Characterizing the microbiota of instrumentation in ophthalmology clinics during and beyond the COVID-19 pandemic.

Heba Mahjoub1, Sean X Zhang2,3, Jiangxia Wang4, Warda Memon3, Heba Mostafa5, Mark P Breazzano6,7,8.   

Abstract

PURPOSE: Increased ophthalmology-specific risk of novel coronavirus 2019 (SARS-CoV-2) transmission is well-established, increasing the fear of infection and causing associated decreased rates of procedures known to save vision. However, the potential transmission from exposure to clinic instrumentation is unknown, including which additional pathogens may be spreading in this context. This study seeks to fill this gap by characterizing the microbiota of instrumentation in ophthalmology clinics during the COVID-19 pandemic and identifying potential sources of pathogenic spread encountered by patients and healthcare workers.
METHODS: Thirty-three samples were captured using standard cultures and media. Ten positive and negative controls were used to confirm proper technique. Descriptive statistics were calculated for all samples. Samples were collected from the retina (N = 17), glaucoma (N = 6), cornea (N = 6), and resident (N = 4) clinics with rigorous disinfection standards at a tertiary academic medical center. Standard media cultures and/or polymerase chain reaction (PCR) was performed for each sample.
RESULTS: From 33 samples, more than half (17/33, 51.5%) yielded bacterial growth. Using two different molecular methods, three samples (3/33, 9%) tested positive for SARS-CoV-2 (cycle thresholds 36.48, 37.14, and 37.83). There was no significant difference in bacterial growth (95% confidence interval [95% CI]: - 0.644-0.358, p = 0.076) among different clinics (retina, glaucoma, cornea, resident). Staphylococcus (S.) epidermidis grew most frequently (12/35, 34%), followed by S. capitis (7/35, 20%), Micrococcus luteus (2/35, 5.7%), Corynebacterium tuberculostearicum (2/35, 5.7%), and Cutibacterium ([C.], Propionibacterium) acnes (2/35, 5.7%). C. acnes growth was more frequent with imaging device forehead rests (2/7, 28.6%) than other surfaces (0/26, 0%, 95% CI: 0.019-0.619, p = 0.040). No samples isolated fungus or adenovirus.
CONCLUSIONS: Most samples across subspecialty clinic instrumentation grew bacteria, and several tested positive for SARS-CoV-2. Many isolated pathogens have been implicated in causing infections such as endophthalmitis, conjunctivitis, uveitis, and keratitis. The clinical implications of the ophthalmology microbiome for transmitting nosocomial infections warrant optimization of disinfection practices, strategies for mitigating spread, and additional study beyond the pandemic.
© 2022. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Entities:  

Keywords:  Bacteria; COVID-19; Instrumentation; Pathogens; SARS-CoV-2

Mesh:

Year:  2022        PMID: 35355117      PMCID: PMC8967689          DOI: 10.1007/s00417-022-05639-0

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.535


Introduction

Decreases in critical procedures known to preserve vision in the pandemic have been attributed to fear of infection [1], intensifying concerns for safety among patients and healthcare workers inherent to ophthalmological care. Maintaining proper disinfection in the clinical setting is crucial for mitigating these concerns as well as safety itself. This concept becomes increasingly relevant with interaction proximity, as occurs between ophthalmologists and patients for example, during slit lamp examination. One study done among 34 ophthalmology clinics in 2005 showed remnant bacteria and fungi on biometry equipment, indicating inadequate elimination during cleaning [2]. Another reported an adenovirus outbreak in neonatal intensive care unit with equipment contaminated from retinopathy of prematurity screening [3]. However, these studies were both performed before the COVID-19 pandemic. Now that disinfection processes are largely replaced by heightened infection control practices [4], it is crucial to evaluate potential risks of pathogenic spread during ophthalmology visits. Increased relative risk of SARS-CoV-2 transmission for ophthalmologists [5, 6] is well-established, but a gap remains in knowledge regarding which microbes may still be spreading. This investigation seeks to fill this gap by characterizing the microbial profile of different ophthalmology clinics.

