Materials with temperature-dependent luminescence can be used as local thermometers when incorporated in, for example, a biological environment or chemical reactor. Researchers have continuously developed new materials aiming for the highest sensitivity of luminescence to temperature. Although the comparison of luminescent materials based on their temperature sensitivity is convenient, this parameter gives an incomplete description of the potential performance of the materials in applications. Here, we demonstrate how the precision of a temperature measurement with luminescent nanocrystals depends not only on the temperature sensitivity of the nanocrystals but also on their luminescence strength compared to measurement noise and background signal. After first determining the noise characteristics of our instrumentation, we show how the uncertainty of a temperature measurement can be predicted quantitatively. Our predictions match the temperature uncertainties that we extract from repeated measurements, over a wide temperature range (303-473 K), for different CCD readout settings, and for different background levels. The work presented here is the first study that incorporates all of these practical issues to accurately calculate the uncertainty of luminescent nanothermometers. This method will be important for the optimization and development of luminescent nanothermometers.
Materials with temperature-dependent luminescence can be used as local thermometers when incorporated in, for example, a biological environment or chemical reactor. Researchers have continuously developed new materials aiming for the highest sensitivity of luminescence to temperature. Although the comparison of luminescent materials based on their temperature sensitivity is convenient, this parameter gives an incomplete description of the potential performance of the materials in applications. Here, we demonstrate how the precision of a temperature measurement with luminescent nanocrystals depends not only on the temperature sensitivity of the nanocrystals but also on their luminescence strength compared to measurement noise and background signal. After first determining the noise characteristics of our instrumentation, we show how the uncertainty of a temperature measurement can be predicted quantitatively. Our predictions match the temperature uncertainties that we extract from repeated measurements, over a wide temperature range (303-473 K), for different CCD readout settings, and for different background levels. The work presented here is the first study that incorporates all of these practical issues to accurately calculate the uncertainty of luminescent nanothermometers. This method will be important for the optimization and development of luminescent nanothermometers.
Nanomaterials
with temperature-dependent
luminescence are one of the most versatile thermometers on the microscopic
scale with applications in biology, electronics, and catalysis.[1−3] The temperature of a nanothermometer is determined by recording
its emission spectrum or its luminescence lifetime. The intensity
ratio between two emission bands is most frequently considered because
this parameter is insensitive to fluctuations in the excitation intensity,
changes in alignment, and scattering of the luminescence. The relative
sensitivity, Sr (in % per K), expresses
how strongly the intensity ratio changes with temperature and is thus
a measure for the measurement accuracy. Sr is an intrinsic property of a thermometer material (which however
depends on temperature) and is easily determined by measuring emission
spectra over a range of temperatures. Newly developed thermometer
materials are therefore often characterized and compared in terms
of this parameter.[4,5]In practice, the reliability
of temperature readout depends not
only on the relative sensitivity but also on the signal-to-noise ratio
of a measurement. These parameters together determine the temperature
uncertainty, σ. Current methods
to determine σ are diverse. The
most direct method is experimentally recording a series of luminescence
spectra and calculating the standard deviation of the extracted temperatures.[6] Alternatively, the noise level on a single spectrum
may be estimated from fluctuations in the baseline. The latter method
underestimates the temperature uncertainty because it fails to take
the noise on detected photons into account.[7] More importantly, both methods are often used in idealized circumstances
where background signal is minimal, a large amount of thermometer
material is measured, luminescence is efficiently collected, and/or
long measurement times are used. The extracted values of σ depend strongly on these circumstances.
In contrast to Sr, σ is not an intrinsic property of a (nano)thermometer
material. Consequently, user-to-user differences have caused variations
in reported uncertainties of several orders of magnitude for the same
thermometer, while some measurement conditions such as the environment
of the thermometer were similar.[4] It is
not clear to what extent these reported values of σ are relevant for actual applications of the (nano)thermometers,
which may put restrictions on the measurement procedure and/or introduce
background fluorescence and blackbody radiation.[8] In addition, undesired emissions from the thermometer itself,
for example, from higher-excited levels, can interfere with temperature
measurements.[9] Although subtracting a reference
spectrum of any background signal removes the systematic error,[10] the influence on the temperature uncertainty
remains. It is currently unclear how these practical complications
affect the performance of luminescent (nano)thermometers. This makes
a fair comparison of potential thermometer materials impossible.In this article, we use the statistics of photon detection to quantify
how noise and background signal affect the temperature uncertainty
of a luminescence (nano)thermometry experiment. Not only the properties
of the thermometer material are important but also the characteristics
of the detector and the sample. We first measure the background-free
upconversion luminescence of NaYF4:Er3+(2%),Yb3+(18%) nanocrystals using a conventional CCD and characterize
the different types of detector noise. Using error propagation, we
quantitatively explain the temperature uncertainties determined by
recording a series of spectra, which increases with the set temperature.
