Literature DB >> 35502628

Evaluation of electrophysiological characteristics and ventricular synchrony: An intrapatient-controlled study during His-Purkinje conduction system pacing versus right ventricular pacing.

Xueying Chen1, Xiaolan Zhou2, Yanan Wang3, Qinchun Jin1, Yufei Chen3, Jingfeng Wang1, Shengmei Qin1, Jin Bai1, Wei Wang1, Yixiu Liang1, Haiyan Chen3, Yangang Su1, Junbo Ge1.   

Abstract

OBJECTIVES TO
BACKGROUND: To compare electromechanical ventricular synchrony when pacing from different sites, including right ventricular apex pacing (RVAP), right ventricular septum pacing (RVSP), His bundle pacing (HBP), left bundle branch pacing (LBBP), and RVSP during unipolar pacing from the ring electrode of LBBP lead (RVSPring ) in each patient and evaluate the correlations between electrophysiological characteristics and ventricular synchrony.
METHODS: Twenty patients with complete atrioventricular block indicated for dual-chamber pacemaker implantation were included in the study. Unipolar pacing at different sites, including RVAP, RVSP, HBP, LBBP, and RVSPring , was successively performed in each patient. The pacing characteristics and echocardiogram parameters were collected and compared among intrinsic rhythm and pacing at different sites.
RESULTS: Similar to HBP (114.84 ± 18.67 ms), narrower paced QRSd was found in LBBP (116.15 ± 11.60 ms) as compared to RVSPring (135.11 ± 13.68 ms), RVSP (141.65 ± 14.26 ms), and RVAP (160.15 ± 19.35 ms) (p < .001). LBBP showed comparable pacing parameters to RVAP or RVSP and was significantly better than HBP, with maintained cardiac function. TS-12-SD was significantly improved in LBBP (41.80 ± 20.97 ms) than RVAP (69.70 ± 32.42 ms, p = .003) and RVSP (63.30.56 ± 32.53 ms, p = .018) but similar to HBP (51.50 ± 25.67 ms, p = .283) or RVSPring (57.80 ± 25.65 ms, p = .198). Among these pacing strategies, negative values of interventricular mechanical delay (IVMD) were only identified in LBBP (-19.25 ± 18.43 ms), significantly different from RVAP (35.00 ± 30.72 ms), RVSP (22.85 ± 22.05 ms), HBP (5.20 ± 18.64 ms), and RVSPring (16.00 ± 26.76 ms (all p < .05). Using Pearson's analysis, Sti-LVAT was positively correlated with QRS duration, IVMD, TS-12-SD, LVEDV, and LVESV, while a negative relationship could be observed for left ventricular ejection fraction.
CONCLUSIONS: His-Purkinje conduction system pacing (HPCSP) achieved better electrical and mechanical synchrony than conventional RV pacing. For interventricular synchrony, only LBBP initiated earlier LV activation than RV, in accordance with the right bundle branch block (RBBB) pattern of paced QRS during LBBP. Sti-LVAT might be a good parameter correlating with LV systolic function and mechanical synchrony.
© 2022 The Authors. Clinical Cardiology published by Wiley Periodicals, LLC.

Entities:  

Keywords:  His bundle pacing; His-Purkinje conduction system pacing; atrioventricular block; electromechanical synchronization; left bundle branch pacing

Mesh:

Year:  2022        PMID: 35502628      PMCID: PMC9286324          DOI: 10.1002/clc.23837

Source DB:  PubMed          Journal:  Clin Cardiol        ISSN: 0160-9289            Impact factor:   3.287


atrioventricular block electrophysiology His bundle pacing His‐Purkinje conduction system pacing intraventricular conduction defect interventricular mechanical delay left bundle branch pacing left ventricular ejection fraction left ventricular end‐diastolic volume left ventricular end‐systolic volume right bundle branch block right ventricular apex pacing right ventricular pacing right ventricular septum pacing RVSP during unipolar pacing from the ring electrode of LBBP lead the stimulus to left ventricular activation time tricuspid annular plane systolic excursion standard deviation of the time‐to‐peak myocardial sustained velocity of 12 left ventricular segments

