Highly selective and sensitive fluorescent probe for the rapid detection of mercury ions.

Zujun Xu1, Wenxiu Shi1, Chengjun Yang1, Jing Xu2, Huapeng Liu1, Jing Xu2, Baocun Zhu2.   

Abstract

Mercury (Hg) is one of the major toxic heavy metals, harmful to the environment and human health. Thus, it is significantly important to find an easy and quick method to detect Hg2+. In this study, we designed and synthesized a simple fluorescent probe with excellent properties, such as high sensitivity and selectivity, rapid response, and outstanding water solubility. When Hg2+ (5 μM) was added to the probe solution, it exhibited a very large fluorescent enhancement (about 350-fold stronger than the free probe) with the help of hydrogen peroxide (H2O2). Probe HCDC could quantitatively detect Hg2+ in the range of 0-10 μM using the fluorescence spectroscopy method and the detection limit was measured to be about 0.3 nM (based on a 3σ/slope). Analytical application was also studied, and the probe HCDC exhibited excellent response to Hg2+ with the addition of H2O2 in real water samples. So, our proposed probe HCDC provided a practical and promising method for determining Hg2+ in the environment. This journal is © The Royal Society of Chemistry.

Entities:  

Year:  2019        PMID: 35515305      PMCID: PMC9062531          DOI: 10.1039/c9ra00622b

Source DB:  PubMed          Journal:  RSC Adv        ISSN: 2046-2069            Impact factor:   3.361


Introduction

Heavy metal ion pollution has become a global environmental problem.[1-3] As a common metal element, mercury (Hg) is extremely toxic and shows a great threat to the environment and human health due to its bioaccumulation, durability, and easy absorption by aquatic organisms.[4-10] Once the mercury enters the organism, it will lead to the dysfunctions of the central nervous systems, the brain, and the kidneys.[11-14] As we all know, the maximum contaminant level of mercury in drinking water is 2 ppb (0.01 μM) which was set by the Environment Protection Agency (EPA).[15-18] It can be seen that mercury could pose a serious threat to humans at very low concentrations. Thus, developing a convenient, rapid, highly selective and sensitive method for the determination of Hg2+ is vitally important. Nowadays, several classical methods have been applied to determine Hg2+ including electrochemical analysis, atomic absorption/emission spectroscopy, electrochemical analysis, and inductively coupled plasma mass spectrometry.[19-23] Though these methods are selective and accurate to detect ions, most of them require long time, expensive instruments, and tedious sample pre-treatment.[24-26] Thus, the techniques of fluorescent probes were widely studied due to the advantages of highly selective, operational simplicity, and relatively low-cost.[27-33] Many fluorescent probes for detecting Hg2+ have been reported; however, most of them were still limited by the disadvantages of low sensitivity, bad water solubility, and long response time.[34-38] So, a simple water-soluble fluorescent probe for determining Hg2+ with high sensitivity and quick response became our primary task. Very recently, we have reported several carbonothioate-based fluorescent probe for detecting Hg2+. These probes respectively chose the 2-(2′-hydroxyphenyl)benzothiazole (HBT),[39] dichlororesorufin,[40] and seminaphthorhodafluor[41] as the fluorophores, and a carbonothioate moiety was used as the recognition receptor. All these showed quick response and high selectivity for detecting Hg2+. In connection with our continuing research, we designed a new probe HCDC (Scheme 1) for detecting Hg2+ with superior properties in this paper, and the property of probe HCDC was compared with other reported studies in Table 1. A dimethyl-thiocarbamic ester group and 7-hydroxycoumarin were respectively chosen as the recognition receptor and the fluorophore. We assumed that the thioester group changed to ester moiety in the presence of Hg2+, and the rapid cleavage of ester group was achieved with the help of H2O2. As a result, this restored the blue fluorescence of 7-hydroxycoumarin. The experimental results demonstrated that this probe possessed the advantages of (1) a very low detection limit of 0.3 nM (on the basis of 3σ/slope); (2) quicker response to Hg2+; (3) excellent water solubility, selectivity and anti-interference; (4) practical applicability in real water samples.
Scheme 1

Synthesis of probe HCDC.

