Jon-Émile S Kenny1, Igor Barjaktarevic2, Andrew M Eibl1, Matthew Parrotta1, Bradley F Long1,3, Mai Elfarnawany1, Joseph K Eibl1,3.
Abstract
PURPOSE: We describe the temporal concordance of 3 hemodynamic monitors.Entities:
Mesh:
Year: 2022 PMID: 35320307 PMCID: PMC8942202 DOI: 10.1371/journal.pone.0265711
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1A and B: The wireless Doppler ultrasound patch.
Fig 2Representative protocol.
Orange dots represent single beat VTI from the wearable carotid Doppler while the orange line depicts 24-second averages of the VTI. Green line is the 24-second average of SV from bioreactance. The blue line is a 20-second SV from non-invasive pulse contour. Y-axis is % change for each metric and x-axis is time.
Diagnostic accuracy summary table.
| AUROC | Threshold | sensitivity | specificity | |
|---|---|---|---|---|
|
| ||||
|
| 0.98 | 10.5% | 100% | 100% |
|
| 0.98 | 2.5% | 100% | 100% |
|
| 0.75 | 13% | 60% | 95% |
|
| 0.98 | 9% | 100% | 100% |
|
| ||||
|
| 0.98 | 5.5% | 100% | 95% |
|
| 0.95 | 3% | 100% | 100% |
|
| 0.77 | 15% | 55% | 100% |
|
| 0.98 | 9% | 100% | 95% |
|
| ||||
|
| 0.97 | 3% | 95% | 95% |
|
| 0.98 | 4% | 100% | 100% |
|
| 0.85 | 7% | 70% | 90% |
|
| 0.98 | 8% | 100% | 100% |
Fig 3Dot-plots for each metric across all 3 temporal windows (early, middle, late).
Horizontal, blue line is baseline for each measurement. The blue arrows represent directional change in preload. Y-axis is fractional change (% Δ) for each metric relative to baseline. A is velocity time integral (VTI) from the wearable Doppler, B. is corrected flow time (FTc) from the wearable Doppler, C. is stroke volume from bioreactance (SVBioreactance) and D. is stroke volume from non-invasive pulse contour (SVPulse Contour). The small, horizontal red arrows represent optimal diagnostic threshold for each metric. See Table 1 for threshold values.