Biomechanical Comparative Study for Osteosynthesis of Pauwels Type III Femoral Neck Fractures: Conventional Devices versus Novel Fixed Angle Devices.

Dae-Hyun Park1, Young-Chae Seo1, Yong-Uk Kwon1, Soo-Hwan Jung1, Seung-Jin Yoo2.   

Abstract

Purpose: Osteosynthesis has been recommended for treatment of Pauwels type III femoral neck fractures in young patients. However, no implant of choice has been reported so far. This study was conducted in order to compare the fixation stability of two conventional fixation methods with three different novel fixed angle devices in this type of fracture. Materials and
Methods: A total of 25 composite femurs (4th Generation Saw bone; Pacific Research Laboratories, USA) corresponding to human bone were used. Pauwels type III fracture type was uniformly reproduced. Specimens were fixed with a cannulated screw, cannulated screw with cable, and Intertan nail, dynamic hip screw, and IKEY nail. Measurement of failure loads and the rotational change of the femoral head fragment was performed for evaluation of fixation stability.
Results: All implants were compared with cannulated screw and dynmaic hip screw. No meaningful improvement was observed for the cannulated screw with cable compared with the cannulated screw and dynamic hip screw. Meaningful improvement in load-to-failure and y-rotation and z-rotation was observed for both the Intertan nail and IKEY nail compared with the cannulated screw. However, compared with the dynamic hip screw, only the IKEY nail showed improvement in the same profile but the Intertan nail did not.
Conclusion: Among novel fixed angle devices, meaningful improvement was observed for the IKEY nail compared with conventional implants. Strengths of this implant include biomechanical stability and simplicity of surgical technique, indicating that it may be another good option for osteosynthesis of Pauwels type III femoral neck fractures.
Copyright © 2022 by Korean Hip Society.

Entities:  

Keywords:  Bone screws; Femoral neck fractures; Femur; Fracture fixation

Year:  2022        PMID: 35355625      PMCID: PMC8931947          DOI: 10.5371/hp.2022.34.1.35

Source DB:  PubMed          Journal:  Hip Pelvis        ISSN: 2287-3260


INTRODUCTION

Femoral neck fractures typically occur in young patients as a result of high-energy trauma or in elderly patients with osteoporosis12). Fractures occurring in the former age group commonly show a pattern of Pauwels type III fractures. Osteosynthesis has been a treatment of choice for this type of fracture due to longer life expectancy and high activity level345). Internal fixation using three cannulated screws, which is sufficient for management of Pauwels type I and II femoral neck fractures36), resulted in high complication rates, including femoral neck shortening, non-union, and avascular necrosis. Fixed angle devices including dynamic hip screws have been introduced in an effort to resist the shearing force, not only the compressive forces78). Currently, cannulated screw fixation and dynamic hip screw fixation have been proposed as a conventional method19). Meanwhile, numerous biomechanical and clinical studies regarding other types of fixed angle devices are being conducted; however, an implant of choice has yet to be identified. In an effort to discover optimal methods, our study was conducted as a biomechanical comparison of two types of conventional devices (cannulated screws, dynamic hip screw) with three types of novel fixed-angle devices; cannulated screws with cable, Intertan nail, and IKEY nail.

MATERIALS AND METHODS

1. Specimen Preparation

A total of 25 composite femurs (4th Generation Saw bone; Pacific Research Laboratories, Vashon, WA, USA) corresponding to human bone were used. All specimens had a 2-mm diameter canal passing through the center of the femoral head and neck, which is used as a landmark for insertion of screws (Fig. 1). In order to ensure similar bone quality, C-arm radiographs were used in evaluation of all specimen femurs to exclude any specimen with pathologic lesions. Pauwels type III fracture type was uniformly reproduced in 25 composite femur models using custom-made saw guides, showing a 70° oblique fracture line (Fig. 2).
Fig. 1

4th generation saw bone (Pacific Research Laboratories, USA) was used for femur specimen. All specimen have 2-mm diameter canal passing through center of femoral head and neck. This figure is image reconstructed by three-dimensional computed tomography, including the canal.

Fig. 2

(A) In order to reproduce uniform fracture line in 25 composite femur, custom-made saw guide is made. It consists of two parts, divided by 70° oblique line which represents Pauwels type III fracture. (B) Custom-made saw guide applied in cephalocervical area of the specimen. (C) After guide application, manual cutting was done along the line.