We study the impact of detector noise by recording upconversion luminescence
of nanocrystals with electron multiplication gain using an electron-multiplying
CCD (EMCCD). An increase in gain boosts the signal to overcome readout
noise and thus reduces the temperature uncertainty. Finally, we examine
the effect of background signal. Even if we subtract the background
signal, the experimental temperature uncertainty increases with higher
background levels as predicted from the larger error on the number
of detected photons. These new methods to calculate the uncertainty
show that not only the relative sensitivity Sr of a thermometer determines its performance but also the
achievable signal-to-noise ratio. The temperature uncertainty σ depends strongly on measurement conditions
and is therefore a poor parameter to compare the potential of thermometer
materials. We propose alternative metrics that could be considered.
Results
We first discuss the uncertainty achieved with a model thermometer
based on two emissive excited levels A and B (Figure a). An increase in temperature affects the
relative intensities emitted by these levels, resulting in a change
in the intensity ratio in the emission spectrum. In a typical experiment,
these emissions are spectrally separated by a grating and captured
by a CCD, photomultiplier tube, or photodiode array. The photosensitive
material in the detector converts the incident photons to photoelectrons.
The number of photoelectrons k recorded in one exposure
will follow the Poisson distributionwhere k̅
is the expected
number of photoelectrons, which is proportional to the product of
the photon flux and the acquisition time and is in general different
for levels A and B. An interesting property of the Poisson distribution
is that the standard deviation is equal to the square root of the
expected value (Figure b). The next step in the detection process is the translation of
photoelectrons to digital counts for each pixel, which enables the
construction of the emission spectrum. For a luminescence thermometry
experiment, the observables of interest in the spectrum are the integrated
counts of the emissions from levels A and B: nA and nB, respectively. As nA and nB are independent
random variables, the measurement error on the intensity ratio R = nB/nA follows from error propagation[11]Here, n̅A,B are the expected counts of A and
B with corresponding variances
σA,B2 and n̅A,B/σA,B are the signal-to-noise ratios on nA and nB. Assuming
that the errors and expected counts are related as described by the
Poisson distribution, we expect lower σ for higher counts.
Figure 1
Temperature uncertainty achieved with a model
thermometer. (a)
Energy level diagram of the model thermometer, in which the solid
black arrows represent the radiative decay pathways. (b) Poisson distribution
with mean k̅ = 100 and standard deviation √k̅ = 10. (c) Simulated luminescence spectrum comprising
two Gaussian emission bands with Poissonian detection noise. The inset
shows a histogram of the temperatures that are extracted from 10 000
simulated spectra using the ratio of integrated counts, a physical
temperature of 298 K, and a relative sensitivity of 1% K–1. The black line is a normal distribution with a mean of 298 K and
a standard deviation that is calculated via eq . (d) Same as in (c) but for a total luminescence
intensity that is 10 times higher.
Temperature uncertainty achieved with a model
thermometer. (a)
Energy level diagram of the model thermometer, in which the solid
black arrows represent the radiative decay pathways. (b) Poisson distribution
with mean k̅ = 100 and standard deviation √k̅ = 10. (c) Simulated luminescence spectrum comprising
two Gaussian emission bands with Poissonian detection noise. The inset
shows a histogram of the temperatures that are extracted from 10 000
simulated spectra using the ratio of integrated counts, a physical
temperature of 298 K, and a relative sensitivity of 1% K–1. The black line is a normal distribution with a mean of 298 K and
a standard deviation that is calculated via eq . (d) Same as in (c) but for a total luminescence
intensity that is 10 times higher.Conversion of R to a temperature value requires
knowledge of the relative sensitivity. This is often obtained by calibrating
the spectral response of the thermometer over a range of temperatures.
Any error in the calibrated relative sensitivity leads to a systematic
difference between the measured and physical temperature. However,
the random error σ on the measured
temperature only depends on the probability distribution function
of the measured R and on the relative sensitivity
of the thermometer[11]This shows that, as expected, the temperature
uncertainty decreases with the increasing signal-to-noise ratio. An
alternative approach of luminescence thermometry relies on the shift
of an emission band at varying temperatures. In the Supporting Information, we use a similar analysis as above
to determine the temperature uncertainty of measurements based on
a spectral shift. Finally, we verify eq by simulating luminescence spectra with two emissions
bands and determine the temperature from the ratio of the simulated
counts (Figure c).
These simulated temperatures follow a normal distribution with a standard
deviation that matches σ calculated
using eq . The distribution
of temperatures would deviate from normal if n̅A,B becomes of order unity, rather than ≫ 1 as
we consider in Figure and is typical in experiments. An increase in counts results in
a narrower distribution of measured temperatures, consistent with eq (Figure d). We thus understand quantitatively how
experiments with higher counts, performed with, for example, longer
acquisition times or brighter thermometers, have a lower temperature
uncertainty.In Figure , we
experimentally study the temperature uncertainty of thermometry measurements
at elevated temperatures. We acquired spectra with a CCD camera because
this is the most frequently used detector in the luminescence thermometry
community. The CCD camera conveniently records an entire spectrum
within one capture. In contrast, step-wise acquisition of a spectrum
with a scanning monochromator and single-point detector such as a
photomultiplier tube leads to additional temperature errors if the
excitation intensity fluctuates during the measurement. We use NaYF4:Er3+(2%),Yb3+(18%) nanocrystals as
thermometers because green upconversion emission from this popular
thermometer material can be excited with 980 nm light, preventing
background fluorescence.[12]Figure a shows the luminescence spectra,
in which the two emission bands at 540 and 520 nm are due to radiative
decay from the thermally coupled levels in Er3+, 4S3/2, and 2H11/2, respectively.
