| Literature DB >> 35281226 |
Gazmend Berisha1,2, Rønnaug Solberg3,4, Claus Klingenberg5,6, Anne Lee Solevåg7.
Abstract
Objectives: Impedance cardiography (ICG) is a non-invasive method for continuous cardiac output measurement and has the potential to improve monitoring and treatment of sick neonates. PhysioFlow® is a signal-morphology ICG-system showing promising results in adults with low and high cardiac output, but no data from neonates or neonatal models exist. The aim of this study was to investigate PhysioFlow® feasibility in asphyxiated newborn piglets.Entities:
Keywords: animal model; asphyxia neonatorum; bioimpedance; cardio-circulatory monitoring of transition; hemodynamics
Year: 2022 PMID: 35281226 PMCID: PMC8913887 DOI: 10.3389/fped.2022.804353
Source DB: PubMed Journal: Front Pediatr ISSN: 2296-2360 Impact factor: 3.418
Figure 1The PhysioFlow® uses 6 snap button electrodes: 2 electrodes on the neck, 2 at the xiphisternum, and 1 on each side of the chest. The blue electrodes are the PhysioFlow® electrodes.
Figure 2Graphical presentation of the study time line.
Characteristics of 15 asphyxiated piglets monitored with PhysioFlow® signal-morphology impedance cardiography.
|
|
| |
|---|---|---|
| Baseline | Sex [female number (%)] | 10 (67) |
| Weight (g) | 2,045 (1,950–2,145) | |
| Baseline hemoglobin (g/dL) | 7.3 (7.0–7.6) | |
| Heart rate (beats per minute) | 172 (154–184) | |
| Mean arterial blood pressure (mm Hg) | 73 (58–74) | |
| Baseline lactate (mmol/L) | 1.6 (1.1–2.0) | |
| Asphyxia, asystole and cardiopulmonary resuscitation | Lactate at asystole (mmol/L) | 15.0 (9.1–17.0) |
| Duration of asphyxia (min) | 30 (20–35) | |
| Duration of cardiopulmonary resuscitation (sec) | 76 (60–201) | |
| Number of adrenaline doses | 4 (2–4) | |
| Stroke volume index (mL/m2) | 25.6 (11.5–38.9) | |
| Cardiac index (L/m2/min) | 4.3 (2.0–5.8) |
IQR, interquartile range.
Summary measure for all piglet 30 min from initiation of asphyxia, cardiopulmonary resuscitation and post-resuscitation observation.
Characteristics of PhysioFlow® signal-morphology impedance cardiography measurements in 15 asphyxiated piglets.
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|
| 1 | Asystole | 100 (85–100) | 114 | 971 | 122 (13) | 86 |
| 2 | Asystole | 40 (20–55) | 43 | 92 | 11 (12) | 11 |
| 3 | 9.5 h | 100 (85–100) | 586 | 6,759 | 5,536 (82) | 8 |
| 4 | Asystole | 65 (35–100) | 150 | 877 | 311 (35) | 286 |
| 5 | Asystole | 45 (25–75) | 82 | 138 | 15 (11) | 14 |
| 6 | 9.5 h | 100 (95–100) | 626 | 3,303 | 2,704 (82) | 67 |
| 7 | 9.5 h | 45 (25–70) | 66 | 2,621 | 603 (23) | 101 |
| 8 | Asystole | 85 (50–100) | 126 | 781 | 134 (17) | 127 |
| 9 | Asystole | 55 (35–85) | 68 | 233 | 80 (34) | 70 |
| 10 | Asystole | 45 (15–80) | 134 | 806 | 71 (9) | 68 |
| 11 | Asystole | 100 (75–100) | 86 | 619 | 196 (32) | 183 |
| 12 | Asystole | 50 (35–60) | 106 | 453 | 37 (8) | 12 |
| 13 | 30 min | 75 (55–90) | 330 | 2,899 | 210 (7) | 149 |
| 14 | Asystole | 55 (35–90) | 73 | 261 | 90 (34) | 56 |
| 15 | Asystole | 40 (25–61) | 61 | 178 | 28 (16) | 16 |
|
|
|
|
|
|
Euthanasia is the practice of ending the life of an animal in a controlled fashion.
Signal quality (%) median (IQR)- Signal quality is given as a number in percentage where above 75% is good signal quality. The median (IQR) indicates median with interquartile range in parenthesis.
The rest of the bold values and their meaning is self-explanatory.
Figure 3All measurements of each piglet are represented by bars and subdivided by signal quality (%). The total number of measurements for each piglet equals 100%.
Figure 4Box plots of cardiac index, pH, lactate, heart rate and arterial blood pressure at baseline, after post-surgical stabilization, at asystole and 9.5 h after return of spontaneous circulation. Within each box, the horizontal black line represents the median value; boxes extend from the 25th to the 75th percentile; while the whiskers represent the minimum and maximum values, respectively.