BACKGROUND

For decades, right ventricular pacing (RVP) has been established as a conventional pacing technique indicated for symptomatic bradycardia for its pacing efficacy and safety. However, RVP is obviously not a physiological pacing strategy and it could lead to interventricular and intraventricular electrical desynchrony, which might result in atrial fibrillation and heart failure, , , especially in ventricular pacing dependent patients. His‐Purkinje conduction system pacing (HPCSP), His bundle pacing (HBP), and left bundle branch pacing (LBBP), have emerged as two physiological conduction system pacing strategies and have been demonstrated to achieve narrower QRS duration and better mechanical synchrony than RVP. Nowadays, HPCSP has been widely adopted from inside and outside the field. Recent studies have also demonstrated that HPCSP significantly improved heart function in heart failure patients with left bundle branch block (LBBB), , , and even better than BVP. , Theoretically, HBP is the most physiological pacing mode and it could achieve the same narrow QRS complex as the intrinsic and correct LBBB or right bundle branch block (RBBB) as well. However, pacing safety concerning high pacing threshold and low R wave amplitude of HBP limits its application in all pacing candidates, especially in infra‐Hisian block cases. While pacing captures the left conduction system more distal than His bundle, LBBP offers a low and stable threshold and high R wave amplitude comparable to RVP and its feasibility has been confirmed in large‐scale studies with mid‐long term follow‐up. , , LBBP activates the left ventricle earlier than the right ventricle through the left conduction system. Although paced QRS of LBBP is narrower than RVP from the literature, it shows an RBBB morphology that is different from the intrinsic. Consequently, it remains unknown that its RBBB shape would result in ventricular mechanical desynchrony as compared to HBP though previous case‐controlled studies have demonstrated its synchrony similar to HBP and superior to RVP. , However, the definitions of LBBP in these studies were not unified and the study populations were not ventricular pacing dependent. It would be more reliable to compare electromechanical synchrony at different pacing sites in each case in typically ventricular pacing dependent patients. Whereas, till now, direct comparisons of electromechanical effects in HPCSP and conventional RVP have not been well estimated. Additionally, an intrapatient analysis focusing on solely atrioventricular block (AVB) patients to balance individual differences has never been performed in previous research. Moreover, it is also unclear which electrophysiological value has the best correlations with ventricular synchrony and left ventricular function. Hence, the purpose of our present study is to draw meaningful comparisons between the electrophysiological and echocardiographic parameters during HPCSP and conventional RVP in AVB from an intrapatient analysis to further assess the relationship between electrical and mechanical synchrony.

METHODS

Study population

Twenty patients with complete AVB indicated for dual‐chamber pacemaker implantation were included in the study. Exclusion criteria included persistent atrial fibrillation, escape beat of intraventricular conduction defect (IVCD) morphology, and left ventricular ejection fraction (LVEF) < 50%. Written informed consent was obtained from all of the participants and the study was approved by the Institutional Review Board of Zhongshan Hospital, Fudan University, Shanghai, China.

Implantation procedure

Unipolar pacing at different sites, including right ventricular apex pacing (RVAP), right ventricular septum pacing (RVSP), HBP, LBBP, and unipolar pacing from the ring electrode of LBBP lead (RVSPring) were successively performed in each patient (Figure 1A−P). HBP and LBBP were attempted according to the literature by the lead (Mode 3830 69 cm; Medtronic) together with the sheath (Mode C315His; Medtronic) during the implantation procedure. , Briefly, His bundle potential was mapped around the atrioventricular grove with the pacing lead connection to the electrophysiology (EP) recording system (GE CardioLab EP Recording System 2000 GE Inc.) through the sheath (Model 3830, C315His; Medtronic Inc.) at RAO 30°. The lead was then turned four to five times clockwise for fixation. Afterward, the paced 12‐lead ECG was the same as the intrinsic during unipolar pacing, which confirmed His bundle capture (Figure 1I−J). LBBP was performed according to the His bundle location at RAO 30°. The pacing lead was moved toward the ventricular side about 1 cm across the tricuspid along the line between His location and RV apex and was deeply screwed into the subendocardium of the left ventricular septum. The criteria of successful LBBP were as follows according to our previous studies: (1) paced QRS of an RBBB pattern; (2) confirmation of selective LBBP (paced QRS of a typical RBBB pattern and a discrete component between stimulus and ventricle activation in intracardiac electrogram); and (3) the stimulus to left ventricular activation time (Sti‐LVAT) (Figure 1), which is defined as the interval from the pacing stimulus to the peak of the R wave in lead V5, shortening abruptly by increasing output or remaining shortest and constant at the final depth. The lead depth inside the interventricular septum was measured via angiography through the sheath at LAO 35° (Figure 1P). During the procedure, the atrial lead placed at the right atrium appendage and ventricular lead at different pacing sites above were temporarily connected to the programmer (Medtronic 2290) with DDD mode, AV delay of 150 ms, and pacing output of 3.5 V/0.5 ms during unipolar configuration at different sites above in each patient.
Figure 1

