Christopher K Akhgar1, Julian Ebner2, Oliver Spadiut2, Andreas Schwaighofer1, Bernhard Lendl1. 1. Institute of Chemical Technologies and Analytics, Technische Universität Wien, Getreidemarkt 9, 1060 Vienna, Austria. 2. Institute of Chemical, Environmental and Bioscience Engineering, Technische Universität Wien, Getreidemarkt 9, 1060 Vienna, Austria.
Abstract
In this study, an external cavity-quantum cascade laser-based mid-infrared (IR) spectrometer was applied for in-line monitoring of proteins from preparative ion-exchange chromatography. The large optical path length of 25 μm allowed for robust spectra acquisition in the broad tuning range between 1350 and 1750 cm-1, covering the most important spectral region for protein secondary structure determination. A significant challenge was caused by the overlapping mid-IR bands of proteins and changes in the background absorption of water due to the NaCl gradient. Implementation of advanced background compensation strategies resulted in high-quality protein spectra in three different model case studies. In Case I, a reference blank run was directly subtracted from a sample run with the same NaCl gradient. Case II and III included sample runs with different gradient profiles than the one from the reference run. Here, a novel compensation approach based on a reference spectra matrix was introduced, where the signal from the conductivity detector was employed for correlating suitable reference spectra for correction of the sample run spectra. With this method, a single blank run was sufficient to correct various gradient profiles. The obtained IR spectra of hemoglobin and β-lactoglobulin were compared to off-line reference measurements, showing excellent agreement for all case studies. Moreover, the concentration values obtained from the mid-IR spectrometer agreed well with conventional UV detectors and high-performance liquid chromatography off-line measurements. LC-QCL-IR coupling thus holds high potential for replacing laborious and time-consuming off-line methods for protein monitoring in complex downstream processes.
In this study, an external cavity-quantum cascade laser-based mid-infrared (IR) spectrometer was applied for in-line monitoring of proteins from preparative ion-exchange chromatography. The large optical path length of 25 μm allowed for robust spectra acquisition in the broad tuning range between 1350 and 1750 cm-1, covering the most important spectral region for protein secondary structure determination. A significant challenge was caused by the overlapping mid-IR bands of proteins and changes in the background absorption of water due to the NaCl gradient. Implementation of advanced background compensation strategies resulted in high-quality protein spectra in three different model case studies. In Case I, a reference blank run was directly subtracted from a sample run with the same NaCl gradient. Case II and III included sample runs with different gradient profiles than the one from the reference run. Here, a novel compensation approach based on a reference spectra matrix was introduced, where the signal from the conductivity detector was employed for correlating suitable reference spectra for correction of the sample run spectra. With this method, a single blank run was sufficient to correct various gradient profiles. The obtained IR spectra of hemoglobin and β-lactoglobulin were compared to off-line reference measurements, showing excellent agreement for all case studies. Moreover, the concentration values obtained from the mid-IR spectrometer agreed well with conventional UV detectors and high-performance liquid chromatography off-line measurements. LC-QCL-IR coupling thus holds high potential for replacing laborious and time-consuming off-line methods for protein monitoring in complex downstream processes.
Preparative
liquid chromatography
(prep-LC) remains an essential unit operation in downstream processing
of complex biopharmaceuticals.[1] The most
widespread detection method for monitoring the corresponding protein
concentrations is single-wavelength UV/vis spectroscopy, offering
a broad dynamic range and excellent sensitivity. The obtained univariate
signal, however, does not give any information about the protein structure
and purity during elution.[2] Thus, the collected
fractions have to be additionally analyzed by laborious and time-consuming
off-line methods in order to measure critical quality attributes (CQAs).