2. Surgical Technique

Specimens were fixed with a cannulated screw (Type 1), cannulated screw with cable (Type 2), Intertan nail (Type 3), dynamic hip screw (Type 4), and IKEY nail system (Type 5) (Fig. 3). The surgical procedure for these models was established by one orthopedic surgeon. Cannulated screw fixation was performed using the partially threaded 7.0 mm IKEY cannulated screw system. In order to reproduce uniform implant fixation, screws are inserted in an inverted triangular shape, each at the point 5 mm away from the canal passing the specimen. The Type 2 method was performed by wrapping a cable after cannulated screw fixation. Following insertion of three cannulated screws, the same as Type 1, a cable was passed through a hole in the head of the cannulated screws. After passing the wire, which eventually formed a round shape, both ends of the cable were tightened and wrapped. Intertan nail and IKEY nail fixation was performed using a 9-mm diameter nail. The starting point was determined to be slightly medial to the tip of the greater trochanter. After nail insertion, two and three cephalocervical screws were fixed in each group, passing through apertures in each nail. Finally, distal locking screws were inserted. The Type 4 method was performed using the CHS Plate System (Orthotech, Hanam, Korea). Three-hole plates were used and secured to the femoral shaft with one lag screw and three cortical screws. After implant fixation in all five types, reduction was confirmed under C-arm radiographs. All specimens showed less than 2 mm of displacement and less than 5° of angulation in anterior-posterior view and axial view, which satisfied our conventional criteria of anatomical reduction10). The distal femoral condyles of all instrumented specimens were removed, and the femoral shafts were potted with zigs at an angle of valgus 25°. This degree of angle was set to simulate normal one-legged weight bearing stance (Fig. 4).
Fig. 3

Five different types of osteosynthesis were done in femur specimen. In order to reproduce uniform implant fixation in type I, cannulated screw was inserted 5 mm away from the central canal in inverted triangular fasion.

Fig. 4

For the experiment, sawbone was fixed with the angle of 25, which presumes anatomical axis of femur. MTS 858 Bionix test system was used for preliminary, dynamic, and static loading. Each settings for loading were chosen to assume fast walking at the period of bone consolidation 6 weeks after surgery.

3. Biomechanical Testing

Tests were performed in the order of preliminary (100 N, 20 N/min), dynamic (75-750 N, 10,000 cycles, 2 Hz), and static (10 mm/min) loading using a universal testing machine (MTS 858 Bionix test system; MTS System, Eden Prairie, MN, USA) (Fig. 4). These loading rates were selected to assume fast walking at the period of bone consolidation six weeks after surgery. Measurement of failure loads and the rotational change of the femoral head fragment was performed for evaluation of fixation stability. Load-to-failure was measured at the time when the fracture of the specimen occurred. Rotatory deformities of the femoral head fragment were measured by tracking the position of three markers (diameter=1 mm) placed in front of the femoral head using a 3-D Coordinate Measuring Machine (Microscribe G2L; Immersion, San Jose, CA, USA). The x-rotation was defined as femoral head rotation occurring in the sagittal plane, viewing the specimen on the medial side. This type of rotation represents the rotation of the femoral head itself around the longitudinal axis of the femoral neck. The y-rotation was defined as femoral head rotation occurring in the coronal plane, viewing the specimen on the anterior side. This type of rotation represents the valgus or varus deformity of the fractured femoral neck. The z-rotation was defined as femoral head rotation occurring in the axial plane, viewing the specimen on the superior side. This type of rotation represents anterecurvatum deformity of the fractured femoral neck. The results were recorded as positive when the direction of the rotation is counterclockwise and vice versa in all three axes (Fig. 5).
Fig. 5

Rotational deformity of the femoral head fragment was measured in three different planes. (A) X-rotation is rotation occured in sagittal plane, representing femoral head rotation around the longitudinal axis of femoral head. (B) Y-rotation is rotation occured in coronal plane, representing valgus or varus deformity. (C) Z-rotation is rotation occured in axial plane, representing anterecurvatum deformity.