The 12‐lead ECG, EGM, and fluoroscopic images during pacing at different sites. (A) The 12‐lead ECG and the EGM recorded during RVAP: The paced QRSd was 154 ms while Sti‐LVAT was 75 ms. (B) The 12‐lead ECG and the EGM recorded during RVSP: The paced QRSd 135 ms was while Sti‐LVAT was 93 ms. (C) PoHis during the intrinsic rhythm with the potential to ventricle interval were 40 ms. (D) The 12‐lead ECG and the EGM recorded during HBP: The paced QRSd was 93 ms while Sti‐LVAT was 74 ms. (E) PoLBB during the intrinsic rhythm with the potential to ventricle interval were 18 ms. (F) The 12‐lead ECG and the EGM recorded by unipolar pacing from the ring electrode of the LBBP lead (RVSP ring): The paced QRSd was 138 ms while Sti‐LVAT was 96 ms. (G) The 12‐lead ECG and the EGM recorded by the LBBP lead (Nonselective LBBP): The paced QRSd was 107 ms while Sti‐LVAT was 68 ms and Sti‐RVAT was 110 ms. (H) The 12‐lead ECG and the EGM recorded by the LBBP lead (Selective LBBP): The paced QRSd was 119 ms while Sti‐LVAT was 63 ms. (I) PoHis obtained during the intrinsic rhythm of a narrow QRS with an HV interval of 40 ms. (J) No PoHis recorded during escape beat of an RBBB morphology. (K) PoLBB obtained with the LBB potential to ventricle interval of 20 ms. Fluoroscopic images at RAO 30° of: (L) RVAP; (M) RVSP; (N) HBP; (O) LBBP; at LAO 35° of (P) LBBP (white arrow depicted the depth of LBBP lead inside the septum via angiography through the sheath). HBP, his bundle pacing; LAO, left anterior oblique; LBB, left bundle branch; LBBP, left bundle branch pacing; RAO, right anterior oblique; RVAP, right ventricular apex pacing; RVSP, right ventricular septum pacing; Sti‐LVAT, stimulus to left ventricular activation time.

The 12‐lead ECG, EGM, and fluoroscopic images during pacing at different sites. (A) The 12‐lead ECG and the EGM recorded during RVAP: The paced QRSd was 154 ms while Sti‐LVAT was 75 ms. (B) The 12‐lead ECG and the EGM recorded during RVSP: The paced QRSd 135 ms was while Sti‐LVAT was 93 ms. (C) PoHis during the intrinsic rhythm with the potential to ventricle interval were 40 ms. (D) The 12‐lead ECG and the EGM recorded during HBP: The paced QRSd was 93 ms while Sti‐LVAT was 74 ms. (E) PoLBB during the intrinsic rhythm with the potential to ventricle interval were 18 ms. (F) The 12‐lead ECG and the EGM recorded by unipolar pacing from the ring electrode of the LBBP lead (RVSP ring): The paced QRSd was 138 ms while Sti‐LVAT was 96 ms. (G) The 12‐lead ECG and the EGM recorded by the LBBP lead (Nonselective LBBP): The paced QRSd was 107 ms while Sti‐LVAT was 68 ms and Sti‐RVAT was 110 ms. (H) The 12‐lead ECG and the EGM recorded by the LBBP lead (Selective LBBP): The paced QRSd was 119 ms while Sti‐LVAT was 63 ms. (I) PoHis obtained during the intrinsic rhythm of a narrow QRS with an HV interval of 40 ms. (J) No PoHis recorded during escape beat of an RBBB morphology. (K) PoLBB obtained with the LBB potential to ventricle interval of 20 ms. Fluoroscopic images at RAO 30° of: (L) RVAP; (M) RVSP; (N) HBP; (O) LBBP; at LAO 35° of (P) LBBP (white arrow depicted the depth of LBBP lead inside the septum via angiography through the sheath). HBP, his bundle pacing; LAO, left anterior oblique; LBB, left bundle branch; LBBP, left bundle branch pacing; RAO, right anterior oblique; RVAP, right ventricular apex pacing; RVSP, right ventricular septum pacing; Sti‐LVAT, stimulus to left ventricular activation time.

Data collection

QRS duration was measured from the onset to the end of the QRS wave and was compared during the intrinsic rhythm and different pacing sites. Sti‐LVAT and pacing stimulus to right ventricular activation time (Sti‐RVAT) (Figure 1G), defined as the interval from the pacing stimulus to the peak of R′ wave in lead V1, were recorded and compared during pacing at different sites. The transthoracic echocardiogram was used during the procedure. Echocardiographic parameters, including LVEF, left ventricular end‐diastolic volume (LVEDV), left ventricular end‐systolic volume (LVESV), interventricular mechanical delay (IVMD), and standard deviation (SD) of the time‐to‐peak myocardial sustained velocity of 12 left ventricular segments (TS‐12‐SD), tricuspid annular plane systolic excursion (TAPSE), were measured after 10 min of pacing at each site with a washout period of 5 min. IVMD was determined as the delay between left and right ventricular pre‐ejection intervals by Doppler (Figure S1). For TS‐12‐SD, pulsed‐wave Doppler and tissue synchronization imaging (Figure S2) were used to measure the left ventricular synchrony. All echocardiograms of our study were assessed by two experienced echocardiographers blinded to our study design. The fluoroscopy time of positioning the ventricular lead at different sites was also collected and compared. After collecting these parameters, LBBP lead was finally implanted together with the atrial lead connected to the ventricular and atrial port of the generator, respectively. Sensed/paced AV delay of 150/180 ms with pacing output of 3.5 V/0.5 ms by unipolar configuration were set in each case.

