Novel surgical technique of sutureless artificial iris and intraocular lens scleral fixation using Yamane technique.

Daniel Rudolf Muth1, Siegfried Georg Priglinger1, Mehdi Shajari1,2, Thomas Christian Kreutzer1, Wolfgang Johann Mayer1.   

Abstract

Purpose: To describe a novel surgical technique of a combined implantation of an artificial iris and a scleral fixated intraocular lens (IOL) using flanged IOL haptics ("Yamane" technique). Observations: The suturelessly implanted artificial iris-IOL-sandwich was stable with good functional as well as aesthetic results. However, our case showed a postoperative intraocular pressure rise. Conclusions: The presented case demonstrates that a visual as well as cosmetical rehabilitation seems to be possible even after severe, penetrating ocular trauma with profound iris defects. Importance: The sutureless IOL scleral fixation technique can also be used in combination with a sutureless artificial iris implantation. Further studies are needed to evaluate the long-term safety profile and rates of postoperative complications.
© 2022 The Authors.

Entities:  

Keywords:  Aniridia; Aphakia; Artificial iris; IOL scleral Fixation; Ocular trauma; Yamane

Year:  2022        PMID: 35387386      PMCID: PMC8978273          DOI: 10.1016/j.ajoc.2022.101502

Source DB:  PubMed          Journal:  Am J Ophthalmol Case Rep        ISSN: 2451-9936


Introduction

Trauma eyes with aniridia and aphakia remain a surgical challenge. In the past, the results have been unrewarding for the patient and the surgeon alike, in terms of aesthetics and visual acuity (VA). With a portfolio of iris protheses, intraocular lens (IOL) types and surgical techniques now available the options for cosmetical and visual rehabilitation have improved. Iris prosthesis were first implanted in the late 1950s by Peter Choyce in form of modified anterior chamber (AC) IOLs that had opaque colored haptics (available in ferrous brown, indocyanine green and cobalt blue)., Forty-four years later, in 1994, Sundmacher et al. developed a polymethyl methacrylate (PMMA) IOL with a black diaphragm (black diaphragm IOL, BDI) produced by Morcher (Morcher GmbH, Stuttgart, Germany).3, 4, 5 Now, a colored version, the Irismatch, is available with 45 colors to choose from. In 2002, Ophtec (Ophtec BV, NR Groningen, The Netherlands) introduced the Model 311 which was a PMMA IOL with a colored (black, blue, brown or green) opaque outer ring. It never achieved approval in the U.S. by the Food and Drug Administration (FDA). Recently, Ophtec launched their Model C0/1 (sulcus fixation with/without optical correction) and Model F0/1 (capsular bag fixation), produced by Reper (Reper-NN Ltd., Nizhniy Novgorod, Russia), which is a foldable acrylic iris-IOL available in a variety of 300 presets to choose from.,, In 2007, Schmidt Intraocularlinsen (Schmidt Intraocularlinsen GmbH, Sankt Augustin, Germany) developed the first flexible iris prosthesis. The company was acquired by HumanOptics (HumanOptics AG, Erlangen, Germany). HumanOptics obtained Conformité Européenne (CE, since 2011) and FDA (since 2018) approval for the Customflex Artificialiris (hereafter referred to as AI) which allows a fully customized colored anterior surface. The HumanOptics AI can be implanted in parts or as a full iris replacement. It can be implanted alone or in combination with any IOL. Even the combination with a toric and a trifocal IOL have been described., The AI, however, is not recommended for implantation in phakic eyes due to the risk of premature cataract formation., An implantation into the anterior chamber (AC) is not recommended because of the high risk of secondary glaucoma. If necessary, an explantation of the AI is possible.

Case report

A 17-year-old boy presented to our hospital with posttraumatic aphakia and aniridia on his left eye (OS) after trauma. The trauma had happened at the age of 10 years, in March 2014, and had been described as an injury with a human finger leading to a corneal perforation with rupture of the iris root and extraocular dislocation of parts of the phakic lens. A primary sclerocorneal wound closure had been performed at the day of the trauma. The following day a pars plana vitrectomy (ppv) with anterior chamber revision, complete phakic lens extraction and retinal endolaser coagulation due to retinal holes had been done. The patient presented now, almost seven years later in January 2021, with a stable aphakia and aniridia (Fig. 1a–c).
Fig. 1

a–c Corneal scars, aniridia and aphakia after trauma, primary wound closure and pars plana vitrectomy including endolaser coagulation.

a–c Corneal scars, aniridia and aphakia after trauma, primary wound closure and pars plana vitrectomy including endolaser coagulation. Despite the peripheral corneal scars, the patient had a corrected distance visual acuity (CDVA) with contact lenses of 0.00 logMAR as the optical zone was clear and intact (Table 1). However, the patient reported that he was heavily impaired by glare due to the aniridia and optical aberrations due to the peripheral corneal scars. The use of an iris-print contact lens was not tolerated.
Table 1

Patient data.