An increase in temperature (T) changes the ratio
of the expected counts within the emission bands of 4S3/2 and 2H11/2, n̅S and n̅H, respectively,
following Boltzmann’s distributionwhere ΔE is the energy
gap between these levels, kB is Boltzmann’s
constant, and C is the pre-exponential factor that
includes the degeneracies and radiative decay rates from the two levels
to the ground state. To use this relation as a calibration of our
thermometer, we average 200 spectra and obtain n̅S and n̅H by summing
the counts of all pixels within the integration boundaries of the
corresponding emission bands. We then fit the ratio measured at various
temperatures to eq and
find a value of 746 cm–1 for ΔE and 15.2 for C (Figure b).[13] We use this
calibration to convert the measured intensity ratios from spectra
with different levels of noise to apparent temperatures.
Figure 2
Experimental
temperature uncertainty. (a) Upconversion luminescence
of dried NaYF4:Er3+(2%),Yb3+(18%)
nanocrystals upon 980 nm excitation at 302 K (blue) and 473 K (red).
(b) Logarithm of the ratio between the 2H11/2 and 4S3/2 emission with integration ranges
516–534 and 538–545 nm (green dots), respectively. The
black line is a fit of the experimental ratios to Boltzmann’s
distribution (eq ),
yielding values of 746 cm–1 for ΔE and 15.2 for C. (c) Temperatures extracted from
200 experimental spectra using the calibration in (b). The thermocouple
in our heating stage measured a temperature of 302 K during the acquisition
of spectra, showing a small deviation with the mean of the temperatures
extracted from the spectra, likely caused by a systematic error in
the calibration. The gray shaded area covers the temperature range
of the mean ± standard deviation. (d) Correlation between the
measured temperature and the total counts within the integration ranges
of the 2H11/2 and 4S3/2 emissions. The correlation coefficient (ρ) is close to zero,
indicating that measured temperature and total counts are uncorrelated.
(e) Distribution of counts per 1000 ms frame, for pixels on our CCD
camera showing an average of 1629 counts/1000 ms when measured over
200 frames. The camera recorded the reflection of a white lamp on
a microscopy slide. The solid line is a fit of the experimental data
to the normal distribution (k̅ = 1629, σ2 = 2112), and the dashed line shows the Poisson distribution
with k̅ = 1629. (f) Plot of the variance against
the mean (purple dots) measured via the procedure in (e) for different
intensities of the white lamp. The black line is a fit of the experimental
data to the model σ2 = k̅/f + σr2, where f = 0.78 is the analog-to-digital conversion factor and σr2 = 57 is the readout
variance of one pixel. (g) Temperature uncertainties at various physical
temperatures obtained via the procedure in (c) (red dots). The black
line is the temperature uncertainty calculated using eq , with no fit parameters.
Experimental
temperature uncertainty. (a) Upconversion luminescence
of dried NaYF4:Er3+(2%),Yb3+(18%)
nanocrystals upon 980 nm excitation at 302 K (blue) and 473 K (red).
(b) Logarithm of the ratio between the 2H11/2 and 4S3/2 emission with integration ranges
516–534 and 538–545 nm (green dots), respectively. The
black line is a fit of the experimental ratios to Boltzmann’s
distribution (eq ),
yielding values of 746 cm–1 for ΔE and 15.2 for C. (c) Temperatures extracted from
200 experimental spectra using the calibration in (b). The thermocouple
in our heating stage measured a temperature of 302 K during the acquisition
of spectra, showing a small deviation with the mean of the temperatures
extracted from the spectra, likely caused by a systematic error in
the calibration. The gray shaded area covers the temperature range
of the mean ± standard deviation. (d) Correlation between the
measured temperature and the total counts within the integration ranges
of the 2H11/2 and 4S3/2 emissions. The correlation coefficient (ρ) is close to zero,
indicating that measured temperature and total counts are uncorrelated.
