Core muscle strength can be reliably measured using a novel isokinetic device: An intra-observer study.

Marit L Mulder1, Elke Van den Steen2, Jan De Neve2, Adam Weir1,3,4.   

Abstract

BACKGROUND: Core strength is an important aspect of physical fitness. A dynamometer was developed to measure isokinetic core muscle strength in multiple planes of motion. Establishing the reliability is needed before it can be used in practice.
OBJECTIVE: Examine the intra-observer reliability of a trunk isokinetic dynamometer.
METHODS: 31 participants were recruited. Tests were performed twice with an interval of 5-9 days by the same observer. Testing included rotation, flexion/extension and lateral flexion at speeds of 90∘/s and 60∘/s. The main outcome measure was peak torque (PT). The secondary outcomes included peak torque angle (PTA), time to peak torque (TTPT) and compensatory torques (CompTQ). The intra-observer reliability was investigated using intraclass correlation coefficients (ICC).
RESULTS: Peak torque reliability was good to excellent (ICC = 0.70-0.91), whereas the reliability of the PTA (ICC =-0.04-0.56) and TTPT (ICC = 0.01-0.68) were poor to moderate. CompTQ reliability was moderate to good (ICC = 0.20-0.88).
CONCLUSIONS: The intra-observer reliability of the isokinetic measurement of core strength peak torque was good. The secondary outcomes peak torque angle and time to peak torque were less reliable and the compensatory torques showed moderate to good reliability. This isokinetic dynamometer could have useful applications in the field of sports medicine and rehabilitation.

Entities:  

Keywords:  Core strength; dynamometer; isokinetic; peak torque; sports medicine

Mesh:

Year:  2022        PMID: 35431226      PMCID: PMC9535570          DOI: 10.3233/BMR-210300

Source DB:  PubMed          Journal:  J Back Musculoskelet Rehabil        ISSN: 1053-8127            Impact factor:   1.456


  20 in total

Review 1.  Core stability and its relationship to lower extremity function and injury.

Authors:  John D Willson; Christopher P Dougherty; Mary Lloyd Ireland; Irene McClay Davis
Journal:  J Am Acad Orthop Surg       Date:  2005-09       Impact factor: 3.020

2.  Trunk extensor and flexor strength measured by the Cybex 6000 dynamometer. Assessment of short-term and long-term reproducibility of several strength variables.

Authors:  O R Madsen
Journal:  Spine (Phila Pa 1976)       Date:  1996-12-01       Impact factor: 3.468

Review 3.  Evaluation isometric and isokinetic of trunk flexor and extensor muscles with isokinetic dynamometer: A systematic review.

Authors:  Jaisson Agne Estrázulas; Jansen Atier Estrázulas; Kelly de Jesus; Karla de Jesus; Rubens A da Silva; João Otacílio Libardoni Dos Santos
Journal:  Phys Ther Sport       Date:  2020-07-17       Impact factor: 2.365

4.  Reliability and validity of trunk flexor and trunk extensor strength measurements using handheld dynamometry in a healthy athletic population.

Authors:  Cedric De Blaiser; Roel De Ridder; Tine Willems; Lieven Danneels; Philip Roosen
Journal:  Phys Ther Sport       Date:  2018-10-12       Impact factor: 2.365

5.  Validity of trunk extensor and flexor torque measurements using isokinetic dynamometry.

Authors:  Gaël Guilhem; Caroline Giroux; Antoine Couturier; Nicola A Maffiuletti
Journal:  J Electromyogr Kinesiol       Date:  2014-07-17       Impact factor: 2.368

6.  Evaluating abdominal core muscle fatigue: Assessment of the validity and reliability of the prone bridging test.

Authors:  C De Blaiser; R De Ridder; T Willems; L Danneels; L Vanden Bossche; T Palmans; P Roosen
Journal:  Scand J Med Sci Sports       Date:  2017-06-28       Impact factor: 4.221

Review 7.  Exercise related syncope, when it's not the heart.

Authors:  C T Paul Krediet; Arthur A M Wilde; Wouter Wieling; John R Halliwill
Journal:  Clin Auton Res       Date:  2004-10       Impact factor: 4.435

8.  Reliability of knee extension and flexion measurements using the Con-Trex isokinetic dynamometer.

Authors:  Nicola A Maffiuletti; Mario Bizzini; Kevin Desbrosses; Nicolas Babault; Urs Munzinger
Journal:  Clin Physiol Funct Imaging       Date:  2007-11       Impact factor: 2.273

9.  Reliability of isokinetic measurements of trunk muscle performance.

Authors:  A Delitto; S J Rose; C E Crandell; M J Strube
Journal:  Spine (Phila Pa 1976)       Date:  1991-07       Impact factor: 3.468

10.  Trunk Stability, Trunk Strength and Sport Performance Level in Judo.

Authors:  David Barbado; Alejandro Lopez-Valenciano; Casto Juan-Recio; Carlos Montero-Carretero; Jaap H van Dieën; Francisco J Vera-Garcia
Journal:  PLoS One       Date:  2016-05-27       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.