Klara Valko1, Silvia Rava2, Shenaz Bunally3, Scott Anderson4. 1. Bio-Mimetic Chromatography Ltd. Business and Technology Centre, Bessemer Drive, Stevenage, SG1 2DX, United Kingdom. 2. University of Pavia, Italy, Erasmus internship at Bio-Mimetic Chromatography Ltd. Business and Technology Centre, Bessemer Drive, Stevenage, SG1 2DX, United Kingdom. 3. Physicochemical Group, GSK, Gunnels Wood Road, Stevenage, SG1 2NY United Kingdom. 4. Regis Technologies Inc. 8210 Austin Avenue Morton Grove, IL US 60053.
Abstract
Immobilized Artificial Membrane (IAM) chromatography columns have been used to model the in vivo distribution of drug discovery compounds. Regis Technologies Inc., the manufacturer, had to replace the silica support and consequently introduced a new IAM.PC.DD2 column that shows slightly different selectivity towards acidic and basic compounds. The application of the new IAM.PC.DD2 columns has been evaluated and the in vivo distribution models have been compared with the previous batches of columns. It was found that due to the improved endcapping of the silica, some of the positively charged drug molecules showed shorter retention than previously published. Therefore, the column system suitability data have been updated. However, these differences do not significantly affect the previously published models for the volume of distribution, brain tissue binding and drug efficiency. Therefore, the published models can be used with the new IAM.PC.DD2 columns.
Immobilized Artificial Membrane (IAM) chromatography columns have been used to model the in vivo distribution of drug discovery compounds. Regis Technologies Inc., the manufacturer, had to replace the silica support and consequently introduced a new IAM.PC.DD2 column that shows slightly different selectivity towards acidic and basic compounds. The application of the new IAM.PC.DD2 columns has been evaluated and the in vivo distribution models have been compared with the previous batches of columns. It was found that due to the improved endcapping of the silica, some of the positively charged drug molecules showed shorter retention than previously published. Therefore, the column system suitability data have been updated. However, these differences do not significantly affect the previously published models for the volume of distribution, brain tissue binding and drug efficiency. Therefore, the published models can be used with the new IAM.PC.DD2 columns.
Pidgeon and Venkataran [1] patented a method for immobilizing phospholipids on HPLC-grade silica stationary phases. The immobilized phospholipids mimic the lipid environment of a fluid cell membrane on a solid matrix. The retention factors (k) of compounds obtained on the IAM stationary phase are proportional to their affinity (partition coefficient) (K) to phospholipids according to the equation (1):where k is the retention factor obtained from the retention time (tR) and the dead time (t0) according to equation (2)and Vs/Vm is the volume ratio of the stationary and mobile phases, respectively. Many drug molecules bind strongly to phospholipids as it is a pre-requisite for membrane permeability. To reduce the retention times various concentrations of acetonitrile can be used in the mobile phase. There is a linear relationship between the log k values and the percentage of acetonitrile in the mobile phase that allows the extrapolation of the retention factor to the zero percentage of acetonitrile, i.e. to the pure aqueous buffer (log k) as shown by equation 3.Equation 3 enables the calculation of the acetonitrile concentration that is necessary to achieve log k = 0, when the retention time is exactly double that of the dead time, which means an equal distribution of the compound in the mobile and the IAM stationary phase [2]. This acetonitrile concentration expressed in volume percentage is called the Chromatographic Hydrophobicity Index (CHI (IAM)) [3]. It has also been shown that the CHI (IAM) values have linear relationships with the gradient retention times of the compounds obtained on the IAM stationary phases. By using an acetonitrile gradient of 0% to a maximum of 85% the obtained gradient retention times can be calibrated from the CHI (IAM) values by using a set of standards, eliminating the need to carry out isocratic measurements. The so obtained CHI (IAM) values show a good correlation to the log k (IAM) values, that are the extrapolated logarithmic retention factors to 0% acetonitrile concentration in the isocratic mode. The chromatographic determination of phospholipid-binding is based on measuring gradient retention times. It is independent of the amount of compound injected onto the column, so there is no need for quantitative analysis. The time measurement is very reproducible, and the calibrated (relative) gradient retention data are suitable for inter-laboratory comparison and application in published models of in vivo drug distribution. The data so obtained can be compiled in databases and is suitable for use in establishing quantitative structure- retention relationships, thus enabling chemists to design compounds with the appropriate IAM binding.The application of IAM chromatography as an aid in drug discovery has been reviewed extensively in recent years [4-8]. The molecular factors that influence IAM retention have also been discussed [9, 10]. It has been found that IAM columns retain strongly positively charged compounds as the IAM columns are negatively charged on the surface similar to cell membranes [11-13]. Several applications of IAM chromatography have also published [14] and this is summarized below.Intestinal absorption and drug distribution depend on a compound’s partitioning into phospholipids. Compounds have to possess a certain degree of affinity to membranes in order to permeate through the biological phospholipid bilayers. Therefore, IAM chromatography can provide an insight into the potential intestinal absorption of compounds [15]. As tissues represent a more non-polar lipid environment relative to the plasma, the human volume of distribution can be modelled by the binding differences between IAM and human serum albumin (HSA) [13, 16], as shown by equation (4):The sum of the albumin and phospholipid-binding inversely correlates to the unbound fraction of the compounds in tissues according to equation (5) [17].The in vivo drug efficiency can be modelled by the sum