| Literature DB >> 35316846 |
Nidhal Kahlaoui1, Thibaut Naninck1, Roger Le Grand1, Catherine Chapon2.
Abstract
INTRODUCTION: A Vereos PET/CT device was adapted to be compatible with the experimentation in large animals within BSL-3 environment. The aim of this study was to investigate the impact of this modification on the performance according to NEMA NU2-2012 standard.Entities:
Keywords: BSL-3 conditions; Infectious diseases; NEMA NU2-2012 standard; PET/CT
Year: 2022 PMID: 35316846 PMCID: PMC8938744 DOI: 10.1186/s40658-022-00450-6
Source DB: PubMed Journal: EJNMMI Phys ISSN: 2197-7364
Fig. 1PET/CT (Vereos, Philips Healthcare) imaging facilities allowing for imaging to be performed in BSL-2 and BSL-3 environment by separating the BSL-1 side (in blue) with the PET/CT device, from the BSL-3 (in red) with the Non-Human Primate (NHP) preparation room. A transparent poly-methyl methacrylate (PMMA) containment tube (2800 mm of length, 680 mm of diameter and 8 mm thickness) extends the partition between BSL-1 and BSL-3 sides into the PET/CT gantry, also extending the biological barrier into imaging system
CT acceptance test without and with tube samples of 10 and 8 mm thickness
| Parameter | Without tube | With 10 mm tube | With 8 mm tube | Acceptable range |
|---|---|---|---|---|
| CT acceptance test | ||||
| Head noise | 4.96 ± 0.09 | 6.15 ± 0.13 | 5.61 ± 0.08 | [4.4,6] |
| Body noise | 13.15 ± 0.29 | 16.24 ± 0.27 | 14.91 ± 0.38 | [11.5,15.5] |
Performance parameters of the PET component of the PET/CT Vereos with or without an 8 mm PMMA biocontainment tube within the bore defined by NEMA NU2-2012 measurements including energy and timing resolution, spatial resolution, sensitivity, count rates and scatter fractions
| Parameter | Without tube | With 8 mm tube | Manufacturer guidelines |
|---|---|---|---|
| Energy resolution (%) | 11 | 11.03 | ≤ 13 |
| Timing resolution (ps) | 311.82 | 312 | ≤ 390 |
| Spatial resolution (FWHM) | |||
| Transverse, 1 cm | 4.03 ± 0.11 | 3.96 ± 0.14 | < 4.9 |
| Transverse, radial, 10 cm | 4.59 ± 0.09 | 4.62 ± 0.09 | < 5.5 |
| Transverse, radial, 20 cm | 5.77 ± 0.04 | 5.8 ± 0.04 | < 7.2 |
| Transverse, tangential, 10 cm | 4.33 ± 0.02 | 4.34 ± 0.03 | < 5.5 |
| Transverse, tangential, 20 cm | 4.94 ± 0.05 | 4.89 ± 0.05 | < 7.2 |
| Axial, 1 cm | 4.06 ± 0.46 | 4.29 ± 0 | < 4.9 |
| Axial, radial, 10 cm | 4.29 ± 0.12 | 4.43 ± 0.10 | < 5.5 |
| Axial, radial, 20 cm | 4.64 ± 0.07 | 4.7 ± 0.13 | < 7.2 |
| Sensitivity (cps/GBq) | |||
| Centered | 6.22 | 5.21 | ≥ 5.1 |
| 10 cm Off-centered | 6.17 | 5.34 | ≥ 5.1 |
| Count rates and scatter fractions | |||
| Peak true count rate (kcps) | 660 @51.7 kBq/mL | 595.59 @51.3 kBq/mL | > 675.000 |
| Peak NECR rate (kcps) | 161,8 @51.75 kBq/mL | 173.63 @51.35 kBq/mL | > 148.000 |
| Location of NECR peak (kBq/mL) | 51.75 | 51.35 | > 25.000 |
| Scatter fraction (%) at low count | 31 @0.14 kBq/mL | 31,2 @0.14 kBq/mL | |
| Scatter fraction (%) at NECR peak | 33.12 | 33 | ≤ 35.00 |
| Max relative count rate error (%) | 6.01 | 5.06 | < 12.00 |
| Image quality (%) | |||
| 10 mm 4:1 | 59.9 | 82.9 | > 26 |
| 13 mm 4:1 | 75.4 | 78.0 | > 40 |
| 17 mm 4:1 | 96.6 | 83.8 | > 50 |
| 22 mm 4:1 | 90.3 | 91.4 | > 60 |
| 28 mm cold | 84.1 | 79.5 | > 68 |
| 37 mm cold | 85.4 | 87.1 | > 70 |
| Variability (%) | |||
| 10 mm 4:1 | 9.5 | 8.8 | < 11 |
| 13 mm 4:1 | 7.7 | 7.3 | < 9 |
| 17 mm 4:1 | 6.1 | 5.9 | < 8 |
| 22 mm 4:1 | 4.6 | 4.5 | < 8 |
| 28 mm cold | 3.3 | 3.5 | < 7 |
| 37 mm cold | 2.2 | 2.8 | < 7 |
| Lung residual | 6.5 | 5.8 | ≤ 15 |
Quality image was also evaluated with the percentage contrast for spheres with different diameters, the percentage background variability, and percentage relative lung error for the images. Measured values were compared to recommended values provided by the manufacturer in the Installation Acceptance Test Document
Fig. 2Axial sensitivity profiles for measurements with line source in center of FOV and at 10-cm radial offset with or without 8 mm thickness PMMA tube
Fig. 3Count rate performance, scatter fraction and count rate accuracy. A Scatter Count Rate, B Random Count Rate, C True Count Rate, D NEC rate, E Scatter fraction, F Relative Count Rate Error: Mean relative count rate errors for the different activity distributions. Mean values are shown as means (full lines) ± maximum/minimum values (dashed lines) with (red) and without (green) the PMMA tube
Fig. 4Central slice of the image quality phantom A without and B with the PMMA tube