Commonly used off-line methods are high-performance liquid chromatography
(HPLC), sodium dodecyl sulphate-polyacrylamide gel electrophoresis,
Western blotting and biological activity assays, such as enzyme-linked
immunosorbent assays. In recent years, quality by design (QbD) principles
were established in the pharmaceutical manufacturing sector.[3] In order to comply with these regulatory guidelines,
process analytical technology (PAT) tools are required, allowing for
in-process control and timely adaption of process parameters. Therefore,
analytical methods able to provide information about CQAs in an in-line
or on-line measurement setup are required.[4,5]Mid-infrared (IR) spectroscopy provides detailed information about
rotational–vibrational transitions of proteins. The established
technique in this spectral region is Fourier transform infrared (FTIR)
spectroscopy, which is routinely used for quantitative and qualitative
analysis of infrared absorption spectra. The most important mid-IR
regions for protein quantification and secondary structure determination
are the amide I (1700–1600 cm–1)[6,7] and amide II (1600–1500 cm–1)[8] band. Coupling of IR spectroscopy and LC were
successfully demonstrated,[9,10] in most cases using
organic solvent gradients. However, flow-through mid-IR transmission
measurements of proteins in aqueous matrix have several challenges.
One of these constraints arises from the overlap of the HOH-bending
band of water near 1643 cm–1 with the protein amide
I band. A second limitation is the low emission power provided by
thermal light sources (Globars) that are used in FTIR spectrometers.
Thus, using FTIR spectroscopy, the optical pathlength for aqueous
protein solutions is restricted to <10 μm for transmission
measurements in order to avoid total IR light absorption of water.[11,12] These low path-lengths are unsuitable for in-line flow-through measurements
due to low robustness and limited sensitivity. Consequently, mid-IR
spectroscopy found its way into in-line detection of preparative protein
chromatography effluents only recently.[13,14] In these studies,
attenuated total reflection-FTIR spectroscopy was applied, offering
robust spectra acquisition, but limited sensitivity. As an alternative
approach, complex solvent-removal setups were developed in order to
enable protein secondary structure analysis in LC effluents.[15] Here, the eluent is evaporated, while the analyte
is (almost) simultaneously deposited onto a suitable substrate prior
to FTIR spectra acquisition. Typical challenges for solvent–evaporation
interfaces are, however, the morphology of certain analytes that can
change over time and possible spatial inhomogeneity.[9] Moreover, this destructive type of sample preparation prevents
fractionation of the effluent after detection.Quantum cascade
lasers (QCLs) are new light sources in the mid-IR
spectral region that provide polarized and coherent light with spectral
power densities higher by a factor of 104 or more compared
with thermal light sources.[16,17] In combination with
an external cavity (EC), QCLs enable tuning over several hundred wavenumbers,
thus offering high potential for mid-IR transmission spectroscopy
of liquids. It was shown that the high available spectral power of
EC-QCLs opens a wide range of applications, including robust in-line
detection of LC effluents.[18] For protein
analysis, the higher spectral power densities enabled to increase
the optical path lengths for transmission measurements by a factor
of 3–4, thus considerably improving the ruggedness by significantly
lowering the backpressure at the cell.[19] In this framework, academic setups were reported that applied EC-QCLs
to investigate the amide I[20,21] and amide I + amide
II band,[22,23] finally achieving a limit of detection almost
10 times lower than high-end FTIR spectroscopy at similar measurement
conditions.[24] These setups were successfully
combined with different chemometric methods in order to quantify individual
protein content in complex mixtures[25−28] and to monitor changes in the
protein secondary structure after denaturation.[21,29,30] Most recently, a commercial EC-QCL-based
spectrometer (ChemDetect Analyzer, Daylight Solutions) was introduced,
offering robust and sensitive spectra acquisition across the wavenumber
region between 1350 and 1770 cm–1.[31]Another challenge in in-line LC-IR is the compensation
of possible
eluent absorbance bands, which can be higher by several orders of
magnitude compared to actual analyte bands. Even though direct subtraction
of a background spectrum is possible under isocratic conditions, background
correction during gradient elution is more challenging. For this purpose,
a method based on a “reference spectra matrix” (RSM)
was introduced.[32] In this approach, the
spectra acquired in-line during the LC run are viewed as “sample
matrix” (SM), whereas the RSM is recorded during a blank run
or the re-equilibration phase of the system. Based on this acquired
information, spectral regions that show absorbance bands characteristic
for eluent composition are identified, which are located in different
spectral regions when compared to the analyte spectrum. Consequently,
each spectrum in the SM is corrected with the RSM spectrum that has
the closest eluent composition. This method was successfully applied
for gradient compensation in a wide range of different reversed-phase
HPLC-FTIR applications, including analysis of carbohydrates,[33,34] nitrophenols,[35,36] and pesticides.[32,37]In the present work, the ChemDetect Analyzer was coupled to
a lab-scale
prep-LC system. The large optical path length of the transmission
cell of 25 μm enabled in-line monitoring of proteins without
solvent evaporation steps. This LC–QCL–IR coupling was
employed to analyze systems based on ion-exchange chromatography (IEX)
and different NaCl gradient profiles. For this application, the laser-based
spectrometer offers advantages regarding acquisition of protein spectra
in aqueous solution. However, due to the limited spectral region of
QCL-IR spectra, gradient compensation is more challenging than with
FTIR spectra, which cover the entire mid-IR region. In the studied
analytical problems, absorption bands from the salt gradient highly
overlap with protein absorptions. Three different case studies were
performed, covering various real-life conditions used in chromatographic
protein downstream processing. In Case I, a reference blank run was
directly subtracted from a sample run with the same linear gradient.