Methods

To assess for the presence of bacteria, fungi, adenovirus, and SARS-CoV-2, we captured samples using COPAN ESwabs™ (Murrieta, CA, USA). The sampling process occurred between 3/9/2021 and 3/30/21 at the Wilmer Eye Institute at Johns Hopkins Hospital in Baltimore, MD, USA. The sampling occurred at 6 am, before any staff or patients entered the rooms, and after being cleaned the previous night. Infection control standards in the clinic included utilizing PDI (Professional Disposables International, Inc.) germicidal disinfectant wipes on surfaces in between patient encounters. Several minutes were allowed for air drying surfaces after PDI wipe use. This study complied with the Declaration of Helsinki and did not require Institutional Review Board approval because it did not involve human subjects. Thirty-three surfaces or instruments were surveyed in four ophthalmology subspecialty clinics and imaging suites including resident (general eye service), retina, glaucoma, and cornea (Table 1). Ten positive and negative controls were also sampled (Table 1). Samples were assessed for bacteria and fungi on standard media including blood agar plates (BAP), chocolate agar (CHOC), colistin nalidixic agar (CNA), MacConkey agar (MAC), inhibitory mold agar (IMA), and brain heart infusion agar (BHI). SARS-CoV-2 presence was determined via two different molecular methods including polymerase chain reaction (PCR). Cycle threshold (Ct) was determined for each positive SARS-CoV-2 sample.
Table 1

Bacteria, fungi, adenovirus, and SARS-CoV-2 presence in controls and samples

SampleAdenovirusSARS-COVID-2Bacterial growthFungal growth
Controls
  S.# aureus cultureN/ANot doneS. aureusNo growth (NG)
  Candida albicans cultureN/ANot doneNGCandida albicans
  Study team member 1, salivaN/ANegativeNeisseria flavescens, Streptococcus mitis, Streptococcus parasanguinis, Rothia aeria, Haemophilus parainfluenzae, Rothia dentocariosaNG
  Study team member 1, R naresN/ANegativeS. aureus, S. epidermidis, Klebsiella pneumoniaeNG
  Study team member 2, salivaN/ANegativeS. aureus, Neisseria flavescens, Streptococcus mitis, Streptococcus salivarius, Rothia mucilaginosa, Rothia aeria, Lautropia mirabilisNG
  Study team member 2, R naresN/ANegativeK.pnemoniae, S.lugdunensis, C.** amycolatum, S. epidermidisNG
  PDI germicidal disinfectant wipeN/ANegativeNGNG
  PDI germicidal disinfectant solutionN/ANegativeNGNG
  Water, retina clinic room sinkN/ANegativeNGNG
  Distilled waterN/ANot doneNGNG
Samples
  Retina clinic, lens kitN/ANegativeC. mucifaciensNG
  Retina clinic, indirect ophthalmoscopeN/ANegativeS. capitis, S. pasteuriNG
  Retina clinic, patient seatN/ANegativeNGNG
  Retina clinic, handheld eye occluderN/ANegativeNGNG
  Retina clinic, keyboard and mouseN/ANegativeNGNG
  Retina clinic, patient room light switchN/ANegativeS. epidermidis, C. minutissimum, C. amycolatum, Micrococcus luteus, C. tuberculostearicum, C. ureicelerivoransNG
  Retina clinic room A, slit lamp chin restN/A

Positive;

Cycle threshold

(CT) = 36.48

NGNG
  Retina clinic room A, slit lamp forehead restN/ANegativeNGNG
  Retina clinic room B, slit lamp chin restNegativeNegativeC. tuberculostearicum, C. singulareNG
  Retina clinic room C, slit lamp chin restNegativeNegativeNGNG
  Retina clinic room D, slit lamp chin restNegativeNegativeMixta calidaNG
  Retina clinic room E, slit lamp chin restNegativeNegativeNGNG
  Retina clinic room F, slit lamp chin restNegativeNegativeNGNG
  Retina imaging suite, SPECTRALIS® Heidelberg OCT chin restNot donePositive; CT = 37.14S. capitis, S. epidermidisNG
  Retina imaging suite SPECTRALIS® Heidelberg OCT forehead restNegativeNegativeS. capitis, S. epidermidis, Cutibacterium acnesNG
  Retina imaging suite Optos® chin restNegativeNegativeS. epidermidisNG
  Retina imaging suite Optos® forehead restNot doneNegativeS. capitis, S. epidermidis, C. kroppenstedtiiNG
  Glaucoma imaging suite OCT chin restNegativeNegativeS. epidermidisNG
  Glaucoma imaging suite OCT forehead restNegativeNegativeS. hominis, Cutibacterium acnes, S. epidermidisNG
  Glaucoma imaging suite Humphrey® Field Analyzer chin restNegativeNegativeS. epidermidis, S. capitis, Micrococcus luteusNG
  Glaucoma imaging suite Humphrey® Field Analyzer forehead restNegativeNegativeMoraxella osloensis, S. capitisNG
  Glaucoma clinic room A slit lamp chin restNegativeNegativeNGNG
  Glaucoma clinic room B slit lamp chin restNegativeNegativeS. epidermidisNG
  GES (general eye service) A slit lamp chin restNot doneNegativeS. epidermidisNG
  GES (general eye service) B slit lamp chin restNegativeNegativeS. epidermidisNG
  GES (general eye service) imaging suite Cirrus OCT chin restNegativeNegativeNGNG
  GES (general eye service) imaging suite Cirrus OCT forehead restNegativeNegativeNGNG
  Cornea imaging suite Pentacam® chin restNot doneNegativeS. epidermidis, S. capitisNG
  Cornea imaging suite Pentacam® forehead restNot done