of the IAM and HSA binding of compounds [18] according to equation 6. The in vitro DRUGeff shows an excellent correlation with the in vivo DRUGeff. Drug efficiency and potency enables early dose estimation as drug efficiency approximates to the proportion of free drug concentration and dose [19].It has been found that basic compounds that have CHI (IAM) value greater than 50 have phospholipidosis potential [20] and also show promiscuous binding, interacting with several targets, causing potential toxicity and side effects [21].At present Regis Technologies Inc. (Morton Grove, IL, USA) is the only manufacturer of the Immobilized Artificial Membrane (IAM) stationary phase. The stationary phase contains only a monolayer of the phosphatidylcholine (PC) with the aliphatic part chemically bonded to the silica while the choline polar head group faces towards the mobile phase. Initially, “Type A” silica was used in the bonding procedure. The PC molecule has only one alkyl chain available with the second alkyl chain directly bonded to the silica surface. In this way, a denser bonding could be achieved which was closer to the natural density of PC molecules on the cell surface. However, in 2018 the company had to change the base silica material to a newer “Type B” as the old supply was no longer available. “Type B” silica has been chosen as the closest in properties to “Type A” regarding the specific surface area, pore size and particle size. When testing the retention selectivity of the new IAM.PC.DD2 columns which were otherwise manufactured in the same way, differences were observed towards acidic and basic molecules. This was somewhat surprising as the new silica support had identical physical specifications to the previously used silica (10 μm particle size, 300 Å pore size and nearly identical surface area). However, given the newer silica is of a much higher purity with only single ppm metal content, these observations are understandable as the new silica provides a much more inert surface and likely contributes to less non-specific binding interactions.The difference in selectivity can also be attributed to the different arrangements of the positive and negative charges of the choline headgroup on the stationary phase surface, again somewhat affected by the higher purity silica during the bonding process. The new batches of columns did not pass the original system suitability test criteria. Hence the aim was to establish new CHI (IAM) values for the system suitability test compounds that could be reproduced on the new batches of IAM columns. In this paper, the investigation of the selectivity difference is reported using over 70 marketed drug molecules and the in vivo distribution models are compared for the old and the new batches of IAM.PC.DD2 stationary phases. The estimated volume of distribution data was compared with the data from the previously published models. The differences of the models for the estimated brain tissue binding and drug efficiency values of the known drug molecules are investigated and compared with the published data obtained using the previous batches of IAM.PC.DD2 columns.
Experimental
All high-performance liquid chromatography measurements were performed on an Agilent 1100 Series HPLC instrument equipped with an ultraviolet diode array detector (DAD).
Measurements of membrane binding using immobilized artificial membrane (IAM) chromatography
The IAM binding was measured using IAM.PC.DD2 HPLC columns (Regis Technologies, Inc., IL, USA) with dimensions of 100 x 4.6 mm, particle size 10 μm with 300 Å pore size.Two batches of IAM.PC.DD2 columns were studied (R20511-014-3 and P20511-014-3) that had been prepared using the new batches of silica support. The selectivity and the models obtained were compared with the published average IAM data obtained on the columns manufactured between 2004 and 2012 [14].Mobile phase A was 50 mM ammonium acetate adjusted to pH 7.4 and mobile phase B was 100% acetonitrile. The mobile phase flow rate was 1.5 mL/min and the run time was 6 min. The acetonitrile gradient was applied from 0 to 85% from 0 to 4.75 min and kept at 85% until 5.15 min. From 5.15 to 5.25 min the acetonitrile concentration dropped back to 0%.The retention times were standardized using the IAM calibration mixture, containing octanophenone, heptanophenone, hexanophenone, valerophenone, butyrophenone, propiophenone, acetophenone, acetanilide and paracetamol. The retention times of the compounds were plotted against the isocratically determined and predefined CHI (IAM) values [4] listed in Table 1. The slope and intercept of the obtained straight line was used to calculate the CHI IAM values of new compounds. The CHI IAM values were converted to the octanol/water lipophilicity scale and expressed as log K (IAM) values using equation (7).
Table 1.
The constant CHI (IAM) and log K (IAM) values for the calibration set of compounds with typical retention times.
Compound
CHI (IAM)
log K (IAM)
tR 7.4 (min)
Octanophenone
49.4
4.97
4.37
Heptanophenone
45.7
4.30
4.17
Hexanophenone
41.8
3.71
3.93
Valerophenone
37.3
3.16
3.66
Butyrophenone
32.0
2.62
3.32
Propiophenone
25.9
2.15
2.90
Acetophenone
17.2
1.67
2.34
Acetanilide
11.5
1.45
2.04
Paracetamol
2.90
1.20
1.58
The retention times were standardized using the calibration set of compounds described previously [23].
The so-obtained log KIAM values express the membrane partition comparable to the octanol/water lipophilicity as described in reference [22]. A typical chromatogram is shown in Figure 1.
Figure 1.
A typical chromatogram obtained on IAM.PC.DD2 100 x 4.6 mm HPLC column. Flow rate: 1.5 mL/min; Mobile phase A: 50 mM ammonium acetate adjusted to pH 7.4, mobile phaseB: acetonitrile. Run time 6 min; Gradient: 0 to 4.75 min 0 to 85% B, 4.75 to 5.15 min 85% B, 5.15 to 5.25 min 0% B; Detection wavelength: 250 nm
The natural state of the phosphatidylcholine headgroup on the IAM stationary phase provides the column selectivity that was tested using the system suitability test compounds containing acidic, basic and neutral compounds.
Protein binding measurements using biomimetic protein stationary phases
The interaction of the compounds listed in Table 2 with human serum albumin (HSA) has been measured using commercially available chemically bonded HSA columns (Chiralpak-HSA from HiChrom, UK) with the dimensions of 50 x 3 mm and a particle size of 5 μM.
