There is an increasing emphasis on the critical evaluation of interbatch purity and physical stability of therapeutic peptides. This is due to concerns over the impact that product- and process-related impurities may have on safety and efficacy of this class of drug. Aspartic acid isomerization to isoaspartic acid is a common isobaric impurity that can be very difficult to identify without first synthesizing isoAsp peptide standards for comparison by chromatography. As such, analytical tools that can determine if an Asp residue has isomerized, as well as the site of isomerization within the peptide sequence, are highly sought after. Ion mobility-mass spectrometry is a conformation-selective method that has developed rapidly in recent years particularly with the commercialization of traveling wave ion mobility instruments. This study employed a cyclic ion mobility (cIMS) mass spectrometry system to investigate the conformational characteristics of a therapeutic peptide and three synthetic isomeric forms, each with a single Asp residue isomerized to isoAsp. cIMS was able to not only show distinct conformational differences between each peptide but crucially, in conjunction with a simple workflow for comparing ion mobility data, it correctly located which Asp residue in each peptide had isomerized to isoAsp. This work highlights the value of cIMS as a potential screening tool in the analysis of therapeutic peptides prone to the formation of isoAsp impurities.
There is an increasing emphasis on the critical evaluation of interbatch purity and physical stability of therapeutic peptides. This is due to concerns over the impact that product- and process-related impurities may have on safety and efficacy of this class of drug. Aspartic acid isomerization to isoaspartic acid is a common isobaric impurity that can be very difficult to identify without first synthesizing isoAsp peptide standards for comparison by chromatography. As such, analytical tools that can determine if an Asp residue has isomerized, as well as the site of isomerization within the peptide sequence, are highly sought after. Ion mobility-mass spectrometry is a conformation-selective method that has developed rapidly in recent years particularly with the commercialization of traveling wave ion mobility instruments. This study employed a cyclic ion mobility (cIMS) mass spectrometry system to investigate the conformational characteristics of a therapeutic peptide and three synthetic isomeric forms, each with a single Asp residue isomerized to isoAsp. cIMS was able to not only show distinct conformational differences between each peptide but crucially, in conjunction with a simple workflow for comparing ion mobility data, it correctly located which Asp residue in each peptide had isomerized to isoAsp. This work highlights the value of cIMS as a potential screening tool in the analysis of therapeutic peptides prone to the formation of isoAsp impurities.
There exists a vast
array of therapeutics centered around biologics,
encompassing antibodies, antibody–drug conjugates, and mRNA
to name but a few.[1−3] Not least of all are peptides; which are relevant
to a number of therapeutic areas and drug discovery programs,[4,5] e.g., glucagon-like peptide-1 receptor agonists used in the treatment
of type 2 diabetes.[6] Noncovalent interactions
and structural complementarity are critical to biological systems
information transfer and intermolecular recognition. It is therefore
unsurprising that peptides, with their innate flexibility and access
to numerous conformational ensembles, are attractive moieties for
drug discovery.A significant difference with small molecule
drugs is that peptides
are more prone to physical instability having, in some cases, a high
propensity for self-assembly and aggregation.[7] Aggregation poses a major concern in the development of peptide
and protein pharmaceuticals due to the complexity of potential adverse
immune responses from aggregates[8,9] and, of course, any
uncertainty over the efficacy or promiscuity of the aggregated species.
In particular, the presence of impurities or seeds (preformed short
fibrils) can trigger and, in some cases, modify fibrillation with
respect to both the rate and the type/structure of fibrils formed.[10−12]Amino acid isomerization can alter a protein or peptide’s
physical stability, conformation, or activity.[13−15] Isomerization
can occur as a byproduct of peptide synthesis[16] or spontaneously; in particular, the formation of isoAsp from either
Asp or deamidated Asn via a common succinimide intermediate[17] is well-known and introduces an additional methylene
into the peptide backbone (Figure ). Isomerized peptides have been shown in many settings
to change the properties of a peptide[18−21] and therefore must be characterized
fully as potential impurities before clinical development. Therefore,
the development of techniques that can identify the location of isoAsp
isomers and differentiate isomers from their Asp counterparts are
crucial.
Figure 1
Isomerization of aspartic acid and location of aspartic acids in
RS19. (a) Isomerization of α-aspartic acid residue that results
in the formation of a β-aspartic acid. (b) Representation of
RS19 (World Patent WO2017153575A1, 2017) with the three potential
isomerization sites (Asp residues) highlighted in green at positions
9 (ISOA), 15 (ISOB), and 21 (ISOC). The site of lipidation is also
shown (*).
