3D Convolutional Neural Network-Based Denoising of Low-Count Whole-Body 18F-Fluorodeoxyglucose and 89Zr-Rituximab PET Scans.

Bart M de Vries1, Sandeep S V Golla1, Gerben J C Zwezerijnen1, Otto S Hoekstra1, Yvonne W S Jauw1,2, Marc C Huisman1, Guus A M S van Dongen1, Willemien C Menke-van der Houven van Oordt3, Josée J M Zijlstra-Baalbergen1,2, Liesbet Mesotten4,5, Ronald Boellaard1, Maqsood Yaqub1.   

Abstract

Acquisition time and injected activity of 18F-fluorodeoxyglucose (18F-FDG) PET should ideally be reduced. However, this decreases the signal-to-noise ratio (SNR), which impairs the diagnostic value of these PET scans. In addition, 89Zr-antibody PET is known to have a low SNR. To improve the diagnostic value of these scans, a Convolutional Neural Network (CNN) denoising method is proposed. The aim of this study was therefore to develop CNNs to increase SNR for low-count 18F-FDG and 89Zr-antibody PET. Super-low-count, low-count and full-count 18F-FDG PET scans from 60 primary lung cancer patients and full-count 89Zr-rituximab PET scans from five patients with non-Hodgkin lymphoma were acquired. CNNs were built to capture the features and to denoise the PET scans. Additionally, Gaussian smoothing (GS) and Bilateral filtering (BF) were evaluated. The performance of the denoising approaches was assessed based on the tumour recovery coefficient (TRC), coefficient of variance (COV; level of noise), and a qualitative assessment by two nuclear medicine physicians. The CNNs had a higher TRC and comparable or lower COV to GS and BF and was also the preferred method of the two observers for both 18F-FDG and 89Zr-rituximab PET. The CNNs improved the SNR of low-count 18F-FDG and 89Zr-rituximab PET, with almost similar or better clinical performance than the full-count PET, respectively. Additionally, the CNNs showed better performance than GS and BF.

Entities:  

Keywords:  18F-FDG; 89Zr-antibody; CNN; denoising; low-count

Year:  2022        PMID: 35328149      PMCID: PMC8946936          DOI: 10.3390/diagnostics12030596

Source DB:  PubMed          Journal:  Diagnostics (Basel)        ISSN: 2075-4418


1. Introduction

18F-fluorodeoxyglucose (18F-FDG) positron emission tomography (PET) is essential in staging of a broad spectrum of malignancies [1,2,3]. Currently, a whole-body 18F-FDG PET scan is acquired using a scan duration of 2 min per bed position and an injected activity of 3.7 MBq/kg. A shorter scan duration per bed position could ideally decrease the total scan duration, and therefore, minimize movement artefacts and increase patient comfort and throughput. A reduction of injected activity would decrease the radiation burden for the patient, and therefore makes it possible to perform more frequent 18F-FDG PET scans per patient for restaging and therapy-response assessments, in case of scanning children and/or for non-oncological cases. However, a shorter scan duration and lower injected activity would result in a lower signal-to-noise ratio (SNR). Poor scan quality due to low-count (LC) is also observed for 89Zr-antibody PET scans, which are obtained after relatively low injected activity imposed by the radiation burden of 89Zr [4]. Therefore, denoising LC whole-body 18F-FDG and 89Zr-antibody PET scans is of interest for improving image quality. Traditionally, Gaussian smoothing (GS) has been used to denoise PET images [5]. However, GS reduces the spatial resolution of the images, and therefore, could impair detectability and quantification of small (tumour) lesions [6]. Bilateral filtering (BF) exhibits superior properties in comparison to the more commonly used GS for noise reduction in PET [7]. BF reduces the noise of PET scans, while preserving spatial information (e.g., edges). However, BF parameters are difficult to optimize in a generic way because both an optimized intensity and spatial parameter need to be determined, which depend on both the tracer and site of interest. Therefore, another adaptive/data-driven denoising method with high accuracy is warranted. Convolutional Neural Networks (CNN) are a specialized type of Neural Networks that use convolution to extract features from the PET scan. This is done by convolution filters which assign importance/weights to (learnable) features present in the PET scan. It can therefore learn and detect features such as PET intensities, edges, shapes, etc. Therefore, CNNs are highly beneficial in various medical image processing/segmentation tasks [8,9,10,11]. A CNN-based deep-learning algorithm may also be superior in denoising tasks since it can learn non-linear latent/hidden (not observable by humans) features (which you want to preserve) from LC PET scans and increase the SNR [12,13,14,15,16,17,18]. Therefore, denoising LC whole-body 18F-FDG and 89Zr-antibody PET scans using a CNN may be performed to improve the SNR and thus their diagnostic value. In previous studies [12,13,14,17,19] a successful application of CNNs for improving 18F-FDG PET scans has been presented. However, these studies were performed in a small (oncology) patient cohort, were based on unsupervised deep learning networks and on improving full-count (FC) 18F-FDG PET scans or a longer scan duration per bed position. Therefore, the aim of this study was to develop, train, and extensively evaluate the performance of CNNs to denoise LC whole-body 18F-FDG and 89Zr-antibody PET scans. A secondary aim was to compare the diagnostic value of the CNNs to that of GS and BF.