Positive;

CT = 37.83

NGNG
  Cornea imaging suite IOL Master® chin restNegativeNegativeNGNG
  Cornea imaging suite IOL Master® forehead restNegativeNegativeNGNG
  Cornea clinic room A slit lamp chin restNegativeNegativeNGNG
  Cornea clinic room B slit lamp chin restNegativeNegativeNGNG

S.# = Staphylococcus

C.** = Corynebacterium

*Imaging machines include ultra-widefield fundus cameras (Optos, Marlborough, MA, USA), OCT machines (including Cirrus, Zeiss, Oberkochen, Germany; Spectralis, Heidelberg, Germany), Humphrey® Field Analyzer (Zeiss), Pentacam® (Oculus, Wetzlar, Germany), and IOL Master® (Zeiss) machines

Bacteria, fungi, adenovirus, and SARS-CoV-2 presence in controls and samples Positive; Cycle threshold (CT) = 36.48 Positive; CT = 37.83 S.# = Staphylococcus C.** = Corynebacterium *Imaging machines include ultra-widefield fundus cameras (Optos, Marlborough, MA, USA), OCT machines (including Cirrus, Zeiss, Oberkochen, Germany; Spectralis, Heidelberg, Germany), Humphrey® Field Analyzer (Zeiss), Pentacam® (Oculus, Wetzlar, Germany), and IOL Master® (Zeiss) machines Descriptive statistics were calculated for all samples using Microsoft Excel (Seattle, WA, USA). The associations between type of surfaces, instruments, or subspecialty clinics with bacterial growth or SARS-CoV-2 positivity were examined using Fisher’s exact tests, and the Pearson correlation coefficients or polychoric correlation coefficients were reported when appropriate. All the analyses were carried out in Stata version 16.1 (College Station, TX, USA), and p ≤ 0.05 was considered significant.

Results

Samples included 33 unique surfaces among 4 clinics. All 10 controls yielded expected results. Saliva and nares from two authors (HM1, MPB) grew an abundance of bacteria but no fungi (Table 1). From 33 samples, more than half (17/33, 51.5%) yielded bacterial growth. eTable 1 shows 15 different species that grew on surfaces 35 times total. Among these, Staphylococcus (S.) epidermidis grew most frequently (12/35, 34%), followed by S. capitis (7/35, 20%), Micrococcus luteus (2/35, 5.7%), Corynebacterium (C.) tuberculostearicum (2/35, 5.7%), and Cutibacterium (Propionibacterium) acnes (2/35, 5.7%) (eTable 1). No fungus or adenovirus was isolated. Three samples (3/33, 9%) tested positive for SARS-CoV-2 virus (retina slit lamp chin rest [Ct 36.48], retina SPECTRALIS [Heidelberg, Germany] Heidelberg optical coherence tomography [OCT] chin rest [Ct 37.14], and cornea Oculus [Wetzlar, Germany] Pentacam forehead rest [Ct 37.83]) (Table 1). Bacterial growth (95% confidence interval [95% CI] − 0.644–0.358, p = 0.076) and SARS-CoV-2 positivity (95% CI − 0.438–0.735, p = 0.948) did not vary by clinic (Table 2). Imaging suites and examination rooms were also similar in bacterial (95% CI − 0.134–0.523, p = 0.220) and SARS-CoV-2 positivity (95%CI − 0.199–0.473, p = 0.389) (Table 2). C. acnes was isolated on imaging device forehead rests more frequently than other surfaces (95% CI 0.019–0.619, p = 0.040) (Table 3). SARS-CoV-2 was identified on 3 surfaces (2 chin rests) (eTable 2).
Table 2