Isomerization of aspartic acid and location of aspartic acids in
RS19. (a) Isomerization of α-aspartic acid residue that results
in the formation of a β-aspartic acid. (b) Representation of
RS19 (World Patent WO2017153575A1, 2017) with the three potential
isomerization sites (Asp residues) highlighted in green at positions
9 (ISOA), 15 (ISOB), and 21 (ISOC). The site of lipidation is also
shown (*).Reversed-phase liquid chromatography
coupled to mass spectrometry
(LC–MS) is an obvious option to determine the presence of impurities,
their molecular formula, and relative hydrophobicity. Fragmentation
of peptides, with MS2, can routinely detect various isomers;
however, impurities in which there has been an isomerization of an
Asp to isoAsp are challenging to detect. MS2 with traditional
collision-induced dissociation (CID) yields fragment ions that remain
isobaric and are therefore ambiguous. Sophisticated and more niche
fragmentation methods have been successful at differentiating isomers
including isoAsp and Asp from each other and deamidation products.[22−24] Electron transfer dissociation (ETD), for example, has been conducted
on Asp- and isoAsp-containing peptides[25] but traditionally uses hazardous reagents, produces low intensity
fragments, and is known to suffer from poor fragmentation efficiency
from analytes with lower charge densities.[26] This and many other techniques lack the appropriate sensitivity,
speed, and resolution required for batch-to-batch quality control.
Pairing MS with a structure-based differentiation technique, like
ion mobility-MS (IMS),[27] is a natural choice
for identifying isomers with the same chemical formula, hence, m/z, but potentially different molecular
conformations. IMS has already demonstrated capability in helping
distinguish trace impurities in pharmaceuticals.[28]Ion mobility experiments involve separating ions
in the gas phase
based primarily on both their charge and their rotationally averaged
collision cross sectional area (CCS, Ω). The technique is often
used in conjunction with time-of-flight mass spectrometry (typically
QTOF) to provide information on both mass and structure. For details
of IMS, its application, and some of the experimental methods available,
interested readers are directed to reviews available elsewhere.[29,30]In traveling wave ion mobility (TWIMS) experiments such as
those
using the Synapt G2 (Waters Corp.) commercial instrument, the ions
pass through a defined length of a stacked ring ion guide (SRIG) composed
of circular electrodes confining ions radially by an applied radio
frequency voltage. Ions are pushed through, against the buffer gas,
by a direct current (DC) pulse which travels along the series of contiguous
electrodes like a wave,[31,32] the principle being
that larger, lower charged ions take longer to travel through the
mobility device. This is because larger ions “tumble over”
DC waves that are sufficient to push smaller ions forward. Hence,
conformations that result in an expanded ion rather than a compact
one exit the drift cell and arrive at the mass detector later. Mobilities
can be represented by arrival time distributions (ATD), with the width,
asymmetry and peak maxima of these ATD providing information on the
relative conformation of the ions. Wider distributions typically represent
one of two things, either multiple ions that have nonresolvable drift
times and hence are likely to be structurally very similar or conformers
of one species that are interconverting on the time scale of the mobility
separation.A new development in IMS is the commercial release
of the SELECT
SERIES Cyclic IMS instrument from Waters Corp. The flagship feature
of cyclic IMS (cIMS) is the potential for passing ions around the
circular “racetrack” of the cyclic T-wave device an
infinite number of times. Improved resolution of the cyclic system
compared to a linear traveling wave device, which is limited by the
length of the mobility cell, has already been highlighted a number
of times despite being a relatively new technology.[33,34] The cyclic system is limited only when ions become so well resolved
that the fastest overtakes the slowest, somewhat analogous to cyclists
being lapped in a velodrome. The technique and instrumental details
are described in depth by Giles et al.,[35] but notably, the SELECT SERIES Cyclic IMS allows ions separated
on account of their mobility (fragments or whole ions) to be selected
and reinjected into the cyclic array to tease out the finer features
of their conformational differences. This technique (termed IMSn)[35] is not possible with the earlier
linear versions of the TWIMS instruments.In this work, we present
the use of cIMS for differentiating a
model system of three synthetic isomers (ISOA, ISOB, and ISOC which
are isoAsp9, isoAsp15, and isoAp21, respectively) of a therapeutic
peptide from their nonisomerized reference standard, RS19.[36] Of the three isomeric peptides, each contained
a single aspartic acid residue as an isoAsp, and the sample set was
chosen with two main objectives in mind. First, the ATD of each isomer
after multiple passes was used as a comparator to differentiate the
isomers from RS19, thereby evaluating cIMS as an analytical technique
sensitive enough to differentiate potential impurities that are chemically
similar. Second, the resolution of cIMS meant that subtle differences
in the ATDs of fragment ions was capable of identifying the position
of the β-aspartic acid. Thus, pseudosequencing of the isomeric
modification was achieved by a pairwise comparison of fragment ion
ATDs which accounted for differences not only in arrival time in a
manner similar to that outlined by Jia et al.[37] and Tomczyk et al.[38] but also peak shape.