2. Materials and Methods

2.1. Participants

We included PET scans of 60 patients with stage I–IV non-small-cell lung carcinoma (NSCLC) (40 patients from Limburg PET-Center Hasselt Belgium (LPC) [20], and 20 patients from Amsterdam UMC, location VUmc), of which five patients with diffuse large B cell lymphoma (DLBCL) non-Hodgkin lymphoma (Amsterdam UMC, location VUmc) [21] (Table 1). The study at LPC was registered at clinical trials.gov, NCT02024113. The data from the patients with lung cancer at Amsterdam UMC were retrospectively obtained from medical records, with a waiver for informed consent from the Medical Ethics Review Committee of Amsterdam UMC, location VUmc. This study was registered as IRB2018.029. The patients with non-Hodgkin lymphoma were included as part of studies performed by Jauw et al. These patients provided written informed consent, and the studies were approved by the Medical Ethics Review Committee of Amsterdam UMC, location VUmc. This study was registered at Dutch Trial Register http://www.trialregister.nl (accessed on 19 January 2022), NTR3392.
Table 1

The patients characteristics included in this study from Amsterdam UMC.

18F-FDG (n = 20) NSCLC–Amsterdam UMC18F-FDG (n = 40) NSCLC–LPC89Zr-Rituximab (n = 5)Non-Hodgkin Lymphoma
Male/female (n)11/925/153/2
Injected dose (MBq)259.4 ± 43.8298.2 ± 49.473.7 ± 0.3
Tumour volume (cubic centimeter (cc))–Test data29.2 ± 54.3-25.6 ± 51.3

2.2. Data Acquisition

Whole-body 18F-FDG PET scans in LPC were acquired with a Gemini Big Bore TF PET/CT scanner, and in Amsterdam UMC with an Ingenuity TF PET/CT and Vereos Digital PET/CT scanners (Philips Medical Systems, Best, The Netherlands). 89Zr-rituximab PET scans (patients with non-Hodgkin lymphoma) were acquired with an Ingenuity TF PET/CT. For 18F-FDG PET scans, 60 min after 259.4 ± 43.8 MBq tracer injection, a low-dose computed tomography (LDCT) scan was performed for attenuation correction and anatomical localisation, and subsequently a 20 min static (exact time depends on patient length) whole-body 18F-FDG PET scan was acquired (2 min per bed position). Six days after the injection of 73.7 ± 0.3 MBq 89Zr-rituximab, an LDCT scan was obtained, directly followed by a 60 min static whole-body PET scan (5 min per bed position). Corrections for decay, dead time, normalization (detector sensitivities), attenuation, random coincidences and scatter were applied. Amsterdam UMC 18F-FDG PET data were reconstructed with a 10 s (super-low-count (SLC), 92% scan time reduction), 30 s (low-count (LC), 75% scan time reduction) and 2 min (full-count (FC)) scan duration per bed-position. The (S)LC PET scans were reconstructed using multiple time points/delays, which was later used for data augmentation during training. These scans were reconstructed using the blob-basis function ordered-subsets time of flight (BLOB-OS-TF) for the Ingenuity TF PET/CT scanner and the novel ordered subset expectation maximization (OSEM 3i15s, 1i6r-PSF, 4 mm FWHM GAUSS, OSEM 3i15s, 3 mm FWHM GAUSS) for the Vereos Digital PET/CT scanner. The 89Zr-rituximab scans, and the 18F-FDG PET data from LPC were reconstructed with a FC 5 min and 2 min scan duration per bed position only using BLOB-OS-TF, respectively. The 18F-FDG and 89Zr-rituximab PET scans from Amsterdam UMC were reconstructed according to current European Association of Nuclear Medicine Research Ltd. (Vienna, Austria) EARL1 standards and settings associated with EARL accreditation [22], respectively. Matrix and voxels sizes were 144 × 144 and 4 mm in all directions, respectively. 18F-FDG PET scans from LPC were reconstructed according to EARL1 standards, with matrix and voxel sizes of 169 × 169 and 4 mm, respectively.

