| Literature DB >> 35607344 |
Justin Phan1,2, Rajesh Subbiah1,2,3, Bruce Walker1,2,3, William Lee1,2,3.
Abstract
Entities:
Keywords: Conduction disease; Giant cell myocarditis; Inappropriate shocks; Over-sensing; Subcutaneous ICD
Year: 2022 PMID: 35607344 PMCID: PMC9123316 DOI: 10.1016/j.hrcr.2022.03.002
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1a: Preimplant subcutaneous implantable cardioverter-defibrillator screening using the Automated Screening Tool with acceptable sensing using the primary vector. The QRS complexes demonstrate a good amplitude compared to the P and T waves. b: Repeat screening with subsequent failure in all vectors. There is now a lower QRS amplitude and a marked change in morphology. Electrocardiograms from the supine position are shown.
Figure 2a: Electrocardiogram (ECG) at subcutaneous implantable cardioverter-defibrillator (ICD) screening with sinus rhythm, inferolateral Q waves, and nonspecific ST-T abnormalities. b: Presenting ECG at the time of admission for inappropriate ICD shocks showing sinus rhythm with right bundle branch block and left anterior fascicular block.
Figure 3Interrogation of subcutaneous implantable cardioverter-defibrillator (ICD) demonstrating an inappropriate ICD shock. The arrow marker demonstrates under-sensing of the QRS complex. Diamond and star markers demonstrate P-wave and T-wave over-sensing, respectively. Note the low-amplitude QRS complexes relative to P wave and T waves. Repeated over-sensing led to fulfillment of the detection counter (T) and a subsequent inappropriate shock (lightning symbol).