BaselinePost-surgery
Age [years] at surgery1616
IOP [mmHg]1817
BCDVA [logMAR]0.00 (contact lens)−0.10
Manifest refraction [D]+13.50/-4.25/10°+1.75/-1.50/10°
SE [D]+11.38+1.00
Axial length [mm]23.90 ± 6.0023.90 ± 6.00
Lens statusAphakiaPseudophakia (Kowa AvanseePreset UV®, +20.50 D)

BCDVA = best-corrected distance visual acuity.

D = diopters.

IOP = intraocular pressure.

SE = spherical equivalent.

Patient data. BCDVA = best-corrected distance visual acuity. D = diopters. IOP = intraocular pressure. SE = spherical equivalent. As the patient had presented with stable retinal findings, normal electrophysiology (pattern electroretinogram, ERG) and color vision (Arden's color test) for years we decided on implantation of an artificial iris (AI) to reduce glare. Furthermore, an artificial iris would allow for cosmetic rehabilitation. Combined with a scleral fixated intraocular lens (IOL), the patient would be independent on contact lenses.

Methods

Surgical rationale

The goal was to correct the aniridia and aphakia in a single surgical session to keep the total surgical trauma for the eye as low as possible. For aniridia correction, we decided to use a made-to-order HumanOptics Customflex Artificialiris Fiber Free (HumanOptics AG, Erlangen, Germany) as it allows the most customized reconstruction of an iris currently available. The fixed pupillary aperture of 3.35mm in diameter acts as a pinhole with positive effects on the contrast sensitivity. We wanted to attach the IOL for aphakia correction suturelessly to the AI by using the IOL haptics. This implied an implantation of the AI and the IOL together as a “sandwich”. Furthermore, we aimed for a sutureless scleral fixation of the AI-IOL-sandwich to avoid suture- and knot-associated complications. Therefore, we decided for a modified version of the IOL scleral fixation technique originally described by Yamane et al., hereafter referred to as ‘Yamane technique’. We used the Kowa AvanseePresetUV 3-piece IOL (Kowa Company Ldt., Nagoya, Aichi, Japan) as we had already made good experience with this IOL type combined with the Yamane technique. As reported by Mayer et al. no biometry correction factor was used for the implanted AI-IOL-sandwich as the IOL, within the sandwich, is supposed to sit at the approximate same position as when implanted in the capsular bag., Trimming of the AI to 10.5mm in diameter was done before hooking the IOL and before implantation to avoid chafing at the ciliary body and iris remnants and to prevent angle closure in case of an IOL tilt. No iridectomy was planned as we wanted to keep the AI as stable as possible as it should be attached to the IOL without sutures. Furthermore, we assumed that after trimming of the AI there should be enough space at the edges of the AI for aqueous humor circulation .

Surgical technique step-by-step

At first, two opposite scleral marks were made 2.0 mm from the limbus on the horizontal 0–180° axis. An additional scleral mark was created at the nasal side 2.0 mm below and at the temporal side 2.0 mm above the first scleral marks (Fig. 2). A pars plana 25-gauge infusion trocar was inserted to keep the intraocular pressure constant especially as the patient had already been vitrectomized previously in 2014 .
Fig. 2

Scleral marks for later IOL haptic externalization. 2.0 mm from limbus on 0–180° axis and 2.0 mm below and above 0–180° axis. 25-gauge pars plana infusion trocar.

Scleral marks for later IOL haptic externalization. 2.0 mm from limbus on 0–180° axis and 2.0 mm below and above 0–180° axis. 25-gauge pars plana infusion trocar. Trimming of the HumanOptics Customflex Artificialiris Fiber Free was made using a single-use trephine with a diameter of 10.5mm (Fig. 3a). The later penetration sites of the IOL haptics were marked at a 0–180° axis 1.0 mm away from the outer iris rim.(see Fig. 3b)
Fig. 3a

Trimming of artificial iris with 10.5 mm diameter trephine.

