Naichang Yu1, Anthony Magnelli1, Danielle LaHurd1, Anthony Mastroianni1, Eric Murray1, Mike Close1, Brian Hugebeck1, John H Suh1, Ping Xia1.
Abstract
PURPOSE/Entities:
Keywords: automation; external beam radiotherapy; patient safety; treatment couch overrides; treatment positioning errors
Mesh:
Year: 2022 PMID: 35506575 PMCID: PMC9278683 DOI: 10.1002/acm2.13629
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
Typical Immobilization devices (manufacturers) used in our department
| Immobilization device | Clinical examples | Manufacture |
|---|---|---|
| Routine immobilization system | HN, breast, pelvis, thorax, prostate | Orfit, Jericho, NY |
| Abdominal compression (with full‐body bag below) | SBRT lung, SBRT liver | CIVCO, Orange City, IA |
| Full body bag | SBRT spine, SBRT pancreas SBRT prostate | BodyFix, Elekta, Stockholm, Sweden |
FIGURE 1A flowchart of patient setup steps for new patients, patients with and without IGRT. IGRT is typically referred to kV‐CBCT. IGRT, imaging guided radiotherapy; kV‐CBCT, kilo‐voltage cone‐beam CT
FIGURE 2Four quality control metrics that measure quality controls of treatment delivery
Couch position tolerance limits based on treatment sites
| Name | Lateral (cm) | Longitudinal (cm) | Vertical (cm) | Angle | Pedestal | Clinical examples |
|---|---|---|---|---|---|---|
| HN | 1.5 | 1.5 | 1.0 | Brain immobilized with mask | ||
| HN‐CBCT | 1.5 | 1.0 | 1.0 | 3 | 0.1 | HN |
| Body | 3.0 | 3.0 | 1.5 | 1.0 | 0.1 | Breast, electron plans based on CT |
| Body‐CBCT | 3.0 | 3.0 | 1.0 | 3 | 0.1 | Thoracic, abdominal, and prostate cases |
| SBRT‐body | 2.0 | 2.0 | 1.5 | 3 | 0.1 | Lung, liver, spine Mets |
| Electron‐clinical setup | 10.0 | 20.0 | 3.0 | Clinical setup without CT planning |
Abbreviations: HN‐CBCT, head and neck cone‐beam CT; SBRT, stereotactic body radiotherapy.
FIGURE 3The frequency of immobilization devices used from 2012 to 2018
FIGURE 4The ranges and frequencies of couch position tolerance limits (in cm) applied in the (a) lateral, (b) longitudinal, and (c) vertical directions from 2012 to 2018
FIGURE 5The annual number of overrides and the number of positioning errors from 2012 to 2018
FIGURE 6The recorded table shifts after kV‐CBCT verifications for the treatment site of (a) HN‐CBCT; (b) body‐CBCT; (c) SBRT‐body, where the corresponding couch position tolerance was applied. HN‐CBCT, head and neck cone‐beam CT; kV‐CBCT, kilovoltage cone‐beam CT; SBRT, stereotactic body radiotherapy
Comparison of our tolerance tables and proposed new tolerance tables (in parenthesis) with three other published tolerance tables
| Lateral (cm) | Longitudinal (cm) | Vertical (cm) | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Name | CCF | ULa | UMb | Surveyc | CCF | UL | UM | Survey | CCF | UL | UM | Survey |
| HN | 1.5 | 0.6 | 1 | 1.5 | 0.7 | 1 | 1.0 | 0.7 | 1 | |||
| HN‐CBCT | 1.5 | 0.9 | 0.9 | 1 | 1.0 | 0.9 | 0.9 | 1 | 1.0 | 0.5 | 0.7 | 1 |
| Body | 3.0 | 3.5 | 2.7 | >5 | 3.0 | 3.8 | 2.5 | >5 | 1.5 | 0.9 | 0.7 | 1 |
| Body‐CBCT |
3.0 (1.5) | >5 |
3.0 (1.5) | >5 |
1.0 (1.5) | 1 | ||||||
| SBRT‐body |
2.0 (1.5) | 0.3 |
2.0 (1.5) | 0.3 |
1.5 (1.5) | 0.3 | ||||||
Abbreviation: CCF, cleveland clinic foundation; HN‐CBCT, head and neck cone‐beam CT; SBRT, stereotactic body radiotherapy; UL, University of Louisville; UM, University of Michigan.
See Ref. [15].
See Ref. [16].
See Ref. [17].