Jirar Topouchian1, Zoya Hakobyan2, Jennifer Asmar3, Svetlana Gurgenian4, Parounak Zelveian2, Roland Asmar5.
Abstract
BACKGROUND: Electronic devices for blood pressure (BP) measurements need to go through independent clinical validation as recommended by different authorities, both in general and in special populations such as pregnancy.Entities:
Keywords: blood pressure; blood pressure measurements; home blood pressure; pre-eclampsia; pregnancy; self-BP measurement; validation
Mesh:
Year: 2018 PMID: 30214220 PMCID: PMC6124447 DOI: 10.2147/VHRM.S165524
Source DB: PubMed Journal: Vasc Health Risk Manag ISSN: 1176-6344
Procedure for reference and test device BP measurements in the same arm sequential validation protocol
| BPA | Entry BP, observers 1 and 2 each with the mercury standard |
| BPB | Device detection BP, supervisor |
| BP1 | Observers 1 and 2 with mercury standard |
| BP2 | Supervisor with the test instrument |
| BP3 | Observers 1 and 2 with mercury standard |
| BP4 | Supervisor with the test instrument |
| BP5 | Observers 1 and 2 with mercury standard |
| BP6 | Supervisor with the test instrument |
| BP7 | Observers 1 and 2 with mercury standard |
Grading criteria achieved with the Omron Evolv® device in pregnancy and pre-eclampsia
| GRADE A, B, C
| ||||
|---|---|---|---|---|
| ≤5 mm | ≤10 mm | ≤15 mm | GRADE | |
| SBP, % | 97% | 100% | 100% | A |
| DBP, % | 99% | 100% | 100% | A |
| SBP, n | 131 | 135 | 135 | |
| DBP, n | 134 | 135 | 135 | |
Notes: Grades represent the cumulative percentage of readings with differences between device and observer falling within 5, 10 and 15 mmHg respectively.
Figure 1Bland–Altman plots of systolic and diastolic BP differences between the Omron Evolv® (HEM-7600T-E) device and the observers’ readings.
Notes: (A) SBP. (B) DBP.
Abbreviations: BP, blood pressure; SBP, systolic blood pressure; DBP, diastolic blood pressure.
Grading criteria achieved with the Omron M3 Comfort® device in pregnancy and pre-eclampsia
| GRADE A, B, C
| ||||
|---|---|---|---|---|
| ≤5 mm | ≤10 mm | ≤15 mm | GRADE | |
| SBP, % | 90% | 100% | 100% | A |
| DBP, % | 97% | 100% | 100% | A |
| SBP, n | 121 | 135 | 135 | |
| DBP, n | 131 | 135 | 135 | |
Notes: Grades represent the cumulative percentage of readings with differences between device and observer falling within 5, 10 and 15 mmHg respectively.
Figure 2Bland–Altman plots of systolic and diastolic BP differences between the Omron M3 Comfort® (HEM-7134-E) device and the observers’ readings.
Notes: (A) SBP. (B) DBP.
Abbreviations: BP, blood pressure; SBP, systolic blood pressure; DBP, diastolic blood pressure.