| Literature DB >> 35484860 |
Megumi Sasaki1, Yasumitsu Hirano1, Hiroki Yonezawa1, Satoshi Shimamura1, Atsuko Kataoka1, Takatsugu Fujii1, Naoto Okazaki1, Shintaro Ishikawa1, Toshimasa Ishii1, Katsuya Deguchi1, Hiroshi Sato1, Shinichi Sakuramoto1, Kojun Okamoto1, Isamu Koyama1.
Abstract
BACKGROUND: The Senhance Digital Laparoscopy System (Asensus Surgical Inc, Morrisville, NC, United States), which was introduced for the first time in Japan by our hospital, is a new surgical assistive robot following the da Vinci Surgical System. We herein report the short-term outcomes of 55 colorectal cancer surgery cases using this system at our hospital to assess the feasibility and safety of our procedures.Entities:
Keywords: Senhance Digital Laparoscopy System; colorectal cancer; robot-assisted surgery
Mesh:
Year: 2022 PMID: 35484860 PMCID: PMC9321837 DOI: 10.1111/ases.13064
Source DB: PubMed Journal: Asian J Endosc Surg ISSN: 1758-5902
FIGURE 1Senhance Digital Laparoscopy System
FIGURE 2The forceps of Senhance Digital Laparoscopy System
FIGURE 3Port placement. (A) Right side colon, (B) left side colon, (C) ileocecal resection, (D) sigmoid colectomy, (E) right hemicolectomy, (F) lower anterior resection. R: 5‐mm port robotic forceps, S: 12‐mm camera port, A: 5‐mm port assistant forceps
Characteristics of the study patients
| 55 cases | |
|---|---|
| Median age, y (range) | 71 (39–91) |
| Gender, male/female | 31/24 |
| Median body mass index (range) | 23.1 (14.0–34.2) |
| History of abdominal surgery, +/− | 15/40 |
| Operation type, A/B | 15/40 |
| ASA, I/II/III | 15/36/4 |
| Tumor location, V/C/A/T/S/R | 2/11/12/2/15/13 |
| cT stage, cTis/1/2/3/4 | 2/8/10/30/5 |
| cN stage, cN0/1/2/3 | 40/11/2/2 |
| cStage, 0/I/II/III | 2/17/21/15 |
Abbreviation: ASA, American Society of Anesthesiologists classification.
Types of operation performed
| Operation | 55 cases (%) |
|---|---|
| Ileocecal resection | 18 (32.7) |
| Right hemicolectomy | 8 (14.5) |
| Transverse colectomy | 1 (1.8) |
| Sigmoid colectomy | 9 (16.3) |
| High anterior resection | 11 (20.0) |
| Low anterior resection | 7 (12.7) |
| Intersphincteric resection | 1 (1.8) |
Surgical results
| 55 cases | |
|---|---|
| Operation time, min (range) | 240 (101–378) |
| Docking time, min (range) | 8 (2–25) |
| Senhance operating time, min (range) | 126 (24–287) |
| Blood loss, mL (range) | 5 (0–167) |
| Lymph node dissection, D2/D3 | 14/41 |
| Number of dissected lymph nodes median (range) | 21 (7–39) |
| fStage, 0/I/II/III/IV | 3/17/18/15/2 |
| Postoperative stay, d (range) | 7 (5–27) |
System characteristics
| Device | da Vinci | Senhance |
|---|---|---|
| Console | Closed type | Open type |
| Optics | 8 mm 3D | 10 mm 3D |
| Forceps size | 8 mm | 3 mm or 5 mm |
| Haptic feedback | No | Yes |
| Field‐of‐view control | Handle + foot pedal | Handle button + eye movement |
| Reuse | Up to 10 times | Possibly without limit |