Table 2.
The name of the marketed drugs, their CAS number, their previously published CHI (IAM) data and the logarithm of their clinical volume of distribution.
Compound
CAS NO
Acid/base character
CHI (IAM) [14]
log Vdss [13]
Acecainide
32795-44-1
basic
23.3
0.10
Acetanilide
103-84-4
neutral
10.7
-0.13
Acetazolamide
59-66-5
neutral
1.7
-0.46
Alclofenac
22131-79-9
acidic
17.8
-1.10
Amoxapine
14028-44-5
basic
56.5
1.67
Bamethan
3703-79-5
basic
18.6
0.01
Betamethasone
378-44-9
neutral
31.7
0.18
Carbamazepine
298-46-4
neutral
26.5
-0.09
Chlorpromazine
50-53-3
basic
61.9
2.15
Cinoxacin
28657-80-9
acidic
1.6
-0.42
Colchicine
64-86-8
neutral
23.7
0.03
Cytarabine
147-94-4
weak base
-15.0
-0.31
Diazoxide
364-98-7
weak acid
24.7
-0.16
Diclofenac
15307-86-5
acidic
33.5
-0.67
Diprophylline
479-18-5
neutral
-4.0
-0.24
Ethinyl estradiol
57-63-6
neutral
46.8
0.39
Famotidine
76824-35-6
weak base
15.7
-0.06
Felbamate
25451-15-4
neutral
19.1
-0.64
Felodipine
72509-76-3
neutral
46.1
0.35
Fenoprofen
31879-05-7
acidic
17.2
-1.27
Finasteride
98319-26-7
neutral
38.9
0.30
Floxacillin
5250-39-5
acidic
23.4
-0.56
Flumazenil
78755-81-4
neutral
18.4
-0.04
Flurbiprofen
5104-49-4
acidic
26.8
-1.56
Furosemide
54-31-9
acidic
21.4
-0.71
Gemfibrozil
25812-30-0
acidic
32.9
-1.34
Glipizide
29094-61-9
acidic
21.1
-0.73
Griseofulvin
126-07-8
neutral
33.1
0.12
Hydrochlorothiazide
58-93-5
weak acid
15.9
-0.13
Hydrocortisone
50-23-7
neutral
27.9
0.12
Imipramine
50-49-7
basic
51.6
1.16
Indomethacin
53-86-1
acidic
25.3
-0.96
Isradipine
75695-93-1
neutral
40.0
0.13
Ketoconazole
65277-42-1
weak base
42.9
0.32
Ketoprofen
22071-15-4
acidic
21.9
-1.01
Labetalol
36894-69-6
amphoteric
43.9
0.71
Lignocaine
137-58-6
basic
32.6
0.40
Methylprednisolone
83-43-2
neutral
32.1
0.16
Metronidazole
443-48-1
weak base
-3.3
-0.25
Minoxidil
38304-91-5
weak base
19.0
-0.04
Nabumetone
42924-53-8
neutral
38.4
0.03
Nadolol
42200-33-9
basic
20.2
-0.33
Nicardipine
55985-32-5
weak base
45.9
0.47
Nifedipine
21829-25-4
neutral
29.0
-0.04
Nitrendipine
39562-70-4
neutral
40.5
0.16
Papaverine
58-74-2
weak base
34.4
0.00
Pentoxifylline
6493-05-6.
neutral
12.0
-0.10
Perphenazine
58-39-9
basic
56.3
1.35
Phenytoin
57-41-0
weak acid
31.6
0.03
Pindolol
13523-86-9
basic
42.0
0.85
Prazosin
19216-56-9
weak base
31.6
-0.05
Prednisolone
50-24-8
neutral
28.0
0.14
Prednisone
53-03-2
neutral
25.9
0.05
Primidone
125-33-7
neutral
8.9
-0.15
Probenecid
57-66-9
acidic
20.1
-0.75
Procainamide
614-39-1
basic
19.9
0.02
Propanolol
525-66-6
basic
50.8
1.29
Propylthiouracil
51-52-5
weak acid
3.9
-0.34
Proxyphylline
603-00-9
neutral
1.1
-0.85
Sulfachlorpyridazine
80-32-0
acidic
6.0
-0.76
Sulfameter
651-06-9
acidic
3.7
-0.61
Sulfamethoxypyridazine
80-35-3
weak acid
9.6
-0.59
Sulfinpyrazone
57-96-5
acidic
26.1
-0.65
Sulfisoxazole
127-69-5
acidic
2.9
-0.72
Sulphadimidine
57-68-1
amphoteric
34.6
0.22
Sulpiride
15676-16-1
basic
25.5
0.12
Tamoxifen
10540-29-1
basic
58.7
1.48
Theobromine
83-67-0
weak acid
-4.1
-0.25
Tolfenamic Acid
13710-19-5
acidic
36.6
-0.76
Trazodone
19794-93-5
weak base
36.3
0.01
Trimethoprim
738-70-5
weak base
20.8
-0.06
Warfarin
81-81-2
acidic
19.9
-0.78
Mobile phase A was 50 mM ammonium acetate adjusted to pH 7.4 and mobile phase B was 2-propanol. The flow rate was 1.5 mL/min and the run time was 6 min. The 2-propanol gradient was applied from 0 to 35% from 0 to 3.00 min and kept at 35% until 4.00 min. From 4.00 to 4.25 min the 2-propanol concentration was dropped back to 0%. Although column manufacturers suggest using pH 7 for chiral separations using the HSA column, in this case, the pH 7.4 mobile phase is used to mimic the plasma pH for the plasma protein binding measurements.