4. Statistical Analysis

Statistical analysis was performed using the IBM SPSS Statistics software (ver. 25; IBM, Armonk, NY, USA). A normality test was performed for screening of normal distribution of load-to-failure and rotational deformities of each five groups. Because the distribution did not approximate a normal distribution, an analysis was performed using the Kruskal–Wallis test. Statistical significance was set at P<0.05, and when a significant probability was found using the Kruskal–Wallis test, a post-hoc analysis was performed using the Bonferroni method. Because load-to-failure and rotational deformity showed statistical significance, we moved on to the next step. Statistical differences between all novel fixed angle devices (cannulated screw with cable, Intertan nail, IKEY nail) and conventional devices (cannulated screw, dynamic hip screw) were compared. Because the distribution did not approximate a normal distribution, statistical differences were assessed using the Wilcoxon rank sum test.

RESULTS

A summary of the mean and standard deviation of load-to-failure and the degree of rotational deformity for each types of device is shown in Tables 1 and 2, respectively. Means are visualized by graphs in Fig. 6 and 7. A summary of the comparison between conventional devices and novel fixed-angle devices is shown in Tables 3, 4, 5, 6.
Table 1

Load-to-Failure of 5 Different Implants

Type 1Type 2Type 3Type 4Type 5P-value*
Load-to-failure (n)2,039.57±252.462,613.72±291.333,549.28±544.202,957.75±87.544,533.28±86.87<0.001

Values are presented as mean±standard deviation.

Type 1: cannulated screw, Type 2: cannulated screw with cable, Type 3: Intertan nail, Type 4: dynamic hip screw, Type 5: IKEY nail.

*Kruskall–Wallis test was done, and post-hoc analysis was performed with the significance level corrected by the Bonferroni method.

Table 2

Rotational Deformity of 5 Different Implants (unit: °, absolute value)

Type 1Type 2Type 3Type 4Type 5P-value*
X-axis0.27±0.210.27±0.160.65±0.240.27±0.150.05±0.03<0.001
Y-axis1.44±1.071.27±0.890.22±0.111.19±0.460.05±0.030.007
Z-axis2.25±1.781.32±0.851.03±0.442.05±0.720.04±0.020.01

Values are presented as mean±standard deviation.

Type 1: cannulated screw, Type 2: cannulated screw with cable, Type 3: Intertan nail, Type 4: dynamic hip screw, Type 5: IKEY nail.

*Kruskall–Wallis test was done, and post-hoc analysis was performed with the significance level corrected by the Bonferroni method.

Fig. 6

Bar and whisker chart for load-of-failure of each devices. Values are presented as mean±standard deviation.

Fig. 7

Bar and whisker chart for rotational deformity of each devices. Three separate charts stands for deformity of x-rotation, y-rotation, and z-rotation, respectively. Values are presented as mean±standard deviation.

Table 3

Comparison of Load-to-Failure between Novel Fixed Angle Devices (Type 2, 3, 5) with Cannlated Screws Fixation (Type 1)

Load-to-failure (n)P-value*
Type 1Type 2
2,035.57 (1,829.01-2,451.78)2,604.78 (2,319.06-3,084.39)0.07
Type 1Type 3
2,035.57 (1,829.01-2,451.78)3,538.18 (2,801.09-4,338.52)0.03
Type 1Type 5
2,035.57 (1,829.01-2,451.78)4,531.28 (4,415.29-4,649.39)0.03

Values are presented as median (interquartile range).

Type 1: cannulated screw, Type 2: cannulated screw with cable, Type 3: Intertan nail, Type 5: IKEY nail.

*Wilcoxon rank sum test.

Table 4

Comparison of Load-to-Failure between Novel Fixed Angle Devices (Type 2, 3, 5) with Dynamic Hip Screw Fixation (Type 4)

Load-to-failure (n)P-value*
Type 4Type 2
2,957.76 (2,837.76-3,073.86)2,604.78 (2,319.06-3,084.39)0.18
Type 4Type 3
2,957.76 (2,837.76-3,073.86)3,538.18 (2,801.09-4,338.52)0.18
Type 4Type 5
2,957.76 (2,837.76-3,073.86)4,531.28 (4,415.29-4,649.39)0.03

Values are presented as median (interquartile range).

Type 2: cannulated screw with cable, Type 3: Intertan nail, Type 4: dynamic hip screw, Type 5: IKEY nail.