Comparison of Fluorescent Probes for Hg2+

ProbeLOD (nM)Solution (v/v)Time (min)References
1.8Ethanol : PBS = 2 : 130 34
10C2H5OH : water = 1 : 140 35
34CH3CN : water = 1 : 1 36
18.8C2H5OH : water = 2 : 820 37
100DMSO : H2O = 9 : 130 38
0.3Aqueous solution10This work

Experimental

Materials and general methods

All the chemicals were gained from commercial suppliers and were used without further purification. The Sartorius Arium 611DI system can produce the ultra-pure water which was used during the entire process of the experiment. Column chromatography using Silica gel 200–300 mesh from the Qingdao Haiyang Chemical Co was used to purify the reaction product. The 1H NMR and 13C NMR were all recorded on the Bruker AV-400 spectrometer with chemical shifts reported as ppm (in DMSO-d6, TMS as internal standard). Fluorescence spectra were gained from a Horiba FluoroMax-4 spectrophotometer. All the fluorescence spectra were uncorrected. Probe HCDC was prepared according to the previous paper.[42]

General procedure for analysis

Stock solutions of probe HCDC (1 mM) were prepared in ethanol. And the parent stock solutions (1 mM) of Hg2+, other metal ions and common anions species (10 mM) were prepared in ultrapure water. The test solution contained 50 μL of the probe stock solution and moderate amount of each stock solution, and eventually diluted to 10 mL in a test tube with 100% ultrapure water containing HEPES (5 mM, pH 7.4). Before Hg2+ and other ions were added to the test solution, H2O2 (3 mM) was added. All the tests were conducted at room temperature (25 °C).

Determination of the detection limit

According to the previous paper,[43,44] the method of fluorescence titration was used to calculate the detection limit. The fluorescence emission spectrum of probe HCDC was measured by fifteen times in the absence of Hg2+ in order to achieve the standard deviation of blank measurement. To gain the slope, the fluorescence intensity at 455 nm was plotted as a concentration of Hg2+. Therefore, the formula for calculating the detection limit is described as following:Detection limit = 3where σ is the standard deviation of the five spectrums among the fifteen spectrums which peaks were nearest, k is the slope between the Hg2+ concentration versus the fluorescence intensity.

Analytical application

Three water samples were chosen to test the analytical application of the probe HCDC. One water sample was collected from Jia Zi Lake, University of Jinan, China, and the other two samples were respectively collected from the Xiu Jiang River and Yellow River, at Jinan, China. All the samples needed to filter through filter paper before the measurement.

Results and discussion

Characteristic spectra

In this paper, the fluorescence and absorption spectra of probe HCDC was shown as Fig. 1. In the absence of Hg2+ and H2O2, the probe solution exhibits a very weak fluorescence emission peak at 455 nm. When only Hg2+ was added to the probe solution, the reaction was very slow. And, the reaction was not carried out when only the H2O2 was added. Once Hg2+ and H2O2 were all added to the probe HCDC solution, the maximum fluorescence emission peak exhibits a very large enhancement at 455 nm (Fig. 1a). The fluorescence intensity peak of the probe solution containing Hg2+ and H2O2 was about 350-fold stronger than that of the blank probe solution. Meanwhile, in the absorption spectra, with the addition of Hg2+ and H2O2, the probe solution exhibited a new absorption peak at 364 nm (Fig. 1b). The result showed that Hg2+ could generate the cleavage of an eater group in the presence of H2O2 (Scheme 2).[40]
Fig. 1

Fluorescence spectra (a) and absorption spectra (b) of probe HCDC (5 μM) toward Hg2+ (5 μM) and H2O2 (3 mM) under the ultra-pure water containing HEPES (5 mM, pH 7.4).

Scheme 2

Reaction Mechanism of probe HCDC for Hg2+ with the help of H2O2.