Compounds studied
In this study, 72 commercial drugs from diverse therapeutic areas have been analysed. 50 μl of 10 mM DMSO solutions of the drugs were diluted down to 150 μl and 10 μl of the diluted solution was injected onto the HPLC columns. The clinical volume of distribution data for all of the drugs is available in the literature [13].Table 2 shows the marketed drug molecules and their clinical volume of distribution data obtained from reference [13] and the previously measured CHI (IAM) data [14] using an IAM column manufactured by Regis Technologies between 2004 and 2012.
Data analysis
The statistical analysis was performed using JMP 13 (SAS Institute, USA).
Results and Discussion
The selectivity and reproducibility of IAM.PC.DD2 columns have been extensively studied. First, the CHI (IAM) values of the system suitability test compounds (see Table 2) were scrutinized. The CHI (IAM) values of the test compounds were then compared with the CHI (IAM) values obtained on the new batches of IAM.PC.DD2 columns. It was found, as shown in Table 3, that carbamazepine, warfarin, indomethacin, nicardipine, ketoprofen, haloperidol and budesonide had very similar CHI (IAM) values. However, imipramine and chlorpromazine showed greater than the 5 CHI (IAM) unit deviations which would be within the experimental error and therefore they did not pass the system suitability criteria. These compounds are strong bases and positively charged at pH 7.4. Most probably the new IAM.PC.DD2 columns have less negative charge on the surface and therefore reduce the retention times of positively charged compounds. It is also possible that more of the free silanol groups were end-capped which would contribute to a shorter retention of basic compounds. Therefore, the system suitability criteria were changed for basic compounds. The “new” CHI (IAM) values are listed in Table 3.
Table 3.
The CHI (IAM) obtained from the literature [13] (CHI (IAM) reference), the CHI (IAM) values obtained on the new IAM.PC.DD2 columns and the acid/base character of the IAM calibration compounds
Compound
Acid/base character
CHI (IAM)
CHI (IAM) “new”
Carbamazepine
neutral
28.9
27.0
Colchicine
neutral
23.7
23.0
Warfarin
acidic
19.9
24.0
Indomethacin
acidic
32.5
30.0
Nicardipine
weak base
45.9
45.0
Propranolol
basic
45.1
42.0
Imipramine
basic
54.1
45.0
Ketoprofen
strong acid
21.9
20.0
Chlorpromazine
basic
61.9
54.0
Haloperidol
basic
44.3
48
Budesonide
neutral
38.7
38
The in vivo distribution models were re-examined using the data from the new IAM.PC.DD2 columns. Because of the selectivity differences, it was essential to investigate the CHI (IAM) values for a larger set of compounds and check how the previously described in vivo distribution models performed using the data from the new columns manufactured with the new silica material.The 72 drug molecules from the original set of marketed drugs for which the models were developed [13] were analysed on the new IAM.PC.DD2 columns and the HSA column. Table 4 shows the CHI (IAM) values and the log k HSA binding data obtained from reference [13] and remeasured in this study.
Table 4.
The experimental CHI (IAM) values of the investigated compounds obtained on the new IAM.PC.DD2 columns and the CHI (IAM) obtained from the literature [13] (reference CHI (IAM))
Compound
CHI (IAM) from ref [13]
R20511-014-3 CHI (IAM)
P20511-014-3 CHI (IAM)
log K HSA from ref [13]
log K HSA
Acecainide
23.30
17.72
17.57
0.69
0.52
Acetanilide
10.70
11.55
11.50
0.62
0.43
Acetazolamide
1.66
0.99
0.74
1.47
1.44
Alclofenac
17.80
20.47
21.63
5.11
5.40
Amoxapine
56.50
45.41
44.74
2.15
2.70
Bamethan
18.60
18.53
18.03
0.49
0.42
Betamethasone
31.70
30.18
30.05
1.09
1.39
Carbamazepine
26.50
27.18
27.07
1.99
1.81
Chlorpromazine
61.90
51.54
50.94
1.18
1.27
Cinoxacin
1.60
5.31
6.67
0.79
0.91
Colchicine
23.70
23.45
23.18
0.69
0.24
Cytarabine
-15.00
-6.76
-6.99
1.59
1.87
Diazoxide
24.70
23.39
23.44
5.42
5.58
Diclofenac
33.50
35.18
35.85
0.63
0.24
Diprophylline
-4.00
-1.49
-1.98
4.1
4.24
Ethinyl Estradiol
46.80
45.30
45.23
0.46
0.50
Famotidine
15.70
16.82
16.08
1.39
3.43
Felbamate
19.07
39.16
39.21
3.46
4.15
Felodipine
46.10
44.37
44.45
5.91
6.11
Fenoprofen
17.20
27.17
28.26
1.58
2.32
Finasteride
38.90
37.15
37.17
3.62
3.55
Floxacillin
23.40
23.76
24.59
0.56
0.65
Flumazenil
18.40
18.27
18.27
7.81
8.50
Flurbiprofen
26.80
30.23
30.89
1.26
4.01
Furosemide
21.41
21.96
22.69
3.64
8.50
Gemfibrozil
32.90
33.70
34.53
3.32
4.05
Glipizide
21.10
22.82
23.22
1.49
1.92
Griseofulvin
33.10
31.06
30.98
0.77
0.80
Hydrochlorothiazide
15.90
13.04
13.08
0.92
1.05
Hydrocortisone
27.90
27.23
27.02
1.95
2.59
Imipramine
51.60
45.58
44.81
6.17
5.59
Indomethacin
25.30
32.49
33.16
2.86
3.36
Isradipine
40.00
39.11
39.16
2.9
3.30
