Non-circular CBCT orbit design and realization on a clinical robotic C-arm for metal artifact reduction.

Yiqun Ma1, Grace J Gang1, Tina Ehtiati2, Tess Reynolds3, Tom Russ4, Wenying Wang1, Clifford Weiss1, Nicholas Theodore1, Kelvin Hong1, Jeffrey Siewerdsen1, J Webster Stayman1.   

Abstract

Metal artifacts have been a difficult challenge for cone-beam CT (CBCT), especially for intraoperative imaging. Metal surgical tools and implants are often present in the field of view and can attenuate X-rays so heavily that they essentially create a missing-data problem. Recently, an increasing number of intra-operative imaging systems such as robotic C-arms are capable of non-circular orbits for data acquisition. Such trajectories can potentially improve sampling and the degree of data completeness to solve the metal-induced missing-data problem, thereby reducing or eliminating the associated image artifacts. In this work, we extend our prior theoretical and experimental work and implement non-circular orbits for metal artifact reduction on a clinical robotic C-arm (Siemens Artis zeego). To maximize the potential for clinical translation, we restrict our implementation to standard built-in motion and data collection functions, also available on other zeego systems, and work within the physical constraints and limitations on positioning and motion. Customized software tools for data extraction, processing, calibration, and reconstruction are used. We demonstrate example non-circular orbits and the resulting image quality using a phantom containing pedicle screws for spine fixation. As compared with a standard circular CBCT orbit, these non-circular orbits exhibit significantly reduced metal artifacts. These results suggest a high potential for image quality improvements for intraoperative CBCT imaging when metal tools or implants are present in the field-of-view.

Entities:  

Keywords:  CBCT; image guidance; metal artifact; non-circular orbits; robotic C-arm; sampling

Year:  2022        PMID: 35599746      PMCID: PMC9119360          DOI: 10.1117/12.2612448

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  8 in total

1.  Normalized metal artifact reduction (NMAR) in computed tomography.

Authors:  Esther Meyer; Rainer Raupach; Michael Lell; Bernhard Schmidt; Marc Kachelriess
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

2.  Task-driven source-detector trajectories in cone-beam computed tomography: I. Theory and methods.

Authors:  J Webster Stayman; Sarah Capostagno; Grace J Gang; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-02

3.  Task-driven source-detector trajectories in cone-beam computed tomography: II. Application to neuroradiology.

Authors:  Sarah Capostagno; J Webster Stayman; Matthew Jacobson; Tina Ehtiati; Clifford R Weiss; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-09

4.  Reduction of CT artifacts caused by metallic implants.

Authors:  W A Kalender; R Hebel; J Ebersberger
Journal:  Radiology       Date:  1987-08       Impact factor: 11.105

5.  Ordered subsets algorithms for transmission tomography.

Authors:  H Erdogan; J A Fessler
Journal:  Phys Med Biol       Date:  1999-11       Impact factor: 3.609

6.  Self-calibration of cone-beam CT geometry using 3D-2D image registration.

Authors:  S Ouadah; J W Stayman; G J Gang; T Ehtiati; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2016-03-10       Impact factor: 3.609

7.  Non-circular CT orbit design for elimination of metal artifacts.

Authors:  Grace J Gang; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-03-16
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.