*Wilcoxon rank sum test.

Table 5

Comparison of Rotational Deformity between Novel Fixed Angle Devices (Type 2, 3, 5) with Cannlated Screws Fixation (Type 1)

Degree of deformity (, absolute value)P-value*
Type 1Type 2
X-axis0.21 (0.03-0.52)0.24 (0.05-0.46)>0.99
Y-axis0.81 (0.59-3.04)0.71 (0.57-2.57)0.69
Z-axis1.50 (0.70-5.08)0.92 (0.46-2.58)0.43
Type 1Type 3
X-axis0.21 (0.03-0.52)0.75 (0.35-0.93)0.13
Y-axis0.81 (0.59-3.04)0.18 (0.10-0.38)0.03
Z-axis1.50 (0.70-5.08)0.82 (0.62-1.63)0.32
Type 1Type 5
X-axis0.21 (0.03-0.52)0.06 (0.01-0.08)0.09
Y-axis0.81 (0.59-3.04)0.06 (0.01-0.08)0.03
Z-axis1.50 (0.70-5.08)0.03 (0.02-0.06)0.03

Values are presented as median (interquartile range).

Type 1: cannulated screw, Type 2: cannulated screw with cable, Type 3: Intertan nail, Type 5: IKEY nail.

*Wilcoxon rank sum test.

Table 6

Comparison of Rotational Deformity between Novel Fixed Angle Devices (Type 2, 3, 5) with Dynamic Hip Screw Fixation (Type 4)

Degree of deformity (, absolute value)P-value*
Type 4Type 2
X-axis0.21 (0.08-0.44)0.24 (0.05-0.46)>0.99
Y-axis1.03 (0.80-1.99)0.71 (0.57-2.57)0.55
Z-axis2.01 (0.98-2.79)0.92 (0.46-2.58)0.13
Type 4Type 3
X-axis0.21 (0.08-0.44)0.75 (0.35-0.93)0.07
Y-axis1.03 (0.80-1.99)0.18 (0.10-0.38)0.03
Z-axis2.01 (0.98-2.79)0.82 (0.62-1.63)0.07
Type 4Type 5
X-axis0.21 (0.08-0.44)0.06 (0.01-0.08)0.07
Y-axis1.03 (0.80-1.99)0.06 (0.01-0.08)0.03
Z-axis2.01 (0.98-2.79)0.03 (0.02-0.06)0.03

Values are presented as median (interquartile range).

Type 2: cannulated screw with cable, Type 3: Intertan nail, Type 4: dynamic hip screw, Type 5: IKEY nail.

*Wilcoxon rank sum test.

Overall, no meaningful improvement in load-to-failure and rotational deformity was observed for the cannulated screw with cable compared with conventional devices. Intertan nail and IKEY nail showed 3,549.28±544.20 N and 4,533.28±86.87 N each; both showing meaningful improvement (P<0.05) compared with the cannulated screw. Meanwhile, compared with the dynamic hip screw, the IKEY nail was the only device showing meaningful improvement in load-to-failure. Regarding rotational deformity, there was no meaningful difference in x-axis rotation. Less degree of deformity was observed for both the Intertan nail and the IKEY nail compared with the cannulated screw in y-axis and z-axis. Compared with the dynamic hip screw, meaningful improvement in y-axis was observed for both, but only the IKEY nail showed improvement in z-axis.