Bacterial growth and SARS-CoV-2 among subspecialty clinics and room types

Bacterial growthPositive for SARS-CoV-2
YesNoP valueCorrelation coefficient (95% confidence interval (CI))YesNoP value95% CI
Subspecialty clinics
  Retina, N = 17980.076 − 0.143 (− 0.644, 0.358)2150.9480.148 (− 0.438, 0.735)
  Glaucoma, N = 65106
  Cornea, N = 61515
  GES*, N = 42204
Room type
  Imaging suite, N = 14950.2200.219 (− 0.134, 0.523)2120.3890.155 (− 0.199, 0.473)
  Clinic room, N = 19811118

*GES general eye service (resident clinic)

Table 3

Presence of Cutibacterium acnes among surfaces

Presence of Cutibacterium acnes
YesNoP value95% CI
Imaging device forehead rest, N = 7250.0400.355 (0.019, 0.619)
All other surfaces, N = 26026
Bacterial growth and SARS-CoV-2 among subspecialty clinics and room types *GES general eye service (resident clinic) Presence of Cutibacterium acnes among surfaces

Discussion

To our knowledge, this surveillance is the first of its kind during the COVID-19 pandemic, an era when disinfection practices have been amplified. PDI (Professional Disposables International, Inc.) germicidal disinfectant wipes are universally used across surfaces in ophthalmology examination and imaging suites between patient encounters. These wipes are bactericidal, viricidal (including SARS-CoV-2), and tuberculocidal, requiring 2 min with complete drying. Despite these measures, numerous pathogens were found in this microbiome. The last surveillance appears documented in 2005 [2] and it is evident that many pathogens remain pervasive. It is still unknown the extent with which these pathogens may be spreading among patients and healthcare workers. The isolation of C. acnes (formerly Propionibacterium acnes) on multiple chin rests is alarming. C. acnes is well-established as a cause of postoperative, chronic endophthalmitis requiring repeated intravitreal antibiotics, and occasionally surgical removal of contaminated intraocular lenses [7, 8]. This bacteria may also cause dacryocystitis [9]. Micrococcus luteus, isolated here, can form biofilms implicated in prosthetic valve endocarditis, a life-threatening condition [10]. Many samples grew Corynebacterium species, which can cause granulomatous mastitis. These infections have reportedly poor outcomes and may be difficult to treat due to the lipophilic nature of associated granulomas [11]. While S. epidermidis was seen among the commensal bacteria in study team member samples, it was also the most frequently grown bacterium among all samples. This bacterium is one of the most common causes of post-intraocular surgical infection [12]. In addition to S. epidermidis, S. capitis is implicated with surface infections such as chronic blepharitis, suppurative keratitis, and purulent conjunctivitis [13]. These findings, then, are suggestive of an uncontrolled vector for these nosocomial infections. While specific adherence to infection control protocols was not assessed here, it is important to evaluate the regularity of execution and effectiveness of individual existing procedures for optimizing antisepsis. One study showed that SARS-CoV-2 half-life on plastic surfaces was 5.3 h with infectivity exceeding 120 h [14]. The alarming finding of 3 positive SARS-CoV-2 samples on plastic surfaces of ophthalmology equipment indicates an unmet need and opportunity for reducing potential infectious spread. For example, copper exhibits antiviral activity by causing irreversible fragmentation of the genome through reactive oxygen species [15]. Therefore, copper chin and forehead rests for slit lamp and imaging machines may be a useful and relatively convenient protective measure. Additionally, UV-C irradiation may decrease SARS-CoV-2 loads on plastic within 21 s [14], and may provide an alternative for maximizing antisepsis. Regardless, current disinfection practices appear deficient for preventing instances of SARS-CoV-2 isolation in ophthalmology clinics. This study has some limitations, including the challenge of directly linking bacteria and SARS-CoV-2 to clinical infection as they can be more insidious than adenovirus2, for example. Just as S. epidermidis does not always cause infection, other pathogens may act similarly. Future studies are needed to elucidate infection rates following ophthalmology visits. Though these findings may indicate a need for increased disinfection, the potential costs such as environmental impact and risks must be weighed, including greater waste products in the form of PDI wipes. Furthermore, samples were derived from one hospital where the disinfection technique is similar between clinics. Some samples were inadequate for testing adenovirus. Expanding surveillance to other sites would be helpful. Still, we would expect negligible microbial growth during the COVID-19 pandemic with increased disinfection practices, which reinforces these findings. In conclusion, we found the majority of sampled surfaces in ophthalmology clinics yielded bacterial growth, and some samples tested positive for SARS-CoV-2 despite rigorous antisepsis. These findings suggest an opportunity for reevaluating disinfection techniques across subspecialties. Future studies are necessary to clarify instances of infection after an ophthalmology appointment. Below is the link to the electronic supplementary material. Supplementary file1 (DOCX 38 KB)
  13 in total