Follow‐up

During follow‐up (1, 3, 6, 12, and 18 months postprocedure), QRS durations and echocardiographic parameters, including LVEF, LVEDV, and LVESV were collected and compared. The procedure‐related complications, including lead dislodgement, perforation, device, or lead infection, pericardial effusion, and thromboembolism were also collected.

Statistical analysis

Continuous variables were expressed as mean ± SD and paired Student's t‐test was used to compare the difference between baseline and 6‐month follow‐up in each group. Analysis of variance test was used to perform comparison among more than two groups and was followed by the least significant difference test for multiple comparisons. Categorical variables were presented as numbers (percentages) by using Pearson's X² test or Fisher's exact test. The correlations between electrophysiological characteristics and echocardiographic parameters were performed by Pearson's analysis. All analyses were performed by SPSS version 22.0 (SPSS, Inc.) and a two‐sided p < .05 was considered statistically significant.

RESULTS

Electromechanical parameters during pacing at different sites

A total of 20 patients (mean age: 66.15 ± 13.65 years, 15 male) with complete AVB were enrolled and their baseline characteristics are shown in Table 1. Dual‐chamber pacemaker implantation with recordings of unipolar pacing at different sites was successfully achieved in all patients (Table 1) without procedure‐related complications. The mean depth of the lead was measured as 12.80 ± 0.89 mm into the interventricular septum. Pacing parameters during different pacing sites are shown in Figure 2B−D. The highest threshold (2.28 ± 1.04 V) and lowest R‐wave amplitude (3.55 ± 1.50 mV), as well as impedance (443.80 ± 105.07 Ω), were achieved in HBP among all the pacing sites (all p < .01). Threshold in LBBP (0.73 ± 0.24 V) was similar to RVAP (0.78 ± 0.16 V, p = .633) and RVSP (0.73 ± 0.13 V, p = .954) but significantly lower than RVSPring (1.34 ± 0.63 V, p < .001). For R‐wave amplitude, there was no difference between LBBP (9.43 ± 4.14 mV) and RVAP (10.21 ± 2.12 mV, p = .474) or RVSP (9.32 ± 2.19 mV, p = .914). With regard to impedance, LBBP (629.61 ± 155.58 Ω) is relatively lower than RVAP (749.26 ± 174.44 Ω, p = .022) but similar to RVSP (730.84 ± 165.63 Ω, p = .064). Moreover, higher impedance was demonstrated in LBBP when compared to RVSPring (493.35 ± 116.36 Ω, p = .012). As for LBBP, LBB potential was recorded in 15 cases (75%) and 16 cases (80%) achieved selective LBBP during the procedure.
Table 1

Baseline characteristics.

VariablesResults
Age (years)66.15 ± 13.65
Male, n (%)15 (75)
Body mass index (kg/m²)24.58 ± 3.14
Hypertension, n (%)10 (50)
Baseline electrocardiogram
QRS durations (ms)118.75 ± 24.63
Complete AVB, (%)20 (100)
Narrow QRS, n (%)9 (45)
LBBB, n (%)2 (10)
RBBB, n (%)9 (45)
Baseline echocardiography
Left atrium (mm)40.30 ± 5.32
LVEDD (mm)49.15 ± 5.46
LVEF (%)62.12 ± 13.83
Ventricular septum (mm)9.80 ± 1.58

Abbreviations: AVB, atrioventricular block; LBBB, left bundle branch block; LVEDD, left ventricular end‐diastolic diameter; LVEF, left ventricular ejection fraction; RBBB, right ventricular branch block.

Figure 2

Fluoroscopic time and pacing parameters during pacing at different sites: (A) Fluoroscopic time; (B) Threshold; (C) R wave amplitude; (D) Impedance.

Baseline characteristics. Abbreviations: AVB, atrioventricular block; LBBB, left bundle branch block; LVEDD, left ventricular end‐diastolic diameter; LVEF, left ventricular ejection fraction; RBBB, right ventricular branch block. Fluoroscopic time and pacing parameters during pacing at different sites: (A) Fluoroscopic time; (B) Threshold; (C) R wave amplitude; (D) Impedance.

