Does cerebral oximetry always measure brain tissue oxygen saturation? An anatomical study utilizing computed tomography.

Rotem Naftalovich1,2, Darrick Chyu3, John T Denny4, Aysha Hasan5, Enrique J Pantin4.   

Abstract

Background and Aims: to quantify the scalp-cortex distance and determine its variation among patients. We hypothesized that in a significant number of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters. Material and
Methods: A retrospective anatomic study using transverse head CT images selected randomly from 102 patients over the age of 18 years without brain swelling, intracranial mass effect, or brain hemorrhage. Scalp-cortex distances were determined at two separate locations along the craniocaudal axis; most cephalad to the frontal sinus (I0) and also 2 cm cephalad to that location (I2). Multiple measurements were obtained bilaterally at 1, 3, 5, 7, and 9 cm from midline.
Results: The average scalp-cortex distance was 14.3 mm and 15 mm at I0 and I2 respectively. Distances varied more in I2 than in I0; from the measurements, 12.8% vs. 6.8% were over 20 mm, 4.4% vs. 2.2% over 25 mm, 1.1% vs. 0.6% over 35 mm and 0.6% vs. none over 40 mm at I2 and I0, respectively. 1.5% of the measurements at I2 were over 30 mm.
Conclusion: Cerebral oximetry manufacturers all claim to measure cerebral tissue up to a depth of 20-25 mm; 20 mm with the EQUANOX and INVOS compared with 25 mm with the FORE-SIGHT. Scalp-cortex distance is within 25 mm in more than 95% of patients. However, even with the probe placed as per the manufacturer's recommendations, in a small but significant subset of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters and hence may not reflect actual brain tissue oxygen saturation. Copyright:
© 2022 Journal of Anaesthesiology Clinical Pharmacology.

Entities:  

Keywords:  Brain; cerebral oximeters; cortex; depth; distance; scalp

Year:  2022        PMID: 35340957      PMCID: PMC8944368          DOI: 10.4103/joacp.JOACP_395_19

Source DB:  PubMed          Journal:  J Anaesthesiol Clin Pharmacol        ISSN: 0970-9185


Introduction

Continuous real-time monitoring of cerebral oxygenation by noninvasive near-infrared spectroscopy (NIRS)-based cerebral oximetry is increasingly used in neurosurgical, cardiac and vascular surgeries to estimate regional tissue oxygen saturation, rSO2, (sometimes referred to as ScO2, SctO2, or rStO2) by transcutaneous measurement of the cerebral cortex using forehead probes. To do so, these devices utilize the process of spatial resolution via multiple detectors located at different distances from the light emitter. The underlining idea is that a longer distance between the light emitter and the detector leads to deeper tissue assessment. Given the multiple detectors, an automated algorithm can then subtract the more superficial reading (i.e., the shorter distance between the light emitter and detector) from the deeper one. Regardless of their proprietary algorithms and calibration, the major cerebral oximetry manufacturers all claim to measure cerebral tissue up to a maximum depth in the range of 20–25 mm (~50% of the emitter-detector distance); 20 mm with the EQUANOX[1] (Nonin Medical Inc; Plymouth, MN, USA) and INVOS (Covidien; Boulder, CO, USA)[2] compared with 25 mm with the FORE-SIGHT[3] (CAS Medical Systems Inc; Brandford, CT, USA). If the scalp-cortex distance, i.e., depth of the brain, is greater than the maximum detection depth of the device then the rSO2 reading would be inaccurate. Anatomical variations among patients,[45] particularly those relating to the distance between the brain cortex and the skin are clinically important to understand since their knowledge can directly guide proper and better use of brain monitors. For example, variations in skull and scalp thickness have numerous effects on electroencephalography (EEG) and magnetoencephalography (MEG) monitoring.[56] Skull thickness varies greatly[7891011]; the thickness of the frontal bone measured by histology from cadaver samples ranged from 5.05–8.13 mm in a study of 7 patients.[12] Similarly, cerebrospinal fluid (CSF) volume varies among patients and appears to be influenced by age likely due to brain tissue atrophy.[13] Given that variations exist among the population in each of the major anatomical components that make up the scalp-cortex distance, it is, therefore, reasonable to assume that the distance between the scalp and the brain also varies among patients. However, quantification of this distance is not available in the literature. This study aimed to quantify the distance between the cortex and the scalp and determine its variation among patients. We did this retrospectively using brain computed tomography (CT) imaging.

