| Literature DB >> 35264042 |
Jeremie Korchia1, Kathleen P Freeman2.
Abstract
Determining a simple quality control (QC) rule for daily performance monitoring depends on the desired total allowable error (TEa) for the measurand. When no consensus TEa exists, the classical approach of QC rule validation cannot be used. Using the results of previous canine serum and urine cortisol validation studies on the Immulite 2000 Xpi, we applied a reverse engineering approach to QC rule determination, arbitrarily imposing sigma = 5, and determining the resulting TEa for the QC material (QCM; TEaQCM) and the resulting probability of error detection (Ped) for each QC rule. For the simple QC rule 12.5S with Ped = 0.96 and probability of false rejection (Pfr) = 0.03, the associated TEaQCM were 20% and 35% for serum and 28% and 24% for urine QCM1 and QCM2. If these levels of TEaQCM are acceptable for interpretation of patient sample results, then users can internally validate the 12.5S QC rule, provided that their QCM CVs and biases are similar to ours. Otherwise, more stringent QC rules can be validated by using a lower sigma to lower the TEaQCM. With spiked samples (relevant cortisol concentrations in the veterinary patient matrix) at 38.6 and 552 nmol/L of cortisol, TEaQCM at sigma = 5 were much higher (54% and 40% for serum; 90.3% and 42.8% for urine). Spiked samples generate TEa that is probably too high to be suitable for daily QC monitoring; however, it is crucial to verify spiked sample observed total error (TEo; 26% and 18% for serum, 60% and 30% for urine) < TEaQCM, and to use spiked sample TEo for patient result interpretation. In the absence of consensus TEa for cortisol in dogs, we suggest the use of a 12.5S rule, provided that users accept the associated level of TEaQCM also as clinical TEa for results interpretation.Entities:
Keywords: Cushing; cortisol; dogs; endocrinology; quality control; total allowable error
Mesh:
Substances:
Year: 2022 PMID: 35264042 PMCID: PMC8921817 DOI: 10.1177/10406387221076129
Source DB: PubMed Journal: J Vet Diagn Invest ISSN: 1040-6387 Impact factor: 1.569
Figure 1.Origin of the observed total error (TEo) formula. TEo is calculated from imprecision (SD) and inaccuracy (bias) by the formula: TEo = 2SD + absolute bias, in which all elements are expressed in units of the assessed measurand. When normalized by the mean of the target value, SD becomes the coefficient of variation (CV), and the formula becomes: TEo(%) = 2CV(%) + absolute bias(%). 4SD includes 95% of the results (mean ± 2SD) when values are normally distributed, which is assumed for measured replicates.
Figure 2.A. Classical approach and B. reverse approach for QC rule validation.
EZ Rules 3 QC design software; OPSpec charts = operational process specification charts; Ped = probability of error detection; Pfr = probability of false rejection; QC = quality control; TEa = total allowable error; TEo = observed total error; σ = sigma metric.
Figure 3.Canine serum cortisol and canine urine cortisol validation study overview. Our study was performed using the databases of 2 previous studies.
ACTHST = ACTH stimulation test; AHDC-CU = Animal Health Diagnostic Center of Cornell University; LDDST = low-dose dexamethasone suppression test; QC = quality control; QCM = quality control material; TEo = observed total error; TVMDL = Texas A&M Veterinary Medical Diagnostic Laboratory; VDL-MSU = Veterinary Diagnostic Laboratory of Michigan State University.
Types of computed observed total error functions of the between-run CV in different media (rows) and different considered biases (columns).
| TEo calculation | Spiking-recovery bias | Average bias | Range-based bias | QCM between-run bias |
|---|---|---|---|---|
| Between-run CV for serum cortisol | TEoSR
| TEoAB
| TEoRB
| TEoQCM
|
| Between-run CV for urine cortisol | TEoSR
| TEoAB
| TEoRB
| TEoQCM.SR
|
AB = average bias; L4 = cortisol concentration of 38.6 nmol/L (1.4 μg/dL); L8 = cortisol concentration of 552 nmol/L (20 μg/dL); QCM = quality control material; RB = range-based bias; SR = spiking-recovery bias; TEo = observed total error.