Electrical and mechanical synchrony

The electrocardiographic parameters during pacing at different sites are summarized in Table S1. RVSP (141.65 ± 14.26 ms, p = .001), RVAP (160.15 ± 19.35 ms, p = .001), as well as RVSPring (135.11 ± 13.68 ms, p = .005) resulted in a remarked increase in QRS duration compared to the baseline (118.75 ± 24.63 ms) (Figure 3A). QRS duration during HBP (114.84 ± 18.67 ms) and LBBP (116.15 ± 11.60 ms) was comparable to the intrinsic and hence, was both significantly narrower than RVSP, RVAP, and RVSPring (all p < .001). Concerning Sti‐LVAT, LBBP (65.47 ± 7.98 ms) showed the shortest Sti‐LVAT compared with RVSP (89.80 ± 14.80 ms, p < .001), RVAP (112.60 ± 8.18 ms, p < .001), HBP (82.25 ± 12.13 ms, p < .001), and RVSPring (90.55 ± 15.85 ms, p < .001) (Figure 3B).
Figure 3

Electrical and mechanical synchrony during pacing at different sites: (A) QRSd; (B) Sti‐LVAT; (C) TS‐12‐SD; (D) IVMD. IVMD, interventricular mechanical delay; QRSd, QRS duration; Sti‐LVAT, stimulus to left ventricular activation time; TS‐12‐SD, standard deviation of the time‐to‐peak myocardial sustained systolic velocity of 12 left ventricular segments.

Electrical and mechanical synchrony during pacing at different sites: (A) QRSd; (B) Sti‐LVAT; (C) TS‐12‐SD; (D) IVMD. IVMD, interventricular mechanical delay; QRSd, QRS duration; Sti‐LVAT, stimulus to left ventricular activation time; TS‐12‐SD, standard deviation of the time‐to‐peak myocardial sustained systolic velocity of 12 left ventricular segments. Among all the different pacing sites, there was no significant difference in TS‐12‐SD as compared to baseline (55.20 ± 27.11 ms) while a decreasing trend could be seen during HBP (51.50 ± 25.66 ms) and LBBP (41.80 ± 20.97 ms) (Figure 3C). Whereas, TS‐12‐SD was significantly lower in LBBP compared with RVSP (p = .029) and RVA (p = .004). Significant negative means of IVMD was demonstrated in LBBP (−19.25 ± 18.43) than RVSP (22.85 ± 22.05 ms), RVAP (35.00 ± 30.72 ms), HBP (5.20 ± 18.64 ms), and RVSPring (16.00 ± 26.76 ms) (all p < .05) (Figure 3D). IVMD during HBP was similar to baseline and were significantly lower than RVSP (p = .009) and RVA (p < .001).

Evaluation of cardiac function during pacing at different sites

LVEF in HBP (62.71 ± 7.69%) and LBBP (62.93 ± 6.09%) was comparable to baseline (62.12 ± 13.83%) while a significantly decreased LVEF was identified in RVSP (59.40 ± 9.81%, p = .016), RVAP (60.47 ± 8.00%, p = .040), and RVSPring (58.50 ± 7.21%, p = .008) (Table S1). For other echocardiographic parameters, a trend toward decreased LVEDV and LVESV compared to baseline was indicated in HBP and LBBP but the difference did not reach statistical significance. TAPSE was significantly lower in RVSP (17.70 ± 3.06 mm, p = .004), RVAP (17.35 ± 2.82 mm, p = .001), and RVSPring (18.37 ± 2.81 mm, p = .033) compared with baseline (20.33 ± 2.54 mm) while HBP (19.53 ± 2.65 mm, p = .352) and LBBP (18.75 ± 2.65 mm, p = .069) remained stable.

Correlation between electrophysiological characteristics and echocardiographic parameters

Correlations between electrophysiological characteristics and echocardiographic parameters are summarized in Table S3. Notable positive linear correlations could be observed between Sti‐LVAT and QRS duration (r = 0.612, p < .001), LVEDV (r = 0.348, p = .003), LVESV (r = 0.338, p = .004), TS‐12‐SD (0.241, p = .016), and IVMD (r = 0.440, p < .001), while a negative relationship between Sti‐LVAT and LVEF (r = −0.245, p = .035) was demonstrated (Figure 4A−F). However, QRS duration was not significantly correlated to the echocardiographic parameters except for IVMD (r = 0.388, p < .001). No significant association between Sti‐RVAT and echocardiographic parameters were confirmed, either.
Figure 4

Linear correlations between electrical synchrony and echocardiographic parameters (A) Sti‐LVAT versus LVEDV; (2)Sti‐LVAT versus LVESV; (3) Sti‐LVAT versus LVEF; (4) Sti‐LVAT versus IVMD; (5) Sti‐LVAT versus TS‐12‐SD; (6) Sti‐LVAT versus QRS duration; (7) Sti‐RVAT versus QRS duration; (8) QRS duration versus IVMD. IVMD, interventricular mechanical delay; LVEF, left ventricular ejection fraction; LVEDV, left ventricular end‐diastolic volume; LVESV, left ventricular end‐systolic volume; Sti‐LVAT, stimulus to left ventricular activation time; TS‐12‐SD, standard deviation of the time‐to‐peak myocardial sustained systolic velocity of 12 left ventricular segments; Sti‐RVAT, stimulus to right ventricular activation time.