Ketoconazole
42.90
39.50
39.26
4.8
5.40
Ketoprofen
21.90
22.93
23.89
1.28
1.80
Labetalol
43.90
38.29
37.48
0.69
0.52
Lignocaine
32.60
24.89
25.53
1.35
1.53
Methylprednisolone
32.10
30.91
30.77
0.76
0.28
Metronidazole
-3.30
-2.71
-3.24
0.66
0.72
Minoxidil
19.00
15.93
16.08
3.16
3.41
Nabumetone
38.40
35.90
35.96
0.49
2.14
Nadolol
20.20
19.89
19.23
2.93
3.55
Nicardipine
45.90
43.82
43.73
1.41
1.95
Nifedipine
29.00
28.11
28.16
3.13
3.37
Nitrendipine
40.50
39.06
39.19
2.35
2.70
Papaverine
34.40
31.22
31.15
0.63
0.39
Pentoxifylline
12.00
14.04
13.87
3.27
4.03
Perphenazine
56.30
49.57
48.97
1.6
2.02
Phenytoin
31.60
30.18
30.20
0.54
0.63
Pindolol
42.00
25.69
25.31
1.81
2.40
Prazosin
31.60
29.82
29.35
0.89
0.95
Prednisolone
28.00
27.31
27.20
0.8
1.08
Prednisone
25.90
25.25
25.13
0.62
0.42
Primidone
8.90
10.31
10.49
3.44
4.01
Probenecid
20.10
23.45
24.38
0.76
0.49
Procainamide
19.90
11.73
11.89
0.73
1.67
Propranolol
50.80
42.24
41.46
0.63
1.02
Propylthiouracil
3.90
6.56
6.52
2.67
3.21
Proxyphylline
1.10
10.05
10.89
1.97
3.02
Sulfachlorpyridazine
6.00
10.08
10.83
2.04
2.20
Sulfameter
3.70
6.92
7.09
4.09
2.44
Sulfamethoxypyridazine
9.60
11.11
11.16
2.33
4.27
Sulfinpyrazone
26.10
26.19
26.89
3.17
2.66
Sulfisoxazole
2.90
6.48
7.37
1.54
3.68
Sulphadimidine
34.55
10.89
11.09
0.75
1.72
Sulpiride
25.50
15.80
16.15
4.84
0.71
Tamoxifen
58.70
54.23
53.78
1.29
4.82
Theobromine
-4.10
-3.16
-3.71
4.68
0.27
Tolfenamic Acid
36.60
40.72
41.53
2.46
6.60
Trazodone
36.30
34.09
33.86
0.8
3.06
Trimethoprim
20.80
20.99
20.52
4.45
0.95
Warfarin
19.90
23.47
24.18
0.7
4.16
Figure 2 shows the plot of the experimental CHI (IAM) values on IAM.PC.DD2 columns P20511-014-3 and R2055-014-3. The CHI (IAM) values obtained on the columns are very similar indicating good batch to batch reproducibility of the new columns.
Figure 2.
The plot of the experimental CHI (IAM) values on IAM.PC.DD2 columns P20511-014-3 and R20511-014-3
Figure 3 shows the plot of the CHI (IAM) from the literature [13], reference CHI (IAM), with the experimental CHI (IAM) values using the IAM.PC.DD2 column R2055-014-3. The CHI (IAM) values of procainamide, sulpiride, zolmitriptan and pindolol were outliers from the correlation with lower CHI (IAM) values on the new batches of IAM columns. Compounds are considered outliers when the difference between the measured and predicted value is greater than the double of the standard error of the estimate, in this case 7.2 CHI IAM units. The slope of the regression line (red) is less than one (0.81) and the regression line is different from the line of unity (green). When these 4 strong basic compounds are left out from the correlation, the correlation coefficient increases to 0.96 and the slope increases to 0.82 as shown in Figure 4.
Figure 3.
The plot of the CHI (IAM) values from the literature [13] (reference CHI (IAM)) with the CHI (IAM) values using IAM.PC.DD2 column R25011-014-3. The green line is the line of unity.
Figure 4.
Comparison of the experimental CHI (IAM) values obtained on the IAM.PC.DD2 (R20511-014-3) column with the reference CHI (IAM) values without the 4 strong basic compounds. The green line is the line of unity
The results shown in Figure 4 suggests that the rank order of the phospholipid binding will be very similar using the CHI IAM data obtained on the new IAM.PC.DD2 column and the old IAM.PC.DD2 column. However, some basic compounds may show significantly weaker phospholipid binding. The absolute CHI IAM values maybe lower when the new batches of IAM columns are used.
The human clinical volume of distribution and protein binding data of the investigated marketed drugs
The volume of distribution model is based on two properties of the compounds, the IAM binding and the human serum albumin (HSA) binding. Therefore, the IAM binding data obtained using the new IAM.PC.DD2 columns were examined in the models of the human clinical volume of distribution together with the re-measured protein binding data.Figure 5 shows the reproducibility of the HSA columns by plotting the newly measured log K (HSA) values as a function of the “reference log K (HSA) “ values published previously [13]. It can be seen that the slope is very close to 1 and the intercept is very close to 0. There are slightly greater discrepancies between the reference and the re-measured log K (HSA) values for strongly bound compounds. This shows that the specific binding sites of the human serum albumin may vary only slightly from column to column. With this excellent agreement between the log K (HSA) values obtained on various Chiral HSA columns, it can be assumed that they would not affect the volume of distribution model which is based on two binding properties for the compounds, namely the IAM and HSA binding.
