Literature DB >> 35452020

MRI Allows Accurate Measurement of Glenoid Bone Loss.

Mirco Sgroi1, Hashuka Huzurudin, Marius Ludwig, Timo Zippelius, Heiko Reichel, Thomas Kappe.   

Abstract

BACKGROUND: Bony Bankart lesions larger than a certain size can lead to a high redislocation rate, despite treatment with Bankart repair. Detection and measurement of glenoid bone loss play key roles in selecting the appropriate surgical therapy in patients with shoulder instability. There is controversy about which diagnostic modalities, using different measurement methods, provide the best diagnostic validity. QUESTIONS/PURPOSES: (1) What are the diagnostic accuracies of true AP radiographs, West Point (WP) view radiographs, MRI, and CT to detect glenoid bone loss? (2) Are there differences in the measurements of glenoid bone loss on MRI and CT? (3) What are the intrarater and interrater reliabilities of CT and MRI to measure glenoid bone loss?
METHODS: Between August 2012 and February 2017, we treated 80 patients for anterior shoulder instability. Of those, we considered patients with available preoperative true AP radiographs, WP radiographs, CT images, and MR images of the affected shoulder as potentially eligible. Based on that, 63% (50 of 80) of patients were eligible for analysis; 31% (25 of 80) were excluded because not all planes or slices (such as sagittal, axial, or frontal) of each diagnostic imaging modalities were available and 7% (5 of 80) because of the insufficient quality of diagnostic images (for example, setting of the layers did not allow adequate en face view of the glenoid). Preoperative true AP radiographs, WP radiographs, CT images and MR images of the affected shoulders were retrospectively assessed for the presence of glenoid bone loss by two blinded observers at a median (range) 25 months (12 to 66) postoperatively. To evaluate sensitivity, specificity, positive predictive value, negative predictive value, accuracy, diagnostic odds ratio, positive likelihood ratio, negative likelihood ratio, and area under the curve (AUC), we compared the detection of glenoid bone loss at follow-up achieved with the aforementioned imaging modalities with intraoperative arthroscopic detection. In all patients with glenoid bone loss, two blinded observers measured the size of the glenoid bone loss on preoperative CT and MR images using six measuring techniques: depth and length of the glenoid bone loss, Bigliani classification, best-fit circle width loss method, AP distance method, surface area method, and Gerber X ratio. Subsequently, the sizes of the glenoid bone loss determined using CT and MRI were compared. To estimate intraobserver and interobserver reliability, measurements were performed in a blinded fashion by two observers. Their level of experience was equivalent to that of orthopaedic residents, and they completed a training protocol before the measurements.
RESULTS: For the ability to accurately diagnose Bankart lesions, the AUC (accuracy of a diagnostic test; the closer to 1.0, the more accurate the test) was good for MRI (0.83 [95% confidence interval 0.70 to 0.94]; p < 0.01), fair for CT (0.79 [95% CI 0.66 to 0.92]; p < 0.01), poor for WP radiographs (0.69 [95% CI 0.54 to 0.85]; p = 0.02) and failed for true AP radiographs (0.55 [95% CI 0.39 to 0.72]; p = 0.69). In paired comparisons, there were no differences between CT and MRI regarding (median [range]) lesion width (2.33 mm [0.35 to 4.53] versus 2.26 mm [0.90 to 3.47], p = 0.71) and depth (0.42 mm [0.80 to 1.39] versus 0.40 mm [0.06 to 1.17]; p = 0.54), and there were no differences concerning the other measurement methods: best-fit circle width loss method (15.02% [2.48% to 41.59%] versus 13.38% [2.00% to 36.34%]; p = 0.66), AP distances method (15.48% [1.44% to 42.01%] versus 12.88% [1.43% to 36.34%]; p = 0.63), surface area method (14.01% [0.87% to 38.25] versus 11.72% [2.45% to 37.97%]; p = 0.68), and Gerber X ratio (0.75 [0.13 to 1.47] versus 0.76 [0.27 to 1.13]; p = 0.41). Except for the moderate interrater reliability of the Bigliani classification using CT (intraclass correlation coefficient = 0.599 [95% CI 0.246 to 0.834]; p = 0.03) and acceptable interrater reliability of the Gerber X ratio using CT (0.775 [95% CI 0.542 to 0.899]; p < 0.01), all other measurement methods had good or excellent intrarater and interrater reliabilities on MRI and CT.
CONCLUSION: The results of this study show that CT and MRI can accurately detect glenoid bone loss, whereas WP radiographs can only recognize them poorly, and true AP radiographs do not provide any adequate diagnostic accuracy. In addition, when measuring glenoid bone loss, MRI images of the analyzed measurement methods yielded sizes that were no different from CT measurements. Finally, the use of MRI images to measure Bankart bone lesions gave good-to-excellent reliability in the present study, which was not inferior to CT findings. Considering the advantages including lower radiation exposure and the ability to assess the condition of the labrum using MRI, we believe MRI can help surgeons avoid ordering additional CT imaging in clinical practice for the diagnosis of anterior shoulder instability in patients with glenoid bone loss. Future studies should investigate the reproducibility of our results with a larger number of patients, using other measurement methods that include examination of the opposite side or with three-dimensional reconstructions. LEVEL OF EVIDENCE: Level I diagnostic study.
Copyright © 2022 by the Association of Bone and Joint Surgeons.