Linear correlations between electrical synchrony and echocardiographic parameters (A) Sti‐LVAT versus LVEDV; (2)Sti‐LVAT versus LVESV; (3) Sti‐LVAT versus LVEF; (4) Sti‐LVAT versus IVMD; (5) Sti‐LVAT versus TS‐12‐SD; (6) Sti‐LVAT versus QRS duration; (7) Sti‐RVAT versus QRS duration; (8) QRS duration versus IVMD. IVMD, interventricular mechanical delay; LVEF, left ventricular ejection fraction; LVEDV, left ventricular end‐diastolic volume; LVESV, left ventricular end‐systolic volume; Sti‐LVAT, stimulus to left ventricular activation time; TS‐12‐SD, standard deviation of the time‐to‐peak myocardial sustained systolic velocity of 12 left ventricular segments; Sti‐RVAT, stimulus to right ventricular activation time.

Pacing parameters and echocardiographic outcomes at follow‐up

Follow‐up echocardiograms were obtained at least 18 months after the initial pacemaker implant in each patient. Stability was maintained in LBBP capture (0.81 ± 0.23 V, p = .300) and sensed R‐wave amplitude (10.98 ± 4.56 mV, p = .295) during unipolar configuration (Table S2). For unipolar pacing impedance, a significant decrease was found 6 months (510.61 ± 88.29 Ω, p = .008) after implantation and it was maintained till 18 months follow‐up (455.78 ± 73.68 Ω, p < .001) (Table S2). Neither LVEF (62.80 ± 6.01%, p = .952) nor QRS duration (113.75 ± 11.36 ms, p = .512) showed significant difference during follow‐up and none of the complications, including lead dislodgement, perforation, device or lead infection, pericardial effusion, or thromboembolism were reported.

DISCUSSION

The present study directly compared electrophysiological characteristics and echocardiographic parameters at different pacing sites in each patient during the procedure and the main findings were as follows: (1) similar to HBP, LBBP preserved better electrical and LV mechanical synchrony compared with conventional RV pacing (RVAP or RVSP); (2) our research provided the initial evidence of earlier LV electrical activation than RV during LBBP in accordance with interventricular synchrony (negative values of IVMD in LBBP); (3) there were significant correlations between Sti‐LVAT and echocardiographic parameters, including LVEDV, LVESV, LVEF, TS‐12‐SD, and IVMD.

Electrical synchrony

Paced QRS duration is demonstrated to be narrower in LBBP , and left ventricular septal pacing as compared to RVP. Wide QRS duration might be associated with ventricular dysynchrony and heart failure. During HBP and LBBP, the heart was activated fast through the conduction system, showing a significantly narrower QRS duration than RVAP or RVSP. When the comparison between HBP and LBBP, HBP showed a narrower QRS duration similar to the intrinsic than LBBP, and LBBP performed an RBBB pattern of paced QRS. Moreover, Sti‐LVAT, which is often used to reflect the lateral precordial myocardium depolarization time, was significantly decreased in HBP and LBBP than RVAP or RVSP. These findings demonstrated that HBP was the most physiological pacing strategy, which activated both ventricles fast through His bundle (LBB and RBB) while LBBP preserved physiological LV activation before RV through LBB. These findings were consistent with the literatures : LBBP showed significantly reduced QRS duration and sti‐LVAT than RVSP in different individuals while our study further demonstrated it in each ventricular pacing dependent patient suffering from complete AVB during the implantation procedure. A novel pacing site of the present study was RVSPring, a specific RVSP during unipolar pacing from the ring electrode of LBBP lead. It might be a better pacing site than conventional RVAP or RVSP since the mean QRS duration during RVSPring was significantly decreased than RVAP and the trend was confirmed when comparing with RVSP but no statistical significance was found. Since the length between the tip and the ring of the lead (Mode 3830) was 10.8 mm, the pacing site of RVSPring was still RVS in most cases and might be a little bit deeper than conventional RVSP when the lead screwed deep inside the septum. Hence, the mean QRS duration and sti‐LVAT during RVSPring were significantly increased than LBBP, though they showed a decreased trend than RVAP or RVSP. And the pacing parameters were good as well during RVSPring. Dual cathodal lead might be designed in the future to give an additional pacing option.

