Ho Seok Chung1, Jinho Lee2, Hun Lee3, Jae Yong Kim3, Hungwon Tchah3.
Abstract
Purpose: In femtosecond laser-assisted cataract surgery (FLACS), capsulorhexis can be performed with an ophthalmic viscosurgical device (OVD) filled in the anterior chamber. We aimed to investigate changes in the laser properties in various optical aspects, such as focal shifting, reflection, and absorption associated with OVD.Entities:
Mesh:
Year: 2022 PMID: 35506930 PMCID: PMC9078068 DOI: 10.1167/tvst.11.5.2
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.Schematic diagram of measurement of laser power and dimensions for each part.
Refractive Index of Water and Ophthalmic Viscosurgical Devices
| Materials | Refractive Index |
|---|---|
| Healon | 1.336 |
| Healon GV | 1.338 |
| Healon 5 | 1.338 |
| Discovisc | 1.340 |
| Cornea | 1.376 |
| Aqueous humor | 1.336 |
| Lens: Cortex | 1.386 |
Provided by the manufacturer.
Measured Laser Power and Absorption Coefficient of Water and the Ophthalmic Viscosurgical Devices Depending on the Thickness of Materials
| 1.5 mm | 3 mm | ||||
|---|---|---|---|---|---|
| Measured Laser Power (µJ) | Absorption Coefficient (mm−1) | Measured Laser Power (µJ) | Absorption Coefficient (mm−1) | Average Absorption Coefficient | |
| Water | 0.918 ± 0.004 | 0.0354 ± 0.0064 | 0.896 ± 0.005 | 0.0258 ± 0.0032 | 0.0306 |
| Healon | 0.888 ± 0.004 | 0.0575 ± 0.0064 | 0.780 ± 0.000 | 0.0720± 0.0034 | 0.0647 |
| Healon GV | 0.898 ± 0.004 | 0.0500 ± 0.0064 | 0.808 ± 0.004 | 0.0620± 0.0034 | 0.0551 |
| Healon 5 | 0.868 ± 0.004 | 0.0727 ± 0.0065 | 0.778 ± 0.004 | 0.0728± 0.0034 | 0.0728 |
| Discovisc | 0.882 ± 0.004 | 0.0620 ± 0.0065 | 0.772 ± 0.004 | 0.0754± 0.0034 | 0.0687 |
Figure 2.(A) Reflectance at normal incidence according to the refractive indices of various ophthalmic viscosurgical devices. (B) Reflectance of perpendicular (R⊥) and parallel (R∥) to the incident plane of laser, according to the various incident angles for aqueous humor and Discovisc.
Figure 3.(A) Beam diameter at the focus. (B) Laser power variation according to the refractive indices.
Figure 4.(A) Measured absorption coefficients of various ophthalmic viscosurgical devices and the water (red line). (B) Input laser power ratio to achieve the same power intensity as aqueous humor.