This new processing method, called (normalized subtraction integration,
NSI), was developed to provide a quantitative measure to describe
how different two ATDs were from each other and was applied to RS19
and its three singly isomerized isoAsp counterparts. The workflow
proposed here is a step toward a more objective analysis of high-resolution
ion mobility data of biomolecules, with the potential to incorporate
it into a more automated workflow through combination with chromatographic
separation to give LC-CID-cIMS.
Experimental Section
Sample
Preparation
All peptides were supplied as lyophilized
powders by AstraZeneca. Peptides were solubilized in 20 mM ammonium
acetate (pH 6.4) prior to filtration through a 0.22 μm syringe
filter (Millex-GV) and dilution to the required concentration (5–9
μM). Solutions were aliquoted and frozen at −80 °C
for storage. Samples were thawed only once prior to infusion by nESI
or injection to LC. Peptide isomer identities were known prior to
data analysis. ISOA was isomerized at Asp9, ISOB was isomerized at
Asp15, and ISOC was isomerized at Asp21.
Instrumentation
Ion mobility mass spectra were acquired
using a SELECT SERIES Cyclic IMS (Waters) instrument using PicoTip
Glass Tip coated capillaries (NewObjective). Native spray conditions
were as follows: capillary 1–1.2 kV, cone 40 V, source temperature
100 °C. All other instrument parameters were used as default.
The cyclic ion mobility cell was operated at a pressure of 1.8 mbar
nitrogen, with a static traveling wave height of 15 V. Single pass
cIMS experiments were performed with an inject time of 10 ms and separate
time of 2 ms. The number of passes was increased up to four by increasing
the separate time accordingly.Liquid chromatographic separation
of the synthetic peptides was performed on an ACQUITY I-Class UPLC
(Waters) equipped with an ACQUITY Premier 1.7 μm, 2.1 ×
100 mm CSH C18 column operated at 60 °C. Mobile phases were water
(A) and acetonitrile (B) each with formic acid added to a final concentration
of 0.1%. The proportion of mobile phase B was increased from 35–45%
over 8 min for separation of the isomeric peptides.For the
infusion-based CID-cIMS experiments, the [M + 3H]3+ ion
(1242 m/z) of each peptide
was selected in the resolving quadrupole before collisional activation
at 75 or 80 V in the trap collision cell. The resulting product ions
were subjected to a single pass only for this analysis. For the LC-CID-cIMS
experiments, fragment ions were generated after quadrupole isolation
of the [M + 4H]4+ ion (933 m/z) using a trap collision energy of 45 V in single pass cIMS mode.
A combination of +1, + 2 and +3 fragments were analyzed for each data
set after a single pass of the cyclic device, but each isomer was
always compared to the corresponding m/z of RS19. On occasion, for less abundant fragments, more than just
the monoisotopic peak had to be extracted to generate an acceptable
ATD.
Software
Mass spectra were processed using Masslynx
v4.2 (SCN 1016 and SCN1007) and DriftScope v3 and v2.8. Arrival time
distributions were replotted in either Microsoft Excel (Office 365)
or GraphPad Prism (v. 9.0.0) after extracting chromatogram lists and
plotting intensity as a percentage of the maximum value, per mobilogram,
in the interval range extracted. Calculations of the fragment ion
ATD similarity (NSI) were conducted using a Python script prepared
in house using Python (v.3.9) and the HTML-based Jupyter Notebook.
(v. 6.4.0) The script was written using pandas[39,40] (v. 1.3.1) and tkinter (v. 8.6) to import mobility data as csv files
which had previously been exported as chromatograms from Masslynx
where the x axis was arrival time (in milliseconds)
and the y axis intensity (I) For each
ion I, the extracted intensity was normalized to a percentage
intensity (relative intensity, RI) at time t by dividing
by the maximum extracted intensity (Imax)
for that specific ion (eq ) across all time points. It was this relative intensity from the
isomer ion that was subtracted from the reference ion relative intensity
at each time point to give the difference, D. As
this number could be positive or negative, dependent on how the two
ion ATDs differed, the absolute value was calculated at each time
point, and finally, this was used to construct a graph of absolute
relative intensity (%) versus time (ms). This plot was integrated
using the trapezoid function from the scipy.integrate library to give
a quantitative value (which we call the NSI) which represents the
difference between that isomer ion ATD and the corresponding RS19
fragment ion ATD (eq ). To reduce fluctuations in these values when plotted per terminal
amino acid along the sequence of the peptide, a running average was
plotted instead. This means that for a fragment ion ending in residue
n, the NSI plotted at that n value was the average NSI for fragment n – 1, n, and n + 1. The workflow, from which NSI values were calculated, can be
seen in Figure S1.