Fig. 3b

Marking of artificial iris at later IOL haptic penetration sites 1.0 mm from the outer iris rim on 0–180° axis.

Trimming of artificial iris with 10.5 mm diameter trephine. Marking of artificial iris at later IOL haptic penetration sites 1.0 mm from the outer iris rim on 0–180° axis. The three-piece Kowa AvanseePresetUV IOL with thin haptics was put behind the artificial iris. A bent 30-gauge thin-walled cannula was pierced through the artificial iris 1.0 mm from the outer rim at the marked sites. The haptics were inserted into the cannula and pierced through the artificial iris with the haptic ends eventually lying on top of the anterior surface of the artificial iris. The artificial iris lied centered and tension-free onto the IOL optic (Fig. 4).
Fig. 4

Artificial iris on top of IOL with pierced IOL haptics at 1.0 mm from the outer iris rim on 0–180° axis.

Artificial iris on top of IOL with pierced IOL haptics at 1.0 mm from the outer iris rim on 0–180° axis. An additional 9-0 nylon “safety suture” was pierced through the artificial iris at 2.0 mm from the outer rim. The safety suture was supposed to stay extraocular throughout the whole implantation process and acted as retractor in case the IOL and iris subluxated into the vitreous cavity during scleral fixation. A 5.5 mm sclerocorneal tunnel was prepared superior (Fig. 5a). The sandwich of IOL and artificial iris was folded using an IOL implantation forceps and was implanted through the tunnel (Fig. 5b). The safety suture ends were kept extraocular.
Fig. 5a

Preparation of sclerocorneal tunnel.

Fig. 5b

Implantation of artificial iris combined with IOL.

Preparation of sclerocorneal tunnel. Implantation of artificial iris combined with IOL. Now, the Yamane technique for scleral fixation of the IOL was completed as follows: A 30-gauge cannula was inserted ab externo starting at the scleral mark that laid below and above the horizontal axis. The cannula was moved intrasclerally to the other scleral mark that laid on the horizontal axis in a curved move penetrating the sclera and entering intraocularly at the 0° and 180° position respectively. The IOL haptics were introduced into the 30-gauge cannula using a forceps and then externalized, again in a curved move. The thin 3-piece IOL haptic was held externally with a forceps and the distal 1.0–2.0 mm were flanged using a mobile cautery device (Fig. 6). The flanged haptic end prevents the haptic to slip back through the sclera. The flange was buried subconjunctivally within the scleral tunnel. The same was done at the other side. At the end, the safety suture was pulled out through the sclerocorneal tunnel, and the tunnel was sutured with a self-resorbing polyglactin (Vicryl) 9-0 suture. The infusion trocar was removed and the sclerotomy was also sutured with a self-resorbing polyglactin (Vicryl) 9-0 suture (Fig. 7). The postoperative therapy was prednisolone acetate 1.0% eye drops, hydrocortisone 2.5% ointment and levofloxacin 5mg/ml eye drops.
Fig. 6

IOL haptic flange using a mobile, single-use cautery device.

Fig. 7

Surgery end.

IOL haptic flange using a mobile, single-use cautery device. Surgery end.

Results

During the 11-month follow-up period an increased intraocular pressure (IOP) of 60 mmHg was noted five days after surgery. The maximum preoperative IOP in our records was 22 mmHg. The increased IOP was suspected to be due to a response reaction to the topical steroids. Therefore, the postoperative treatment topical regime was changed from prednisolone acetate 10mg/ml to loteprednole 0.5% and dorzolamide 20mg/ml and timolol 5mg/ml eye drops. With this treatment the IOP could be lowered to 23 mmHg. One week after, the patient presented to our emergency department with a slight vitreous hemorrhage and an IOP increase up to 46 mmHg. As cause, we suspected a bleeding around the sclerotomy site that had slowly leaked postoperatively. Therefore, we decided for a pars plana vitrectomy (ppv) with removal of the intraocular blood. During the second procedure, a surgical iridectomy was created with intraocular scissors at the temporal-superior rim of the artificial iris. The IOP lowered to normal levels (17–22 mmHg) under timolol 5mg/ml eye drops and remained stable. The topical therapy was phased out slowly. During the 11-month follow-up period, the patient did not show any irido-pseudophakodonesis. There was a slight inferior decentration of the AI pupil aperture. The best-corrected visual acuity after surgery increased up to −0.10 logMAR and the spherical equivalent reduced from 11.38 to 1.00 diopters. Eventually, the patient was very happy with the cosmetic as well as the functional results. (Fig. 8a and b).
Fig. 8

a, b Results at follow up 1 week after surgery.

a, b Results at follow up 1 week after surgery.