Figure 5.
The plot of the log K (HSA) values from the literature [13] and the re-measured log K (HSA) for this study
Estimating the volume of distribution using the old and new IAM.PC.DD2 columns
The logarithmic values of the HSA and IAM binding data have been used to estimate the human clinical volume of distribution. Using the published equation (equation 4), the estimated values are compared using the IAM data obtained on columns R20511-014-3 and P20511-014-3 and the IAM data from the literature [12].The estimated log Vdss values have been calculated using equation 4 and the measured log K (HSA) and log K (HSA) data obtained on the new IAM.PC.DD2 columns. The estimated log Vdss values are listed in Table 5 together with the acid/base character of the compounds.
Table 5.
The estimated log Vdss values of the investigated compounds using equation 4 and their acid/base character
Drug name
Acid/base character
R20511-014-3 log Vdss
P20511-014-3 log Vdss
reference log Vdss
Acecainide
basic
-0.03
-0.03
0.1
Acetanilide
neutral
-0.12
-0.12
-0.13
Acetazolamide
neutral
-0.47
-0.47
-0.46
Alclofenac
acidic
-1.04
-1.02
-1.1
Amoxapine
basic
0.62
0.57
1.67
Bamethan
basic
0.01
0
0.01
Betamethasone
neutral
0.12
0.12
0.18
Carbamazepine
neutral
-0.07
-0.07
-0.09
Chlorpromazine
basic
0.88
0.82
2.15
Cinoxacin
acidic
-0.38
-0.37
-0.42
Colchicine
neutral
0.02
0.01
0.03
Cytarabine
weak base
-0.26
-0.26
-0.31
Diazoxide
weak acid
-0.19
-0.19
-0.16
Diclofenac
acidic
-0.6
-0.57
-0.67
Diprophylline
neutral
-0.22
-0.23
-0.24
Ethinyl estradiol
neutral
0.28
0.27
0.39
Famotidine
weak base
-0.04
-0.05
-0.06
Felbamate
neutral
0.07
0.07
-0.64
Felodipine
neutral
0.23
0.23
0.35
Fenoprofen
acidic
-1.02
-0.98
-1.27
Finasteride
neutral
0.21
0.21
0.3
Floxacillin
acidic
-0.55
-0.53
-0.56
Flumazenil
neutral
-0.04
-0.04
-0.04
Flurbiprofen
acidic
-1.44
-1.42
-1.56
Furosemide
acidic
-0.7
-0.68
-0.71
Gemfibrozil
acidic
-1.31
-1.27
-1.34
Glipizide
acidic
-0.69
-0.68
-0.73
Griseofulvin
neutral
0.04
0.03
0.12
Hydrochlorothiazide
weak acid
-0.18
-0.17
-0.13
Hydrocortisone
neutral
0.1
0.09
0.12
Imipramine
basic
0.66
0.6
1.16
Indomethacin
acidic
-0.72
-0.69
-0.96
Isradipine
neutral
0.08
0.08
0.13
Ketoconazole
weak base
0.12
0.1
0.32
Ketoprofen
acidic
-0.98
-0.96
-1.01
Labetalol
amphoteric
0.38
0.34
0.71
Lignocaine
basic
0.14
0.16
0.4
Methylprednisolone
neutral
0.12
0.11
0.16
Metronidazole
weak base
-0.24
-0.25
-0.25
Minoxidil
weak base
-0.11
-0.1
-0.04
Nabumetone
neutral
-0.09
-0.09
0.03
Nadolol
basic
-0.34
-0.35
-0.33
Nicardipine
weak base
0.32
0.32
0.47
Nifedipine
neutral
-0.07
-0.07
-0.04
Nitrendipine
neutral
0.08
0.08
0.16
Papaverine
weak base
-0.13
-0.13
0
Pentoxifylline
neutral
-0.07
-0.07
-0.1
Perphenazine
basic
0.66
0.61
1.35
Phenytoin
weak acid
-0.02
-0.02
0.03
Pindolol
basic
0.14
0.13
0.85
Prazosin
weak base
-0.11
-0.13
-0.05
Prednisolone
neutral
0.12
0.12
0.14
Prednisone
neutral
0.03
0.03
0.05
Primidone
neutral
-0.13
-0.13
-0.15
Probenecid
acidic
-0.66
-0.64
-0.75
Procainamide
basic
-0.13
-0.12
0.02
Propanolol
basic
0.64
0.59
1.29
Propylthiouracil
weak acid
-0.31
-0.31
-0.34
Proxyphylline
neutral
-0.75
-0.74
-0.85
Sulfachlorpyridazine
acidic
-0.71
-0.7
-0.76
Sulfameter
acidic
-0.57
-0.57
-0.61
Sulfamethoxypyridazine
weak acid
-0.57
-0.56
-0.59
Sulfinpyrazone
acidic
-0.64
-0.62
-0.65
Sulfisoxazole
acidic
-0.68
-0.67
-0.72
Sulphadimidine
amphoteric
-0.41
-0.41
0.22
Sulpiride
basic
-0.11
-0.1
0.12
Tamoxifen
basic
0.94
0.89
1.48
Theobromine
weak acid
-0.24
-0.25
-0.25
Tolfenamic Acid
acidic
-0.54
-0.49
-0.76
Trazodone
weak base
-0.09
-0.1
0.01
Trimethoprim
weak base
-0.05
-0.06
-0.06
Warfarin
acidic
-0.69
-0.68
-0.78
Figure 6 shows the plot of the estimated log Vdss values using the CHI (IAM) values obtained on the IAM.PC.DD2 column P20511-014-3 and those derived from the reference CHI (IAM) values. A good correlation was found between the estimated log Vdss values. In this study equation 4 was used with the same log K (HSA) values and different log K (IAM) values (depending on which IAM column was used).