Results and Discussion
Differentiation
of Precursor Ions using cIMS
RS19 and
the isomers were individually solubilized in 20 mM ammonium acetate
(pH 6.4) to concentrations between 5 and 9 μM before direct
infusion by nESI into the SELECT SERIES Cyclic IMS. The monomeric
[M + 3H]3+ species (the most abundant ion observed for
each isomer Figure S2) was quadrupole selected
and investigated, first after a single pass and then multiple passes.There are three Asp residues in the sequence with each peptide
having a single isomerization to isoAsp, except RS19, which was known
to be completely in the Asp form. Even after one pass (Figure a), the isomers were more distinct
from each other than they appeared using a linear TWIMS instrument
(Figure S3). The arrival time distributions
of the major peak maxima ranged from 44.28 to 47.72 ms, and shoulders
were apparent in the ATD for ISOA, ISOC, and RS19. The major population
for each isomer did not have the same arrival time, indicating a different
conformational preference for each.
Figure 2
Comparison of the ATDs of individual intact
isomers and the reference
standard (RS19) after a single pass and four passes around the cyclic
T-wave device. Subtle features in the ATDs after a single pass (a)
are exaggerated after four passes (b) where it is possible to clearly
differentiate one isomer from another. (c) For ISOA two distinct conformer
populations were observed, but for ISOB and ISOC only one was observed.
Each isomer was observed to have a different arrival time for its
major conformation. The bottom left panel shows the position of the
isomerized Asp residue (green) in each isomer.
Comparison of the ATDs of individual intact
isomers and the reference
standard (RS19) after a single pass and four passes around the cyclic
T-wave device. Subtle features in the ATDs after a single pass (a)
are exaggerated after four passes (b) where it is possible to clearly
differentiate one isomer from another. (c) For ISOA two distinct conformer
populations were observed, but for ISOB and ISOC only one was observed.
Each isomer was observed to have a different arrival time for its
major conformation. The bottom left panel shows the position of the
isomerized Asp residue (green) in each isomer.Figure shows the
difference between one and four cIMS passes for the [M + 3H]3+ ion (m/z 1242.9) with replicate
data shown in Figure S4. The four pass
data revealed distinct and unique ATDs in the range 130–160
ms for the [M + 3H]3+ monomer ions of each of the isobaric
peptides (ISOA, ISOB, and ISOC). Here, the greater resolving power
of four passes revealed differentiating features between isomers that
were not as distinct after only a single pass (Figure S5). The most noticeable difference was that the shoulders
on both ISOA and RS19 were now two distinct peaks. This is in contrast
to the shoulder observed with ISOC which was not as apparent as it
had been after one pass. In this case, either the second conformer
population of ISOC was structurally more similar to its major population
for ISOA and RS19 or the conformational populations in ISOC may have
been interconverting between the two populations faster than the mobility
measurement. This suggests that the populations of ISOA and RS19 were
not exchanging as fast, if at all. Significantly more work is required
to probe the exact energy landscape and dynamics of the multiple populations
observed for each isomer, but this result exemplifies the potential
of the instrument to differentiate the conformational ensembles of
closely related peptide therapeutics, even after a single pass.The shorter arrival time of the ISOA major population after four
passes (139.85 ms) compared to that of the other peptides, in particular
RS19 (153.01 ms), indicated a significant proportion of molecules
were present in a more compact conformation. It became strikingly
clear after four passes how different the conformational ensembles
of ISOB and ISOC were from their regiomeric ISOA. ISOB had an arrival
time most similar to RS19 and ISOC had a conformational preference
for a more compact conformation (ATD 144.03 ms) compared to ISOB (149.08
ms).cIMS provided strong experimental evidence that the site
of isomerization
can noticeably impact the conformational ensemble of a therapeutic
peptide. Four passes around the cyclic device demonstrated that the
three different isomers of RS19 are not conformationally equivalent,
resulting in dissimilar ATDs. The change in conformation was not entirely
unexpected as isomerization to a β-peptide introduces an additional
backbone methylene which could disrupt intramolecular hydrogen bonding
and could conceivably alter the preferred torsion angles of the backbone.
Computational analyses of the conformational/energy landscape for
the isomers have not been conducted, preventing definitive conclusions
on their possible conformations, and the authors note that further
investigation is required to explain why the conformations
appear as they do. Here, we have shown how cIMS is capable of differentiating
a therapeutic peptide from its potential impurities after only minutes
of data acquisition.
