| Literature DB >> 35465196 |
Daiki Sakai1,2,3, Sakiko Demoto1,2, Yukako Iwai1,2, Yasuhiko Hirami1,2, Makoto Nakamura3, Yasuo Kurimoto1,2.
Abstract
Purpose: To compare the accuracy of intraoperative aberrometry (IA) for predicting postoperative refraction between eyes with emmetropia and myopia targets. Patients andEntities:
Keywords: axial myopia; cataract surgery; intentional myopia; intraocular lens calculation formula; postoperative refraction
Year: 2022 PMID: 35465196 PMCID: PMC9020572 DOI: 10.2147/OPTH.S363228
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Characteristics of the Patients and Implanted Intraocular Lenses
| Target | |||
|---|---|---|---|
| Emmetropia (n = 39) | Intentional Myopia (n = 22) | ||
| Age (years), mean ± SD | 64.7 ± 8.6 | 70.7 ± 11.1 | 0.027* |
| Female, n (%) | 24 (61.5%) | 10 (45.5%) | 0.225 |
| Axial length (mm) | |||
| Mean ± SD | 26.8 ± 1.5 | 26.7 ± 1.4 | 0.881 |
| Median (IQR) | 26.4 (2.1) | 26.4 (2.1) | |
| Range | 25.1–31.3 | 25.2–30.9 | |
| Average keratometry (mm) | |||
| Mean ± SD | 7.78 ± 0.43 | 7.64 ± 0.26 | 0.235 |
| Median (IQR) | 7.71 (0.44) | 7.61 (0.32) | |
| Range | 7.01–9.65 | 7.12–8.14 | |
| White to white (mm) | |||
| Mean ± SD | 12.0 ± 0.4 | 11.9 ± 0.4 | 0.249 |
| Median (IQR) | 12.0 (0.6) | 11.9 (0.6) | |
| Range | 11.2–12.9 | 11.2–12.7 | |
| IOL power (D) | |||
| Mean ± SD | 11.4 ± 3.7 | 16.2 ± 3.6 | <0.001* |
| Median (IQR) | 11.0 (5.0) | 16.8 (3.9) | |
| Range | 1.0–21.0 | 7.5–21.0 | |
| Postoperative refraction (spherical equivalent, D) | |||
| Mean ± SD | 0.003 ± 0.42 | −3.65 ± 0.85 | <0.001* |
| Median (IQR) | 0 (0.38) | −3.44 (1.16) | |
| Range | −1.00 to 1.25 | −5.88 to −2.50 | |
Notes: *Significant at P < 0.05 (Mann–Whitney U-test).
Abbreviations: IOL, intraocular lens; SD, standard deviation; IQR, interquartile range.
Figure 1Box and whisker plots of the numerical (A) and absolute (B) errors of intraoperative aberrometry in eyes with target of emmetropia and intentional myopia. The horizontal line in the box represents the median. The top and bottom of the box represent the upper and lower quartiles respectively. The whisker above the box represents the values within 1.5 times the of interquartile range plus the upper quartile. The whisker below the box represents the values within 1.5 times of the interquartile range minus the lower quartile. Values, which are not included between whiskers were plotted as outliers.
Prediction Error of Intraoperative Aberrometry
| Target | |||
|---|---|---|---|
| Emmetropia (n = 39) | Intentional Myopia (n = 22) | ||
| Numerical error (D) | |||
| Mean ± SD | 0.04 ± 0.39 | −0.37 ± 0.54† | 0.001* |
| Median (IQR) | 0.06 (0.39) | −0.38 (0.60) | |
| Range | −1.11 to 1.00 | −1.50 to 0.79 | |
| Hyperopic error, n (%) | 21 (53.8%) | 5 (22.7%) | |
| Myopic error, n (%) | 18 (46.2%) | 17 (77.3%) | |
| Absolute error (D) | |||
| Mean ± SD | 0.28 ± 0.27 | 0.51 ± 0.41 | 0.014* |
| Median (IQR) | 0.18 (0.37) | 0.40 (0.46) | |
| Range | 0.01–1.11 | 0.04–1.50 | |
| ±0.25 D, n (%) | 21 (53.8%) | 8 (36.4%) | |
| ±0.50 D, n (%) | 33 (84.6%) | 14 (63.6%) | |
| ±0.75 D, n (%) | 37 (94.9%) | 17 (77.3%) | |
| ±1.00 D, n (%) | 38 (97.4%) | 19 (86.4%) | |
Notes: *Significant at P < 0.05 (Mann–Whitney U-test). †Significantly different from zero at P < 0.05 (one-sample t-test).
Abbreviations: SD, standard deviation; IQR, interquartile range.
Comparison Between Accuracy of Intraoperative Aberrometry and Intraocular Lens Calculation Formulas in 39 Eyes with Target of Emmetropia
| IA | SRK/T | Holladay 1 | Hoffer Q | Holladay 2 | Haigis | Barrett Universal II | |
|---|---|---|---|---|---|---|---|
| Absolute error (n = 39) | |||||||
| Mean ± SD | 0.28 ± 0.27 | 0.38 ± 0.36 | 0.59 ± 0.40 | 0.56 ± 0.39 | 0.47 ± 0.37 | 0.44 ± 0.35 | 0.35 ± 0.33 |
| Median (IQR) | 0.18 (0.37) | 0.26 (0.46) | 0.58 (0.57) | 0.47 (0.61) | 0.36 (0.44) | 0.37 (0.42) | 0.27 (0.29) |
| Range | 0.01–1.11 | 0.01–1.64 | 0.02–1.84 | 0.01–1.66 | 0.03–1.69 | 0.02–1.47 | 0.01–1.46 |
| 0.121 | <0.001* | <0.001* | 0.015* | 0.311 | 0.513 |
Notes: The mean absolute errors varied significantly between the 7 groups (P < 0.001, Friedman test). *Significant at P < 0.05 vs intraoperative aberrometry (Wilcoxon signed-rank test with Bonferroni correction).
Abbreviations: IA, intraoperative aberrometry; SD, standard deviation; SRK/T, Sanders, Retzlaff, and Kraft/theoretical.
Comparison Between Accuracy of Intraoperative Aberrometry and Intraocular Lens Calculation Formulas in 22 Eyes with Target of Intentional Myopia
| IA | SRK/T | Holladay 1 | Hoffer Q | Holladay 2 | Haigis | Barrett Universal II | |
|---|---|---|---|---|---|---|---|
| Absolute error (n = 22*) | |||||||
| Mean ± SD | 0.51 ± 0.41 | 0.52 ± 0.38 | 0.67 ± 0.52 | 0.57 ± 0.40 | 0.47 ± 0.35 | 0.50 ± 0.40 | 0.39 ± 0.38 |
| Median (IQR) | 0.40 (0.46) | 0.43 (0.59) | 0.51 (0.74) | 0.48 (0.62) | 0.36 (0.44) | 0.45 (0.50) | 0.26 (0.43) |
| Range | 0.04–1.50 | 0.03–1.36 | 0.03–1.94 | 0.09–1.59 | 0.03–1.69 | 0.06–1.40 | 0.01–1.40 |
Notes: The mean absolute errors were not significantly different between the 7 groups (P = 0.051, Friedman test). *Holladay 2: n = 21.
Abbreviations: IA, intraoperative aberrometry; SD, standard deviation; SRK/T, Sanders, Retzlaff, and Kraft/theoretical.