Carla R Lamb1,2, Ekaterina Yavarovich3, Vincent Kang3, Elliot L Servais4, Lori B Sheehan5,6, Sara Shadchehr3,6, James Weldon7, Matthew J Rousseau7, Gregory P Tirrell7.
Abstract
BACKGROUND: Single-use flexible bronchoscopes eliminate cross contamination from reusable bronchoscopes and are cost-effective in a number of clinical settings. The present bench study aimed to compare the performance of a new single-use bronchoscope (Boston Scientific EXALT Model B) to a marketed single-use comparator (Ambu aScope 4), each in slim, regular and large diameters.Entities:
Keywords: Bronchoscopy; Disposable equipment; Interventional pulmonology; Medical technology
Mesh:
Substances:
Year: 2022 PMID: 35550062 PMCID: PMC9095814 DOI: 10.1186/s12890-022-01982-4
Source DB: PubMed Journal: BMC Pulm Med ISSN: 1471-2466 Impact factor: 3.320
Technical specifications of the tested devices
| Attribute | Slim | Regular | Large | |||
|---|---|---|---|---|---|---|
| aScope 4a | EXALT Model B | aScope 4a | EXALT Model B | aScope 4a | EXALT Model B | |
| Field of view | 85° | 90° | 85° | 90° | 85° | 90° |
| Articulation angle | 180° up, 180° down | 180° up, 180° down | 180° up, 180° down | 180° up, 180° down | 180° up, 160° down | 180° up, 180° down |
| Insertion tube outer diameter (OD) | 3.8 mm | 3.8 mm | 5.0 mm | 5.0 mm | 5.8 mm | 5.8 mm |
| Working length | 60 cm | 60 cm | 60 cm | 60 cm | 60 cm | 60 cm |
| Average working channel diameter | 1.2 mm | 1.2 mm | 2.2 mm | 2.2 mm | 2.8 mm | 2.8 mm |
| Minimum working channel diameter | 1.2 mm | 1.0 mm | 2.0 mm | 2.0 mm | 2.6 mm | 2.6 mm |
| Minimum endotracheal tube size compatibility | 5.0 mm | 5.0 mm | 6.0 mm | 6.0 mm | 7.0 mm | 7.0 mm |
| Minimum double lumen tube size compatibility | 35Fr | 35Fr | 41Fr | 41Fr | N/A | N/A |
aFrom: https://www.ambu.com/endoscopy/pulmonology/bronchoscopes/product/ambu-ascope-4-slim
https://www.ambu.com/endoscopy/pulmonology/bronchoscopes/product/ambu-ascope-4-regular
https://www.ambu.com/endoscopy/pulmonology/bronchoscopes/product/ambu-ascope-4-large
Accessed on October 29, 2021
Fig. 1EXALT Model B single-use bronchoscope tested in the current study. Yellow arrow shows the working channel diameter
Fig. 2Bronchoscopy bench simulation. Investigators clear a “mucus” plug (a) and view “bronchi” (b and c) during bench model simulation
Primary endpoint: “mucus” mass suctioned in 30 s in Task 2 (N = 15)
| Comparison of aScope 4 size versus EXALT Model B size | Mean mass of “mucus” suctioned in 30 s (g) | ||
|---|---|---|---|
| aScope 4 | EXALT Model B | ||
| Large versus Regulara | 41.5 ± 5.7 | 41.8 ± 8.3 | 0.914 |
| Slim versus Slim | 1.5 ± 0.6b | 4.3 ± 1.3 | < .001 |
| Regular versus Regular | 15.4 ± 2.9 | 41.8 ± 8.3 | < .001 |
| Large versus Large | 41.5 ± 5.7 | 72.9 ± 9.0 | < .001 |
aBased on a preliminary experiment, the study hypothesis was that the EXALT Model B regular scope would suction significantly more “mucus” than the aScope 4 large scope (not confirmed)
bThere were 14 observations for aScope 4 Slim because one physician was called away for patient care and could not complete this task
“Mucus” plug clearance in Task 3 (N = 15)
| Task 3 measure | Bronchoscope model | ||
|---|---|---|---|
| aScope 4 | EXALT Model B | ||
| Median percentage of mucus plug cleared | |||
| Slim | 10.0 (0.0–100.0)a | 45.0 (5.0–100.0) | 0.043 |
| Regular | 80.0 (5.0–100.0) | 95.0 (60.0–100.0) | 0.095 |
| Large | 95.0 (50.0–100.0) | 99.0 (80.0–100.0) | 0.066 |
| Mucus plug cleared | |||
| Slim | 7.1% (1/14)a | 13.3% (2/15) | 0.999 |
| Regular | 60.0% (9/15) | 93.3% (14/15) | 0.080 |
| Large | 93.3% (14/15) | 100.0% (15/15) | 0.999 |
| Median completion time (seconds) | |||
| Slim | 60.0 (60.0–60.0)a | 60.0 (30.0–60.0) | 0.164 |
| Regular | 55.0 (30.0–60.0) | 42.0 (13.0–60.0) | 0.002 |
| Large | 30.0 (16.0–60.0) | 23.0 (10.0–45.0) | 0.005 |
| Mean number of passes | |||
| Slim | 2.0 ± 0.0a | 1.9 ± 0.3 | 0.899 |
| Regular | 1.9 ± 0.4 | 1.6 ± 0.5 | 0.584 |
| Large | 1.5 ± 0.5 | 1.2 ± 0.4 | 0.532 |
a14 observations for aScope 4 Slim after one physician was called away for patient care and could not complete task
Visualization ratings (N = 15)
| Attribute | Median rating (range) | ||
|---|---|---|---|
| aScope 4 / aView 2 Advance monitor | EXALT model B/EXALT monitor | ||
| Definition | 8.0 (7.0−10.0) | 9.0 (8.0−10.0) | 0.001 |
| Color | 8.0 (7.0−10.0) | 9.0 (7.0−10.0) | 0.009 |
| White balance | 8.0 (6.0−10.0) | 9.0 (7.0−10.0) | 0.029 |
| Far field resolution | 7.0 (6.0−10.0) | 9.0 (7.0−10.0) | 0.001 |
| Near field resolution | 9.0 (7.0−10.0) | 10.0 (8.0−10.0) | 0.018 |
Median performance ratings for “mucus” suction
| Peformance characteristic | Median rating (range)a | ||
|---|---|---|---|
| aScope 4 | Model B | ||
| Ease of use | |||
| Slim | 10.0 (7.0−10.0)b | 10.0 (6.0−10.0) | 0.530 |
| Regular | 10.0 (8.0−10.0) | 10.0 (8.0−10.0) | 0.562 |
| Large | 9.0 (7.0−10.0) | 10.0 (8.0−10.0) | 0.195 |
| Perception of efficacy | |||
| Slim | 1.0 (1.0−9.0)b | 2.0 (1.0−7.0) | 0.045 |
| Regular | 5.0 (3.0−8.0) | 8.0 (5.0−10.0) | < .001 |
| Large | 7.0 (3.0−10.0) | 9.0 (4.0−10.0) | 0.237 |
| Operator comfort | |||
| Slim | 9.5 (7.0−10.0)b | 10.0 (8.0−10.0) | 0.425 |
| Regular | 10.0 (7.0−10.0) | 10.0 (8.0−10.0) | 0.600 |
| Large | 9.0 (6.0−10.0) | 10.0 (6.0−10.0) | 0.469 |
aRated on a scale of one (worst) to 10 (best)
b14 observations for aScope 4 Slim after one physician was called away for patient care and could not complete task