For Case II and Case III, two different gradient profiles were employed
when compared to the one of the reference blank run. Here, a modified
RSM-based approach was devised by incorporating the signal of the
conductivity detector. With this approach, each spectrum in the SM
was corrected with the RSM spectrum that had the closest conductivity
value. Thus, a singular measurement of the RSM blank run was sufficient
to correct sample runs with different salt gradient profiles, revealing
high quality protein spectra. The obtained results demonstrate the
successful coupling of a laser-based IR spectrometer with a LC system
and present a novel approach for LC-IR-based gradient correction of
highly overlapping eluent and analyte mid-IR spectra, indicating high
flexibility for future in-line monitoring of the protein secondary
structure in chromatographic effluents.
Experimental Section
Reagents
and Samples
Tris, hydrochloric acid (HCl),
and NaCl used for eluents of the prep-LC were purchased from Carl
Roth (Karlsruhe, Germany). Hemoglobin (Hemo) from bovine blood and
β-lactoglobulin (β-LG) from bovine milk (≥90%)
were purchased from Sigma-Aldrich (Steinheim, Germany). For LC-IR
and protein secondary structure reference measurements, proper amounts
of lyophilized protein powder were dissolved in the corresponding
buffer. Ultrapure water (MQ) was from a Milli-Q system from Merck
Millipore (Darmstadt, Germany). HPLC-Grade acetonitrile (ACN) and
trifluoroacetic acid (TFA) were both purchased from AppliChem (Darmstadt,
Germany).
LC–QCL–IR Setup
The flow path of LC-IR
measurements is illustrated in Figure . An ÄKTA pure system (Cytiva Life Sciences,
MA, USA) equipped with an U9-M UV monitor, a C9 conductivity monitor,
and a F9-C fraction collector was used for the prep-LC runs. All runs
were performed with a 1 mL HiTrap Capto Q column (Cytiva Life Sciences,
MA, USA). Laser-based mid-IR spectra were acquired with a ChemDetect
Analyzer (Daylight Solutions Inc., San Diego, USA). The delay volume
between the conductivity detector and ChemDetect Analyzer was determined
by injection of 1 M NaCl solution.
Figure 1
Schematic of the flow path in the LC-QCL-IR
setup.
Schematic of the flow path in the LC-QCL-IR
setup.