Mechanical synchrony

Better LV mechanical synchrony has been previously confirmed during LBBP than RVSP using SPECT MPI and echocardiogram. , The latter study by Cai et al. demonstrated that LV mechanical synchrony during LBBP was similar to that of native‐conduction concerning LV systolic dyssynchrony index and the SD of TS in the 12 segments, and the LV synchrony in LBBP was superior to the RVSP significantly in sick sinus syndrome patients. Consistently, our results showed the similarity of TS‐12‐SD between HBP, LBBP, and the intrinsic while increased TS‐12‐SD during RVP indicated LV dysynchrony, showing preserved LV mechanical synchrony in HPCSP. Whereas, the biggest advantage of the present intro‐patient‐controlled study in AVB was that the baseline difference between individuals had been minimized to the utmost extent to facilitate the measurement and comparison of the echocardiographic results at different pacing sites. However, there was no significant difference concerning TS‐12‐SD between LBBP and RVSPring. It might be attributed to the study population (AVB patients without heart failure) and small‐sample‐sized study design. Furthermore, IVMD was measured and compared in the study and showed a significant difference between LBBP and HBP or RVAP or RVSP or RVSPring. IVMD evaluates the mechanical synchrony between LV and RV and is recognized as a predictive factor in CRT response. During LBBP, earlier LV activation than RV was confirmed by pacing activating LV before RV through LBB, showing an RBBB morphology of paced QRS. Consequently, significant negative means of IVMD during LBBP were confirmed as compared to positive ones at other pacing sites, indicating mechanical contraction of LV earlier than RV. Earlier LV electrical activation than RV during LBBP as an RBBB paced morphology in accordance with interventricular synchrony (shown as negative values of IVMD) was initially demonstrated in our research. Concerning our findings of TS‐12‐SD and IVMD, HBP maintained inter and intraventricular synchrony while LBBP preserved LV synchrony with delayed RV activation. Conversely, RVP resulted in earlier RV activation and deteriorated LV synchrony.

Correlations between electrophysiological characteristics and echocardiographic parameters

To maintain satisfied heart function and mechanical synchrony after implantation, it is of significant importance to confirm a simple and reliable electrophysiological value correlated to echocardiographic parameters during pacing. Sti‐LVAT is indicated for the depolarization duration of the LV wall. Hence, a shorter Sti‐LVAT may represent rapid propagation of LV activation leading to synchronous LV contraction. During LBBP, Sti‐LVAT has been reported to be a useful parameter to determine LBB capture according to electrophysiological mapping while QRS duration failed to act as an ideal diagnostic value for LBB capture due to its delayed RV activation. As a main and novel finding, our research further confirmed that Sti‐LVAT was also notably correlated with LV systolic function (LVEDV, LVESV, and LVEF) and mechanical synchrony (TS‐12‐SD and IVMD), while QRS duration failed except for IVMD, indicating that Sti‐LVAT might be a better variable correlated to LV‐related echocardiographic parameters than QRS duration. Since a positive relationship was also shown between Sti‐LVAT and QRS duration, shorter Sti‐LVAT could be reasonably proposed to depict both favorable electrical synchrony and LV systolic function and mechanical synchrony. As for RV function, no remarkable correlations were shown in our study between Sti‐RVAT and TAPSE. Since our study was performed based on small sample size and merely TAPSE was included as RV‐related echocardiographic parameters, the correlation between Sti‐RVAT and RV function might be highly underestimated. In the future, further analysis of RV function (RV fractional area change, Tei index, speckle tracking derived free wall strain, and so forth.) was warranted to specifically evaluate the correlation with Sti‐RVAT during LBBP.

Limitations

The study focused on the electrical and mechanical synchrony at different pacing sites during the procedure. These findings were acute hemodynamic results. Thus, the long‐term hemodynamic effects of LBBP remain uncertain. In addition, it is a single‐center, self‐controlled observational study, with relatively small sample size. The main study population was AVB with narrow QRS or RBBB and normal LVEF. Consequently, the results of the study could not be generalized to patients with IVCD or heart dysfunction.

CONCLUSIONS

HPCSP provided better electrical and mechanical left ventricular synchrony than conventional RVP. While interventricular synchrony during LBBP was significantly different as compared to HBP, RVAP, and RVSP, showing earlier LV activation than RV, which was consistent with the RBBB pattern of paced QRS during LBBP. Sti‐LVAT might be a good electrophysiological parameter correlated to LV systolic function and mechanical synchrony.

AUTHOR CONTRIBUTIONS

All authors have read and approved the manuscript. Methodology and manuscript writing: Xueying Chen, Qinchun Jin, Yanan Wang, Xiaolan Zhou. Data collection and pacing perfomance: Yufei Chen, Jingfeng Wang, Shengmei Qin, Jin Bai, Wei Wang, Yixiu Liang. Conceptualization and supervision: Haiyan Chen, Yangang Su, Junbo Ge.