Discussion

The presented case shows a good postoperative CDVA of the affected left eye. On the postoperative photos an exotropia of the left eye can be noted. However, the patient did not complain about double vision. The patient had presented to our hospital on the day of trauma for the first time. Therefore, we do not have reports about pre-trauma CDVA and strabismus status. The achievable CDVA and the fact that the trauma took place at the age of ten let us draw the conclusion that the patient had undergone a normal development of vision. Consequently, the patient should be capable of full stereo vision. An orthoptics examination is advisable during future follow-ups. Review of current literature identified inter alia 14 publications with combined HumanOptics Artificial iris-IOL implantation, 6 publications with combined Morcher Black Diaphragm Intraocular Lens (BDI) implantation, 1 publication with a combined Morcher IrisMatch implantation, 7 publications with a combined Ophtec Model 311 Iris Reconstruction Lens implantation, 1 publication with a personalized Ophtec Black Diaphragm Intraocular Lens (BDI) implantation, 2 publications with a combined Ophtec Artisan Iris Reconstruction Intraocular Lens implantation, 3 publications with a combined Ophtec/Reper Artificial Iris implantation (see Table 2).
Table 2

Literature overview of selected publications.

YearPublicationArtificial iris typeFixation of IOL to AIFixation of combined AI-IOL-sandwichMean follow-up rounded in [months]Total clinical cohort (n)
2022Mayer CS et al. 13HumanOptics Customflex Artificialiris (with fiber)Sutured: at optic-haptic junctionSulcus ciliaris, sutured transsclerally (scleral tunnel)3n = 1
2021Frisina R et al. 9Ophtec/Reper Artificial Iris Model CCombined implant by manufacturerAnterior chamber, sutured transsclerally (10-0 polypropylene, through eyelets at each of the 3 haptics, scleral flaps)12n = 4
2021Jakob-Girbig J et al. 6Ophtec/Reper Artificial Iris Model CCombined implant by manufacturerSulcus ciliaris, sutured transsclerally (10-0 polypropylene, through eyelets at each of the 3 haptics, Z-suture)Sulcus ciliaris, sutured transsclerally (10-0 polypropylene, cow hitch loop through eyelets at each of the 3 haptics, scleral flap)3–4n = 2
2021Jiang S et al. 33Ophtec/Reper Artificial Iris Model F (?)Combined implant by manufacturerAnterior chamber, sutured transsclerally (Gore-Tex suture)24n = 1
2021Mayer C et al. 19HumanOptics Customflex Artificialiris (n/a)Various techniques inter alia:Sutured: at IOL optic (haptics removed) (“Technique 4”)Various techniques inter alia:Sulcus ciliaris, sutured transsclerally (9-0 polypropylene)5n = 44
2020Mayer CS et al. 27HumanOptics Customflex Artificialiris (with fiber)Various techniques inter alia:Sutured: through proximal haptic (distal haptic ends removed) (“Technique E”)Various techniques inter alia:Sulcus ciliaris, sutured transsclerally (10-0 polypropylene)5n = 59
2020Parikakis E et al. 34Ophtec Model 311 Iris Reconstruction LensCombined implant by manufacturerSulcus ciliaris (?) sutured transsclerally (CV-8 Gore-Tex, cow hitch knot at haptic eyelets, 4-point fixation)6n = 1
2020Riedl JC et al. 35HumanOptics Customflex Artificialiris (n/a)Various techniques inter alia:Sutureless: enclavation of iris fixated IOL (“Sandwich-Implantat”)Sutured? (“Hybridimplantation”)Various techniques inter alia:Sulcus ciliaris, sutureless behind residual irisSulcus ciliaris, sutured transsclerally (10-0 polypropylene)41n = 51
2020Sousa Silva R et al. 10HumanOptics Customflex Artificialiris (with fiber)Sutured: through IOL eyelets (10-0 polypropylene)Sulcus ciliaris, sutured transsclerally (7-0 CV-8 Gore-Tex)6n = 4
2020Wang et al. 7Ophtec Model 311 Iris Reconstruction LensCombined implant by manufacturerVarious techniques inter alia:Sulcus ciliaris, suturelessSulcus ciliaris sutured transsclerally (9-0 polypropylene, scleral flaps as needed)60n = 38
2020Wolf A et al. 12HumanOptics Customflex Artificialiris (n/a)Sutured: pierced through IOL plate haptic edges (4-point fixation)Sulcus ciliaris, sutured transsclerally (Hoffman scleral pockets)1–6n = 2
2019Yoeruek E et al. 25HumanOptics Customflex Artificialiris (n/a)Sutureless: 3-piece IOL haptics directly through 0.9mm stab incisions (“belt-loop”)Sulcus ciliaris (?), sutured transsclerally (10-0 polypropylene, Z-suture, 3-point fixation)25n = 5