Site of Isomerization Located Using CID-cIMS
Fragmentation
techniques, such as CID, cannot differentiate between fragments containing
α- or β-Asp, as all isomers result in fragment pairs (e.g., b/y ions) with the same m/z to the nonisomerized peptide. Currently, potential
impurities are synthesized as “standards” and chromatographic
techniques used to compare a drug product with an impurity. This is
time-consuming and resource intensive; thus, we combined CID with
cIMS, allowing fragment ions to be differentiated on the basis of
their mobilities (shape) despite having the same m/z ratio. As with previous studies on shorter peptides,[37,38] only fragments containing the isomer should be different from the
nonisomerized reference. It was expected that the conformational change
induced by an Asp isomerization in fragment ions should be relatively
more significant than in longer precursor ions. However, unlike previous
studies, this analysis investigated peptides that had 30 residues,
contained a conformationally flexible lipid, and, crucially also accounted
for differences in ATD asymmetry as well as absolute arrival times.
Accounting for ATD asymmetry may aid with interlaboratory comparisons
given that arrival times can be liable to variation between instruments.
Additionally, isomerization of larger molecules may induce only subtle
changes, resulting in ATDs with multiple, less intense maxima instead
of significant changes in the absolute arrival time.Consequently,
a pairwise comparison of the ATDs of isomer fragment ions with the
ATDs of the nonisomerized RS19 fragment ions was undertaken, initially
in a qualitative manner (by eye), to decide whether the ATD was “equivalent”
or “different”. Figure illustrates a theoretical fragment similarity pattern
for an isomer at position 15, which in this case corresponds to ISOB.
ATDs are predicted to be similar for b- and y-ions up to, but not including, fragments that contain
the isoAsp, b1–14, or y1–15
(green, Figure ).
For isomerization at Asp9 (ISOA) and Asp21 (ISOC), the point at which
“matching” fragments switch to “different”
is also indicative of the fragmentation site proximity to the isomeric/structural
change, and this is different for each isomer (Figure ). A schematic of the instrument used in
this study is shown in Figure S6 and depicts
the SELECT SERIES Cyclic IMS with the cyclic T-wave device located
upstream of the TOF mass analyzer. After quadrupole isolation (1) of the desired ions, CID in the trap generated a series
of b/y-fragments (2). These fragments entered the cyclic device and were separated according
to mobility (3) before reaching the high-field pusher
of the TOF mass analyzer. Representative regions of fragmentation
mass spectra are illustrated Figure S7.
For ISOA, y fragments up to y21
were expected to be equivalent, for ISOB fragments up to y15 were expected to be equivalent, and for ISOC fragments up to y9 were expected to be equivalent.
Figure 3
Schematic to illustrate
the theoretical similarity of fragments
in an isomerized peptide in comparison with a nonisomerized reference
standard. Green squares are fragments that have an equivalent ATD
to the RS19, and fragments with a red square are those that differ.
Figure 4
ATDs of fragment ions for RS19 and each isomer. For y11 it was expected and observed that the ATD for ISOA (red)
and ISOB
(green) would match RS19 (blue) but that ISOC (orange), containing
the isoAsp, would have a different ATD. For fragment y18, it was expected and observed that only ISOA would match the ATD
of RS19. For fragments y28 and y15 the change in the shape of the ATD was more pronounced than any
change to the absolute arrival time.
Schematic to illustrate
the theoretical similarity of fragments
in an isomerized peptide in comparison with a nonisomerized reference
standard. Green squares are fragments that have an equivalent ATD
to the RS19, and fragments with a red square are those that differ.ATDs of fragment ions for RS19 and each isomer. For y11 it was expected and observed that the ATD for ISOA (red)
and ISOB
(green) would match RS19 (blue) but that ISOC (orange), containing
the isoAsp, would have a different ATD. For fragment y18, it was expected and observed that only ISOA would match the ATD
of RS19. For fragments y28 and y15 the change in the shape of the ATD was more pronounced than any
change to the absolute arrival time.Absolute arrival time differentiation (i.e., the ATD maximum) was
not always observed, and in some cases fragments expected to be “similar”
and “different” had approximately a 0.37 ms difference
in arrival time. (Figure S8). This variability
observed in absolute arrival times meant that this metric was not
always sufficient to discern between “equivalent” and
“different”. Figure illustrates some mobilograms for fragments y11, y15, y18, and y28 of RS19 and the three isoAsp peptides detailing how
isomer position changed the ATDs of the fragment ions. The fragment
ions from y11 were all expected to be similar in
conformation, and hence ATD, to RS19 except for that which contained
an isoAsp. In this case, ISOC was the only fragment to be different
which is consistent with the peptide being isoAsp21. The arrival time
of ISOC y11 was 43.79 ms, and it was therefore proposed
to be a larger, less compact ion than RS19 which had an arrival time
of 42.68 ms. It is interesting to note that this is different to the
results obtained for full-length precursor ions where RS19 had the
least compact major conformation. For y18 both ISOC
and ISOB were not equivalent to RS19; both had shorter arrival times
and hence more compact structures. Interestingly, despite both containing
an isoAsp, isoAsp21, and isoAsp15, respectively, the two y18 fragments of ISOB and ISOC were not equivalent to each other either,
showing that the position of isomerization mattered even after CID