2.3. Image Processing

For each FC whole-body 18F-FDG PET scan from LPC, SLC PET, scans were simulated using the SiMulAtion and ReconsTruction (SMART)-PET package [23]. Simulation-reconstruction settings were chosen so that the simulated noisy 18F-FDG PET scans from LPC showed an almost similar coefficient of variation as the SLC-reconstructed 18F-FDG PET scans from Amsterdam UMC. The simulated PET images were used to initially train the CNN, while parts of the actual reconstructed images were used for further fine-training of the CNN, details are explained later.

2.4. Model Architecture

A supervised U-Net based [11] 3D-CNN (Figure 1 and Appendix B) was used to denoise the (S)LC 18F-FDG PET scans while maintaining their diagnostic value. However, instead of the max-pooling layer that is traditionally used, in this study the down sampling layers consisted of convolution layers with a stride of two [24]. Although the convolution layer compresses the feature image just as is the case for a max-pooling layer, it does not exclude voxels by only looking at the maximum values. It therefore, not only reduces computation time (although less than max-pooling), but most importantly increases the model its ability to learn [24]. Additionally, in contrast to conventional CNNs, a kernel size of 6 × 6 × 6 instead of 3 × 3 × 3 was applied to learn inter-slice morphological features [25].
Figure 1

Architecture of the U-net shaped 3D-CNN used in the study. It consists of an encoding and decoding path, which are connected with concatenation layers at each resolution block.

2.5. Model Performance

2.5.1. Quantitative Performance

The simulated SLC whole-body 18F-FDG PET scans from LPC were used to pre-train a 3D-CNN. Next, the reconstructed SLC and LC 18F-FDG PET data from Amsterdam UMC were used for fine-training (transfer-learning) the pre-trained model, which generated two additional models (SLC-CNN and LC-CNN) that are tailored to manage low or super low count/quality images. These two models were subsequently used for further evaluation. Training of the CNN model on the simulated noisy LPC 18F-FDG data was performed to avoid overfitting due to the small dataset. Noise characteristics of 89Zr-rituximab and LC 18F-FDG PET scans were almost similar. However, we used the SLC-CNN to denoise the 89Zr-rituximab PET scans instead of the LC-CNN, because of the higher level of noise reduction. The 18F-FDG PET data from LPC was split into a training (80%, n = 32) and a validation (20%, n = 8) set. Thereafter, for further refinement, validation and testing, the two CNN models, 18F-FDG and 89Zr-rituximab data from Amsterdam UMC, were used. During this training, the data were split into a training (32%, n = 8), a validation (8%, n = 2) and an independent test (60%, n = 15) set. The training and validation set from Amsterdam UMC consisted of only 18F-FDG PET scans from the Ingenuity TF PET/CT scanner. The test set, however, consisted both of 18F-FDG PET scans from the Ingenuity TF PET/CT scanner, the Vereos Digital PET/CT scanner, and 89Zr-rituximab PET scans from the Ingenuity TF PET/CT scanner. PET data augmentation was applied during each training epoch (train-data only) by randomly sampling the different (time points/delays) (S)LC 18F-FDG PET scans for each patient during training. In other words, instead of traditional augmentation (shifts, zoom, translation, rotation, etc.), in each training epoch, minor differences in noise characteristics were present. To compare the performance of the CNNs with other denoising methods, the (S)LC test PET scans were also denoised using traditional GS (18F-FDG) and more advanced BF [17] (18F-FDG and 89Zr-rituximab) denoising methods (Table 2). A Mann–Whitney U test (p < 0.05) was used to compare tumour recovery coefficients (TRC) and levels of noise in the images after applying the denoising methods.
Table 2