2017Mayer C et al. 16HumanOptics Customflex Artificialiris (with fiber)Various techniques inter alia:Sutured: through proximal haptic (distal haptic ends removed) (“Technique E”)Sulcus ciliaris, sutured transsclerally (10-0 polypropylene)NAn = 51
2017Villemont AS et al. 36Morcher Black Diaphragm Intraocular Lens (BDI) Model 67F and 67FMorcher IrisMatch Model 30BCombined implant by manufacturerSulcus ciliaris sutured transsclerally (10-0 polypropylene, scleral flaps)32n = 17
2016Gunenc U et al. 37HumanOptics Customflex Artificialiris (n/a)Sutured: through haptic eyelets of PMMA IOL (9-0 polypropylene)Sulcus ciliaris (?), sutured transsclerally (9-0 polypropylene, scleral tunnel)0,25n = 1
2016Spitzer MS et al. 14HumanOptics Customflex Artificialiris (n/a)Various techniques inter alia:Sutureless: 3-piece IOL haptics directly through 0.9mm stab incisions (“belt-loop”)Various techniques inter alia:Sulcus ciliaris, sutured transsclerally (10-0 polypropylene, Z-suture, 2-point fixation)24n = 34
2015Mayer CS et al. 38HumanOptics Customflex Artificialiris (with fiber)Various techniques inter alia:SuturedVarious techniques inter alia:Sulcus ciliaris, sutured transsclerally14n = 36
2014Gooi P et al. 24HumanOptics Customflex Artificialiris (with fiber)Sutureless: 3-piece IOL haptics directly through stab incisions (“belt-loop”)Sulcus ciliaris, sutured transsclerally (scleral flapsl, optionally fibrin glue)4n = 1
2013Forlini C et al. 29HumanOptics Customflex Artificialiris (n/a)Various techniques inter alia:Sutureless: retropupillary and prepupillary enclavation of iris-claw IOL (both unsuccessful)SuturedVarious techniques inter alia:Sulcus ciliaris, sutured transsclerally (10-0 polypropylene. scleral pockets, 3- and 4-point fixation)6n = 4
2012De Grande V et al. 39Ophtec Model 311 Iris Reconstruction LensCombined implant by manufacturerPosterior chamber, sutured transsclerally (CV-8 Gore-Tex, 3-point fixation at 2 haptic eyelets and 1 “sling suture” at haptic-optic junction, scleral patch, fibrin glue)n/an = 1
2012Spitzer MS et al. 28HumanOptics Customflex Artificialiris (n/a)Sutureless: 3-piece IOL haptics directly through 0.9mm stab incisions (“belt-loop”)Sulcus ciliaris, sutured transsclerally (10-0 polypropylene, Z-suture, 2-point fixation)4–48n = 4
2011Mashor RS et al. 40Morcher Black Diaphragm Intraocular Lens (BDI) Model 67G and 67FOphtec Model 311 Iris Reconstruction LensCombined implant by manufacturerSulcus ciliaris (?), sutured transsclerally2–44n = 11
2009Roman S et al. 41Various devices inter alia:Ophtec Model 311 Iris Reconstruction LensMorcher Black Diaphragm Intraocular Lens (BDI)Combined implant by manufacturerVarious techniques inter alia:Sulcus ciliaris, suturelessSulcus ciliaris, sutured transsclerally (10-0 polypropylene)44n = 7
2008Aslam SA et al. 42Morcher Black Diaphragm Intraocular Lens (BDI) Model 67FCombined implant by manufacturerSulcus ciliaris, suturelessSulcus ciliaris, sutured transsclerally42n = 40
2008Sminia ML et al. 43Ophtec Artisan Iris Reconstruction Intraocular LensCombined implant by manufacturerAnterior chamber, sutureless, enclavated to residual iris106n = 5
2007Miller AR et al. 44Ophtec Model 311 Iris Reconstruction LensCombined implant by manufacturerSulcus ciliaris, sutured transsclerally (10-0 polypropylene)3–36n = 9
2005Menezo Jl et al. 45Various implants inter alia:Morcher Black Diaphragm Intraocular Lens (BDI) (Model 67G and 67F)Ophtec Black Diaphragm Intraocular Lens (BDI) (personalized)Combined implant by manufacturerCapsular bag, suturelessSulcus ciliaris, sutured transsclerally (10-0 polypropylene, scleral flap)22n = 7
2004Price MO et al. 46Ophtec Model 311 Iris Reconstruction LensCombined implant by manufacturerCapsular bag, suturelessSulcus ciliaris, suturelessSulcus ciliaris, sutured transsclerally12n = 10
2003Hanumanthu S et al. 47Ophtec Artisan Iris Reconstruction Intraocular Lens (custom-made, colored polymethyl methacrylate)Combined implant by manufacturerAnterior chamber, sutureless, enclavated to residual iris10n = 1
2000Reinhard T et al. 48Morcher Black Diaphragm Intraocular Lens (BDI) Model 67 (various versions)Combined implant by manufacturerSulcus ciliaris, sutureless46n = 19
Literature overview of selected publications.