(e) Distribution of counts per 1000 ms frame, for pixels on our CCD
camera showing an average of 1629 counts/1000 ms when measured over
200 frames. The camera recorded the reflection of a white lamp on
a microscopy slide. The solid line is a fit of the experimental data
to the normal distribution (k̅ = 1629, σ2 = 2112), and the dashed line shows the Poisson distribution
with k̅ = 1629. (f) Plot of the variance against
the mean (purple dots) measured via the procedure in (e) for different
intensities of the white lamp. The black line is a fit of the experimental
data to the model σ2 = k̅/f + σr2, where f = 0.78 is the analog-to-digital conversion factor and σr2 = 57 is the readout
variance of one pixel. (g) Temperature uncertainties at various physical
temperatures obtained via the procedure in (c) (red dots). The black
line is the temperature uncertainty calculated using eq , with no fit parameters.Figure c shows
the temperatures that we extracted from a series of spectra using
the calibration of Figure b.[14] These values are evenly distributed
around the mean, which is a sign of a stable physical temperature
during the measurement. Correlations between the extracted temperature
and the total green luminescence counts could indicate that the laser
heats the sample as variations in laser intensity would result in
higher count rates coinciding with more laser heating. Experiments
at higher laser powers do show such correlations (Figure S2), which indicates that increasing the excitation
intensity to reduce the uncertainty can induce a systematic error
on temperature readout. This is not observed in our experiments shown
in Figure d. We therefore
used the standard deviation of the measured temperatures as the experimental
temperature uncertainty at a fixed sample temperature. Fluctuations
in excitation intensity below the heating threshold do not affect
the intensity ratio nor the signal-to-noise ratio (Figure S3).To understand the magnitude of the variations
in measured temperature
(Figure c), we must
consider the noise generated by our detector. The main noise sources
in a CCD measurement are counting noise due to the statistics of incident
photons and readout noise due to the translation of photoelectrons
to digital counts by the analog-to-digital converter (ADC).[15] We characterize these by acquiring a large set
of 200 reference images on our CCD camera, illuminating it with a
white lamp, and histogramming the digital counts of pixels with the
same mean (Figure e). The distribution of digital counts approximates a normal distribution
with a variance that is, in our case, slightly larger than the corresponding
Poisson distribution would have, taking the single-pixel readout variance
into account. This difference is due to the conversion of photoelectrons
to digital counts, which changes the variance on the output counts
by the ADC factor, f.[16,17]Figure f shows a fit of the experimental
variances to a model that includes the ADC factor and the readout
noiseWe find a value of 0.78 for the ADC
factor,
which is specific to our camera in the used settings. If an emission
band is integrated over N camera pixels, the variance
of the total readout noise on this band is N times
larger than the single-pixel readout variance σr2 = 57. We can now
insert this expression and the relative sensitivity of a Boltzmann
thermometer, ΔE/kBT2,[18] into eq to calculate the expected
temperature uncertainty from a luminescence spectrum.Figure g shows
the temperature dependence of the uncertainty by comparing experiments
with the theoretical trend (eqs –5). Using the method presented
in Figure c, we determine
the temperature uncertainty at various physical temperatures and find
values of 0.2 K at room temperature increasing to more than 0.4 K
at 473 K.[19,20] Between the different physical temperatures
of the experiment, the total counts within the spectrum varied slightly—a
decrease is likely due to thermal quenching and an increase could
indicate water desorption from the surface of the dried nanocrystals.[21] This affects the signal-to-noise ratio and thus
obscures the impact of the intensity ratio on the temperature uncertainty.
We therefore kept the sum of n̅S and n̅H roughly constant at 2
× 106 counts, which allows us to separately calculate n̅H and n̅S using eq in a range
of physical temperatures. We can therefore use eq to calculate n̅H and n̅S separately, depending
on the physical temperature. Inserting these values, along with the
detector characteristics (eq ), into eq yields
the theoretical uncertainty (black line in Figure g). The calculated uncertainties agree well
with the experimental values, without any fit parameters. We therefore
conclude that after proper characterization of the photodetector,
error propagation correctly predicts the experimental uncertainty
and its temperature dependence.As a further illustration of
the effect of the detector noise characteristics
on a temperature measurement, we consider the effect of electron multiplication
(EM) in an EMCCD. Emerging applications of luminescent nanomaterials,
such as single-particle thermometry, require photodetectors that are
able to record extremely weak signals.[22,23] EMCCDs could
offer a solution as they enhance the signal by orders of magnitude,
compared to conventional CCDs, but the electron multiplication process
causes additional noise.[15,16] We start by considering
the detection of photons and generation of photoelectrons, which in
both a conventional CCD and an EMCCD follows Poisson statistics. Both
types of cameras then transfer the photoelectrons to the ADC via the
serial readout register and convert them to digital counts. In an
EMCCD camera, the readout register is extended with additional registers
that, depending on the applied voltage, multiply the number of photoelectrons
and thus boost the signal. In practice, the output electrons pass
through hundreds of multiplication registers, resulting in a total
EM gain G. The number of output electrons nout follows the gamma distributionHere, k is the number of
photoelectrons generated by a CCD pixel, which enter the multiplication
registers. Figure a shows the probability distribution of nout as a function of the expected number of input photoelectrons k̅ and the EM gain G. EM produces
an expected number of counts of n̅out = k̅G with a variance that approximates σ2 = 2k̅G2/f (eq S1 and Figure S4). The signal-to-EM-counting-noise ratio , where readout noise is excluded,
is thus
independent of the gain factor. This derivation shows that EMCCD measurements
have an additional counting noise of √2, commonly referred
to as the excess noise factor. Therefore, EM gain can only improve
a temperature measurement if nout is small
with respect to other noise sources.