DISCUSSION

In Pauwels type III femoral neck fractures, vertical orientation of the pattern is known to be susceptible to greater shear forces rather than compression forces1211). Internal fixation using three cannulated screws, which is sufficient for management of Pauwels type I and II femoral neck fractures36), resulted in high complications, including femoral neck shortening, non-union, and avascular necrosis. Fixed angle devices including dynamic hip screws have been introduced in an effort to resist the shearing force, not only the compressive forces78). Despite its complications, the cannulated screw is still regarded as a conventional surgical technique for treatment of Pauwels type III femoral neck fractures due to short operation time and technical ease. When using cannulated screws, fixation of three parallel screws at least 6.5 mm in diameter and in an inverted triangular fashion has been proposed as ideal121314). Among fixed angle devices, the dynamic hip screw is regarded as conventional. Dynamic hip screw fixation is a method that utilizes a single large screw at a fixed angle with a side-plate1516). However, the clinical result is still not satisfying and many studies regarding the proper implant for Pauwels type III femoral neck fractures have been reported11171819). Wrapping a cable around the screw head was proposed and biomechanically tested by Kuan et al.17). In this study, the cable was wrapped by passing through the space between three fixated cannulated screws, which was then tightened. In this study addition of the cable resulted in improved mechanical behaviors compared with only fixating three cannulated screw in terms of axial stiffness, cyclic displacement, and failure load and the cable is thought to perform as a fixed angle device. The Intertan nail was introduced in 2005 for treatment of intertrochanteric fractures, showing good clinical outcomes and low complication rates20). This second generation intramedullary nail uses two cephalocervical screws, which can allow linear intraoperative compression and rotational stability of the head or neck fragments. Several biomechanical studies have suggested that the Intertan nail has a superior profile compared with other implants in Pauwels type III femoral neck fractures. Rupprecht et al.18) reported that the Intertan nail showed less inferior head displacement, higher loads to failure, and a longer survival under physiologic loads compared with the dynamic hip screw and cannulated screws. However, it still shows relative fragility in mechanical strength compared with intact femur. The IKEY nail, which was introduced in 2021, allows three cephalocervical screws. There are three apertures allowing the insertion of one lag screw and two additional cannulated screws. Apertures are designed so that three screws are not aligned in a single plane. These profiles are expected to stabilize femoral head rotation sufficiently and enable easier performance of the surgical procedure. Because there is no implant of choice in osteosynthesis of Pauwels type III fractures, introduction of many different types of novel fixed angle devices, including the aforementioned implants, continues. Many studies have reported superior outcome of novel fixed angle devices. The current study differs from other studies on three points. First, the cannulated screw with wiring was compared with the dynamic hip screw, not only the cannulated screw without wiring. Second, this study is the first to analyze biomechanical stability of supplementary cannulated screw fixation around an intramedullary nail in treatment of isolated Pauwels type III femoral neck fracture. Third, this study is the first to analyze the IKEY nail, which includes additional apertures allowing insertion of a supplementary cannulated screw. Cannulated screw with wiring is technically simple. A previous study comparing this technique with the cannulated screw reported improved stability in a 7° valgus and 25° valgus setting17). In the current study, load-to-failure and rotational deformity on the x-, y-, and z-axes were checked in a 25° valgus setting, and both cannulated screw and dynamic hip screw were used as controls. None of the superior results were observed compared with conventional devices. In brief, the results regarding the use of the cannulated screw with wiring in the 25° valgus setting differed significantly from those of the previous experiment when compared with the cannulated screw. It is considered that there is no biomechanical advantage compared with the dynamic hip screw. Supplementary cannulated screw fixation around the intramedullary nail has been reported to show superior biomechanical stability in intertrochanteric fractures and combined ipsilateral femoral neck and shaft fractures2122). Although biomechanical superiority has not been demonstrated in isolated Pauwels type III femoral neck fractures, supplementary cephalocervical screw fixation naturally enhances stability as in anti-rotational screw insertion in dynamic hip screws. Insertion of screws in free-handle fashion is the traditional method. For insertion of supplementary screws, it is important to secure sufficient space on the anterior side of the cephalocervical area. In order to achieve this, the authors prefer to insert the nail slightly posterior to the tip of the greater trochanter. After insertion of the nail and the cephalic screw, the guide-wire is inserted in the point anterior and proximal to the site of cephalic screw insertion, heading posteriorly. The cannulated screw is inserted subsequently, just adjacent to the intramedullary nail. However, making a supplementary screw insertion around the nail in free-handle fashion is technically demanding232425). Despite continued advancement in nail design and the fact that the Intertan nail has two proximal apertures allowing screw insertion, Tan et al.26) and Watson and Moed27) argued that two screws lie on a single plane in this type and are still insufficient for fixation of Pauwels type III femoral neck fracture. The IKEY nail has three apertures on the proximal portion, so that screws do not lie on a single plane when screws are inserted in all apertures, unlike the Intertan nail. This not only provides technical simplicity but also biomechanical stability similar to that of supplementary screw insertion with free-handle fashion. In the current study, improved stability in load-to-failure, y- and z-axes was observed for the IKEY nail compared with all conventional devices, while other novel fixed angle did not. To the best of our knowledge, the current study is the first comparative analysis for supplementary cannulated screw around intramedullary nail in treatment of Pauwels type III femoral neck fractures. Additional differentiation from previous study is that supplementary screw insertion was performed through additional apertures in the proximal portion of the nail, which was specially designed. However, because this study is the first to analyze the IKEY nail, its effectiveness is difficult to generalize. Follow-up study is required in order to determine whether the IKEY nail can show reproducible outcomes in other experimental settings. There are some limitations of the current study. First, five femur specimens were used for each fixation group, which is relatively small. Second, although there was no breakage around proximal apertures using the IKEY nail in this study, further study is needed in order to determine the potential of this device. Third, all intramedullary nail tested in this study were the screw type of lag screw, not blade type. Because the blade type screw is one of the novel designs for anti-rotation of the femoral head, comparative study with this type of implant may have induced other meaningful results. Last, although promising biomechanical results have been obtained using this type of intramedullary nail, further study is needed in order to evaluate the clinical outcomes.