Preparative Chromatography
Conditions
For all four
prep-LC runs, the setup described in Figure was used. The flowrate was kept constant
at 75 cm/h, with Buffer A (50 mM Tris/HCl, pH 8.5) being used for
equilibration and wash and a gradient of Buffer A and Buffer B (50
mM Tris/HCl, pH 8.5, 1 M NaCl) being used for elution. Injections
were performed using a 1 mL loop, and absorbance at 280 nm as well
as conductivity were recorded for all runs. Fractions of 1 mL were
collected over the whole run-time and the respective protein concentrations
of each fraction measured using the described reversed phase (RP)-HPLC
method. The specific elution profiles are shown in Figure S1 and described in detail below.Blank run:
a volume of 1 mL Buffer A was injected. A linear gradient elution
was performed over 60 min (30 CVs) from 0% Buffer B to 100% Buffer
B.Case I: as load, 1 mL of 10 mg/mL Hemo and 10 mg/mL β-LG
in Buffer A was injected. A linear gradient elution identical to the
blank run was performed (0–100% Buffer B in 60 min).Case II: as load, 1 mL of 5 mg/mL Hemo and 5 mg/mL β-LG in
Buffer A was injected. A linear gradient elution was performed over
30 min (15 CVs) from 0% Buffer B to 100% Buffer B.Case III:
as load, 1 mL of 5 mg/mL Hemo and 5 mg/mL β-LG
in Buffer A was injected. A step gradient elution was performed with
25% Buffer B 6 min after injection, 50% Buffer B 20.5 min after injection,
and 100% Buffer B 30.5 min after injection.
EC-QCL-Based Mid-IR Measurements
The ChemDetect Analyzer,
equipped with an EC-QCL (tunable between 1350 and 1750 cm–1) and a diamond transmission flow cell (25 μm optical path
length), was used for acquisition of laser-based mid-IR spectra. An
external water-cooling unit was operated at 17 °C in order to
ensure thermal stabilization of the laser head during operation. Spectra
acquisition was performed with the ChemDetect Software package. For
flow-through LC-QCL-IR measurements, first, a reference background
spectrum was recorded by averaging 121 scans (60 s), followed by continuous
acquisition of 20 averaged scans, resulting in a measurement time
of 10 s per spectrum. Protein secondary structure off-line reference
measurements were acquired by averaging 91 scans during a measurement
time of 45 s.
HPLC Reference Measurements
As an
off-line analytical
method for the collected fractions, a previously published RP-HPLC
method was applied.[38] In short, a BioResolve
RP mAb Polyphenyl column (Waters, MA, USA) was used with MQ as mobile
phase A and ACN as mobile phase B, both supplemented with 0.1% v/v
TFA. Column temperature was kept constant at 70 °C and an injection
volume of 2 μL was used for all samples. Total run time for
one injection was 18 min with a flow of 0.4 mL/min and UV/vis absorbance
at 214, 280, and 404 nm was recorded for the whole run. A linear gradient
from 25% B to 75% B (10 min) was performed, followed by 2 min with
75% B, after which the column was re-equilibrated with 25% B for 6
min.
Quantification of IR and UV Signals
Protein concentrations
(c) across the chromatographic run were calculated from mid-IR and
UV signals, according to the Beer–Lambert lawwhere A denotes the recorded
absorbance values, d is the transmission path, and
ε indicates the molar decadic absorption coefficient.For laser-based IR spectroscopy, values for A were obtained by integrating
the spectra across the amide II region (1500–1600 cm–1). The absorption coefficients (ε) for the two proteins were
received by integrating the same area region from off-line recorded
reference spectra with known protein concentrations.For quantification
of UV/vis spectra, absorption coefficients of
the proteins were obtained by performing reference measurements using
a Cary 50 Bio UV/vis spectrometer (Agilent Technologies, Santa Clara,
California, USA). Spectra were recorded using the Cary WinUV software.
Cuvettes with a path length of 10 mm were used to measure 0.5 mg/mL
protein solutions. The thereby obtained absorption coefficients at
280 nm agreed well with those calculated via ExPASy ProtParam tool[39] and were used to calculate protein concentrations.
Data Processing
Data processing and analysis was conducted
with in-house codes developed in Matlab R2020b (MathWorks, Inc., Natick,
MA, 2020). During the first preprocessing step, absorbance bands of
water vapor from the atmosphere were automatically corrected by subtraction
of a water vapour reference spectrum, if necessary. Gradient correction
was performed by direct blank run subtraction (Case I) and a modified
procedure based on RSM (Case II and Case III).[32] Finally, absorption spectra were smoothed with a second
order Savitzky–Golay filter (window = 15 points). The spectral
resolutions of 1.2 and 3.6 cm–1 were determined
for smoothed and unsmoothed spectra by comparing the bandwidth of
water vapor spectra to FTIR reference spectra.