Gaussian smoothing (GS) and Bilateral Filtering (BF) settings evaluated for the denoising of the low-count whole-body PET scans.

GS (FWHM)BF (FWHM; SUV)
18F-FDG PET
SLC8 mm, 10 mm and 12 mm4 mm and 5 mm; SUV2.5
LC4 mm, 6 mm and 8 mm3 mm and 4 mm; SUV2.5
89Zr-rituximab PET6 mm, 8 mm and 10 mm2 mm, 3 mm and 4 mm; SUV2.5
We calculated TRC for the 18F-FDG and 89Zr-rituximab PET scans (test data) using Equation (1). TRC was computed for the test data post-processed with a 3D-CNN, GS, or BF denoising method and compared this to the FC data. PET uptake features from the tumour volumes were extracted (U, X = average, maximum and 3Dpeak) for both the denoised as the FC  PET scans, using the in-house built and open-access ACCURATE tool (quAntitative onCology moleCUlaR Analyses SuiTE) [26,27]. From the 18F-FDG scans, only the primary lung tumour was extracted using a 50% SUV3Dpeak isocontour (Table 1 and Figure A1). For the 89Zr-rituximab PET scans, tumours were extracted using manual delineation (Table 1 and Figure A1). For patients with non-Hodgkin lymphoma with more than three tumours, bootstrapping was applied to randomly choose three tumours for analysis.
Figure A1

Volume (cc) and SUVBW distribution of the tumours in the 18F-FDG PET and the 89Zr-rituximab PET scans (test data).

The level of noise was presented as the coefficient of variance (COV; Equation (2)). Four spherical volume of interest (VOIs) were drawn in the liver (because the liver showed homogeneous tracer uptake in this cohort, and therefore, could be used to reliably assess the level of noise). Average standard deviation and average uptake were extracted using these four VOIs.

2.5.2. Qualitative Performance

For a qualitative assessment of the denoising methods (CNN and BF), the images after denoising were independently evaluated by two experienced nuclear medicine physicians (BZ and OH). GS was not included in this assessment due to a mostly significant (p < 0.05) lower quantitative performance in comparison to the CNNs and BF. The questionnaire was drafted to assess the reliability and effectiveness of the denoising methods. The assessment was blinded, i.e., the scans presented to the physicians were a random combination (without labels) of the FC, SLC, LC (with and without denoising) PET scans per patient. The 18F-FDG and 89Zr-rituximab PET scans were scored per patient (1–5: low to high) based on the level of noise, tumour detectability, overall scan quality, clinical acceptability (yes/no), and overall best performance (1st/2nd/3rd/4th/(5th)). A Mann–Whitney U test (p < 0.05) was used to compare the performance of the denoising methods.

3. Results

3.1. Quantitative Assessment

3.1.1. 18F-FDG

The BF and the proposed CNN (SLC- and LC-CNN) denoised PET scans have an overall higher TRC and more similar COV to the FC PET scans than GS (Figure 2 and Table A1). In contrast with BF, the SLC-CNN denoised PET scans showed a higher average uptake TRC, a higher 3Dpeak uptake TRC, but a lower maximum uptake TRC. With regard to the LC scans, the LC-CNN denoised PET scans showed a trend (0.05 < p < 0.1) of a higher TRC than the BF denoised PET scans for the average uptake, and 3Dpeak uptake. Additionally, the LC-CNN showed a higher but not significant maximum uptake TRC than the BF. In addition, the LC-CNN denoised PET scans showed a similar COV as the FC PET scans. The SLC-CNN had the second closest COV to the FC PET scans.
Figure 2