Figure 3
Effect of electron multiplication on temperature
uncertainty. (a)
Distribution of output electrons for different input photoelectrons k̅ and EM gains G. This distribution
is obtained by convolution of the gamma distribution and the Poisson
distribution. (b) Experimental temperature uncertainties obtained
from 200 upconversion spectra for various levels of EM gain ranging
from 2 to 90 (yellow dots). The average numbers of photoelectrons
generated by a pixel are 5 and 30 for the 2H11/2 and 4S3/2 emissions, respectively. The solid
black line is the temperature uncertainty calculated via eqs and 7, with
no fit parameters. (c) Color map of the temperature uncertainty as
a function of the expected photoelectrons k̅
per pixel and the EM gain. The temperature uncertainties in the color
map were calculated via the expected value of the output electrons,
excluding spurious electrons, and the variance of all output electrons
(eq S1). All uncertainties were normalized
to the minimum value within the map. The contour lines correspond
to σ values of 1.2, 2, and 8 K.
We obtain the simulated spectra in (c) by drawing random numbers of
output electrons from the distribution of eq S1, with as an input a spectrum consisting of two peaks with Gaussian
shape of equal amplitude covering a total of 400 pixels. The four
insets show example spectra simulated for the experimental settings
to which they are linked in the color map. We set the readout noise
for each pixel to σr = 26, matching that of our EMCCD
detector at a 30 MHz readout rate and preamplifier gain 2. We assume
a probability of spurious charges of ps = 0.0004 to give the simulated spectra the characteristic background
noise of electron multiplication.
Effect of electron multiplication on temperature
uncertainty. (a)
Distribution of output electrons for different input photoelectrons k̅ and EM gains G. This distribution
is obtained by convolution of the gamma distribution and the Poisson
distribution. (b) Experimental temperature uncertainties obtained
from 200 upconversion spectra for various levels of EM gain ranging
from 2 to 90 (yellow dots). The average numbers of photoelectrons
generated by a pixel are 5 and 30 for the 2H11/2 and 4S3/2 emissions, respectively. The solid
black line is the temperature uncertainty calculated via eqs and 7, with
no fit parameters. (c) Color map of the temperature uncertainty as
a function of the expected photoelectrons k̅
per pixel and the EM gain. The temperature uncertainties in the color
map were calculated via the expected value of the output electrons,
excluding spurious electrons, and the variance of all output electrons
(eq S1). All uncertainties were normalized
to the minimum value within the map. The contour lines correspond
to σ values of 1.2, 2, and 8 K.
We obtain the simulated spectra in (c) by drawing random numbers of
output electrons from the distribution of eq S1, with as an input a spectrum consisting of two peaks with Gaussian
shape of equal amplitude covering a total of 400 pixels. The four
insets show example spectra simulated for the experimental settings
to which they are linked in the color map. We set the readout noise
for each pixel to σr = 26, matching that of our EMCCD
detector at a 30 MHz readout rate and preamplifier gain 2. We assume
a probability of spurious charges of ps = 0.0004 to give the simulated spectra the characteristic background
noise of electron multiplication.In Figure b, we
compare the experimental temperature uncertainty at various levels
of EM gain with theoretical predictions. First, we acquired 200 experimental
spectra with an EM gain of only a factor 2 while keeping the number
of incident photons per pixel low. This resulted in an extremely high
uncertainty of 30 K. Increasing the EM gain to values of 25 causes
a sharp decline of the uncertainty to 3 K. The effect of even higher
EM gains is weak. We again explain this trend using eqs and 5 (solid
line) by realizing that the variance of the 4S3/2 and 2H11/2 counts is due to a combination
Poissonian counting noise amplified by the ADC factor f and the EM gain factor G2 and readout
noise. [For photon fluxes relevant for luminescence thermometry, we
can neglect noise due to spurious electrons created during shifting
of charges through the multiplication register (Figure S5).[15,16]] EM gain increases the signal
(as well as the counting noise) with respect to the readout noiseFigure c illustrates the effect of EM gain for a range of expected
photoelectrons. We observe the highest uncertainty in the bottom left
of the map, where both the EM gain and the number of expected photoelectrons
are low—the corresponding simulated spectrum barely shows emission.
For all numbers of expected photoelectrons, we observe a rapid decrease
of the uncertainty with increasing gain, although this effect becomes
negligible once readout noise is overcome (Figure S6). Indeed, the signal-to-total-noise ratio (eq ) approaches a constant value of for large G and EM cannot
improve it further (Figure S7). In practice,
EM gain is thus useful if and only if the signal is weak compared
to readout noise.Besides the photodetector, background emission
by the surroundings
of the thermometer can be another source of uncertainty, which is
relevant when the thermometer is used in realistic experimental conditions.[24] We discuss how such a distortion of the spectrum
affects the temperature uncertainty even after subtraction of the
background. Figure a shows how we have mimicked this experimental issue: we have measured
the upconversion luminescence with and without an additional broadband
background signal from a white lamp. Subtracting a reference measurement
of the lamp recovers a clean thermometer spectrum from the experiment
with background. However, the noise on the background signal cannot
be removed. The corrected spectrum therefore contains more noise compared
to the background-free upconversion emission spectrum (Figure b). This translates to an increased
temperature uncertainty (Figure c). We can further understand this from the expression
for the variance σb,2 in the counts from emitting state i after background
removalwhere n̅b, is the expected number of background counts
removed.