fragmentation.Fragment y28 (Figure ) illustrated a typical challenge
encountered
when analyzing the data acquired in these experiments and, as mentioned
before, some differences only manifested as changes to the shape of
ATD and not the actual arrival time. An initial close inspection by
eye of each fragment ion pair, across the entire series of each isomer,
and extensive knowledge of the cIMS system, meant that the y28 fragment ions of RS19 and ISOB were initially qualitatively
regarded as “different”. However, there was some ambiguity.Hence, two mathematically simple approaches, subtraction and integration,
were used to incorporate changes to the ATD shape into the analysis
in order to distinguish isomer fragments further. Here, we propose
that intensity normalization of isomer fragment ATDs (percentage intensity
relative to maximum intensity of that fragment ion) followed by subtraction
from the normalized RS19 mobilogram, then integration of the resulting
curve to give an NSI value (normalization, subtraction integration,
as described in the Experimental Section)
is an appropriate metric. By normalizing the mobilograms to the highest
peak intensity per fragment, the overall shape and arrival times were
conserved enabling a quantitative comparison. This approach had the
advantage of avoiding complications surrounding particular fragmentation
efficiencies at different points in the sequence or slightly different
peptide concentrations which might affect the overall number of ions
produced per fragment.The NSI values were calculated (Tables S1–S3) for each fragment pair (isomer
and RS19) and were added to a running
average value (fragment n, n –
1, and n + 1), and this average value was then plotted
for each fragment. In a plot of y fragment terminal
residue vs NSI (Figure ), an inflection point was apparent (sudden increase in NSI value)
with one major transition observed per peptide indicating a significant
conformational change had occurred in fragments containing that Asp
residue only. Regarding the b fragment series, not
as many of the fragments were observed but an increase in NSI was
also seen for one replicate of ISOA (Figure S9). Thus, y ions were the focus of this study, given
that they spanned almost the entire peptide sequence whereas b ions did not. The values around inflection points (Figure ) were inspected
to provide a guide toward establishing an arbitrary “threshold”
value such that fragments with a running average NSI value above this
were proposed to be significantly different. This analysis revealed
that ISOA switched from “matching” to “different”
at Asp9, ISOB at Asp15, and ISOC at Asp21. This was in agreement with
the known identity of the peptides which had been synthesized with
an isoAsp at each of these positions.
Figure 5
Pair-wise comparison of the running average
of NSI per fragment
plotted versus the residue at the terminus of each y fragment length. Example plot (top panel) illustrates an example
with an arbitrary threshold dictating whether fragment ion ATDs were
different or similar to the reference standard fragment ion. The six
plots indicate that upon working from y1 – y30 (left to right), the first inflection in NSI value indicated
correctly the fragment lengths which were conformationally different
from the corresponding fragment in nonisomerized RS19. Hence, these
were the fragments that contained the isoAsp. The experiment was run
twice with two aliquots of the same samples, and data were acquired
after only a single pass through the cyclic T-wave array. Red dashed
vertical lines show the position of the isomerized Asp residue; therefore,
fragments to the right of the line (longer y fragments)
were expected, and observed, to have NSI values higher than this arbitrary
threshold value (red horizontal line). The threshold value for replicate
one is 40, and the value for replicate two is 90.
Pair-wise comparison of the running average
of NSI per fragment
plotted versus the residue at the terminus of each y fragment length. Example plot (top panel) illustrates an example
with an arbitrary threshold dictating whether fragment ion ATDs were
different or similar to the reference standard fragment ion. The six
plots indicate that upon working from y1 – y30 (left to right), the first inflection in NSI value indicated
correctly the fragment lengths which were conformationally different
from the corresponding fragment in nonisomerized RS19. Hence, these
were the fragments that contained the isoAsp. The experiment was run
twice with two aliquots of the same samples, and data were acquired
after only a single pass through the cyclic T-wave array. Red dashed
vertical lines show the position of the isomerized Asp residue; therefore,
fragments to the right of the line (longer y fragments)
were expected, and observed, to have NSI values higher than this arbitrary
threshold value (red horizontal line). The threshold value for replicate
one is 40, and the value for replicate two is 90.To assess day-to-day reproducibility of the approach, we performed
the experiment twice at different times, months apart. Both experiments
were performed as a single cIMS pass under similar conditions. Replicates
1 and 2 (Figure )
displayed the expected NSI inflection points for all three isomers
confirming the position of the isoAsp residue. However, some differences
were observed between the replicates that are worth discussing. First,
the threshold values are not quantitatively consistent between replicates
1 and 2, being 40 and 90, respectively. This can be accounted for
by considering the high sensitivity of the NSI approach to small differences
in the appearance of the ATD profiles. Identical ATDs would give NSI
values of zero, but in reality the ion mobility data, even for equivalent
ions, will have subtle differences that contribute to the NSI value.
