| Literature DB >> 35572423 |
Ofra A Pottorf1, Daniel J Lee1, Paul-Neil Czujko2.
Abstract
Background: Forearm pronation and supination are important for everyday functional tasks and some recreational activities. Healthcare providers use reliable and valid tools during the physical rehabilitation process to measure joint range of motion (ROM), assess functional mobility, guide decisions for skilled interventions, and progress a plan of care. Since the onset of the COVID-19 pandemic, both healthcare providers and patients benefited from mobile health technologies that have emerged, which can be used by patients in the home to monitor ROM and assist the healthcare provider in guiding the rehabilitation process when utilizing telehealth. Purpose: The goal of this study was to investigate the reliability and concurrent validity of a smartphone application for obtaining goniometric measurements of forearm pronation and supination.Entities:
Keywords: COVID-19; Mobile health; Rehabilitation; Reliability; Self-monitoring; Smartphone; Validity
Year: 2022 PMID: 35572423 PMCID: PMC9091929 DOI: 10.1016/j.jseint.2022.02.002
Source DB: PubMed Journal: JSES Int ISSN: 2666-6383
Figure 1(A) Clinometer application on the iPhone, (B) Standard half-circle goniometer used to measure forearm ROM.
Figure 2(A) Calibration procedure between subjects, (B) Blinding procedure for testers.
Comparison of smartphone and standard goniometer for pronation and supination with ICC at 95% and SEM.
| Standard goniometer (degrees) | iPhone (degrees) | |
|---|---|---|
| Pronation (mean) | 77.5 (5.5) | 81.5 (6.5) |
| Supination (mean) | 79.6 (5.3) | 81.9 (6.9) |
ICC, intraclass correlation coefficient; SEM, standard error of measurement.
Figure 3Relationship for (A) pronation on the smartphone and standard goniometer. (r = 0.71, P < .001) and (B) Supination on the smartphone and standard goniometer (r = 0.71, P < .001).
Intrarater reliability between the smartphone and standard goniometer for pronation and supination shown as ICC, 95% confidence interval.
| Goniometer ICC (95% CI) | iPhone ICC (95% CI) | |
|---|---|---|
| Pronation | ||
| Tester 1 | ICC = 0.68 (0.50-0.81) | ICC = 0.76 (0.61-0.85) |
| Tester 2 | ICC = 0.39 (0.06-0.64) | ICC = 0.66 (0.42-0.82) |
| Supination | ||
| Tester 1 | ICC = 0.64 (0.44-0.78) | ICC = 0.77 (0.63-0.86) |
| Tester 2 | ICC = 0.80 (0.63-0.90) | ICC = 0.60 (0.33-0.78) |
ICC, intraclass correlation coefficient; CI, confidence interval.