Entities:  

Mesh:

Year:  2022        PMID: 35452020      PMCID: PMC9384911          DOI: 10.1097/CORR.0000000000002215

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.755


  45 in total

1.  Articular arc length mismatch as a cause of failed bankart repair.

Authors:  S S Burkhart; S M Danaceau
Journal:  Arthroscopy       Date:  2000-10       Impact factor: 4.772

2.  Receiver operating characteristic curves and their use in radiology.

Authors:  Nancy A Obuchowski
Journal:  Radiology       Date:  2003-10       Impact factor: 11.105

3.  Can we improve radiological analysis of osseous lesions in chronic anterior shoulder instability?

Authors:  C Charousset; V Beauthier; L Bellaïche; R Guillin; N Brassart; H Thomazeau
Journal:  Orthop Traumatol Surg Res       Date:  2010-10-28       Impact factor: 2.256

4.  Influence of Glenoid Defect Size and Bone Fragment Size on the Clinical Outcome After Arthroscopic Bankart Repair in Male Collision/Contact Athletes.

Authors:  Shigeto Nakagawa; Tatsuo Mae; Kenji Yoneda; Kazutaka Kinugasa; Hiroyuki Nakamura
Journal:  Am J Sports Med       Date:  2017-04-20       Impact factor: 6.202

5.  The importance of CT for the pre-operative surgical planning in recurrent anterior shoulder instability.

Authors:  Philipp Moroder; Herbert Resch; Silke Schnaitmann; Thomas Hoffelner; Mark Tauber
Journal:  Arch Orthop Trauma Surg       Date:  2012-11-23       Impact factor: 3.067

6.  Techniques to evaluate glenoid bone loss.

Authors:  Hiroyuki Sugaya
Journal:  Curr Rev Musculoskelet Med       Date:  2014-03

7.  The meaning and use of the area under a receiver operating characteristic (ROC) curve.

Authors:  J A Hanley; B J McNeil
Journal:  Radiology       Date:  1982-04       Impact factor: 11.105

8.  3-D CT is the most reliable imaging modality when quantifying glenoid bone loss.

Authors:  Julie Y Bishop; Grant L Jones; Michael A Rerko; Chris Donaldson
Journal:  Clin Orthop Relat Res       Date:  2012-09-21       Impact factor: 4.176

Review 9.  Classification of glenohumeral joint instability.

Authors:  Christian Gerber; Richard W Nyffeler
Journal:  Clin Orthop Relat Res       Date:  2002-07       Impact factor: 4.176

10.  Is the Bare Spot reliable for the bone loss measurement?

Authors:  Max Rogerio Freitas Ramos; Yonder Archanjo Ching San Junior; Luiz Henrique Pereira Alves; Flavio Carvalho Cruz; Henrique Mansur
Journal:  Shoulder Elbow       Date:  2018-03-04
View more
  1 in total

1.  Calculating glenoid bone loss based on glenoid height using ipsilateral three-dimensional computed tomography.

Authors:  Johnny Rayes; Jian Xu; Sara Sparavalo; Jie Ma; Lauren Jonah; Ivan Wong
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2022-06-08       Impact factor: 4.342

  1 in total