The performance of the denoising methods for low-count 18F-FDG PET. The (A) average, (B) maximum, (C) 3Dpeak TRC of the SLC-CNN, GS (8 mm, 10 mm and 12 mm) and BF (4 mm and 5 mm) denoising methods of the SLC 18F-FDG PET from the Ingenuity TF PET/CT and the Vereos Digital PET/CT scanner. The (D) average, (E) maximum, (F) 3Dpeak TRC of the LC-CNN, GS (4 mm, 6 mm and 8 mm) and BF (3 mm and 4 mm) denoising methods of the LC 18F-FDG PET from the Ingenuity TF PET/CT and the Vereos Digital PET/CT scanner.

Table A1

Overview of the performance of the GS, the BF and the proposed CNNs denoising methods using the external test 18F-FDG PET scans from the Ingenuity TF PET/CT and Vereos Digital PET/CT scanners. For the FC (green), SLC and LC (with (blue) and without (orange) post-processing) PET scans, the TRC and COV values are shown in each column.

TRC–AverageTRC–MaximumTRC–3DPeakCOV
FC 0.09
SLC0.991.111.030.29
GS (12; 10; 8 mm)[0.69 *; 0.74 *; 0.81 **][0.59 *; 0.66 *; 0.74][0.68 *; 0.74 *; 0.80][0.09; 0.11; 0.14]
BF (5; 4 mm)[0.79; 0.82][0.85; 0.87][0.82; 0.84][0.01; 0.11]
SLC-CNN0.860.850.870.10
LC0.981.020.990.17
GS (8; 6; 4 mm)[0.80 *; 0.85 **; 0.90][0.71 *; 0.79 *; 0.86 *][0.79 *; 0.84 *; 0.90 *][0.09; 0.11; 0.13]
BF (4; 3 mm)[0.81 **; 0.88][0.78; 0.87][0.82 **; 0.89][0.07; 0.11]
LC-CNN0.950.940.960.10

* significant (p < 0.05) higher/lower than (S)LC-CNN. ** trend (0.05 < p < 0.1).

3.1.2. 89Zr-Rituximab

The SLC-CNN denoised PET scans showed a predominant trend of a TRC higher than the 3 mm and 4 mm BF (Figure 3 and Table A2). The SLC-CNN even showed a significantly (p < 0.05) higher average uptake TRC than the 3 mm and 4 mm BF. The 3 mm and 4 mm BF presented a comparable COV as the SLC-CNN. The 2 mm BF had a significantly (p < 0.05) higher COV than the 3 mm and 4 mm BF and the SLC-CNN.
Figure 3

The performance of the denoising methods for low-count 89Zr-rituximab PET. The (A) average, (B) maximum, (C) 3Dpeak TRC of the SLC-CNN and BF (2 mm, 3 mm and 4 mm) denoising methods of the FC 89Zr-rituximab PET from the Ingenuity TF PET/CT scanner.

Table A2

Overview of the performance of the BF and the SLC-CNN on the external test 89Zr-rituximab PET scans from the Ingenuity TF PET/CT scanner. For the FC (orange), and post-processed (blue) PET scans, the TRC and COV values are shown in each column.

TRC–AverageTRC–MaximumTRC–3DpeakCOV
FC 0.14
BF (4; 3; 2 mm)[0.88 *; 0.94 *; 0.98][0.78 **; 0.86; 0.94 *][0.84 **; 0.92 **; 0.97 *][0.07; 0.10; 0.13]
SLC-CNN0.960.880.940.10

* significant (p < 0.05) higher/lower than SLC-CNN. ** trend (0.05 < p < 0.1).

3.2. Qualitative Assessment

For the 18F-FDG scans, the observers found lower levels of noise, better tumour detectability, better overall scan quality and higher clinical acceptability for all the CNN models in comparison to the BF denoising methods (Figure 4 and Figure A2, Table 3), with the only exception being SLC-CNN in terms of noise levels and tumour detectability.
Figure 4

Illustration of a (A) FC, (B) SLC, (C) SLC-CNN, (D) BF 4 mm, (E) GS 10 mm denoised 18F-FDG PET scan (axial orientation) from the Ingenuity TF PET/CT scanner. Illustration of a (F) LC, (G) LC-CNN, (H) BF 4 mm, (I) GS 6 mm denoised 18F-FDG PET scan (axial orientation) from the Ingenuity TF PET/CT scanner.