CONFLICTS OF INTEREST

The authors declare no conflicts of interest. Supporting information. Click here for additional data file. Supporting information. Click here for additional data file. Supporting information. Click here for additional data file.
  22 in total

1.  Evaluation of the Criteria to Distinguish Left Bundle Branch Pacing From Left Ventricular Septal Pacing.

Authors:  Shengjie Wu; Xueying Chen; Songjie Wang; Lei Xu; Fangyi Xiao; Zhouqing Huang; Rujie Zheng; Limeng Jiang; Pugazhendhi Vijayaraman; Parikshit S Sharma; Lan Su; Weijian Huang
Journal:  JACC Clin Electrophysiol       Date:  2021-04-22

2.  A beginner's guide to permanent left bundle branch pacing.

Authors:  Weijian Huang; Xueying Chen; Lan Su; Shengjie Wu; Xue Xia; Pugazhendhi Vijayaraman
Journal:  Heart Rhythm       Date:  2019-06-22       Impact factor: 6.343

3.  The feasibility and safety of left bundle branch pacing vs. right ventricular pacing after mid-long-term follow-up: a single-centre experience.

Authors:  Xueying Chen; Qinchun Jin; Jin Bai; Wei Wang; Shengmei Qin; Jingfeng Wang; Yixiu Liang; Yangang Su; Junbo Ge
Journal:  Europace       Date:  2020-12-26       Impact factor: 5.214

Review 4.  Heart failure due to right ventricular apical pacing: the importance of flow patterns.

Authors:  Vincenzo Cicchitti; Francesco Radico; Francesco Bianco; Sabina Gallina; Gianni Tonti; Raffaele De Caterina
Journal:  Europace       Date:  2016-05-31       Impact factor: 5.214

5.  Electrophysiological parameters and anatomical evaluation of left bundle branch pacing in an in vivo canine model.

Authors:  Xueying Chen; Qinchun Jin; Bingyu Li; Jianguo Jia; Parikshit S Sharma; Weijian Huang; Yangang Su; Junbo Ge
Journal:  J Cardiovasc Electrophysiol       Date:  2019-12-18

6.  Left Bundle Branch Pacing for Cardiac Resynchronization Therapy: Nonrandomized On-Treatment Comparison With His Bundle Pacing and Biventricular Pacing.

Authors:  Shengjie Wu; Lan Su; Pugazhendhi Vijayaraman; Rujie Zheng; Mengxing Cai; Lei Xu; Ruiyu Shi; Zhouqing Huang; Zachary I Whinnett; Weijian Huang
Journal:  Can J Cardiol       Date:  2020-05-07       Impact factor: 5.223

7.  Improvement in Clinical Outcomes With Biventricular Versus Right Ventricular Pacing: The BLOCK HF Study.

Authors:  Anne B Curtis; Seth J Worley; Eugene S Chung; Pei Li; Shelly A Christman; Martin St John Sutton
Journal:  J Am Coll Cardiol       Date:  2016-05-10       Impact factor: 24.094

8.  The characteristics of the electrocardiogram and the intracardiac electrogram in left bundle branch pacing.

Authors:  Xueying Chen; Shengjie Wu; Lan Su; Yangang Su; Weijian Huang
Journal:  J Cardiovasc Electrophysiol       Date:  2019-05-15

9.  Evaluation of electrophysiological characteristics and ventricular synchrony: An intrapatient-controlled study during His-Purkinje conduction system pacing versus right ventricular pacing.

Authors:  Xueying Chen; Xiaolan Zhou; Yanan Wang; Qinchun Jin; Yufei Chen; Jingfeng Wang; Shengmei Qin; Jin Bai; Wei Wang; Yixiu Liang; Haiyan Chen; Yangang Su; Junbo Ge
Journal:  Clin Cardiol       Date:  2022-05-03       Impact factor: 3.287

10.  Evaluation of cardiac synchrony in left bundle branch pacing: Insights from echocardiographic research.

Authors:  Binni Cai; Xinyi Huang; Linlin Li; Jincun Guo; Simei Chen; Fanqi Meng; Huimin Wang; Biqin Lin; Maolong Su
Journal:  J Cardiovasc Electrophysiol       Date:  2020-01-20
View more
  1 in total

1.  Evaluation of electrophysiological characteristics and ventricular synchrony: An intrapatient-controlled study during His-Purkinje conduction system pacing versus right ventricular pacing.

Authors:  Xueying Chen; Xiaolan Zhou; Yanan Wang; Qinchun Jin; Yufei Chen; Jingfeng Wang; Shengmei Qin; Jin Bai; Wei Wang; Yixiu Liang; Haiyan Chen; Yangang Su; Junbo Ge
Journal:  Clin Cardiol       Date:  2022-05-03       Impact factor: 3.287

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.