Additional counts from dark current in the photodectector have an
equivalent impact on the temperature uncertainty as background emissions.
Again, inserting this expression into eq gives the theoretical temperature uncertainty after
background removal.
Figure 4
Influence of background subtraction on temperature uncertainty.
(a) Reflection spectrum of the white lamp illuminating dried NaYF4:Er3+(2%), Yb3+(18%) nanocrystals without
(gray) and with simultaneous 980 nm excitation (green). Subtraction
of the gray spectrum from the green spectrum yields the corrected
upconversion spectrum (red). The total signal (thermometer) and background
(lamp) counts within the integration ranges of the 2H11/2 and 4S3/2 emission are 2 ×
106 and 2 × 107, respectively. (b) Zoom-in
on the noise of a spectrum acquired without background (blue) and
with a broadband background that is subsequently subtracted (red).
(c) Calibration curve of Figure b was used to convert a series of 200 spectra with
no background (blue) and with lamp background subtracted (red) to
temperatures. The red and blue dots show a systematic difference of
the mean, which we attribute to an error introduced by recording and
subtracting the reference spectrum, for example, because the lamp
spectrum fluctuated slightly over time. (d) Temperature uncertainties
as a function of set temperature, measured by comparing the apparent
temperature of 200 background-subtracted spectra. The solid black
lines are the expected temperature uncertainties for 2 × 106 (yellow dots) and 2 × 107 (purple dots) subtracted
background counts, calculated using eq . The expected background counts on the 2H11/2 and 4S3/2 emissions are obtained
by taking the sum of the counts in the subtracted spectrum between
516–534 and 538–545 nm, respectively, averaged over
200 spectra. The dashed black line marks the minimum of the temperature
uncertainty for all subtracted background counts.
Influence of background subtraction on temperature uncertainty.
(a) Reflection spectrum of the white lamp illuminating dried NaYF4:Er3+(2%), Yb3+(18%) nanocrystals without
(gray) and with simultaneous 980 nm excitation (green). Subtraction
of the gray spectrum from the green spectrum yields the corrected
upconversion spectrum (red). The total signal (thermometer) and background
(lamp) counts within the integration ranges of the 2H11/2 and 4S3/2 emission are 2 ×
106 and 2 × 107, respectively. (b) Zoom-in
on the noise of a spectrum acquired without background (blue) and
with a broadband background that is subsequently subtracted (red).
(c) Calibration curve of Figure b was used to convert a series of 200 spectra with
no background (blue) and with lamp background subtracted (red) to
temperatures. The red and blue dots show a systematic difference of
the mean, which we attribute to an error introduced by recording and
subtracting the reference spectrum, for example, because the lamp
spectrum fluctuated slightly over time. (d) Temperature uncertainties
as a function of set temperature, measured by comparing the apparent
temperature of 200 background-subtracted spectra. The solid black
lines are the expected temperature uncertainties for 2 × 106 (yellow dots) and 2 × 107 (purple dots) subtracted
background counts, calculated using eq . The expected background counts on the 2H11/2 and 4S3/2 emissions are obtained
by taking the sum of the counts in the subtracted spectrum between
516–534 and 538–545 nm, respectively, averaged over
200 spectra. The dashed black line marks the minimum of the temperature
uncertainty for all subtracted background counts.The measured uncertainties as a function of physical temperature
match the predicted values for a range of different background levels
(Figure d). We observe
higher absolute values of the uncertainty with an increasing background.
This effect is as large as a factor of 3 for the background level
in Figure a, even
though we could subtract the background using a reference measurement.
We further observe that the minimum of the temperature uncertainty
(dashed line) shifts to higher set temperatures with increasing background
counts. In practice, low levels of background are challenging to completely
avoid, especially at elevated temperatures where blackbody radiation
becomes an issue. Figure shows how this affects the absolute value of temperature
uncertainty as well as the optimal operating temperature of a thermometer
compared to idealized measurement conditions without background signal.Our work clearly demonstrates how the precision of a temperature
measurement depends not only on the relative sensitivity of the thermometer
but also on the measurement conditions. As these measurement conditions
will be different for different applications, this raises the question
of how to define a relevant metric to compare thermometers. Currently,
the achieved temperature uncertainty is frequently reported in the
literature. However, our results show that the temperature uncertainty
is not a fundamental property of the thermometer.
Temperature uncertainties measured under idealized experimental conditions
are difficult to compare and may not be relevant for applications.
In particular, different experimental settings can yield wildly different
contributions of various noise sources. However, whatever the specific
experimental noise contributions are, the temperature uncertainty
is always minimal for a high relative sensitivity Sr and for a strong luminescence signal (eq ). Although Sr of newly developed thermometer materials is commonly reported,[26] the potential signal strength is hardly considered.The realization that signal strength is essential for precise temperature
measurement makes it possible to identify relevant parameters, in
addition to Sr, that define a good thermometer.