AI trimming

Preoperative biometry of the left eye showed a horizontal white-to-white (HWTW) distance of 11.9mm. The axial length of both eyes was normal (right: 24.1, left: 23.9 mm). Trephination of the AI makes an angle closure unlikely as there should be sufficient flow of aqueous humor around the peripheral rim of the artificial iris. In literature, there are different opinions on whether to trephine and on how much to trephine. Bonnet et al. suggest to use larger AI diameters and to avoid trephination if possible. Ayliffe et al. suggest to add +0.5mm to the HWTW diameter. Mayer et al. would aim for an AI diameter of 9.5mm for an in-the-bag implantation and 11.5–12.5mm for sulcus-ciliaris implantation. Rickman et al. would subtract 0.5–1.0mm from the WTW distance and rather aim for a smaller AI diameter with a target of 11.0mm and 10.0mm in small eyes. For children, even smaller AI diameters of 8.0mm should be considered as Jusufovic et al. report. In another publication Mayer et al. give the advice of trimming the AI diameter to WTW minus 0.5mm and point out that a removal of the residual iris in sutureless sulcus implantation of the AI might have a positive effect in the prevention of secondary glaucoma. Gooi et al. recommend to assure that there is enough clearance of the outer AI rim to the IOL haptics to avoid iris and ciliary body chafing. Yoeruek et al. suggest to subtract 1.0mm from the WTW distance. Investigations of trimmed artificial iris suggest that AI with fiber show sharp fiber ends sticking out of the trimmed AI edge, being a possible cause for prolonged postoperative inflammation and bleeding., As our patient showed some residual iris, we aimed for smaller AI diameter of 10.5mm to avoid contact of the artificial iris with the ciliary body in the event of an IOL tilt. Furthermore, we worked with a single-use trephine to assure best-possible sharpness of the instrument and to create a cutting edge of the fiber-free AI that was as smooth as possible. The AI without fiber has shown less risk for secondary glaucoma.