Ranged-based bias from groups of the comparison study of the closest ranges from those spiked levels.
There is no target value for urine cortisol QCM with the Immulite 2000 Xpi; thus, to calculate TEo, we used 3 other types of biases: SR and RB at a roughly similar concentrations to those of QCM1 and QCM2, and AB. The latter was elected for QC rule determination for urine cortisol.
Results from canine serum and canine urine cortisol validation studies, used as materials in our current study.
| Spiked serum samples | ||||||||
|---|---|---|---|---|---|---|---|---|
| Cortisol level | Target values, nmol/L (μg/dL) | Precision (%) | Bias (%) | TEo (%) | ||||
| Between-run CV | SR | RB | AB | TEoSR | TEoRB | TEoAB | ||
| L4 | 38.6 (1.4) | 9.53 | −6.50 | −13.8 | −2.93 | 25.6 | 32.8 | 22.0 |
| L8 | 552 (20) | 7.42 | 3.08 | −4.47 | 17.9 | 19.3 | 17.7 | |
| QCM | ||||||||
| Cortisol level | Target values, | Precision (%) | Bias (%) | TEoQCM (%) | ||||
| Between-run CV | ||||||||
| QCM1 | 193 (7.0) | 4.08 | −0.27 | 8.5 | ||||
| QCM2 | 389 (14.1) | 7.01 | −0.52 | 14.5 | ||||
| Spiked urine samples | ||||||||
| Cortisol level | Target values, nmol/L (μg/dL) | Precision (%) | Bias (%) | TEo (%) | ||||
| Between-run CV | SR | RB | AB | TEoSR | TEoRB | TEoAB | ||
| L4 | 38.6 (1.4) | 9.55 | 42.3 | 8.67 | −1.09 | 61.4 | 27.8 | 20.2 |
| L8 | 552 (20) | 4.60 | 19.8 | −10.7 | 28.9 | 19.9 | 10.3 | |
| QCM | ||||||||
| Cortisol level | Target values, | Precision (%) | Bias (%) | TEoQCM (%) | ||||
| Between-run CV | SR | RB | AB | TEoQCM.SR | TEoQCM.RB | TEoQCM.AB | ||
| QCM1 | 193 (7.0) | 5.42 | 35.8 | 8.67 | −1.09 | 46.6 | 19.5 | 11.9 |
| QCM2 | 389 (14.1) | 4.51 | 19.8 | −10.7 | 28.7 | 19.7 | 10.1 | |
For section A, we used the commercial QCM K9CON (Immulite systems control; Siemens), which has target values available. In serum, TEoSR, TEoRB, and TEoAB are roughly similar; TEo is ~30% for L4 and ~20% for L8.
For section B, we used the commercial QCM Liquicheck (Bio-Rad), for which target values for Immulite analyzers are not available. Means are provided to give an idea of the tested level. Because of the absence of target values, the QCM bias could not be assessed. Instead, TEo was calculated using the biases from the spiking-recovery study (SR) and the comparison study (RB and AB). Unlike in serum, in urine 10 TEoSR, TEoRB, and TEoAB are mismatching because of widely different corresponding biases. The SR bias may be too high and irrelevant in a clinical setting for which comparisons are made between analyzers. The AB may be too low as resulting from the average of opposite biases, and thus irrelevant as neglecting the impact on the measurand concentration on the bias. The RB bias may be the most relevant clinically and would warrant further investigation; of note, the RB bias yields TEo in urine roughly equivalent to serum for L4 (~30%) and L8 (~20%).
AB = average bias; L4 = cortisol concentration of 38.6 nmol/L (1.4 μg/dL); L8 = cortisol concentration of 552 nmol/L (20 μg/dL); RB = range-based bias; SR = spiking-recovery bias; TEo = observed total error.* Closest available bias.† There were no target values provided by the manufacturer of the QCM for canine urine cortisol.