The expected counts on emissions A and B can be written asHere, absorption of excitation light is determined
by the number of luminescent centers in the excitation volume (NC) and the absorption cross section σabs per luminescent center. The luminescence further scales
with the photoluminescence quantum yield ηPL of the
relevant thermometer emission lines and the temperature-dependent
populations that define the fractions ϕA,B of emission
coming from A or B. The spectroscopic equipment sets the detection
efficiency ηdet. The integration time t, the excitation power used Pexc, and
the EM gain G can be chosen by the experimentalist.In eq , we can distinguish
the experimental factors (ηdet, Pexc, t, NC, and G) from the thermometer properties (σabs, ηPL, and ϕA,B). The
experimental factors will depend on the available equipment and the
type of sample. The freedom to choose a long t, high NC, or high Pexc may
be restricted if the sample is not static, if the sample volume is
small, or if strong excitation induces laser heating (which also depends
on heat dissipation in the sample). The values of these parameters
are not intrinsic thermometer properties but depend strongly on the
application. Another factor that affects the temperature uncertainty
is the emission wavelength of the thermometer because it determines
the required type of detector and therefore the amount of dark current.
Infrared detectors typically have a high dark current due to the small
band gap of the photosensitive material, resulting in a relatively
high uncertainty for infrared-emitting thermometers. Equation also contains some intrinsic
properties that can vary by orders of magnitude between different
materials. We propose that a fair comparison between potential thermometer
materials should consider these intrinsic parameters. For practical
applications, high σabs and ηPL for
a given doping content NC and excitation
power Pexc are as important as a high
thermal sensitivity of ϕA,B.[27] These intrinsic properties of the thermometer determine the achievable
signal compared to various application-related issues, like background
fluorescence or blackbody radiation.Studies of new thermometers
often include measurements of the sensitivity,
but experimental values of ηPL and σabs are rare. Depending on the doping concentration, the synthesis procedure,
and the excitation power, ηPL can vary over a few
orders of magnitude, and it is therefore an important parameter to
report. A well-established method to measure ηPL is
to determine the number of absorbed and emitted photons of a sample
using an integrating sphere. This has already improved the design
and synthesis of thermometer materials. For example, recent studies
on NaYF4:Er3+,Yb3+ nanocrystals have
optimized the quantum yield of NaYF4:Er3+,Yb3+ in a range of excitation powers, reaching values comparable
to bulk material.[28]Characterizing
σabs can be more challenging, especially
for microcrystalline samples where strong light scattering prevents
measuring optical absorption over a well-defined path length. We study
a clear dispersion of Yb3+-doped NaYF4 nanocrystals
with absorption spectroscopy to show that for a specific ion–host
combination σabs simply follows Lambert–Beer’s
law (Figure a).[29−31] We find a maximum value of 7.5 × 10–21 cm2 at 977 nm, which matches literature values obtained from
absorption measurements of single crystals[32,33] and from the kinetics of upconversion luminescence.[34] These values are an order of magnitude lower than the value
that Sui et al. obtain from Judd–Ofelt parameters and an experimental
photoluminescence decay rate.[35,36] Nonradiative processes,
indicated by the multiexponential decay curve, have likely resulted
in an overestimation of the spontaneous emission rate and, thus, of
the absorption cross section. Another method has recently been developed
by our group, which extracts σabs from the luminescence
saturation characteristics.[37,38] In contrast to absorption
measurements, it works well on microcrystalline samples and is thus
a suitable alternative for samples that cannot be synthesized in nanocrystalline
form. This method requires standard spectroscopic equipment, a continuous-wave
laser, a lens to achieve sufficiently high excitation intensities,
and careful characterization of the excitation spot on the sample
(Figure S8). We used this method to acquire
the luminescence of Yb3+-doped NaYF4 microcrystals
at various excitation intensities (Iexc), revealing clear signs of saturation above 5 kW cm–2 (Figure b).[39] We fit this trend to the steady-state emission
intensity of an excited two-level system that suffers from ground-state
depletion.where A is a scaling constant, hν is the energy of
an excitation photon, and kdecay is the
total decay rate of the excited
ion.[40] We find a σabs value
of 7.8 × 10–21 cm2 at 980 nm, which
perfectly matches the result of Figure a. This demonstrates that the methods presented here
provide a reliable σabs. Together with existing methods
to determine ηPL and Sr, it should now be possible to predict the uncertainty of temperature
measurements for any particular experimental setting. This shows how
valuable it is to consider absorption cross sections and the quantum
yields in the design of future and existing thermometers.
Figure 5
Experimental
methods to determine the absorption cross section.