These differences might arise from arrival time drift between acquiring
data for the RS19 peptide and the isomer peptide, possibly in response
to environmental changes in the laboratory. Such a drift could be
corrected either by CCS or “lock-drift” processing,
but the format of the data makes this less facile than direct arrival
time comparison. A second contributor to the threshold could be variations
in ion statistics. If, for example, a particular ATD has fewer ion
events to describe its shape, it may result in a larger NSI value
when compared to an ATD for the same ion constituting a greater number
of ion events. Such a situation might arise as a result of differences
in acquisition time between RS19 and the isomer or differences in
other instrumental conditions that may vary day-to-day. It is recommended,
therefore, that each experiment should be analyzed as an individual
data set and comparisons between replicates not be made, apart from
the inflection point indicating the site of isomerization, which is
consistent between replicates. It would not be appropriate to merge
for example, some fragments from one acquisition with those from a
separate acquisition of the same isomer.Second, significant
differences are observed between replicates
for some NSI values for the same terminating residues. An example
of this is exhibited most notably by ISOC y15 and y16 ions which in replicate 1 have running average NSI values
of 197 and 162 and in replicate 2 have values of 90 and 85, respectively
(Figure ). This phenomenon
is due to different product ion charge states in the different replicates
being chosen for NSI processing. In replicate 1, only singly charged
ions of y15 and y16 ions are compared
between RS19 and ISOC, whereas in replicate 2 the doubly charged ions
were chosen for these product ions (due to signal-to-noise considerations).
Singly charged ions will likely have drastically different structures
than doubly charged ions meaning the NSI values between any pair of
ions will not be equivalent. It should be noted that we always selected
the same ion to generate NSI values for each isomer within a replicate.When evaluating trends across each sequence to
identify the point
of a conformational change, the use of NSI data without taking the
running average along the fragment ion series, was considered. In
this case, a threshold was difficult to determine because past the
site of isomerization, the difference in NSI between isomer and RS19
decreased with increasing fragment length (Figure ). This can be rationalized by considering
that the optimal ion for differentiating fragments is when there are
sufficient residues after the β-amino acid to translate into
a conformational shift; but not too many residues such that the sensitivity
to conformational changes is lost. This is a delicate balance to maintain
as changes in conformation are relatively more significant in shorter
fragments than longer ones.Another significant point of note
was regarding the lipidated fragments.
Upon inspection of the two replicates for ISOA, a conformational change
between isomer and reference was still detected despite the fact that
the isomerized residue was directly adjacent to the lipidation site
at Lys10. It was thought that fragments containing the lipid may not
show any conformational differences arising from the isoAsp due to
the high degree of flexibility of the lipid chain. However, this was
not the case and for all three isomers, fragments that contained both
isoAsp and the lipid were differentiated from their corresponding
reference fragment. These results highlight the potential use of cIMS
for peptide pharmaceuticals which are often lipidated analogues or
conjugated to various polymers to optimize pharmacokinetics.[41,42]Therefore, combining MS2 with the structure-based
cIMS
described here, enabled samples of four individual isomeric peptides
to be identified from the ATDs of key fragment ions.
LC-CID-MS Analysis
of isoAsp Peptides
The characterization
of therapeutic peptides and their impurities often relies on liquid
chromatographic separation. However, retention time alone is not always
sufficient for unambiguously assigning impurities such as isoAsp.
While separation by LC is possible,[43,44] subtle changes
to mobile phases, e.g., pH, means the elution order from a mixed sample
is not always diagnostic without first confirming peak identity using
synthesized standards. There are, however, advantages to combining
the above cIMS workflow with LC, not least of all is the automatability
of LC as well as the potential for increased throughput for screening
of many samples. Liquid chromatographic separation of peptides, prior
to mobility separation and MS2 to generate an LC-CID-MS
workflow, would allow more complex mixtures to be analyzed in a single
experiment. Thus, determining if there were peptide impurities in
a sample (LC), whether they were isoAsp isomers or chemical degradation
products (MS) and also the location of the isomerized residue (CID-cIMS),
if present.The same NSI analysis was conducted for samples
of each peptide submitted to LC prior to CID-cIMS individually as
well as a sample of each peptide mixed (25% v/v) (Figure ). For one of the UPLC peaks
in the mixed sample (ISOA) there was the characteristic spike in the
running average NSI cIMS data for y fragments, as
one would expect for isomerization at Asp9 (Figure S10). Thus, ISOA was easily identified from the heterogeneous
sample. The retention time of RS19 was known, as would be the case
in a typical drug development program and by extension the remaining
choice of which peak was which between ISOB and ISOC (Figure ). Inspection of the y fragment data for the remaining two peaks was ambiguous,
as not all of the key fragments were observed. With pure samples of
each isomer available, comparison of retention times was used to decipher
peak 4 from peak 2, although, inspection of the running average NSI
values for peak 4 did give an inflection point where it would be expected
for isomerization at Asp15 residue. Despite this, after LC separation
it was not possible to unambiguously assign ISOC without observing
fragments in the range of y6–14.
