Francisco Rincon1,2, Julien Pidoux3, Srini Murali1,2, Jean-Jacques Goy4.
Abstract
BACKGROUND: Our goal was to evaluate the performance of a new wearable arm located pulse oximeter.Entities:
Keywords: Masimo device; Nellcor device; Oximetry; SmartCardia; SpO2 measurement
Mesh:
Substances:
Year: 2022 PMID: 35331157 PMCID: PMC8943352 DOI: 10.1186/s12871-022-01604-w
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.376
Fig. 1SmartCardia device especially dedicated for SpO2 measurement
Mean ± SD SpO2 values at different plateaus with the 3 oximeters for the 12 subjects. Mean differences between the 3 oximeters are shown as delta
| Oxymeter | H | SC | Ma | Ne | Delta H/SC | Delta H/Ma | Delta H/Ne |
|---|---|---|---|---|---|---|---|
| Baseline | 97,83 ± 0,43 | 97,85 ± 0,85 | 98 ± 2,92 | 97,9 ± 1,2 | 0,02 | 0,17 | 0,07 |
| Plateau 1 | 94,81 ± 1,33 | 94,58 ± 1,39 | 96,75 ± 2,38 | 94,83 ± 1,45 | −0,23 | 1,94 | 0,02 |
| Plateau 2 | 91,25 ± 2,48 | 90,78 ± 3,45 | 93,75 ± 3,11 | 91,54 ± 3,02 | −0,47 | 3,5 | 0,29 |
| Plateau 3 | 87,14 ± 3,04 | 87,09 ± 3,32 | 89,29 ± 3,99 | 87,25 ± 3,84 | −0,05 | 2,15 | 0,11 |
| Plateau 4 | 82,96 ± 3,25 | 83,16 ± 3,43 | 86,04 ± 4,62 | 84 ± 4,75 | 0,20 | 3,08 | 1,04 |
| Plateau 5 | 76,90 ± 3,5 | 77,49 ± 3,67 | 79,5 ± 5,95 | 78,5 ± 4,93 | 0,59 | 2,6 | 1,6 |
| Plateau 6 | 72,22 ± 3,42 | 73,03 ± 3,56 | 74,64 ± 5,86 | 72,79 ± 4 | 0,81 | 2,42 | 0,63 |
| Break | 99,28 ± 0,31 | 98,75 ± 0,59 | 98,791,28 | 98,75 ± 1,23 | −0,53 | −0,49 | −0,53 |
| Plateau 7 | 92,73 ± 2,81 | 92,49 ± 2,53 | 95,08 ± 2,84 | 93,15 ± 2,29 | −0,24 | 2,35 | 0,42 |
| Plateau 8 | 87,65 ± 2,28 | 87,81 ± 2,36 | 90,5 ± 3,32 | 88,42 ± 2,02 | 0,16 | 2,85 | -0,23 |
| Plateau 9 | 82,43 ± 2,77 | 82,82 ± 3,2 | 85,29 ± 5,09 | 84,05 ± 2,44 | 0,39 | 2,86 | 1,62 |
| Plateau 10 | 76,26 ± 2,36 | 76,72 ± 2,21 | 78,71 ± 6,1 | 77,95 ± 2,33 | 0,46 | 2,45 | 1,69 |
| Plateau 11 | 71,71 ± 2,59 | 73,06 ± 2,16 | 73,29 ± 7,92 | 73,45 ± 3,86 | 1,35 | 1,58 | 1,84 |
| Mean | 0,42 ± 0,24 | 2,12 ± 1,01 | 0,78 ± 0,58 |
H hemoximeter, Ma the Masimo system, Ne the Nellcor system, SC the SmartCardia system. Delta is the mean difference between SaO2 and the tested system
Fig. 2Plots showing agreement between SaO2 (H) and SpO2 detected by the SmartCardia device (SC), the Masimo device (Ma) and the Nellcor (N). Results +/− SD are presented as the difference between the device and the gold standard
Bias and ARMS at different oximetry ranges with the 3 oximeters compared with SaO2
| Hemoximeter-Range (%) | 60–70 | 70–80 | 80–90 | 90–100 | 70–100 |
|---|---|---|---|---|---|
Mean ± Standard deviation Range: (Minimum, Maximum) | 1.82 ± 0.94 (0.3;3.4) | 0.57 ± 1.81 (−2.9; 4.8) | 0.17 ± 1.23 (−3.6;3.1) | −0.32 ± 1 (−3.3;1.5) | 0.17 ± 1.4 (− 3.6;4.8) |
| Count | 12 | 86 | 91 | 109 | 286 |
| Missing Data | 0 | 0 | 1 | 1 | 2 |
| Root Mean Square | 2.03 | 1.88 | 1.23 | 1.04 | 1.4 |
Mean ± Standard deviation Range: (Minimum, Maximum) | 1.93 ± 3.6 (−7.7;6.4) | 2.27 ± 6.55 (−7.8;18.6) | 2.68 ± 4.19 (−7.9;13.7) | 1.35 ± 2.63 (−5.8;9.5) | 2.02 ± 4.6 (− 7.9;18.6) |
| Count | 12 | 86 | 91 | 109 | 286 |
| Missing Data | 0 | 0 | 0 | 2 | 2 |
| Root Mean Square | 2.96 | 4.9 | 6.9 | 3.95 | 5.00 |
Mean ± Standard deviation Range: (Minimum, Maximum) | 1.51 ± 1.52 (−1.7;3.9) | 1.43 ± 2.94 (−3.9;11.1) | 1.0 ± 2.1 (−2.0;8.0) | 0 ± 1.3 (−4.3;5.5) | 0.76 ± 2.19 (−4.3;11.1) |
| Count | 12 | 86 | 91 | 109 | 286 |
| Missing Data | 0 | 9 | 3 | 2 | 14 |
| Root Mean Square | 1.29 | 2.32 | 3.26 | 2.1 | 2.31 |
SD standard deviation, CI confidence interval
Fig. 3Bland-Altman curve showing agreement between a: SmartCardia, b: Masimo and c: Nellcor, SpO2 and arterial SpO2. Data is plotted as Hemoximeter data (SaO2) vs. pulse oximeter bias (SpO2 – SaO2). A different marker is used for each study subject. Linear regression is shown for all subjects combined, and the equation with R2 is shown on the plot. Mean bias is displayed as a solid horizontal line, and the upper and lower LoA (mean bias ±1.96•SD*) are shown by dashed horizontal lines