Effect of H2O2

Different concentrations of H2O2 was respectively added to the solution of probe HCDC (5 μM) and Hg2+ (5 μM) in the aqueous solution which contained HEPES (5 mM, pH 7.4). It was shown in inset of Fig. 2, the fluorescence intensity was increased first and then decreased with the increasing concentrations of H2O2 (the range of 1 to 40 mM). The enhancement of fluorescence reached the maximum when the concentration of H2O2 was 3 mM. The results implied that the low concentration of H2O2 (1 to 3 mM) could promote the cleavage of ester group and amplify the response of probe HCDC to Hg2+. But when the concentration is over 3 mM (4 to 40 mM), H2O2 will inhibit the response of probe HCDC to Hg2+. Thus, we chose 3 mM H2O2 as the amplification reagent in the detection of Hg2+.
Fig. 2

Fluorescence spectra of probe HCDC (5 μM) in the presence of Hg2+ (5 μM) and different concentrations of H2O2 (final concentration: 0, 1, 2, 3, 4, 5, 10, 15, 20, 25, 30, 35, 40 mM) under HEPES (5 mM, pH 7.4) aqueous solution. Excitation wavelength = 330 nm. Error bar = RSD (n = 3).

Time dependence of detecting Hg2+

Response time is a significant element to test the effect of the probe. So, the response time of probe HCDC was investigated under the analytical conditions which were mentioned above. The result was shown in Fig. 3. When only Hg2+ was added to the probe solution, the reaction was very slow. While, the fluorescence intensity at 455 nm was increased as soon as Hg2+ and H2O2 were added, and increased gradually until the reaction time exceeded 10 min. Thus, all the measurements of the spectrum were carried out 10 minutes after the addition of Hg2+. The results showed that the probe HCDC could provide a rapid analytical method for detecting Hg2+.
Fig. 3

Reaction time for probe HCDC (5 μM) with the addition of only Hg2+ (5 μM) (a), and Hg2+ (5 μM) and H2O2 (3 mM) (b).

Quantification of Hg2+

As shown in Fig. 4, the continuous enhancement of fluorescence intensity at 455 nm can be seen with the continuous addition of Hg2+ (the range of 0 to 10 μM). And, there was a good linearity (y = 225439x + 3899.2, R2 = 0.998) between the Hg2+ concentrations of 0–2 μM and the fluorescence intensity with a detection limit of 0.3 nM (based on 3σ/slope). The detection limit of the probe HCDC was far lower than the standard of Hg2+ (10 nM) in drinking water which set up by the Environmental Protection Agency (EPA). The above results showed that the probe HCDC could detect Hg2+ quantitatively by the fluorescence spectrometry method with excellent sensitivity.
Fig. 4

Fluorescence spectra of probe HCDC (5 μM) toward the various concentrations of Hg2+ (final concentration: 0, 0.2, 0.4, 0.6, 0.8, 1, 1.5, 2, 2.5, 3, 3.5, 4, 4.5, 5, 6, 7, 8, 9, 10 μM) and H2O2 (3 mM). The fluorescence intensity at 455 nm of probe HCDC increased with the increasing concentrations of Hg2+. Excitation wavelength = 330 nm, emission wavelength = 455 nm. Conditions: in HEPES (5 mM, pH 7.4). Solution: ultrapure water. Error bar = RSD (n = 3).