CONCLUSION

To the best of our knowledge so far, there is no fixation of choice in treatment of Pauwels type III fracture. This study biomechanically compared two conventional methods with three novel fixed angle devices. Among all, the IKEY nail, the newly developed intramedullary nail, showed meaningful improvement in this study. Further study is needed in order to evaluate the outcome on a larger scale and regarding the clinical outcome.
  25 in total

1.  Intracapsular fractures of the neck of femur. Parallel or crossed garden screws?

Authors:  M J Parker; K M Porter; D M Eastwood; M Schembi Wismayer; A A Bernard
Journal:  J Bone Joint Surg Br       Date:  1991-09

2.  Fracture fixation in the operative management of hip fractures (FAITH): an international, multicentre, randomised controlled trial.

Authors: 
Journal:  Lancet       Date:  2017-03-03       Impact factor: 79.321

Review 3.  Femoral neck fractures: current management.

Authors:  Anthony V Florschutz; Joshua R Langford; George J Haidukewych; Kenneth J Koval
Journal:  J Orthop Trauma       Date:  2015-03       Impact factor: 2.512

Review 4.  Proximal Femoral Fractures: What the Orthopedic Surgeon Wants to Know.

Authors:  Scott E Sheehan; Jeffrey Y Shyu; Michael J Weaver; Aaron D Sodickson; Bharti Khurana
Journal:  Radiographics       Date:  2015-07-17       Impact factor: 5.333

5.  A comparative biomechanical analysis of fixation devices for unstable femoral neck fractures: the Intertan versus cannulated screws or a dynamic hip screw.

Authors:  Martin Rupprecht; Lars Grossterlinden; Andreas H Ruecker; Alexander Novo de Oliveira; Kay Sellenschloh; Jakob Nüchtern; Klaus Püschel; Michael Morlock; Johannes Maria Rueger; Wolfgang Lehmann
Journal:  J Trauma       Date:  2011-09

6.  Treatment of ipsilateral femoral neck and shaft fractures with the Russell-Taylor reconstructive nail.

Authors:  P Randelli; S Landi; F Fanton; G K Hoover; M Morandi
Journal:  Orthopedics       Date:  1999-07       Impact factor: 1.390

7.  Cephalomedullary interlocked nail for ipsilateral hip and femoral shaft fractures.

Authors:  Pankaj Jain; Lalit Maini; Puneet Mishra; Ashish Upadhyay; Ajay Agarwal
Journal:  Injury       Date:  2004-10       Impact factor: 2.586

8.  Ipsilateral femoral neck and shaft fractures. Retrospective study of 33 cases.

Authors:  C C Wu; C H Shih
Journal:  Acta Orthop Scand       Date:  1991-08

9.  Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques.

Authors:  Arash Aminian; Fan Gao; Wasyl W Fedoriw; Li-Qun Zhang; David M Kalainov; Bradley R Merk
Journal:  J Orthop Trauma       Date:  2007-09       Impact factor: 2.512

10.  The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression.

Authors:  Andreas H Ruecker; M Rupprecht; M Gruber; M Gebauer; F Barvencik; D Briem; J M Rueger
Journal:  J Orthop Trauma       Date:  2009-01       Impact factor: 2.512