Dorota Maria Borowicz1, Agnieszka Skrobała1,2, Marta Kruszyna-Mochalska1,2, Julian Malicki1,2.
Abstract
BACKGROUND: Hypofractionated stereotactic body radiation therapy (SBRT) for prostate cancer (PCa) can be delivered with the robot-assisted CyberKnife (CK) system or on a linear accelerator using dynamic intensity-modulated radiotherapy (IMRT) or volumetric arc radiotherapy (VMAT). This retrospective study was performed to determine whether CK offers better dose distribution than IMRT and/or VMAT.Entities:
Keywords: CyberKnife; intensity-modulated radiotherapy; prostate carcinoma; stereotactic radiotherapy
Mesh:
Year: 2022 PMID: 35344646 PMCID: PMC9122296 DOI: 10.2478/raon-2022-0010
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 4.214
Treatment plan acceptance criteria
| Structure | Criteria |
|---|---|
| PTV: IMRT and VMAT | DMax:107% of the prescribed dose |
| DMin: 95% of the prescribed dose | |
| Prescribed isodose: 78% | |
| PTV: Cyber Knife | Total dose of prescribed isodose covering at least 95% of PTV volume |
| V18 Gy < 50% Vol | |
| V29 Gy < 20% Vol | |
| Rectum | V32.6 Gy < 10% Vol |
| V36.25 Gy < 5% Vol | |
| V36 Gy < 1 mm3 | |
| V18 Gy < 55% Vol | |
| V29 Gy < 25% Vol | |
| Bladder | V32.6 Gy < 15% Vol |
| V36.25 Gy < 10% Vol | |
| V37 Gy < 10 mm3 | |
| Femoral heads: left and right | V25 Gy < 45% Vol |
IMRT = intensity-modulated radiotherapy; PTV = planning target volume; VMAT = volumetric arc radiotherapy
Figure 1Example dose distributions on a transversal scan for (A) IMRT plan, (B) VMAT plan and (C) CyberKnife plan for the same patient.
Dosimetric parameters for VMAT, IMRT and CyberKnife
| VMAT | IMRT | CyberKnife | p | |
|---|---|---|---|---|
| D mean PTV | 36.25 | 36.25 | 39.59±1.11 | - |
| D min PTV | 34.6±0.41 | 34.67±0.52 | 31.48±0.99 | p < 0.001 |
| D max PTV | 38.76±0.47 | 37.57±0.4 | 45.89±0.94 | p < 0.001 |
| Target coverage D98% | 97.36±1.02 | 99.25±0.75 | 96.08±0.76 | p < 0.001 |
D = dose; PTV = planning target volume; SD = standard deviation
Mean dosimetric parameters for the three techniques
| VMAT | IMRT | CyberKnife | p | |
|---|---|---|---|---|
|
| ||||
| V18 Gy [%] | 15.33 ± 2.97 | 15.17 ± 2.79 | 22.58 ± 6.6 | p |
| V29 Gy [%] | 7.13 ± 2 | 7.25 ± 1.89 | 7.18 ± 2.82 | p |
| V32.6 Gy [%] | 4.75 ± 1.51 | 4.82 ± 1.58 | 3.23 ± 1.66 | p |
| V36.25 Gy [%] | 0.54 ± 0.29 | 0.32 ± 0.25 | 0.24 ± 0.25 | p |
| V36 Gy in mm3 | 591.59 ± 226.53 | 422.62 ± 236.64 | 219.78 ± 235.02 | p |
|
| ||||
| V18 Gy [%] | 11.98 ± 5.33 | 14.93 ± 5.56 | 19.25 ± 8.62 | p |
| V29 Gy [%] | 5.87 ± 2.32 | 8 ± 2.73 | 7.35 ± 3.27 | p |
| V32.6 Gy [%] | 4.38 ± 1.72 | 5.7 ± 2.08 | 4.71 ± 2.14 | p |
| V36.25 Gy [%] | 2 ± 0.93 | 1.22 ± 0.81 | 2.16 ± 1.09 | p |
| V37 Gy in mm3 | 1090.75 ± 610.81 | 4.5 ± 15.7 | 3256 ± 1518.84 | p |
| Left femoral head V25 Gy | 0 | 0.12 ± 0.25 | 0 | |
| Right femoral head V25 Gy | 0 | 0.12 ± 0.3 | 0 |
p values > 0.05 were considered non-significant
Mean conformity and homogeneity index values
| VMAT | IMRT | CyberKnife | p | |
|---|---|---|---|---|
| CI | 1.17 ± 0.04 | 1.25 ± 0.07 | 1.07 ± 0.03 | p |
| HI | 1.07 ± 0.01 | 1.04 ± 0.01 | 1.25 ± 0.03 | p |
Figure 2Gamma index for verification plans.
Figure 3IMRT gamma index (2%/2mm) values for each patient.
Figure 4Gamma Index for IMRT for each treatment beam and all plans.