Results and Discussion
Relation
between Salt Gradient, Conductivity, and Mid-IR Spectra
In-line
monitoring of prep-LC effluents was performed with the
ChemDetect Analyzer as well as with conventional UV/vis and conductivity
detectors (Figure ). In addition to the routinely recorded signals, laser-based mid-IR
spectroscopy offered robust spectra acquisition in the most important
wavenumber range for protein secondary structure analysis.In
the present LC–QCL–IR application, IEX was employed
as the separation mechanism. Here, the pI of the analyte and the pH
of the mobile phase are decisive for the degree of retention. Bound
analytes are commonly eluted by utilizing a gradient of increasing
salt concentration. Conductivity detection is a widespread method
for measuring the salt concentration in IEX.[40]Figure displays
the signals of the mid-IR and conductivity detector on the example
of an IEX blank run with a linear NaCl gradient, and thus, the corresponding
signal (A) linearly increases with the NaCl concentration.
Figure 2
Detector signals
from an ion-exchange preparative-liquid chromatography
blank run with a NaCl gradient (0–1 M between 10 and 70 min).
(A) Conductivity detector response. (B) Heatmap of the corresponding
laser-based mid-IR spectra. White areas indicate negative absorbance
and dark blue areas positive absorbance. (C) Absorbance spectra of
0.3 M NaCl (retention time: 31.3 min, blue line) and 5 mg/mL hemoglobin
(red, dashed line).
Detector signals
from an ion-exchange preparative-liquid chromatography
blank run with a NaCl gradient (0–1 M between 10 and 70 min).
(A) Conductivity detector response. (B) Heatmap of the corresponding
laser-based mid-IR spectra. White areas indicate negative absorbance
and dark blue areas positive absorbance. (C) Absorbance spectra of
0.3 M NaCl (retention time: 31.3 min, blue line) and 5 mg/mL hemoglobin
(red, dashed line).The blue line in Figure C shows the distinctive
mid-IR spectrum of NaCl (blue line)
with the NaCl-free starting buffer as the reference. Increasing absorbance
can be observed between approximately 1610 and 1680 cm–1, while the absorbance in the remaining spectrum decreases with higher
salt concentrations. Even though Na+ and Cl– are not IR active, reorganization of the water molecules due to
the presence of these ions is responsible for a change in the mid-IR
spectrum compared to the NaCl-free initial buffer.[41] The maximum of the HOH bending band of water (at approximately
1643 cm–1) is not shifted in position; however,
its intensity increases, accompanied by narrowing of the band with
increasing NaCl concentrations.[42] The negative
absorbance in the other spectral regions can be related to partial
displacement of H2O molecules with salt.Figure B shows
a 2D heat map depicting the response of the NaCl gradient recorded
by the ChemDetect analyzer. White areas indicate negative absorbance,
whereas dark blue areas highlight positive absorbance.The red
dashed line in Figure C shows the IR spectrum of Hemo, illustrating the spectral
overlap of protein bands with the one of NaCl (blue line). Particularly,
the amide I band, representing the most important mid-IR region for
protein secondary structure determination, is impeded by the modified
HOH bending band of water. The described overlap of analyte and changing
solvent band indicate significant challenges when employing the ChemDetect
Analyzer for in-line monitoring of IEX. In order to obtain feasible
mid-IR protein spectra, the influence of the changing salt concentration
has to be eliminated. Consequently, feasible gradient compensation
strategies are discussed and demonstrated based on model protein systems
in the following subchapters.
Case I: Direct Blank Run
Subtraction to Enable In-Line Monitoring
of Proteins
In this approach, for background compensation,
an analyte-free reference blank run was performed with the same gradient
as the sample run. Based on the retention time, the reference spectra
were subsequently subtracted from the sample spectra.In the
sample run, Hemo and β-LG were included, due to their difference
in the pI and secondary structure. Figure shows the spectral 3D plots of the (A) uncorrected
sample run and the (B) reference blank run, as well as the (C) corrected
sample run after direct background subtraction. The mid-IR spectra
of the blank run were already discussed above. Even though the amide
I and amide II bands are visible in the uncorrected sample run, the
effect of the NaCl gradient on the absorbance spectra is clearly dominating.
