Design, Development and Validation of a Knee Brace to Standardize the US Imaging Evaluation of Knee Osteoarthritis.

A Sorriento1,2, A Cafarelli1,2, P Spinnato3, A Russo4, G Lisignoli5, F Rabusseau6, P Cabras6, E Dumont6, L Ricotti1,2.   

Abstract

OBJECTIVE: A repeatable and reliable follow-up of knee injuries would be desirable to prevent delayed diagnosis and to monitor the efficacy of the applied treatment over time. Ultrasound (US) techniques are an attractive option to this purpose, since they are safe, low-cost and non-invasive. However, its use in the clinical practice is limited by the high dependency on the operator's experience. Hence, the objective of this study is to provide a standardization of the US image acquisition process for knee osteoarthritis (OA) allowing an extended clinical use of US technologies in this domain.
METHODS: Clinical specifications were provided by expert musculoskeletal radiologists thus identifying the subject poses and the US probe positions needed to evaluate the cartilage structure, signs of synovitis and joint effusion. Such considerations were used to derive the technical requirements needed for the development of a wearable brace equipped with specific openings to guide the correct placement of the probe. The feasibility of the developed wearable brace was tested on three healthy volunteers, which were asked to acquire informative US images, similar to the reference images performed by the musculoskeletal radiologist.
RESULTS: Thanks to the knee brace, the untrained subjects were able to self-acquire informative B-mode images comparable to the corresponding images acquired by an expert clinician. DISCUSSION/
CONCLUSION: The use of a knee brace intended for knee OA US diagnosis demonstrated the possibility to standardize the acquisition protocol and make its application achievable also for untrained subjects, representing a key step toward tele-ultrasonography.

Entities:  

Keywords:  Acquisition standardization; osteoarthritis; tele-ultrasonography; ultrasound imaging

Mesh:

Year:  2021        PMID: 35391755      PMCID: PMC8979410          DOI: 10.1109/JTEHM.2021.3137628

Source DB:  PubMed          Journal:  IEEE J Transl Eng Health Med        ISSN: 2168-2372


  13 in total

1.  A system for simultaneously measuring contact force, ultrasound, and position information for use in force-based correction of freehand scanning.

Authors:  Michael R Burcher; J Alison Noble; Lianghao Han; Mark Gooding
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2005-08       Impact factor: 2.725

Review 2.  Ultrasound in the study and monitoring of osteoarthritis.

Authors:  I Möller; D Bong; E Naredo; E Filippucci; I Carrasco; C Moragues; A Iagnocco
Journal:  Osteoarthritis Cartilage       Date:  2008-08-29       Impact factor: 6.576

3.  Diagnostic performance of knee ultrasonography for detecting degenerative changes of articular cartilage.

Authors:  S Saarakkala; P Waris; V Waris; I Tarkiainen; E Karvanen; J Aarnio; J M Koski
Journal:  Osteoarthritis Cartilage       Date:  2012-02-01       Impact factor: 6.576

Review 4.  Osteoarthritis: research update and clinical applications.

Authors:  Annamaria Iagnocco; Esperanza Naredo
Journal:  Rheumatology (Oxford)       Date:  2012-12       Impact factor: 7.580

Review 5.  Quantitative assessment of articular cartilage using high-frequency ultrasound: research findings and diagnostic prospects.

Authors:  Heikki J Nieminen; Y Zheng; S Saarakkala; Q Wang; J Toyras; Y Huang; J Jurvelin
Journal:  Crit Rev Biomed Eng       Date:  2009

6.  Quantitative Parameters for the Degeneration in Cartilage and Subchondral Bone of Human Knee by 3-D Ultrasound Scanning System.

Authors:  Wataru Kiyan; Yasuaki Nakagawa; Shogo Mukai; Akira Ito; Tatsuo Arai; Hiroshi Kuroki
Journal:  Ultrasound Med Biol       Date:  2020-09-24       Impact factor: 2.998

Review 7.  Weight and osteoarthritis.

Authors:  D T Felson
Journal:  Am J Clin Nutr       Date:  1996-03       Impact factor: 7.045

8.  Potential Use of Remote Telesonography as a Transformational Technology in Underresourced and/or Remote Settings.

Authors:  Linping Pian; Lawrence M Gillman; Paul B McBeth; Zhengwen Xiao; Chad G Ball; Michael Blaivas; Douglas R Hamilton; Andrew W Kirkpatrick
Journal:  Emerg Med Int       Date:  2013-01-28       Impact factor: 1.112

9.  Comparison of Diagnostic Performance of Semi-Quantitative Knee Ultrasound and Knee Radiography with MRI: Oulu Knee Osteoarthritis Study.

Authors:  Jana Podlipská; Ali Guermazi; Petri Lehenkari; Jaakko Niinimäki; Frank W Roemer; Jari P Arokoski; Päivi Kaukinen; Esa Liukkonen; Eveliina Lammentausta; Miika T Nieminen; Osmo Tervonen; Juhani M Koski; Simo Saarakkala
Journal:  Sci Rep       Date:  2016-03-01       Impact factor: 4.379

10.  Recent Developments in Tele-Ultrasonography.

Authors:  E C Constantinescu; C Nicolau; A Săftoiu
Journal:  Curr Health Sci J       Date:  2018-03-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.