Selectivity to Hg2+

High selectivity is also an important factor for an excellent probe. The selectivity of HCDC toward Hg2+ was evaluated under the above-mentioned conditions. The effect of the metal ions (Mg2+, Zn2+, Ni2+, Ca2+, Na+, K+, Pb2+, Cd2+, Cr6+, Fe2+, Fe3+, Sn2+) and the common anions species (NO2−, NO3−, HCO3−, H2PO4−, F−, CO32−, Cl−, SO42−) on fluorescence spectra of probe HCDC were investigated. Firstly, Hg2+, other metal ions and common anions species were respectively added to the solution, and only Hg2+ resulted in a considerable fluorescence enhancement (Fig. 5a and c). The results exhibited that HCDC possesses high selectivity toward Hg2+.
Fig. 5

(a) Fluorescence responses of probe HCDC (5 μM) toward Hg2+ (5 μM), and other metal ions (100 μM). (b) Fluorescence responses of probe HCDC toward Hg2+ and in presence of other metal ions. (1) Hg2+; (2) Mg2+; (3) Zn2+; (4) Sn2+; (5) Ca2+; (6) Na+; (7) Fe3+; (8) Pb2+; (9) Cd2+; (10) Fe2+; (11) K+; (12) Ni2+. (c) Fluorescence responses of HCDC (5 μM) toward Hg2+ (5 μM), and common anions species (100 μM). (d) Fluorescence responses of probe HCDC toward Hg2+ and in presence of other common anions species (1) Hg2+; (2) NO2−; (3) NO3−; (4) HCO3−; (5) H2PO4−; (6) F−; (7) CO32−; (8) Cl−; (9) SO42−. The solution: in H2O2 (3 mM), HEPES (5 mM, pH 7.4) aqueous solution. Error bar = RSD (n = 3).

The interference of the metal ions and common anions species on the detection of Hg2+ was also studied. Even though the mental ions and common anions species were respectively existed in the solution, the probe HCDC also had a good response to Hg2+ (Fig. 5b and d). These studies clearly indicated that the probe HCDC could be used for the selective detection of Hg2+ without interference from anions. Based on the above results, the probe HCDC owned the excellent properties containing a very low detection limit, quicker response, excellent water solubility, selectivity and anti-interference. So, we attempted to investigate the practical application of the probe for the selective detection of Hg2+ in three real water samples. The results were shown in Table 2. At first, there were no Hg2+ in three water samples, and then 2 μM, 5 μM, or 10 μM Hg2+ was respectively added to the solution of real water (5 mM, HEPES, pH 7.4) containing 5 μM probe HCDC and 3 mM H2O2. Every sample was repeatedly configured and tested for 3 times. Therefore, we obtained a good method for detecting Hg2+ in the environment.

Analysis results of Hg2+ in three real water samples a

Real water sampleFound Hg2+Addition Hg2+(μM)Found (μM)Recovery (%)RSD (n = 3) (%)
Sample ANo21.82 ± 0.0691.062.89
54.52 ± 0.1190.472.10
109.61 ± 0.4096.064.02
Sample BNo21.85 ± 0.1192.525.36
54.97 ± 0.2999.335.83
1010.11 ± 0.25101.112.51
Sample CNo22.24 ± 0.13111.936.56
55.15 ± 0.21102.964.19
109.73 ± 0.3797.283.72

Sample A, lake water from JiaZi Lake, University of Jinan; sample B, from the Xiu Jiang River, at Jinan, China; sample C, from the Yellow River, at Jinan, China.

Sample A, lake water from JiaZi Lake, University of Jinan; sample B, from the Xiu Jiang River, at Jinan, China; sample C, from the Yellow River, at Jinan, China.

Conclusion

In a word, we have designed and synthesized a highly sensitive and selective probe for detecting Hg2+ by the fluorescence spectrum. The enhancement of fluorescence intensity of the probe solution containing Hg2+ and H2O2 was very significant, which could reach 350-fold stronger than that of the blank probe solution. There was a good linear relationship in the range of 0–2 μM Hg2+, and the low detection limit was 0.3 nM (on the basis of 3σ/slope). The probe showed excellent solubility in water. More importantly, the response of probe HCDC to Hg2+ in real water samples was also satisfactory. All these remarkable advantages of this new probe suggest that this probe will have excellent application prospect in the detection of mercury.