Figure 6
TIC chromatograms
of peptide isomers in either a mixed sample (left)
which would not be unambiguously identified by MS2 alone
and the TIC chromatogram of samples individually submitted to LC (right)
which allowed correlation of cIMS fragment ion assignments to known
retention times for each isomer.
Figure 7
ATDs generated
from fragment ions of each peak in the LC chromatogram
corresponding to each peptide fragment y17 (left)
or y15 (right). After elution from the columns the
samples were subjected to CID fragmentation (45 V) followed by a single
pass around the cIMS T-wave array.
TIC chromatograms
of peptide isomers in either a mixed sample (left)
which would not be unambiguously identified by MS2 alone
and the TIC chromatogram of samples individually submitted to LC (right)
which allowed correlation of cIMS fragment ion assignments to known
retention times for each isomer.ATDs generated
from fragment ions of each peak in the LC chromatogram
corresponding to each peptide fragment y17 (left)
or y15 (right). After elution from the columns the
samples were subjected to CID fragmentation (45 V) followed by a single
pass around the cIMS T-wave array.However, differentiation was indeed possible by analysis of specific
fragment ion ATDs alone, rather than the running average method described
above. Inspection of fragment ion peaks for y15 and y17 allowed speculation as to which peak belonged to which
isomer as a result of the sensitivity of the cIMS instrument after
a single pass. If the peptide was isomerized at Asp9, then both y15 and y17 would match; this corresponds
to the red trace in Figure . If the isomerized peptide was isoAsp21, then it was expected
that both fragments would be different. and this was observed; see
the orange trace (peak 2). However, the green trace for peak 4 (Figure ) was seen to change
from matching at the y15 fragment length but was
considerably different for the y17 fragment. This
would be consistent with what was expected for isomerization at Asp15.
Therefore, while the fragmentation post LC requires optimization to
observe the entire y fragment series, there is significant
promise for using LC-CID-cIMS for the identification of unknown isoAsp
peptides from a sample containing a mixture of all isomers and the
nonisomerized reference standard for the therapeutic peptide of interest.
Conclusion
Despite their propensity for aggregation, peptides
play an increasingly
important role as therapeutics sitting between the pharmacokinetic
stereotypes of small molecules and large biomolecules. As such, analytical
platforms able to support these modalities are becoming increasingly
important. IMS technology has advanced to such an extent that even
conformational subtleties, like those imparted by amino acid isomerization,
can be measured. Thus, there is an opportunity to now use such techniques
to develop accurate and powerful methods to detect and characterize
even small differences in therapeutic peptides. As part of this process,
the search for simple yet accurate methods for comparison of ATD ion
mobility data from different peptides is of paramount importance.
In this study, we chose a mathematically facile approach to transition
from qualitative assessment of ATD differences to a more objective
and quantitative framework. Here, we have demonstrated that the high
resolution and sensitivity of cIMS, in conjunction with a simple normalization-subtraction-integration
(NSI) approach to data processing, has the ability to differentiate and identify the site of isomerization in a series of three
synthetic isobaric peptides based on an important therapeutic scaffold.
This approach has significant benefit over other conventional solution
ensemble techniques which cannot always identify these impurities
with sufficient speed or resolution. Movement toward higher resolution
techniques like cIMS will vastly improve our understanding of the
potential impact of degradants/impurities on the physical stability
of a peptide and factors affecting its shelf life. This capability
for conformation-based differentiation using cIMS together with mass
spectrometry is a major step forward for the characterization of chemically
similar and isobaric but structurally different biopharmaceutical
impurities, which is crucial for ensuring safety and quality control
of peptide therapeutics.
Authors: Hester A Doyle; Jing Zhou; Martin J Wolff; Bohdan P Harvey; Robert M Roman; Renelle J Gee; Raymond A Koski; Mark J Mamula Journal: J Biol Chem Date: 2006-09-01 Impact factor: 5.157
Authors: Peter B O'Connor; Jason J Cournoyer; Sharon J Pitteri; Paul A Chrisman; Scott A McLuckey Journal: J Am Soc Mass Spectrom Date: 2005-12-09 Impact factor: 3.109