing of the LASIL cell. The gray parts represent PEEK, the PFA tubing pushed in the PEEK body is depicted in red, the PDMS spacer in black, and the fused silica window in green. (Right): Cross-section of the LASIL cell showing the fluid path and the flows through the sample cell. The LASIL cells consisted of a PEEK body with a cavity for the sample. Through the body, three tubes pushed the flow for two inlets and one outlet. The carrier flowed over the sample, the make-up solution was added behind the sample in the fluid path, so it did not come into contact with the sample. The two flows mixed just before they left the LASIL cell at the outlet to the ICP-MS instrument. The cell was sealed by a PDMS spacer and a fused silica window.
ICP-MS measurement parameters for online-LASIL.
| Parameter | Value |
|---|---|
| RF power | 1550 W |
| Auxiliary gas flow (Ar) | 1.0 L/min |
| Cooling gas flow (Ar) | 14 L/min |
| Nebulizer gas flow (Ar) | 0.8 L/min |
| CCT bias | −21 V |
| Pole bias | −18 V |
| KED gas flow (7% H2 in He) | 5 mL/min |
| Monitored ions | 11B, 27Al, 28Si, 52Cr, 53Cr, 115In, |
| Dwell time | 0.01 s for 27Al, 52Cr, 53Cr, 115In |
Laser parameters for the online-LASIL measurements.
| Parameter | Value |
|---|---|
| Laser energy depth profile | 0.17 mJ |
| Laser energy survey run | 0.51 mJ |
| Spot size | 100 µm |
| Scan speed | 500 µm/s |
| Carrier solution flow | 0.53 mL/min |
| Makeup solution flow | 0.28 mL/min |
| Repetition rate | 10 Hz |
| Investigated sample area | 0.1 mm2 |