Mekides Abera Seyoum1, Menen Teshome Belachew1, Meron Berihun Dessale1, Meti Wakjira Adugna1,2, Tizita Yohannes Keto1, Yezbalem Getnet Wassie1, Habtamu Abafoge Dobamo3, Melese Uma Getaneh1, Hundessa Daba Nemomssa1.
Abstract
Background: Currently, hospital-acquired infections in healthcare workers and patients are a major concern. On the other hand, sexually transmitted infections and diseases, unwanted pregnancies, and unsafe abortions continue to be a public health concern, particularly in developing countries. Gloves are among the most commonly used personal protective equipment to safeguard healthcare workers' hands from contagious infections, and using a condom is strongly advised for people who have sexual relations with more than one partner. However, the quality of gloves and condoms in developing countries is a subject of concern. The usage of quality test instruments such as glove leakage test apparatus (GLTA), leakage testers by water level, the Ammonia leak testing method (ALTM), conductivity-based leakage testers, and water hang testers in developing countries is limited owing to cost, accessibility, and safety. The main purpose of this study was to develop and test a low-cost integrated device to test glove and condom leakage that is safe and easily accessible in resource poor settings. Method: In this study, an integrated glove and condom leakage testing device for detecting pin holes and leakages is proposed. The device automatically fills a randomly selected condom and glove with a predetermined volume of water based on International Organization for Standardization (ISO) criteria.Entities:
Keywords: Condom leakage; Glove leakage; Hospital acquired infection; Low-cost tester; Resource constrained areas; Sexually transmitted diseases; Unwanted pregnancy
Year: 2022 PMID: 35620039 PMCID: PMC9127158 DOI: 10.1016/j.amsu.2022.103791
Source DB: PubMed Journal: Ann Med Surg (Lond) ISSN: 2049-0801
Fig. 1Block diagram of low-cost glove and condom leak test device.
Fig. 2The result of the controlling system simulation.
Fig. 3Final prototype of low-cost glove and condom leakage tester.
Materials used for construction of the prototype.
| S·N | Components | Specification |
|---|---|---|
| 1 | Microcontroller | Arduino Mega 2560 |
| 2 | Ultrasonic sensor | HC-SR04 |
| 3 | Solenoid valve | 2-way, 0.5-inch diameter, 12 V |
| 4 | Relay | 5 V |
| 5 | Switch | 10A, 125 V |
| 6 | LCD | 20 × 4 Alphanumeric |
| 7 | Water pump | 5 V |
| 9 | Transformer | 220 V to 12-V step down transformer |
| 10 | Diode | 1N5408 rectifier diode |
| 11 | Voltage regulator | 7805 and 7812 |
| 12 | Capacitor | 2200 μF |
| 13 | Resistors | 10 kΩ |
| 14 | Potentiometer | 10 kΩ |
| 15 | Wires | Male and female jumper wires |
Performance evaluation of our device with standard measurement.
| Iteration number | Volume measurement test for Condom sample | Volume measurement test for Glove sample | ||||
|---|---|---|---|---|---|---|
| Using a measuring cylinder (in ml) | Using our Device (in ml) | % Error | Using a measuring cylinder (in ml) | Using our Device (in ml) | % Error | |
| 1 | 295 | 300 | 1.69 | 994 | 1000 | 0.60 |
| 2 | 296 | 300 | 1.33 | 992 | 1000 | 0.81 |
| 3 | 296 | 300 | 1.33 | 995 | 1000 | 0.50 |
| 4 | 298 | 300 | 0.67 | 993 | 1000 | 0.70 |
| 5 | 296 | 300 | 1.33 | 991 | 1000 | 0.91 |
| 6 | 295 | 300 | 1.67 | 994 | 1000 | 0.60 |
| 7 | 297 | 300 | 1.00 | 992 | 1000 | 0.81 |
| 8 | 295 | 300 | 1.67 | 993 | 1000 | 0.70 |
| 9 | 296 | 300 | 1.33 | 994 | 1000 | 0.60 |
| 10 | 296 | 300 | 1.33 | 992 | 1000 | 0.81 |
Results of tests formed on our device.
| Design Criteria | Test methodology | Result |
|---|---|---|
| Accuracy | Comparing the water volume measurement by our device with gold standard | Accuracy of 98.66% for condom sample and 99.29% for glove sample were obtained |
| Cost | Calculating the cost of items used to construct the prototype | $104.49 |
| User safety | By comparing the testing method used in our device with the methods used in existing devices and ISO standards | Water is used for testing the leakage of gloves and condoms in our device. Hence it is safe for the user. |
| User friendliness | The time taken for a first-time user to use the device | It takes 2–3 min to test one sample by first time user |
| Portability | Device weight | <10 Kg |