Thus, no reliable information about the protein secondary structure
can be obtained. In contrast, the corrected sample run shows a stable
baseline and distinct protein spectra. Figure depicts the absorbance spectra extracted
from the peak maxima of the differently corrected sample runs as well
as off-line recorded reference IR spectra of the investigated proteins.
For Case I (blue lines), the first chromatographic peak with a maximum
at approximately 17 min can be related to Hemo due to its higher pI
of 7.1.[43] Hemo mainly contains α-helical
secondary structures and shows the characteristic narrow amide I band
with a maximum at approximately 1656 cm–1 and a
narrow amide II band at 1545 cm–1.[7,44,45] β-LG is predominantly composed
of β-sheet secondary structures[46] and has a pI of 5.1[47] and thus elutes
second under these IEX conditions. The corresponding absorbance spectra
show broader amide I and II bands with β-sheet typical maxima
at 1632 and 1550 cm–1 and a shoulder at 1680 cm–1.[22,48,49] The obtained high-quality protein spectra thus indicate excellent
long-term stability of the ChemDetect Analyzer.
Figure 3
Spectral 3D plots for
Case I: (A) uncorrected sample run, (B) reference
blank run, and (C) corrected sample run after direct background subtraction.
Figure 4
Laser-based mid-IR absorbance spectra, extracted from
the (A) first
and (B) second peak maxima of the corrected chromatographic runs from
Case I (blue lines), Case II (red lines), Case III (green lines),
and off-line recorded reference IR spectra (black lines) of 5 mg/mL
hemoglobin and β-lactoglobulin. The individual spectra were
offset for better visibility. The pointed black lines indicate zero
absorbance for every spectrum.
Spectral 3D plots for
Case I: (A) uncorrected sample run, (B) reference
blank run, and (C) corrected sample run after direct background subtraction.Laser-based mid-IR absorbance spectra, extracted from
the (A) first
and (B) second peak maxima of the corrected chromatographic runs from
Case I (blue lines), Case II (red lines), Case III (green lines),
and off-line recorded reference IR spectra (black lines) of 5 mg/mL
hemoglobin and β-lactoglobulin. The individual spectra were
offset for better visibility. The pointed black lines indicate zero
absorbance for every spectrum.A drawback of this direct gradient compensation approach is the
need for a blank run that is performed identical to the sample run.
Consequently, any adaptions in the gradient profile require acquisition
of an additional blank run and very high reproducibility of the chromatographic
system is required in general. The additional time consumption of
the blank run and the re-equilibration phase of the column make this
approach methodically rigid and impractical to realize for bench-scale
as well as industrial applications. Hence, a more flexible strategy
is presented in the next subchapter.
Case II and III: Flexible
Gradient Compensation with Adapted
RSM
Even though direct blank run subtraction revealed excellent
protein spectra, more flexible and less laborious gradient compensation
strategies are favorable. Gradient compensation based on RSM requires
only a single blank run, which can be used for correcting various
gradient profiles.[32] This approach was
successfully applied for HPLC-FTIR, where the entire mid-IR range
and eluent specific absorbance bands were available. These specific
bands were used as mobile phase identification parameter (IP) in order
to select the blank spectrum to be subtracted. This approach, however,
is not feasible for monitoring proteins in IEX due to lack of specific
absorbance bands and limited sensitivity of FTIR spectroscopy. Due
to significantly higher spectral power densities of the EC-QCL, the
ChemDetect Analyzer overcomes the typical limitations of FTIR spectroscopy
and offers robust and highly sensitive flow-through measurements in
the most important wavenumber range for protein analysis.In
the present study, a modified RSM method based on an external IP is
proposed and successfully applied. For this purpose, the signal from
the conductivity detector was related to the mid-IR spectra. The measured
conductivity represents an adequate IP, due to its high dependency
on the NaCl concentration, whereas the influence of the protein concentration
on the conductivity is negligible (Figure S2). For this purpose, the blank run with the linear NaCl gradient
(same as in Case I) was taken as the reference. In order to evaluate
the applicability of this approach, two chromatographic sample runs
with different gradients compared to the one from the reference run
were conducted: Case II included a linear gradient with a steeper
profile, while Case III featured a 3-step gradient. A specific conductivity
value was assigned to each mid-IR spectrum in the SM and RSM, based
on the retention time. Figure depicts the principle of the applied compensation method
on the example of a baseline and protein spectrum in Case III. Based
on the comparison of the measured conductivity in the (C) sample run
and (A) the reference blank run, the corresponding (B) blank run spectra
are identified. Finally, the (D) sample spectra are corrected by subtracting
the identified blank run spectra with the most similar conductivity
value.
