Parounak Zelveian1, Jirar Topouchian2, Zoya Hakobyan1, Jennifer Asmar3, Heghine Gharibyan1, Roland Asmar4.
Abstract
Background: Electronic devices for blood pressure (BP) measurements require independent clinical validation as recommended by various authorities/societies, both in general and special populations such as pregnancy. Objective: To assess the accuracy of the Withings BPM Connect device in pregnancy and pre-eclampsia according to the Universal Standard Validation Protocol.Entities:
Keywords: Withings; accuracy; blood pressure measurements; home blood pressure; pre-eclampsia; pregnancy; validation
Mesh:
Year: 2022 PMID: 35370407 PMCID: PMC8967988 DOI: 10.2147/VHRM.S351313
Source DB: PubMed Journal: Vasc Health Risk Manag ISSN: 1176-6344
Figure 1The Withings BPM Connect device.
Characteristics of the Study Participants (n=45)
| Mean ± SD | Range | |
|---|---|---|
| Age (years) | 31 ± 5 | 21–42 |
| Arm circumference (cm) | 28.8 ± 2.6 | 23.5–35.5 |
| Entry SBP R0 (mmHg) | 135 ± 21 | 89–164 |
| Entry DBP R0 (mmHg) | 89 ± 14 | 59–111 |
| Normotensive women | 15 | |
| Hypertensive women | 15 | |
| Pre-eclampsia women | 15 | |
Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure; SD, standard deviation.
Validation Study Results According to Protocol Requirements (Criterion 1)
| Pass Requirement | SBP | DBP | |
|---|---|---|---|
| Mean BP difference (mmHg) | ≤ 5 | −0.5 | −0.8 |
| SD (mmHg) | ≤ 8 | 5.7 | 3.8 |
| Pass | Pass | ||
Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure; SD, standard deviation.
Blood Pressure Differences According to Subgroup of Patients
| Pregnancy Status | Number | Mean SBP Difference ± SD (mmHg) | Mean DBP Difference ± SD (mmHg) |
|---|---|---|---|
| Normotensive | 15 | 0.40 ± 5.64 | −1.54 ± 3.11 |
| Hypertensive | 15 | −1.99 ± 5.89 | −1.13 ± 4.30 |
| Pre-eclampsia | 15 | 0.14 ± 5.50 | 0.39 ± 3.68 |
Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure; SD, standard deviation.
Figure 2Standardized Bland-Altman scatter plots of the Test-Reference BP differences against their mean values. Panel A SBP = Systolic blood pressure, Panel B DBP = Diastolic blood pressure.