Material and Methods

Trigonometric corrections have been applied to imaging in previous studies to more accurately infer actual anatomic size[1415] and are even used to make preoperative corrections for guiding orthopedic surgery.[1617] Because the forehead lies obliquely relative to the standard axial imaging plane, CT measurements of brain depth need to be trigonometrically corrected [Figure 1] to more accurately reflect brain depth. The oximeter probe lays flush to the skin on the forehead, i.e., in the plane tangential to the forehead skin and emits light in the plane perpendicular to it.
Figure 1

Schematic of the trigonometric correction factor used to compensate for the curvature of the patient’s forehead. The black Cartesian plane represents the imaging with its x and y axes representing the coronal and transverse (axial) planes, respectively. The coronal axis is shown even though we only studied transverse images. The uncorrected CT depth measurements are represented by the segment labeled d which overestimates the true scalp cortex distance represented by segment d’. Using the x-axis of the red Cartesian plane, the line tangential to the patient’s forehead, we derive d’ = d cos Θ

Schematic of the trigonometric correction factor used to compensate for the curvature of the patient’s forehead. The black Cartesian plane represents the imaging with its x and y axes representing the coronal and transverse (axial) planes, respectively. The coronal axis is shown even though we only studied transverse images. The uncorrected CT depth measurements are represented by the segment labeled d which overestimates the true scalp cortex distance represented by segment d’. Using the x-axis of the red Cartesian plane, the line tangential to the patient’s forehead, we derive d’ = d cos Θ After approval from the Rutgers Health Sciences Institutional Review Board, data were collected from 102 head CTs selected randomly from the institution’s radiology database of CT images taken after October 2012 and their accompanying official final reads from the radiologists. CT was chosen because it is the gold standard for structural evaluation of the cranium.[5] The only inclusion criteria was age of at least 18 years. We excluded patients with a history of cerebral edema, intracranial mass effect, intracranial hemorrhage, or head trauma. For each patient, transverse scalp-cortex distances (d) were determined at two separate locations along the craniocaudal axis; the first location being the most cephalad aspect of the frontal sinus (I0) and the second (I2) being 2 cm cephalad to I0. The scalp location on the I0 transverse plane is superior to the eye-brow landmark and roughly correlates to the cerebral oximeter probe location whereby I2 would represent a high placement of the oximeter probe as may occur if it is used concurrently with a Bispectral Index (BIS) monitor or with a Mayfield skull clamp. Since the thickness of the calvarium varies at different points,[18] multiple measurements were obtained at both the I0 and I2 levels. To do this, we used the imaging software measuring tool and measured d bilaterally at the scalp locations of 1, 3, 5, 7, and 9 cm from midline which was defined by the falx cerebri [Figure 2]. The transverse CT slice thickness (i.e. the distance along the craniocaudal axis between consecutive CT images) was 2 mm. To determineΘ, the consecutive cephalad image was utilized whereby the 2 mm distance represented the segment adjacent to Θ on a triangle laying in the coronal plane with a hypotenuse being the segment connecting I0 scalp point with the scalp point of the next consecutive cephalad image. The distance of this hypotenuse segment was measured (the software enables the user to connect points laying on different planes using the mouse cursor; in this case consecutive image slices) and Θ was then calculated from the equation cos Θ = adjacent/hypotenuse. The above process for computing Θ was repeated for each of the measurements at 1, 3, 5, 7, and 9 cm bilaterally. The entire process described above was then repeated for I2.
Figure 2

A schematic demonstrating the various points of measurement in each image

A schematic demonstrating the various points of measurement in each image After all the d and Θ values were obtained, the respective d’ for each was then calculated using the equation cos Θ = d/d’. The parameter of interest, d’, does not represent the scalp-cortex distance along the I0 or I2 planes (that distance is d which is an overestimation of the true distance) but rather the orthogonal distance from the scalp on those planes to the point on the cortex of the consecutive cephalic image.

Results

On average the patients studied were 64 years of age (18–94), 165.7 cm (132.1–195.6) tall, and weighed 78.42 kg (26–185.8). The average orthogonal scalp-cortex distance was 14.3 mm and 15 mm at I0 and I2, respectively. These orthogonal scalp-cortex distances varied more in I2 then in I0; from the measurements, 12.8% vs. 6.8% were over 20 mm, 4.4% vs. 2.2% over 25 mm, 1.1% vs. 0.6% over 35 mm and 0.6% vs. none over 40 mm in I2 and I0, respectively. Of note, 1.5% of the measurements at I2 were over 30 mm. If the trigonometric correction is not applied (i.e. examining d instead of d’), 9.5% of measurements at I0 were over 20 mm and 2.3% of them over 25 mm. At I2, 21% were over 20 mm and 8.7% over 25 mm when uncorrected.