Figure 5
Principle of the applied RSM-based background compensation method.
(A) Conductivity detector signals of the reference blank run and (B)
sample runs of Case II and Case III. Dotted lines indicate the relation
between selected conductivity values and the corresponding mid-IR
spectra of background (orange) and protein (purple) of Case III. The
absorbance spectra of the (C) blank run were subtracted from (D) the
3-step gradient sample run spectra, based on the most similar conductivity
value, revealing distinctive protein spectra (purple) and a flat baseline
(orange).
Principle of the applied RSM-based background compensation method.
(A) Conductivity detector signals of the reference blank run and (B)
sample runs of Case II and Case III. Dotted lines indicate the relation
between selected conductivity values and the corresponding mid-IR
spectra of background (orange) and protein (purple) of Case III. The
absorbance spectra of the (C) blank run were subtracted from (D) the
3-step gradient sample run spectra, based on the most similar conductivity
value, revealing distinctive protein spectra (purple) and a flat baseline
(orange).Figure displays
the spectral 3D plots of the uncorrected (A,C) and corrected (B,D)
sample runs. The uncorrected run from Case II constitutes a similar
profile as the reference blank run with a shorter measurement time,
whereas the run from Case III shows a clearly different profile due
to the 3-step gradient. Both corrected runs comprise stable baselines
and excellent protein spectra, comparable to those from direct blank
run subtraction (Figure ). Consequently, acquisition of a singular blank run is sufficient
for correcting sample runs of widely different gradient profiles.
Due to high flexibility and little additional time consumption, the
applied gradient compensation approach shows high potential for industrial
application.
Figure 6
Spectral 3D plots for Case II and Case III: (A,C) uncorrected
sample
runs and (B,D) corrected sample runs after background correction with
adapted RSM.
Spectral 3D plots for Case II and Case III: (A,C) uncorrected
sample
runs and (B,D) corrected sample runs after background correction with
adapted RSM.
Comparison of Mid-IR In-Line
Measurements to Conventional Off-Line
Analysis
In order to verify the signal of the ChemDetect
Analyzer, protein concentrations calculated from mid-IR absorbance
were compared to those calculated from the UV detector signal. Figure displays the concentrations,
determined from the ChemDetect Analyzer (red line) and UV detector
(black line) over the chromatographic run from Case III. In mid-IR
spectroscopy, the total protein content is best represented by the
amide II band because it is less influenced by water absorption than
the amide I band. Thus, the wavenumber region between 1500 and 1600
cm–1 was integrated to obtain the absorbance values.
For comparison, protein concentrations were also calculated from the
UV signals at 280 nm recorded by the UV detector of the LC instrument.
UV absorption at this wavelength detects aromatic protein residues
and disulfide bonds and is most commonly used to monitor proteins
in chromatographic applications.[1] Absorption
coefficients for Hemo and β-LG for IR and UV spectroscopy were
obtained from reference measurements with known protein concentrations.
Because these values are different for the two proteins, for conversion
of the recorded signals to concentration values, the chromatogram
was split into two parts and the absorption coefficients of Hemo were
used from 0 to 19 min, whereas the coefficients of β-LG were
applied between 19 and 40 min. Figure depicts the highly overlapping protein concentrations
derived from the signals of ChemDetect and UV detector in the time
axis as well as for the protein concentration, demonstrating that
equivalent quantitative signals can be obtained by the two methods.
Figure 7
Comparison
of protein concentrations obtained from mid-IR amide
II band (red line) and UV detector signal at 280 nm (black line) across
the chromatographic run from Case III and protein reference concentrations
obtained by measuring the collected fractions with reversed-phase
HPLC (green and blue bars).