Figure 6.
The plot of the estimated log Vdss values using the CHI (IAM) values obtained on the IAM.PC.DD2 column P20511-014-3 and from the reference [13]. The green line is the line of unity.
Figure 7 shows the plot of the estimated log Vdss values using the CHI (IAM) values obtained on the IAM.PD.DD2 column P20511-014-3 and those obtained from the reference but without the 4 strong basic compounds (procainamide, sulpiride, zolmitriptan and pindolol). It can be seen that the agreement is slightly improved with an R2 value of 0.96, however, the slope is still less than one, while the intercept is close to zero. This means that the rank order of the volume of distribution data would be the same when using the new batches of IAM.PC.DD2 columns as it was when using the old batches of IAM columns. However, absolute values may be slightly different for the neutral, acidic and basic compounds and greater discrepancies can be expected for some strong basic compounds which are similar to those left out from the plot.
Figure 7.
The plot of the estimated log Vdss values using the CHI (IAM) values obtained on column P20511-014-3 and from the reference values [12] without the four strong basic compounds. The green line is the line of unity.
It was important to investigate how the estimated volume of distribution using the CHI (IAM) data from the new batches of the IAM.PC.DD2 columns correlated with the actual clinical volume of distribution data. Figures 8 and 9 show the correlation of the estimated log Vdss values using the same coefficients as in equation 4 with the clinical log Vdss values. Figure 9 shows that the CHI (IAM) values obtained on the new batches of IAM.PC.DD2 column can be used to estimate the human clinical volume of distribution data by using equation 4. There is better agreement with the clinical data when using the log Vdss obtained on the new batch of the IAM.PC.DD2 column. However, the absolute values may be slightly different and a correction factor (slope 0.62) will be required. In a drug discovery setting for ranking and differentiating compounds using an estimated log Vdss, these values can be used with confidence. However, it is important to acknowledge that certain strong basic compounds may appear to have a lower volume of distribution than expected.
Figure 8.
The plot of the estimated and the measured clinical volume of distribution data using the old batches of IAM columns (reference [13] reference log Vdss) using equation 4.
Figure 9.
The plot of the estimated and measured clinical volumes of distribution using the new batch R20511-014-3 IAM.PC.DD2 column versus the reference log Vdss
[.
It was also important to investigate how the CHI (IAM) and log K (IAM) values perform in other in vivo distribution models.
Estimating brain tissue binding using the old and new batches of IAM.PC.DD2 columns
The brain tissue binding can be estimated from the sum of the albumin and the phospholipid binding as described by equation 8. The model was built on 135 drug discovery compounds from several neuroscience projects using rat brain tissue binding data obtained by equilibrium dialysis and measured HSA and IAM binding data as described in reference [12].where n is the number of compounds, r is the correlation coefficient, s is the standard error of the estimate and F is the Fisher-test value.Figure 10 shows the comparison of the estimated brain tissue binding (%BTB) when the old (reference [13]) and the new column P20511-014-3 IAM.PC.DD2 data was used in the model.
Figure 10.
The plot of the estimated brain tissue binding (%BTB) obtained using the IAM data obtained from the reference [13] and that obtained using IAM.PC.DD2 column P20511-014-3
The agreement is very good with an R2 value of 0.95, the slope is close to 1 and the intercept is close to zero using the CHI (IAM) values obtained on P20511-014-3 from the latest batch of IAM.PC.DD2 columns. However, the four basic compounds that showed significantly lower CHI (IAM) values are significant outliers resulting in these compounds being estimated to have lower brain tissue binding than their actual brain tissue binding.
Estimating the unbound volume of distribution using the old and new batches of IAM.PC.DD2 columns
The agreement between the estimated unbound volume of distribution data using the reference [13] CHI (IAM) values and the CHI IAM values obtained on column R25011-014-3 are shown in Figure 11.
Figure 11.
A plot of the estimated unbound volume of distribution using the IAM data from the reference (reference [12]) and on the IAM.PC.DD2 R20511-014-3 column
It can be seen that the estimated unbound volume of distribution using the CHI (IAM) values obtained on column R20511-014-3 shows good agreement with the logVdu from the reference CHI (IAM) values. It can be observed from the plot that basic compounds tend to show a lower unbound volume of distribution using the CHI (IAM) values obtained on the new column R20511-014-3.
Estimating drug efficiency using the old and new batches of IAM.PC.DD2 columns
Figure 12 shows the comparison of the estimated drug efficiency values using the old and a new batch (R20511-014-3) of IAM.PC.DD2 column.
Figure 12.