Figure A2

Illustration of a (A) FC, (B) SLC, (C) SLC-CNN, (D) BF 4 mm, (E) BF 5 mm, (F) GS 10 mm denoised 18F-FDG PET scan (coronal orientation) from the Ingenuity TF PET/CT scanner. Illustration of a (G) LC, (H) LC-CNN, (I) BF 3 mm, (J) BF 4 mm, (K) GS 6 mm denoised 18F-FDG PET scan (coronal orientation) from the Ingenuity TF PET/CT scanner.

Table 3

Scores provided by the Nuclear Medicine Physicians as part of qualitative assessment of the denoised 18F-FDG PET scans. The best performing method is indicated in bold based on the average score of both physicians.

SLCLCFC
Metrics[1–5: Low–High]SLC-CNNBF-4 mmBF-5 mmLC-CNNBF-3 mmBF-4 mm
Level of noise 3.0–3.2 4.0–4.0 *3.6–4.0 * 1.8–2.0 2.6–2.0 **2.2–3.8 *1–1 *
Tumour detectability2.0–3.0 2.2–3.0 2.0–3.0 4.0–4.0 3.6–4.02.8–3.0 *5–5 *
Overall scan quality 2.4–2.6 1.6–2.0 *1.8–2.0 * 4.4–4.0 3.8–4.02.8–2.2 *5–5 *
Clinically acceptable? [%] 0–80 0–0 *0–0 * 100–100 80–10020–0 *100–100
Best scan (1/2/3/4) 2–2 3–34–4 2–3 3–2 4–41–1

* significant (p < 0.05) higher/lower than (S)LC-CNN. ** trend (p < 0.1).

For the 89Zr-rituximab scans, the observers found a comparable level of noise, but similar/better tumour detectability, better overall scan quality and higher clinical acceptability for the SLC-CNN in comparison to the BF denoising methods (Figure 5 and Table 4).
Figure 5

Illustration of a (A) FC, (B) SLC-CNN, (C) BF 3 mm, (D) BF 4 mm, (E) BF 5 mm denoised 89Zr-rituximab PET scan (coronal orientation) from the Ingenuity TF PET/CT scanner.

Table 4

Qualitative assessment of the 89Zr-rituximab PET scans. Scores were given for the PET scans with (SLC-CNN and BF) and without (FC) denoising by both Nuclear Medicine Physicians. In bold the best performing method (or scan) is indicated based on the average score of both physicians.

Metrics[1–5: Low–High]SLC-CNNBF-2 mmBF-3 mmBF-4 mmFC
Level of noise2.4–2.63.8–4.6 *2.8–2.6 1.4–1.2 * 4.6–4.8 *
Tumour detectability3.4–3.84.4–4.0 *2.4–2.4 *1.4–1.4 * 4.6–4.2 *
Overall scan quality 3.8–3.8 3.6–3.83.4–3.0 *2.0–1.4 *3.4–3.0 *
Clinical acceptable? [%] 100–100 100–100 80–80 0–0 * 80–80
Best scan (1/2/3/4/5) 1–1 3–22–44–53–3

* significant (p < 0.05) higher/lower than SLC-CNN.