Comparison
of protein concentrations obtained from mid-IR amide
II band (red line) and UV detector signal at 280 nm (black line) across
the chromatographic run from Case III and protein reference concentrations
obtained by measuring the collected fractions with reversed-phase
HPLC (green and blue bars).For protein identification, IR spectra from the two chromatographic
peak maxima were extracted and compared to pure Hemo and β-LG
reference spectra, respectively. Figure reveals excellent agreement between in-line
and off-line ChemDetect measurements. Absorbance band positions as
well as band shapes show excellent comparability between mid-IR spectra
of the chromatographic peak maxima and those of the pure reference
solutions. In contrast, with conventional LC instrumentation employing
an UV detector, real-time information about the protein secondary
structure, and consequently type of protein cannot be obtained. In
practical applications where the protein identity or purity are usually
unknown, highly biased results would be obtained from direct quantitation
of the UV signal because absorption coefficients at 280 nm can be
different by more than an order of magnitude between proteins.[50] Then, protein identification must be performed
by off-line RP-HPLC measurements of the collected fractions. In this
regard, the green bars shown in Figure indicate the concentration of the protein that was
identified as Hemo by RP-HPLC, and the blue bars show the concentration
of β-LG. Consequently, the obtained LC-QCL-IR chromatograms
agree well with the conventionally applied quantification methods,
while offering the additional advantage of providing real-time information
about the protein secondary structure.
Conclusions and Outlook
In this work, a laser-based mid-IR spectrometer was successfully
applied for in-line monitoring of proteins from preparative IEX. A
major challenge was the highly overlapping absorbance bands of proteins
and NaCl gradient that dominated the recorded sample run spectra.
Advanced background compensation strategies based on adapted RSM were
developed and their implementation in three different case studies
resulted in high-quality protein spectra. In Case I, a reference blank
run was directly subtracted from a sample run with the same gradient
profile. The obtained protein spectra indicated excellent long-term
stability of the ChemDetect Analyzer. Case II and III included sample
runs with a steeper linear gradient and a 3-step gradient, respectively.
Here, a novel gradient compensation approach based on RSM and conductivity
IP was introduced. With this method, a single blank run was sufficient
for compensating the NaCl gradient in sample runs with distinctively
different profiles. The thereby obtained protein spectra showed excellent
comparability to off-line reference measurements.It was shown
that the protein concentrations evaluated from signals
obtained by the ChemDetect Analyzer are equivalent to UV spectroscopy
at 280 nm, which is the standard quantification method for proteins
in chromatographic systems. Furthermore, compared to conventional
LC detectors, the laser-based mid-IR spectrometer offers the major
advantage of providing real-time information about the protein secondary
structure, comparable to high-end off-line measurements. The ChemDetect
Analyzer thus holds high potential for complementing laborious and
time-consuming off-line methods and provides an easily accessible
in-line method. In combination with the presented adapted RSM method
to compensate for varying gradient profiles, QCL-IR presents a powerful
tool for in-process monitoring and control. Especially, in the light
of QbD principles, a near real-time PAT tool able to give information
about protein secondary structures and corresponding CQAs presents
high potential. In the future, LC-QCL-IR coupling can be employed
for chemometrics-based analysis of possible impurities and individual
quantification of co-eluting proteins.
Authors: Milagros Montemurro; Andreas Schwaighofer; Anatol Schmidt; María J Culzoni; Helmut K Mayer; Bernhard Lendl Journal: Analyst Date: 2019-09-09 Impact factor: 4.616
Authors: Andreas Schwaighofer; Milagros Montemurro; Stephan Freitag; Christian Kristament; María J Culzoni; Bernhard Lendl Journal: Anal Chem Date: 2018-05-24 Impact factor: 6.986
Authors: Christopher K Akhgar; Julian Ebner; Mirta R Alcaraz; Julian Kopp; Héctor Goicoechea; Oliver Spadiut; Andreas Schwaighofer; Bernhard Lendl Journal: Anal Chem Date: 2022-08-04 Impact factor: 8.008