Conflicts of interest

There are no conflicts of interest to declare.
  21 in total

1.  Highly sensitive and selective turn-on fluorescent chemosensors for Hg2+ based on thioacetal modified pyrene.

Authors:  Yunyan Gao; Tuotuo Ma; Zhize Ou; Wenjiao Cai; Guoqiang Yang; Yi Li; Moheng Xu; Qingqing Li
Journal:  Talanta       Date:  2017-10-03       Impact factor: 6.057

2.  A multi-signal mitochondria-targeted fluorescent probe for real-time visualization of cysteine metabolism in living cells and animals.

Authors:  Xiaopeng Yang; Wenya Liu; Jun Tang; Ping Li; Haibo Weng; Yong Ye; Ming Xian; Bo Tang; Yufen Zhao
Journal:  Chem Commun (Camb)       Date:  2018-10-09       Impact factor: 6.222

3.  Nanomolar Hg2+ Detection Using β-Lactoglobulin-Stabilized Fluorescent Gold Nanoclusters in Beverage and Biological Media.

Authors:  Jiachen Zang; Changan Li; Kai Zhou; Haisheng Dong; Bin Chen; Fudi Wang; Guanghua Zhao
Journal:  Anal Chem       Date:  2016-09-28       Impact factor: 6.986

4.  A highly selective and ultrasensitive ratiometric fluorescent probe for peroxynitrite and its two-photon bioimaging applications.

Authors:  Zuokai Wang; Liu Wu; Yawei Wang; Meng Zhang; Ziyang Zhao; Caiyun Liu; Qingxia Duan; Pan Jia; Baocun Zhu
Journal:  Anal Chim Acta       Date:  2018-05-26       Impact factor: 6.558

5.  A Switch-On NIR Probe for Specific Detection of Hg2+ Ion in Aqueous Medium and in Mitochondria.

Authors:  Hridesh Agarwalla; Pankaj S Mahajan; Debashis Sahu; Nandaraj Taye; Bishwajit Ganguly; Santosh B Mhaske; Samit Chattopadhyay; Amitava Das
Journal:  Inorg Chem       Date:  2016-11-10       Impact factor: 5.165

6.  Thymine Functionalized Graphene Oxide for Fluorescence "Turn-off-on" Sensing of Hg2+ and I- in Aqueous Medium.

Authors:  Diptiman Dinda; Bikash Kumar Shaw; Shyamal Kumar Saha
Journal:  ACS Appl Mater Interfaces       Date:  2015-07-02       Impact factor: 9.229

7.  Nitrogen-doped carbon quantum dots: facile synthesis and application as a "turn-off" fluorescent probe for detection of Hg2+ ions.

Authors:  Ruizhong Zhang; Wei Chen
Journal:  Biosens Bioelectron       Date:  2013-12-10       Impact factor: 10.618

8.  Rational Design of a Hepatoma-Specific Fluorescent Probe for HOCl and Its Bioimaging Applications in Living HepG2 Cells.

Authors:  Qingxia Duan; Pan Jia; Zihan Zhuang; Caiyun Liu; Xue Zhang; Zuokai Wang; Wenlong Sheng; Zilu Li; Hanchuang Zhu; Baocun Zhu; Xiaoling Zhang
Journal:  Anal Chem       Date:  2019-01-10       Impact factor: 6.986

9.  A highly specific and ultrasensitive near-infrared fluorescent probe for imaging basal hypochlorite in the mitochondria of living cells.

Authors:  Baocun Zhu; Liu Wu; Meng Zhang; Yawei Wang; Caiyun Liu; Zuokai Wang; Qingxia Duan; Pan Jia
Journal:  Biosens Bioelectron       Date:  2018-02-09       Impact factor: 10.618

10.  A mitochondria-targeted ratiometric fluorescent probe for rapid, sensitive and specific detection of biological SO2 derivatives in living cells.

Authors:  Junchao Xu; Jian Pan; Xiaoming Jiang; Caiqin Qin; Lintao Zeng; Huan Zhang; Jun Feng Zhang
Journal:  Biosens Bioelectron       Date:  2015-10-19       Impact factor: 10.618

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.