Discussion

This anatomical variance in brain depth likely contributes to the lack of consensus and the poorly defined role that cerebral oximetry devices currently have in cardiovascular and neurosurgical anesthesia.[19] Furthermore, this variation in brain depth is expected to contribute to the varying error magnitude observed among subjects with all of the available cerebral oximeters.[20] The above results indicate that in a significant number of patients, their brain lays deeper than the maximum detection depth of the device. This is particularly the case with the EQUANOX and INVOS and if the probe is placed higher on the forehead. Consequently, we recommend a lower placement of the probe. This is important to consider in elderly patients who may have brain atrophy. A slight difference in brain depth can mean that the brain tissue is deeper than the maximum depth detected by the device and hence the rSO2 reading is not measuring cortex oxygenation in which case the anesthesiologist would be under the impression that the monitor reading reflects brain tissue oxygenation when in fact it does not. This is a clinical problem and identification of the subset of patients for which the rSO2 reading is not indeed rSO2 may be of interest. Some would incline to believe that the incidence of false rSO2 reading is lower than the ~ 5% incidence where the brain tissue is deeper than the maximum depth detected by the device (2.2–6.8% with the probe placed slightly above the eyebrow). Regardless, a false rSO2 may very well go unnoticed by the clinician. Even though the study excluded patients with a history of cerebral edema, intracranial mass effect, intracranial hemorrhage, or head trauma, it did not investigate the medical record extensively to identify medical conditions that could lead to calvarial thickening. For example, calvarial thickening can occur in chronic ventricular shunting, acromegaly, in chronic antiepileptic use such as with phenytoin, in hyperparathyroidism, Paget’s disease and even with anemia. We do not believe that the typical user of these devices would routinely evaluate such conditions in their decision of whether or not to use the device. Our opinion is based on the fact that the INVOS manual published by Covidien claims no contraindications for the use of their cerebral oximetry (e.g. Operations Manual INVOS® System, Model 5100C). Another limitation of this purely retrospective anatomic study is that we did not correlate the anatomical variation of brain depth with rSO2. Such correlation can be done in a prospective study of patients undergoing perioperative cerebral oximetry monitoring with preoperative CTs and should take into consideration hemoglobin concentration, the thickness of the CSF layer, and other factors affecting brain tissue saturation.[21] An even more ideal imaging modality would be a high-resolution 3D reconstruction in which case trigonometric correction would not be needed. Light can only travel in a straight line, with angles for reflections and refractions. Even though the device emits light perpendicularly to the skin, the light path is not perpendicular to the skin throughout since the receiver is located on the forehead, lateral to the emitter. Therefore, at least some of the light penetrates the brain tissue at an angle. The brain surface is not a smooth sphere but rather rugged due to gyri and sulci, the effects of this on the device readings are difficult to predict due to certain properties of the light such as scattering and a non-collimated beam. We only quantified total scalp-cortex distance. The effect of anatomical variations on rSO2 readings is complicated by light beam refraction which occurs as photons travels across different mediums (e.g., skin-bone or CSF-brain tissue).[18] As a result, it is not solely the linear distance between the probe and the cortex that matters but also the proportion of each medium from the total distance. Nonetheless, in patients where the brain cortex lays deeper than the maximum detection depth of the cerebral oximeter, total depth, as opposed to the proportion of each medium, is a more critical factor to the device efficacy. The decision to use cerebral oximetry depends on the surgical case involved, clinical judgment, and clinician preference. Broadly speaking, it is used in elderly patients with neurovascular disease or stroke risk. We did not select for such patients in our study. As a result, the group of patients we studied differs from the population of patients who typically undergo surgery with cerebral oximetry but it is a fair representative sample. The average age of our patients was 64 years which is lower than the average age of patients in these surgeries; 68.8 years for cardiovascular surgery patients in Germany[22] and over 70 years for CABG patients in the USA.[23] Furthermore, the age distribution of our patients was likely wider than those undergoing surgery. Given the normal cerebral atrophy associated with age,[24] it is hence reasonably possible that patients undergoing cardiovascular surgery are older than our studied population, have more brain atrophy and hence their scalp-cortex distances are even larger than our results. We anticipate that rSO2 readings, or errors, would vary whether the probe is placed at the recommended location above the eyebrow vs. on the eyebrow itself. Moreover, rSO2 readings would be more accurate if the probe is placed on the eyebrow but we did not measure the scalp-cortex distance at the eyebrow, nor did we study the effect of eyebrow hair on the device light. Our findings show that the anatomical scalp to cortex distance is within 25 mm in more than 95% of patients; i.e., the scalp-cortex distance is greater than the maximum penetration depth of current devices in less than 5% of patients. This distance is larger when measured higher on the forehead, 2 cm above the cephalad edge of the frontal sinus, as compared to at the cephalad aspect of the sinus. Such higher forehead placement is likely to decrease the accuracy of the device, at least in a subset of patients with deeper brains. It appears that, even with probe placement as per the manufacturer’s recommendations, in a small subset of patients the cerebral oximeter reading may not reflect actual brain tissue oxygen saturation.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  22 in total

