Jasmine A Graham1, Gage Redler1, Kirby B Delozier1, Hsiang-Hsuan Michael Yu1, Daniel E Oliver1, Stephen A Rosenberg1.
Abstract
PURPOSE/OBJECTIVE(S): Whole brain radiotherapy with hippocampal avoidance (HA-WBRT) is a technique utilized to treat metastatic brain disease while preserving memory and neurocognitive function. We hypothesized that the treatment planning and delivery of HA-WBRT plans is feasible with an MRI-guided linear accelerator (linac) and compared plan results with clinical non-MRI-guided C-Arm linac plans. MATERIALS/Entities:
Keywords: Dosimetry; HA-WBRT; MRI linac; Treatment planning
Mesh:
Year: 2022 PMID: 35344266 PMCID: PMC9195042 DOI: 10.1002/acm2.13587
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
Optimization approach for MRI‐guided radiotherapy (MRgRT) plans
| Structure | Purpose | Type | Importance | Power | Threshold |
|---|---|---|---|---|---|
| PTV | Increase 30 Gy target coverage | Increase dose | 66 (10–250) | 1.6 (1.5–2.0) | 30.1 Gy (30–31 Gy) |
| PTV | Decrease 33 Gy heterogeneity | Decrease dose | 66 (10–250) | 1.6 (1.5–2.0) | 33.1 Gy (33–34 Gy) |
| Optic nerves/chiasm | Limit Dmax | Decrease dose | 6.5 (1–4) | 1 (1–1.3) | 31.7 Gy (28–33 Gy) |
| Hippocampi | Limit Dmax | Decrease dose | 28.3 (1–100) | 1 | 15.4 Gy (13–16 Gy) |
| Hippocampi | Limit D100% | Decrease dose | 13.9 (1–100) | 1 | 8.5Gy (7.9–9.0 Gy) |
| Lenses | Approach as low as reasonably achievable (ALARA) dose | Decrease dose | 1.3 (1–3) | 1 | 4.6 Gy (4.3–5.0 Gy) |
| Normal tissue/PTV ring structures | Provide conformality and reduce scalp dose | Decrease dose | 1 | 1 | 29 Gy |
Abbreviation: PTV, planning target volume.
FIGURE 1Representative beam arrangements for (a) MRI‐guided radiotherapy (MRgRT) co‐planar fixed‐field IMRT and (b) noncoplanar volumetrically modulated arc therapy (VMAT) on a conventional c‐arm linear accelerator
FIGURE 2Comparison of achieved dose distributions for hippocampal avoidance (HA) whole brain radiotherapy (WBRT) in a representative patient plan. (a) MRI‐guided radiotherapy (MRgRT) with co‐planar fixed‐field IMRT. (b) Clinical plan with noncoplanar volumetrically modulated arc therapy (VMAT) on a conventional c‐arm linear accelerator
FIGURE 3Planning target volume (PTV) dose volume histogram (DVH) comparisons for all cases. MR‐linac plans are in blue and clinical volumetrically modulated arc therapy (VMAT) plans are in gold
Average dosimetric values for clinical VMAT plans and MRgRT (static field IMRT) plans
| Average values | |||
|---|---|---|---|
| Clinical VMAT | MRgRT |
| |
|
| 95.02% ± 0.00% | 95.00% ± 0.03% |
|
|
| 0.00% ± 0.00% | 0.03% ± 0.03% |
|
|
| 99.50% ± 0.66% | 99.59% ± 0.16% |
|
|
| 96.26% ± 1.52% | 96.08% ± 0.41% |
|
|
| 14.19 Gy ± 1.29 Gy | 15.00 Gy ± 1.51 Gy |
|
|
| 8.62 Gy ± 0.51 Gy | 7.92 Gy ± 0.33 Gy |
|
|
| 30.94 Gy ± 1.10 Gy | 31.26 Gy ± 1.97 Gy |
|
|
| 5.64 Gy ± 2.53 Gy | 6.82 Gy ± 4.21 Gy |
|
Note: PTV V37.5 and Hippocampi D100% are statistically significant.
Abbreviations: CTV, clinical target volume; MRgRT, MRI‐guided radiotherapy; PTV, planning target volume; VMAT, volumetrically modulated arc therapy.