| Literature DB >> 35402036 |
Ichiro Ogino1, Hidetaka Seto1,2, Daisuke Shigenaga1, Masaharu Hata3.
Abstract
Background: The most common secondary cancer is contralateral breast (CLB) cancer after whole breast irradiation (WBI). The aim of this study was to quantify the reduction of CLB dose in tangential intensity modulated radiotherapy (t-IMRT) for WBI using flattening-filter-free (FFF) beams. Materials and methods: We generated automated planning of 20 young breast cancer patients with limited user interaction. Dose-volume histograms of the planning target volume (PTV), ipsilateral lung, heart, and CLB were calculated. The dose of PTV, the most medial CLB point, and the CLB point below the nipple was measured using an ionization chamber inserted in a slab phantom. We compared the two t-IMRT plans generated by FFF beams and flattening-filter (FF) beams.Entities:
Keywords: contralateral breast; flattening-filter-free; tangential intensity modulated radiotherapy; whole breast irradiation; young breast cancer
Year: 2022 PMID: 35402036 PMCID: PMC8989456 DOI: 10.5603/RPOR.a2022.0006
Source DB: PubMed Journal: Rep Pract Oncol Radiother ISSN: 1507-1367
Figure 1Representative axial CT image of a right breast cancer patient; A. Medial marker (purple), lateral marker (light blue), isocenter (red), and breast point (green): 2 cm from isocenter perpendicular to the posterior edge of the tangent beams; B. markers for the center of the phantom surface (purple), planning target volume (PTV) point (yellow), the most medial contralateral breast (CLB) point (brown), and CLB below the nipple point (orange) for the phantoms
Figure 2Measurement points in the phantom; A. planning target volume (PTV) point; B. The most medial contralateral breast (CLB) point; C. CLB below the nipple point
Figure 3Mean dose-volume histograms for the planning target volume (PTV) (A); ipsilateral lung (ILL) (B), and contralateral breast (CLB) (C) for all patients and heart for left breast cancer patients (D)
Comparison of calculated parameters (mean ± standard deviation) between two tangential intensity modulated radiotherapy (IMRT) plans
| Parameter | Flattening-filter | Flattening-filter-free | Difference | p-value |
|---|---|---|---|---|
| HI | 0.1 ± 0.01 | 0.1 ± 0.02 | −0.04 ± 0.02 (−52.8) | < 0.001 |
| CI | 0.8 ± 0.1 | 0.8 ± 0.1 | 0 ± 0.1 (−0.5) | 0.841 |
| PTV D98% [cGy] | 4761.7 ± 40 | 4601.8 ± 74 | 159.9 ± 66.1 (3.4) | < 0.001 |
| PTV D2% [cGy] | 5174.2 ± 39.8 | 5234.1 ± 54.2 | −59.9 ± 34.4 (−1.2) | < 0.001 |
| PTV Dmax [cGy] | 5337.5 ± 93.3 | 5385.9 ± 115.1 | −48.4 ± 93.1 (−0.9) | 0.031 |
| PTV Dmin [cGy] | 4320.1 ± 145.5 | 4184 ± 138.3 | 136.2± 107.7 (3.2) | < 0.001 |
| ILL Dmax [cGy] | 4917.5 ± 143 | 5001.1 ± 138 | −83.7 ± 63.1 (−1.7) | < 0.001 |
| ILL V20 (%) | 14.1 ± 5.3 | 13.9 ± 5.5 | 0.2 ± 0.4 (1.3) | 0.051 |
| Heart Dmax [cGy] | 4402.3 ± 1031.3 | 4455.2 ± 1074.6 | −52.9 ± 105.9 (−1.2) | 0.149 |
| Heart V30 (%) | 1.9 ± 2.3 | 1.7 ± 1.6 | 0.2 ± 1.4 (−11.7) | 0.631 |
| CLB Dmax [cGy] | 278.6 ± 219 | 320.3 ± 266.1 | −41.7 ± 55.8 (−15) | 0.003 |
| CLB Dmean [cGy] | 41.4 ± 13.7 | 55.4 ± 17.4 | −14 ± 4 (−33.8) | < 0.001 |
Difference between with filter and without filter;
ten left breast cancer patients;
HI — homogeneity index; CI — conformal index; PTV — planning target volume; ILL — ipsilateral lung; CLB — contralateral breast
Comparison of phantom measured parameters (mean ± standard deviation) between two tangential intensity modulated radiotherapy (IMRT) plans
| Parameter | Flattening-filter | Flattening-filter-free | Difference | (%) p-value |
|---|---|---|---|---|
| Monitor Units | 268.4 ± 11 | 342.5 ± 21.7 | −74.1 ± 18.4 (−27.6) | < 0.001 |
| Delivery time (second) | 49.8 ± 5.5 | 43.7 ± 7.4 | 6.1 ± 7.1 (12.2) | 0.001 |
| Treatment time (second) | 138.8 ± 12.8 | 131.2 ± 10.1 | 7.6 ± 11 (5.5) | 0.006 |
|
| ||||
| PTV [cGy] | 208.7 ± 5.9 | 212.8 ± 7.3 | −4.1 ± 2.4 (−2) | < 0.001 |
| Medial CLB [cGy] | 7.2 ± 1.6 | 5.7 ± 1.2 | 1.5 ± 0.4 (21.1) | < 0.001 |
| CLB below nipple [cGy] | 2.9 ± 0.3 | 2.3 ± 0.3 | 0.6 ± 0.1 (20) | < 0.001 |
|
| ||||
| Medial CLB/PTV (%) | 3.4 ± 0.7 | 2.7 ± 0.5 | 0.8 ± 0.2 (22.6) | < 0.001 |
| CLB below nipple/PTV (%) | 1.4 ± 0.1 | 1.1 ± 0.1 | 0.3 ± 0.1 (21.5) | < 0.001 |
Difference between with filter and without filter;
PTV — planning target volume; CLB — contralateral breast; Medial CLB/PTV — dose ratios in points of medial CLB to PTV; CLB below nipple/PTV — dose ratios in points below nipple of CLB to PTV