Reverse approach (Fig. 2B) for QC rule validation for commercial QCM (K9CON for serum, Liquicheck for urine), at low (insufficient), intermediate (optimized), and high (excessively high) TEa for the system, using EZ Rules 3, for n = 2 QCM levels.
| Increasing TEaQCM | Low TEaQCM: 20% | Optimized TEaQCM (σ = 5) | High TEaQCM: 50% | |||||
|---|---|---|---|---|---|---|---|---|
| Serum | QCM1 | Sigma | σ = 4.8 | TEa = 20.8% | σ = 12.1 | |||
| CV = 4.1% | QC rule | Ped | Pfr | Ped | Pfr | Ped | Pfr | |
| Bias = 0.3% | 12S | 0.99 | 0.09 | 1 | 0.09 | 1.00 | 0.09 | |
| TEo = 8.5% | 12.5S |
|
|
|
|
|
| |
| 13S | 0.80 | 0.00 | 0.86 | 0.00 |
|
| ||
| 13S/22S/R4S |
|
|
|
|
|
| ||
| QCM2 | Sigma | σ = 2.8 | TEa = 35.5% | σ = 7.1 | ||||
| CV = 7% | QC rule | Ped | Pfr | Ped | Pfr | Ped | Pfr | |
| Bias = 0.5% | 12S | 0.35 | 0.09 | 1 | 0.09 | 1.00 | 0.09 | |
| TEo = 14.5% | 12.5S | 0.16 | 0.03 |
|
|
|
| |
| 13S | 0.06 | 0.00 | 0.86 | 0.00 | >0.98 | 0.00 | ||
| 13S/22S/R4S | 0.08 | 0.01 |
|
|
|
| ||
| Urine | QCM1 | Sigma | σ = 3.5 | TEa = 28.1% | σ = 9.1 | |||
| CV = 5.4% | QC rule | Ped | Pfr | Ped | Pfr | Ped | Pfr | |
| Bias | 12S | 0.72 | 0.09 | 1 | 0.09 | 1.00 | 0.09 | |
| TEo = 11.9% | 12.5S | 0.48 | 0.03 |
|
|
|
| |
| 13S | 0.25 | 0.00 | 0.86 | 0.00 | >0.98 | 0.00 | ||
| 13S/22S/R4S | 0.59 | 0.01 |
|
|
|
| ||
| QCM2 | Sigma | σ = 4.2 | TEa = 23.6% | σ = 10.9 | ||||
| CV = 4.5% | QC rule | Ped | Pfr | Ped | Pfr | Ped | Pfr | |
| Bias | 12S | 0.91 | 0.09 | 1 | 0.09 | 1.00 | 0.09 | |
| TEo = 10.1% | 12.5S | 0.77 | 0.03 |
|
|
|
| |
| 13S | 0.57 | 0.00 | 0.86 | 0.00 | >0.98 | 0.00 | ||
| 13S/22S/R4S | 0.69 | 0.01 |
|
|
|
| ||
Optimized TEa of the system, usually the intermediate one, has been set up for sigma approaching 5; one exception is QCM1 for serum, for which sigma already approaches 5 at “low” TEa (20%). The validated QC rules are those with bold Ped and Pfr, verifying at the same time Ped > 0.9 and Pfr < 0.05.
AB = average bias; Ped = probability of error detection (by the QCM); Pfr = probability of false rejection (by the QCM); QC = quality control; QCM = quality control material; RB = range-based bias; σ = sigma metric; SR = spiking-recovery bias; TEo = total allowable error; TEo = observed total error; K9CON (Immulite Systems Control; Siemens); Liquicheck (Bio-Rad).
For serum QCM1, performance parameters (CV, bias, TEo) were so good that the elected low TEa was already enough to generate a sigma approaching 5.
For the urine QCM, given that no target values were provided by the manufacturer for this method, TE calculations have been done with the global bias of the comparison study (see discussion for justification).