Sutured vs. sutureless

We favored a completely sutureless approach to avoid suture- and knot-associated complications such as suture erosion, suture breaks and suture loosening especially with regards to the long-term stability considering the young age of our patient.,,, Therefore, even established suture techniques as well as novel suture-based approaches in combination with an AI such as the Z suture described by Spitzer et al., Rickann et al. and Yoeruek et al. or the “slip-and-slide” technique by Wolf et al. were not amongst our options. Forlini et al. reported the attempt of enclavating an iris-claw IOL to an AI. An appealing approach because of its technical simplicity. However, the execution was reported unsuccessful. Thus, we decided for a 3-piece IOL. A 3-piece IOL held the possibility to suturelessly attach the IOL to the AI by the IOL haptics which would keep all possibilities for scleral fixation of the AI-IOL-sandwich., Spitzer et al. and Gooi et al. were amongst the first to publish a direct sutureless fixation of a 3-piece IOL to the AI using the IOL haptics held by created “belt-loops” within the AI., To reduce manipulation of the AI we chose a facilitated approach by simply piercing the IOL haptics through the peripheral AI with the help of a 30-gauge thin-walled cannula. In our opinion, the advantage of the Yamane technique in combination with a haptic fixated AI-IOL sandwich technique lies in its relative technical easiness and robustness. The Yamane technique keeps the scleral trauma to a minimum by not using intrascleral sutures (Z suture), tissue glue or scleral flaps (e.g. Hoffman scleral pockets) which may be impossible to create in severely traumatized eyes.

Fiber vs. fiber-free

The fiber-free version of the HumanOptics Customflex Artificial Iris is less rigid and hence easier to fold and to implant, especially when combined with an IOL. Furthermore, the fiber-free artificial iris is lighter and thinner which reduces the risk of an IOL tilt and thereby reduces the risk of irritation and chafing of the iris remnants. A healthy phakic crystalline lens shows a mean unaccommodated thickness of around 4mm., The total thickness of the AI-IOL sandwich is estimated as follows: AI thickness 0.25mm (peripheral) to 0.40mm (central), IOL thickness 0.50mm–1.00mm and residual iris thickness approximately 0.50mm which is in sum less than a crystalline lens. A HumanOptics Artificial Iris fiber-free approximately weighs 31–34 g when trimmed to a diameter of 10.5mm according to the manufacturer. The manufacturer Kowa states an approximate mass for the IOL type AvanseePreset UV of 1,9 g. Consequently, the package of artificial iris and IOL weighs approximately 17 times more than the IOL alone. Moreover, trimming the AI seems to bear less complications with the fiber-free version as described above. With the presented technique no sutures remain within the AI. In this regard the fiber-containing version would not offer additional benefit. Previous studies have shown good results using the fiber-free version of the artificial iris.,, However, the AI is supposed to stay centered on top of the IOL via the IOL haptics. Spitzer et al. used both, the fiber and the fiber-free, version of the AI and they did not see the necessity to use the fiber-containing version as long as complete iris prostheses are implanted. Gooi et al. used the fiber version for their “belt-loop” technique.

IOP and iridectomy

The postoperative IOP of our patient increased. He was under topical dorzolamide 20mg/ml and timolol 5mg/ml and loteprednole 0.5%. Due to the vitreous hemorrhage and the conservatively difficult to control IOP we made the decision for a pars plana vitrectomy. Our main concern was that an IOL tilt might push the AI anteriorly causing an angle closure. Even though the artificial iris did not show a bombée configuration, we performed a surgical iridectomy of the artificial iris during that same intervention. As the AI was stable and required no further manipulation the performed scissors-iridectomy should not decrease the stability of the AI as reported elsewhere. Even without fiber, the AI material is comparatively rigid. We therefore assume that a clinically visible iris bombée indicating for a pupillary block will not form in the same manner as it would do with a normal iris. Consequently, we did not dare to rely on the clinical presence of an iris bombée especially as we had already decided for the second intervention. The iridectomy was done as safety precaution also with regards to a possible future tilt of the Yamane IOL that might push the artificial iris forward and might cause an angle narrowing.

Alternative combined implants

As mentioned in the introduction, besides the HumanOptics Customflex Artificial Iris there are other iris prostheses that can be combined or are already combined with an IOL. Apart from the Ophtec/Reper Artificial Iris the HumaOptics Artificial Iris is the only foldable implant allowing small incisions 2.8–3.0mm. The Ophtec/Reper AI offers a color range of 300 presets to choose from whereas the HumanOptics AI is individually manufactured based on photographs of the partner eye and of iris remnants. Furthermore, the Ophtec/Reper Model A, C and F devices are only available with one IOL type whereas the HumanOptics AI theoretically can be combined with any IOL in the market. Therefore, we preferred the HumanOptics AI above the other combined AI-IOL devices that would have been available.