4. Discussion

CNN models to denoise (S)LC 18F-FDG and 89Zr-rituximab PET scans were trained and extensively evaluated. Overall, the CNN models performed better than the conventional GS and the more advanced BF denoising methods for both the 18F-FDG and 89Zr-rituximab PET scans. As such, the CNN models show promise for reducing the acquisition time and injected activity of 18F-FDG PET scans and increasing the image quality of 89Zr-rituximab PET scans. In this study, we trained noise-specific CNN models, to address the difference in noise levels seen for different scan acquisition times and injected activity in 18F-FDG PET scans. However, in case of PET tracers such as 89Zr-antibody PET, training a noise specific CNN model was not feasible. Due to the dose limits of 89Zr, the overall image quality was impaired (low SNR), and therefore, no high quality 89Zr-antibody PET images were available for training a CNN. So, the only possible solution was to directly apply the SLC-CNN (trained using SLC 18F-FDG PET scans) to the 89Zr-rituximab PET scans and test its performance. The main advantage of this approach is that this validation is the ultimate way of externally testing the CNN on data that are obtained with a different tracer. Although the SLC-CNN is not trained on 89Zr-rituximab PET scans, it obtained a higher TRC than the 3 mm and 4 mm BF denoising methods (Figure 3 and Table A2). With regard to 18F-FDG PET scans with a low injected tracer activity, such as scans with shorter scan duration, a lower SNR will be observed, which impairs both the quantitative and qualitative value of these scans. The CNNs could therefore also be useful to maintain a good image quality when reducing injected 18F-FDG activity in whole-body 18F-FDG PET studies, and therefore, reduce radiation burden for the patient, but maintain diagnostic value. However, further assessment is necessary to evaluate the performance of CNNs when used for a reduction in the injected activity for whole-body 18F-FDG PET acquisitions. The qualitative assessment also showed that the proposed CNNs were preferred over BF. However, the CNN denoised (S)LC 18F-FDG PET scans did show an overall lower qualitative performance than the FC 18F-FDG PET scans. Yet, the LC-CNN denoised LC 18F-FDG PET scans obtained a similar clinical acceptability score as the FC 18F-FDG PET scans (Table 3), while for 89Zr-rituximab PET scans, the SLC-CNN increased the overall image quality of the FC 89Zr-rituximab PET scans (Table 4). The observers preferred the SLC-CNN denoised 89Zr-rituximab PET scans over both the BF denoised and FC 89Zr-rituximab PET scans. This can be explained by a higher ratio between tumour signal and background signal present in the SLC-CNN denoised 89Zr-rituximab PET scans (Figure 5). This indicates that the SLC-CNN shows promise for establishing an optimal denoising setup for 89Zr-antibody PET scans. In this study, several strategies to prevent overfitting were applied. First, data augmentation was applied by randomly sampling the different (S)LC 18F-FDG PET scans for each patient. By using traditional augmentation, interpolation may be different between the training data ((S)LC) and training labels (FC), and therefore, this was not applied in this study. Another method by which overfitting was reduced is by using the symmetric connections in the U-Net based 3D-CNN [28]. As shown in previous studies [12,13], training a model using a small dataset could result in overfitting. Since acquiring sufficient real (S)LC 18F-FDG PET data were not feasible, SLC 18F-FDG PET data were generated using the already available LPC data. SLC 18F-FDG PET data from LPC were simulated using SMART, which facilitated the development of a pre-trained model familiar with morphological features. This resulted in a shorter learning time, lower probability of overfitting, and a more accurate and robust model. Even though pre-training of the model was only performed on SLC 18F-FDG PET data from LPC, the fine-trained LC-CNN showed a higher performance than a LC-CNN without a pre-trained model. The main limitation of this study is the size of the patient cohort. Small-sized tumours in the (S)LC PET scans are more prone to being underestimated by the proposed CNNs. This is because the training data were devoid of small tumours. Therefore, it could be that the model specifies this signal as noise rather than a tumour-specific signal [29]. However, the proposed CNNs showed better correspondence with the FC PET scans than GS and BF. So, although small tumours were present in a small number in the training data, by using the proposed CNNs, more quantitative information was retained in comparison to GS and BF. Even though the differences in performance between the proposed CNNs and BF were small, contrary to BF, a CNN still has the ability to learn and improve by incorporating more patients. Thus, further evaluation in a larger and more heterogeneous cohort could further improve CNNs performances. However, although the proposed method showed promising results for denoising low-count 18F-FDG PET scans, obtaining a similar quantitative and qualitative value as the FC 18F-FDG PET scans may not be fully feasible and we therefore foresee that the main applications of the CNNs are denoising and improving image quality of 89Zr-antibody PET studies.