1.  Antimicrobial Treatment Options for Granulomatous Mastitis Caused by Corynebacterium Species.

Authors:  Hazel C Dobinson; Trevor P Anderson; Stephen T Chambers; Matthew P Doogue; Lois Seaward; Anja M Werno
Journal:  J Clin Microbiol       Date:  2015-07-01       Impact factor: 5.948

2.  Microbial contamination of immersion biometry ultrasound equipment.

Authors:  Leonardo J Velázquez-Estades; Audrey Wanger; Judianne Kellaway; David R Hardten; Thomas C Prager
Journal:  Ophthalmology       Date:  2005-05       Impact factor: 12.079

3.  The microbiologic profile of dacryocystitis.

Authors:  Stella Y Chung; Leon Rafailov; Roger E Turbin; Paul D Langer
Journal:  Orbit       Date:  2018-05-11

4.  New York City COVID-19 resident physician exposure during exponential phase of pandemic.

Authors:  Mark P Breazzano; Junchao Shen; Aliaa H Abdelhakim; Lora R Dagi Glass; Jason D Horowitz; Sharon X Xie; C Gustavo de Moraes; Alice Chen-Plotkin; Royce Ws Chen
Journal:  J Clin Invest       Date:  2020-09-01       Impact factor: 14.808

5.  Outbreak of Adenovirus in a Neonatal Intensive Care Unit: Critical Importance of Equipment Cleaning During Inpatient Ophthalmologic Examinations.

Authors:  Julia S Sammons; Erin H Graf; Sara Townsend; Cindy L Hoegg; Sarah A Smathers; Susan E Coffin; Katie Williams; Stephanie L Mitchell; Ursula Nawab; David Munson; Graham Quinn; Gil Binenbaum
Journal:  Ophthalmology       Date:  2018-09-01       Impact factor: 12.079

6.  Frequency of Urgent or Emergent Vitreoretinal Surgical Procedures in the United States During the COVID-19 Pandemic.

Authors:  Mark P Breazzano; Archana A Nair; J Fernando Arevalo; Mark R Barakat; Audina M Berrocal; Jonathan S Chang; Andrew Chen; Dean Eliott; Sunir J Garg; Quraish Ghadiali; Dan Gong; Dilraj S Grewal; James T Handa; Matthew Henderson; Yannek I Leiderman; Theodore Leng; Amar Mannina; Thomas A Mendel; Debarshi Mustafi; Lisa C Olmos de Koo; Shriji N Patel; Tapan P Patel; Jonathan Prenner; Paige Richards; Rishi P Singh; Charles C Wykoff; Nicolas A Yannuzzi; Hannah Yu; Yasha S Modi; Stanley Chang
Journal:  JAMA Ophthalmol       Date:  2021-04-01       Impact factor: 7.389

7.  A Measured Approach to Inpatient Ophthalmologic Screening in the COVID-19 Era: A Multicenter Perspective.

Authors:  Tavish Nanda; John B Bond; Royce W S Chen; Srilaxmi Bearelly; H Russell Day; George A Cioffi; James T Handa; J Fernando Arevalo; Sean P Donahue; Mark P Breazzano
Journal:  Ophthalmology       Date:  2020-08-07       Impact factor: 12.079

8.  Advances in medicine and positive natural selection: Prosthetic valve endocarditis due to biofilm producer Micrococcus luteus.

Authors:  Guillermo Rodriguez-Nava; Abdelnasir Mohamed; Maria Adriana Yanez-Bello; Daniela Patricia Trelles-Garcia
Journal:  IDCases       Date:  2020-03-09

9.  Postoperative Endophthalmitis Caused by Cutibacterium (Formerly Propionibacterium) Acnes: Case Series and Review.

Authors:  Benjamin J Fowler; Darlene Miller; Xiaohe Yan; Nicolas A Yannuzzi; Harry W Flynn
Journal:  Case Rep Ophthalmol       Date:  2021-01-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.