| Literature DB >> 35393127 |
Srikar Anagoni1, Nagaraja Mudhigeti2, Mohan Alladi3, Verma Anju4, Padmalatha Am5, Usha Kalawat6.
Abstract
PURPOSE: A large number of new molecular or virology laboratories have been established to increase the testing capacity for SARS-CoV-2. Due to heavy workload, there is delay in testing of samples. In order to avoid the negative effect of delayed testing on RTPCR results guidelines are issued from WHO and CDC to transport samples in cold chain. However, in pandemic situations the recommended guidelines for transport and storage conditions are often compromised. This study was conducted to evaluate the effect of sample storage conditions at different temperatures on the results of RT PCR test.Entities:
Keywords: Covid-19; RTPCR; Viral transport medium
Mesh:
Year: 2022 PMID: 35393127 PMCID: PMC8979488 DOI: 10.1016/j.ijmmb.2022.03.005
Source DB: PubMed Journal: Indian J Med Microbiol ISSN: 0255-0857 Impact factor: 1.347
Comparison between 4 °C and Room temperature in all five days of testing.
| Days & Temp | Sensitivity | Specificity | PPV | NPV | Diagnostics accuracy |
|---|---|---|---|---|---|
| 4 °C | 98.36% | 100% | 100% | 99.26% | 99.26% |
| RT | 98.39% | 100% | 100% | 98.66% | 99.26% |
| 4 °C | 90.83% | 98.65% | 98.20% | 92.99% | 95.15% |
| RT | 83.19% | 91.72% | 89.19% | 86.93% | 87.88% |
| 4 °C | 82.64% | 95.92% | 94.34% | 87.04% | 89.93% |
| RT | 91.58% | 91.67% | 87.88% | 94.29% | 91.63% |
| 4 °C | 79.20% | 100% | 100% | 90.51% | 90.51% |
| RT | 79.67% | 99.33% | 98.99% | 85.55% | 90.44% |
| 4 °C | 75.63% | 100% | 100% | 83.71% | 89.18% |
| RT | 75.83% | 93.92% | 91% | 82.74% | 85.82% |
Note: RT: Room Temperature, 4 °C: 4 degree temperature.
Fig. 1Diagrammatic representation of decreasing no. of positive results among subsequent days at both temperatures compare to initial true results.
Fig. 3A & B Representing Ct Values of False Positive Samples compared to initial true negatives results at both storage Temperatures.
Types of Errors among 4 °C and Room Temperature with initial results. Very major error: Positives become negative, Major error: Negatives become positive, minor error: positive/negative to repeat sample due to internal control failure and/or single target gene amplified.
| 4 °C Temperature | |||||
|---|---|---|---|---|---|
| Very major error | 2 | 11 | 21 | 26 | 29 |
| Major error | 0 | 2 | 6 | 0 | 0 |
| Minor error | 4 | 7 | 7 | 1 | 7 |
| Room temparature | |||||
| Very major error | 2 | 20 | 17 | 25 | 29 |
| Major error | 0 | 12 | 12 | 1 | 9 |
| Minor error | 4 | 11 | 13 | 3 | 7 |
Fig. 2A & B representing Ct Value differences in Very major error samples (False negative results) from Initial true positive results in subsequent days with both storage temperatures.