(a) Absorption spectrum of NaYF4:Yb3+(18%) nanocrystals
dispersed in cyclohexane. The spectrum shows a clear absorption band
due to the 2F7/2 → 2F5/2 transition of Yb3+. Rayleigh scattering caused
a background of roughly 0.0015 absorbance units, which was subtracted
from the spectrum. The concentration of the bare nanocrystals without
ligands is 10 mg/mL, which corresponds to a concentration of Yb3+ ions of 5.4 × 1018 cm–3. Taking the inhomogeneous dielectric surroundings of the Yb3+ ions inside the dispersed nanocrystal into account,[25] this translates into an absorption cross section
of σabs = 7.5 × 10–21 cm2 at 977 nm. (b) Intensity of the Yb3+ luminescence
measured on microcrystalline NaYF4:Yb3+(18%)
for various excitation intensities Iexc of 980 nm light (green dots). The solid black line is a fit of the
experimental data to eq , which yields a σabs value of 7.8 × 10–21 cm2 at 980 nm. The dashed line is a linear
fit to the low-excitation-intensity data to clearly visualize the
nonlinear trend of the high-excitation-intensity data.
Experimental
methods to determine the absorption cross section.
(a) Absorption spectrum of NaYF4:Yb3+(18%) nanocrystals
dispersed in cyclohexane. The spectrum shows a clear absorption band
due to the 2F7/2 → 2F5/2 transition of Yb3+. Rayleigh scattering caused
a background of roughly 0.0015 absorbance units, which was subtracted
from the spectrum. The concentration of the bare nanocrystals without
ligands is 10 mg/mL, which corresponds to a concentration of Yb3+ ions of 5.4 × 1018 cm–3. Taking the inhomogeneous dielectric surroundings of the Yb3+ ions inside the dispersed nanocrystal into account,[25] this translates into an absorption cross section
of σabs = 7.5 × 10–21 cm2 at 977 nm. (b) Intensity of the Yb3+ luminescence
measured on microcrystalline NaYF4:Yb3+(18%)
for various excitation intensities Iexc of 980 nm light (green dots). The solid black line is a fit of the
experimental data to eq , which yields a σabs value of 7.8 × 10–21 cm2 at 980 nm. The dashed line is a linear
fit to the low-excitation-intensity data to clearly visualize the
nonlinear trend of the high-excitation-intensity data.In conclusion, we have characterized how experimental conditions
affect the uncertainty of temperature measurements through (nano)thermometry
based on luminescence intensity ratios. We first measured all noise
sources associated with photon detection and developed statistical
models to quantitatively predict the temperature uncertainty in a
wide range of temperatures and for various experimental settings.
We observed that enhancement of the luminescence signal by applying
EM gain significantly reduces the uncertainty until readout noise
is overcome. In addition, we studied the impact of background emissions,
which is a realistic practical issue. Background increases the uncertainty
of a temperature measurement even if it is properly subtracted from
the measurement. Our work demonstrates that the temperature uncertainty
is not an intrinsic property of a luminescent (nano)thermometer but
instead strongly depends on the photodetector and measurement conditions.
We propose a guideline of how to compare different thermometers in
a way that is relevant irrespective of the spectroscopic equipment
used or of the sample under consideration. Such new ways of comparing
luminescent (nano)thermometers are essential to develop and choose
the ideal thermometer for the desired application.
Authors: Fiorenzo Vetrone; Rafik Naccache; Alicia Zamarrón; Angeles Juarranz de la Fuente; Francisco Sanz-Rodríguez; Laura Martinez Maestro; Emma Martín Rodriguez; Daniel Jaque; José García Solé; John A Capobianco Journal: ACS Nano Date: 2010-06-22 Impact factor: 15.881
Authors: Christian Homann; Lisa Krukewitt; Florian Frenzel; Bettina Grauel; Christian Würth; Ute Resch-Genger; Markus Haase Journal: Angew Chem Int Ed Engl Date: 2018-06-14 Impact factor: 15.336
Authors: Jonathan De Roo; Maria Ibáñez; Pieter Geiregat; Georgian Nedelcu; Willem Walravens; Jorick Maes; Jose C Martins; Isabel Van Driessche; Maksym V Kovalenko; Zeger Hens Journal: ACS Nano Date: 2016-01-26 Impact factor: 15.881
Authors: Thomas P van Swieten; Tijn van Omme; Dave J van den Heuvel; Sander J W Vonk; Ronald G Spruit; Florian Meirer; H Hugo Pérez Garza; Bert M Weckhuysen; Andries Meijerink; Freddy T Rabouw; Robin G Geitenbeek Journal: ACS Appl Nano Mater Date: 2021-03-30
Authors: Robin G Geitenbeek; P Tim Prins; Wiebke Albrecht; Alfons van Blaaderen; Bert M Weckhuysen; Andries Meijerink Journal: J Phys Chem C Nanomater Interfaces Date: 2017-01-20 Impact factor: 4.126
Authors: Andrea D Pickel; Ayelet Teitelboim; Emory M Chan; Nicholas J Borys; P James Schuck; Chris Dames Journal: Nat Commun Date: 2018-11-21 Impact factor: 14.919
Authors: Thomas Hartman; Robin G Geitenbeek; Caterina S Wondergem; Ward van der Stam; Bert M Weckhuysen Journal: ACS Nano Date: 2020-04-20 Impact factor: 15.881