5. Conclusions

The 3D-CNNs used in this study to denoise (S)LC whole body 18F-FDG and 89Zr-rituximab PET scans were constructed and tested. The CNN denoised (S)LC 18F-FDG and 89Zr-rituximab PET scans showed almost similar or better clinical performance than the FC scans, respectively. Therefore, the proposed CNNs show promise for reducing PET scan duration or lowering injected activity of whole-body 18F-FDG PET scans but are particularly useful to increase the quantitative and qualitative image quality of 89Zr-rituximab PET scans.
  20 in total

Review 1.  Partial-volume effect in PET tumor imaging.

Authors:  Marine Soret; Stephen L Bacharach; Irène Buvat
Journal:  J Nucl Med       Date:  2007-05-15       Impact factor: 10.057

2.  Full-count PET recovery from low-count image using a dilated convolutional neural network.

Authors:  Karl Spuhler; Mario Serrano-Sosa; Renee Cattell; Christine DeLorenzo; Chuan Huang
Journal:  Med Phys       Date:  2020-08-06       Impact factor: 4.071

3.  An investigation of quantitative accuracy for deep learning based denoising in oncological PET.

Authors:  Wenzhuo Lu; John A Onofrey; Yihuan Lu; Luyao Shi; Tianyu Ma; Yaqiang Liu; Chi Liu
Journal:  Phys Med Biol       Date:  2019-08-21       Impact factor: 3.609

Review 4.  18F-FDG PET/CT imaging in oncology.

Authors:  Ahmad Almuhaideb; Nikolaos Papathanasiou; Jamshed Bomanji
Journal:  Ann Saudi Med       Date:  2011 Jan-Feb       Impact factor: 1.526

5.  Suitability of bilateral filtering for edge-preserving noise reduction in PET.

Authors:  Frank Hofheinz; Jens Langner; Bettina Beuthien-Baumann; Liane Oehme; Jörg Steinbach; Jörg Kotzerke; Jörg van den Hoff
Journal:  EJNMMI Res       Date:  2011-10-05       Impact factor: 3.138

6.  Performance of 89Zr-Labeled-Rituximab-PET as an Imaging Biomarker to Assess CD20 Targeting: A Pilot Study in Patients with Relapsed/Refractory Diffuse Large B Cell Lymphoma.

Authors:  Yvonne W S Jauw; Josée M Zijlstra; Daphne de Jong; Danielle J Vugts; Sonja Zweegman; Otto S Hoekstra; Guus A M S van Dongen; Marc C Huisman
Journal:  PLoS One       Date:  2017-01-06       Impact factor: 3.240

7.  Classification of Alzheimer's Disease by Combination of Convolutional and Recurrent Neural Networks Using FDG-PET Images.

Authors:  Manhua Liu; Danni Cheng; Weiwu Yan
Journal:  Front Neuroinform       Date:  2018-06-19       Impact factor: 4.081

8.  Plausibility and redundancy analysis to select FDG-PET textural features in non-small cell lung cancer.

Authors:  Elisabeth Pfaehler; Liesbet Mesotten; Ivan Zhovannik; Simone Pieplenbosch; Michiel Thomeer; Karolien Vanhove; Peter Adriaensens; Ronald Boellaard
Journal:  Med Phys       Date:  2021-02-06       Impact factor: 4.071

9.  SMART (SiMulAtion and ReconsTruction) PET: an efficient PET simulation-reconstruction tool.

Authors:  Elisabeth Pfaehler; Johan R De Jong; Rudi A J O Dierckx; Floris H P van Velden; Ronald Boellaard
Journal:  EJNMMI Phys       Date:  2018-09-18

10.  Classification of negative and positive 18F-florbetapir brain PET studies in subjective cognitive decline patients using a convolutional neural network.

Authors:  Bart Marius de Vries; Sandeep S V Golla; Jarith Ebenau; Sander C J Verfaillie; Tessa Timmers; Fiona Heeman; Matthijs C F Cysouw; Bart N M van Berckel; Wiesje M van der Flier; Maqsood Yaqub; Ronald Boellaard
Journal:  Eur J Nucl Med Mol Imaging       Date:  2020-09-02       Impact factor: 9.236