The plot of the drug efficiency values for the investigated drugs using the IAM data obtained from the reference (reference [12]) and by R20511-014-3 column
There is good agreement between the drug efficiency values obtained on the new batch of the IAM column and the literature data. However, the basic compounds showed greater drug efficiency than expected when using the new batch (R20511-014-3) of IAM.PC.DD2 column.In conclusion, some differences between the old and the latest batches of IAM.PC.DD2 columns were observed when estimating the in vivo distribution of compounds. The discrepancies are more significant when the models are based on the differences between the measured IAM and HSA binding (volume of distribution). When the models are based on the sum of the two types of binding, the correlations between the estimated values (brain tissue binding, the unbound volume of distribution and drug efficiency) are much stronger. However, systematic differences could be observed for basic compounds in the estimated drug efficiency values. Polar basic (log P less than 2) compounds with over 6 rotatable bonds were estimated to have greater dug efficiency values than previously estimated using the old batches of IAM.PC.DD2 columns. These discrepancies are due to the weaker interaction of the new batches of IAM.PC.DD2 stationary phases with positively charged basic compounds. This could be the result of better endcapping procedures to block the free silanol groups and it could also be due to the higher purity of the silica providing less overall negatively charged silanols.Based on these observations it was decided to investigate the potential benefit of developing new models using the data obtained on the new silica based IAM stationary phase and to compare the new models with the existing model equation (equation 4).
Developing new distribution models using the new batches of IAM.PC.DD2 columns
The correlations between the old and the new batches of IAM retention data have been investigated. It was found that the reference CHI (IAM) data can be estimated using the CHI (IAM) data obtained on the new batches (P20511-014-3 and R20511-014-3) of IAM.PC.DD2 columns when the retention difference on C-18 columns obtained at pH 10.6 and pH 2.6 are included in the equation (equation (9)). This second variable significantly increased the explained variance of the reference CHI (IAM).The correlation was slightly improved (from 0.92 to 0.97) between the reference CHI (IAM) values and the CHI (IAM) values obtained on the new batch of IAM.PC.DD2 columns. However, in order to be able to use equation 9, the CHI measurements on C-18 columns should be carried out at high and low pH. This requires two additional measurements for each compound. When resources are limited, the calculated molecular descriptor can be used. The following parameters have been calculated for the investigated drug molecules: log P, molecular weight, number of H-bond donor and acceptor groups, total polar surface area, flexibility and the number of rotatable bonds (NRB). It was found that the only significant variables among the calculated parameters that improved the correlation between the CHI (IAM) values obtained using the new and old batches of IAM.PC.DD2 columns were clogP and the total polar surface area (TPSA) as shown in equation 10. However, the correlation coefficient improved only slightly from 0.92 to 0.96.The other option was to build a new model for the estimation of the volume of distribution using the new batches of IAM.PC.DD2 columns and the clinical volume of distribution data.The new equation for modelling clinical steady-state volume of distribution data is shown in equation 11. The mathematical-statistical parameters are very similar to the original equation (Equation 4); however, the coefficients of the IAM and HSA binding data are very different.The plot of the estimated and measured clinical steady-state volume of distribution data are shown in Figure 13. When comparing the coefficients of equation 4 (the published Vdss model equation) and equation 10, it can be seen that the coefficients for the HSA binding do not differ significantly (0.24 and 0.22 with an error of 0.02). However, the coefficients of log K (IAM) are different at the 95% confidence interval and agree only within a 90% confidence interval: 0.54 ± 0.04 and 0.44 ± 0.02.
Figure 13.
The plot of the calculated log Vdss values using the new model with the CHI (IAM) values obtained on column (R20511-014-3) as a function of the clinical log Vdss values.
Conclusions
The use of a new silica support provided a slightly different selectivity for the new IAM.PC.DD2 columns regardless of the bonding procedure used. Therefore, new CHI (IAM) values have been proposed for the system suitability test compounds. New IAM.PC.DD2 columns should be evaluated using the new CHI (IAM) values (Table 4.). The reproducibility of the new batches of IAM.PC.DD2 columns are very good and are within 2 CHI (IAM) units. The effect of the selectivity change has been investigated on previously developed in vivo models for drug distribution that are based on CHI (IAM) values. A set of 72 marketed drugs with clinical distribution data (volume of distribution, unbound volume of distribution and drug efficiency) have been analyzed on the new batches of IAM columns. The CHI (IAM) values were compared with the values obtained from the previous batches of IAM columns and it was observed that small polar basic compounds showed significantly shorter retention times on the new IAM.PC.DD2 columns. The correlation coefficient improved significantly when the data from these compounds are removed from the analysis. The possible reason for these discrepancies may be due to the better endcapping procedure on the new IAM columns which is blocking the silanol effects that increase retention of basic compounds. It is also likely that the newer “Type B” silica is higher purity silica with a more inert surface and creates fewer negatively charged silanols decreasing the amount of non-specific binding. It was found that using the new CHI (IAM) data in the previously published model equations, a good correlation was obtained with the original values. However, the absolute values were different but the rank order for the in vivo distribution values remain the same using the old and the new columns. New models have also been developed for estimating the clinical volume of distribution of compounds using the data from the new IAM columns. The mathematical statistical characteristics of the new models are similar, but the coefficients of the CHI (IAM) and log K (HSA) parameters are slightly different. Since the correlation between the estimated and clinical volume of distribution data was the same or slightly better with the data obtained from the new batches of IAM columns, it is expected that the in vivo estimates can be as accurate with the data obtained from the new batches of IAM columns as with the older batches of IAM columns. The newly derived equation (eqation 11) for the estimation of the volume of distribution is suggested when someone begins to use biomimetic chromatographic measurements with the new columns. However, reference can still be made to the existing data previously published using the older columns.
Authors: Eugen Lounkine; Michael J Keiser; Steven Whitebread; Dmitri Mikhailov; Jacques Hamon; Jeremy L Jenkins; Paul Lavan; Eckhard Weber; Allison K Doak; Serge Côté; Brian K Shoichet; Laszlo Urban Journal: Nature Date: 2012-06-10 Impact factor: 49.962