Conclusion

The sutureless scleral fixation technique described by Yamane et al. is a currently very popular and successful but comparatively young technique. Only very limited data on long-term stability and tilt resistance of these scleral fixated IOLs exists. Moreover, to our best knowledge, this report is the first to describe an IOL with flanged haptics that piggybacks an artificial iris. This technique is as minimally invasive and technically straight-forward as possible on the premise of avoiding sutures and their inherent long-term complication risks. However, especially stability has to be closely followed. A limitation of this report is the short follow-up time of 11 months. The patient will be seen and monitored during future appointments. Further improvements of the described technique are possible, such as using a modified 2.8–3.0 mm IOL cartridge to fold and shoot the IOL with the attached artificial iris through a clear cornea incision. This would have the advantage of a surgical technique being completely sutureless. Long-term follow-ups of this case and similar cases are needed to prove stability and low complication rates.

Patient consent

This report does not contain any personal information that could lead to the identification of the patient. Approval of the Ethics Committee of the Ludwig-Maximilians-University (LMU) Munich was obtained (reference number: 21–0747).

Funding

No funding or grant support.

Conflicts of interest

The following authors have no financial disclosures regarding this publication: DRM, SGP, MS, TCK, WJM.

Authorship

All authors attest that they meet the current ICMJE criteria for authorship.

CRediT author statement

DRM: Validation, Writing - Review & Editing, Supervision, Writing - Original Draft, Visualization, Data Curation. SGP: Validation, Writing - Review & Editing, Supervision. MS: Validation, Writing - Review & Editing. TCK: Validation, Writing - Review & Editing. WJM: Validation, Writing - Review & Editing, Supervision, Writing - Original Draft, Visualization, Conceptualization.
  47 in total

1.  Scleral fixation without conjunctival dissection.

Authors:  Richard S Hoffman; I Howard Fine; Mark Packer
Journal:  J Cataract Refract Surg       Date:  2006-11       Impact factor: 3.351

2.  Residual Iris Retraction Syndrome After Artificial Iris Implantation.

Authors:  Christian S Mayer; Andrea E Laubichler; Michael Masyk; Philipp Prahs; Daniel Zapp; Ramin Khoramnia
Journal:  Am J Ophthalmol       Date:  2018-09-17       Impact factor: 5.258

3.  Implantation of iris devices in congenital and traumatic aniridias: surgery solutions and complications.

Authors:  J L Menezo; R Martínez-Costa; A Cisneros; M C Desco
Journal:  Eur J Ophthalmol       Date:  2005 Jul-Aug       Impact factor: 2.597

4.  Ophtec iris reconstruction lens United States clinical trial phase I.

Authors:  Marianne O Price; Francis W Price; David F Chang; Kathy Kelley; Michael D Olson; Kevin M Miller
Journal:  Ophthalmology       Date:  2004-10       Impact factor: 12.079

5.  Customized humanoptics silicone iris prosthesis in eyes with posttraumatic iris loss: outcomes and complications.

Authors:  Martin Stephan Spitzer; Anja Nessmann; Julia Wagner; Efdal Yoeruek; Karl Ulrich Bartz-Schmidt; Peter Szurman; Gesine B Szurman
Journal:  Acta Ophthalmol       Date:  2016-01-25       Impact factor: 3.761

6.  Implantation of the Artisan iris reconstruction intraocular lens in 5 children with aphakia and partial aniridia caused by perforating ocular trauma.

Authors:  Marije L Sminia; Monica Th P Odenthal; Nitza Gortzak-Moorstein; Liesbeth J J M Wenniger-Prick; Hennie J Völker-Dieben
Journal:  J AAPOS       Date:  2008-03-10       Impact factor: 1.220

7.  Treatment of congenital aniridia associated with subluxated infantile cataract.

Authors:  Vahid Jusufovic; Emir Cabric; Allen Popovic-Beganovic; Zlatko Musanovic; Jasmin Zvornicanin
Journal:  Med Arch       Date:  2014-05-31

8.  Functional outcomes after combined iris and intraocular lens implantation in various iris and lens defects.

Authors:  Christian S Mayer; Andrea M Hoffmann; Phillipp Prahs; Lukas Reznicek; Ramin Khoramnia
Journal:  BMC Ophthalmol       Date:  2020-09-15       Impact factor: 2.209

9.  Challenges and Complication Management in Novel Artificial Iris Implantation.

Authors:  Christian S Mayer; Andrea E Laubichler; Ramin Khoramnia; Tamer Tandogan; Philipp Prahs; Daniel Zapp; Lukas Reznicek
Journal:  J Ophthalmol       Date:  2018-09-23       Impact factor: 1.909