Reverse approach (Fig. 2B) of QC rule validation for 2 cortisol concentrations L4 (38.6 nmol/L [1.4 μg/dL]) and L8 (552 nmol/L [20 μg/dL]), using the spiking-recovery bias, the between-run CV, and optimized TEa (σ = 5) by EZ Rules 3.
| Spiked sample QCM | QC rules | 12S | 12.5S | 13S | 13S/22S/R4S | ||||
|---|---|---|---|---|---|---|---|---|---|
| TE for σ = 5 | Ped | Pfr | Ped | Pfr | Ped | Pfr | Ped | Pfr | |
| Serum cortisol | |||||||||
| L4 | 54% | 1 | 0.09 | 0.96 | 0.03 | 0.86 | 0 | 0.94 | 0.01 |
| L8 | 40% | 1 | 0.09 | 0.96 | 0.03 | 0.86 | 0 | 0.94 | 0.01 |
| Urine cortisol | |||||||||
| L4 | 90% | 1 | 0.09 | 0.96 | 0.03 | 0.86 | 0 | 0.94 | 0.01 |
| L8 | 43% | 1 | 0.09 | 0.96 | 0.03 | 0.86 | 0 | 0.94 | 0.01 |
L4 = sample at a cortisol concentration of 38.6 nmol/L (1.4 μg/dL); L8 = sample at a cortisol concentration of 552 nmol/L (20 μg/dL); Ped = probability of error detection (by the QCM); Pfr = probability of false rejection (by the QCM); QC = quality control; QCM = quality control material; σ = sigma metric; TE = total error.
Summary of spiked samples (L4: 38.6 nmol/L = 1.4 µg/dL; L8: 552 nmol/L = 20 µg/dL) and QCM samples (QCM1 and QCM2; K9CON for serum, Liquicheck for urine) performance for cortisol measurement in dogs with the Immulite 2000 Xpi: between-run CV, bias (SR bias for spiked samples, target-value bias for serum commercial QCM, average bias from interlaboratory comparison-study for urine commercial QCM), and total error for sigma = 2 (TEo) and sigma = 5 (optimized TEaQCM).
| Level | Serum cortisol | Urine cortisol | ||||||
|---|---|---|---|---|---|---|---|---|
| CV(%) | Bias(%) | TEo (σ = 2) | TEaQCM (σ = 5) | CV(%) | Bias(%) | TEo (σ = 2) | TEaQCM (σ = 5) | |
| L4 | 9.5 | −6.50 | 25.6 | 54 | 9.6 | 42.3 | 61.4 | 90.3 |
| L8 | 7.4 | 3.08 | 17.9 | 40 | 4.6 | 19.8 | 28.9 | 42.8 |
| QCM1 | 4.1 | −0.27 | 8.5 | 20.8 | 5.4 | −1.09 | 11.9 | 28.1 |
| QCM2 | 7 | −0.52 | 14.5 | 35.5 | 4.5 | −1.09 | 10.1 | 23.6 |
K9CON (Immulite Systems Control; Siemens); Liquicheck (Bio-Rad).
Summary of within-run CV(%) for spiked samples (n = 20), between-run CV(%) for spiked samples (n = 20 over 5 consecutive days), and QCM (n = 17 for urine and n = 22 for serum, over 1 mo).
| Within-run spiked sample | Between-run spiked sample | QCM sample | |
|---|---|---|---|
| Serum low | 1.4 μg/dL | 1.4 μg/dL | K9CON QCM1: 7 μg/dL |
| CV | 7.5% | 9.5% | 4.1% |
| Serum high | 20 μg/dL | 20 μg/dL | K9CON QCM2: 14 μg/dL |
| CV | 4.7% | 7.4% | 7% |
| Urine low | 1.4 μg/dL | 1.4 μg/dL | Liquicheck QCM1: 6.7 μg/dL |
| CV | 6.4% | 9.6% | 5.4% |
| Urine high | 20 μg/dL | 20 μg/dL | Liquicheck QCM2: 23.5 μg/dL |
| CV | 2.5% | 4.6% | 4.5% |
K9CON (Immulite Systems Control; Siemens); Liquicheck (Bio-Rad).
Figure 4.Influences of QC rule determinants on each other. Red plain single arrow: increases; red plain double arrow: vary in similar sense; black dotted single arrow: decreases; black dotted double arrow: vary in opposite sense, which means that when QC rules increase (from 12S to 12.5S to 13S), Ped decreases (not desirable) and Pfr decreases (desirable). EZ Rules 3 QC design software; Ped = probability of error detection; Pfr = probability of false rejection; QC = quality control; σ = sigma metric; TEa = total allowable error; TEo = observed total error.