1.  Near-infrared light propagation in an adult head model. II. Effect of superficial tissue thickness on the sensitivity of the near-infrared spectroscopy signal.

Authors:  Eiji Okada; David T Delpy
Journal:  Appl Opt       Date:  2003-06-01       Impact factor: 1.980

Review 2.  Practical use of near-infrared spectroscopy in carotid surgery.

Authors:  Djordje Radak; Vuk Sotirovic; Milan Obradovic; Esma R Isenovic
Journal:  Angiology       Date:  2013-10-31       Impact factor: 3.619

3.  Skull thickness and magnitude of EEG alpha activity.

Authors:  Dirk Hagemann; Johannes Hewig; Christof Walter; Ewald Naumann
Journal:  Clin Neurophysiol       Date:  2008-04-01       Impact factor: 3.708

4.  Variation of bone layer thicknesses and trabecular volume fraction in the adult male human calvarium.

Authors:  Sourabh Boruah; Glenn R Paskoff; Barry S Shender; Damien L Subit; Robert S Salzar; Jeff R Crandall
Journal:  Bone       Date:  2015-04-25       Impact factor: 4.398

5.  Femoral imaging artifacts associated with dorsal recumbency craniocaudal radiographic positioning. Description of a modified bisecting angle technique.

Authors:  P Maguire; M Siclari; A Lesser
Journal:  Vet Comp Orthop Traumatol       Date:  2014-07-03       Impact factor: 1.358

6.  Quantitative volumetric analysis of brain MR: normative database spanning 5 decades of life.

Authors:  D D Blatter; E D Bigler; S D Gale; S C Johnson; C V Anderson; B M Burnett; N Parker; S Kurth; S D Horn
Journal:  AJNR Am J Neuroradiol       Date:  1995-02       Impact factor: 3.825

7.  Measurement of skull bone thickness for bone-anchored hearing aids: an experimental study comparing both a novel ultrasound system (SonoPointer) and computed tomographic scanning to mechanical measurements.

Authors:  Philipp A Federspil; Steffen H Tretbar; Friederike H Böhlen; Stefan Rohde; Simon Glaser; Peter K Plinkert
Journal:  Otol Neurotol       Date:  2010-04       Impact factor: 2.311

8.  Effects of hemoglobin concentration, skull thickness, and the area of the cerebrospinal fluid layer on near-infrared spectroscopy measurements.

Authors:  Kenji Yoshitani; Masahiko Kawaguchi; Norikazu Miura; Takashi Okuno; Tomoko Kanoda; Yoshihiko Ohnishi; Masakazu Kuro
Journal:  Anesthesiology       Date:  2007-03       Impact factor: 7.892

9.  Clinical outcomes, angiographic patency, and resource utilization in 200 consecutive off-pump coronary bypass patients.

Authors:  J D Puskas; V H Thourani; J J Marshall; S J Dempsey; M A Steiner; B H Sammons; W M Brown; J P Gott; W S Weintraub; R A Guyton
Journal:  Ann Thorac Surg       Date:  2001-05       Impact factor: 4.330

Review 10.  The evolution of cardiovascular surgery in elderly patient: a review of current options and outcomes.

Authors:  Francesco Nicolini; Andrea Agostinelli; Antonella Vezzani; Tullio Manca; Filippo Benassi; Alberto Molardi; Tiziano Gherli
Journal:  Biomed Res Int       Date:  2014-04-10       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.