Mass spectrometry (MS) allows for automated analysis of complex samples at high resolution without the need for labeling/derivatization. Liquid atmospheric pressure matrix-assisted laser desorption/ionization (LAP-MALDI) enables rapid sample preparation and MS analysis using microtiter-plate formats and high-performing mass spectrometers. We present a step change in high-speed, large-scale MS sample analysis of peptides at 20 samples/s and an enzymatic assay at 40 samples/s, i.e., an order of magnitude faster than current MS platforms. LAP-MALDI requires only low amounts of sample volume (<2 μL), of which only a fraction (<1%) is typically consumed, and allows for multiplexing and high-speed MS/MS analysis, demonstrated at ∼10 samples/s. Its high ion signal stability and similarity to electrospray ionization enables CVs below 10% and the analysis of multiply charged peptide ions at these extreme speeds. LAP-MALDI MS fulfills the speed requirements for large-scale population diagnostics and compound screening with the potential of analyzing >1 million samples per day.
Mass spectrometry (MS) allows for automated analysis of complex samples at high resolution without the need for labeling/derivatization. Liquid atmospheric pressure matrix-assisted laser desorption/ionization (LAP-MALDI) enables rapid sample preparation and MS analysis using microtiter-plate formats and high-performing mass spectrometers. We present a step change in high-speed, large-scale MS sample analysis of peptides at 20 samples/s and an enzymatic assay at 40 samples/s, i.e., an order of magnitude faster than current MS platforms. LAP-MALDI requires only low amounts of sample volume (<2 μL), of which only a fraction (<1%) is typically consumed, and allows for multiplexing and high-speed MS/MS analysis, demonstrated at ∼10 samples/s. Its high ion signal stability and similarity to electrospray ionization enables CVs below 10% and the analysis of multiply charged peptide ions at these extreme speeds. LAP-MALDI MS fulfills the speed requirements for large-scale population diagnostics and compound screening with the potential of analyzing >1 million samples per day.
Label-free, high-throughput
MS analysis has recently pushed the limits of sample throughput for
compound library screening and inhibitor studies. Especially ambient
ionization techniques such as AMI (Acoustic Mist Ionization)[1] (2–3 samples/s), MAI (Matrix-Assisted
Ionization)[2] (1 sample/s), DESI (Desorption
ElectroSpray Ionization)[3] (2.7 samples/s),
ESI (ElectroSpray Ionization)[4] (0.4 samples/s),
and ADE (Acoustic Droplet Ejection)[5] (0.45
samples/s), as well as conventional MALDI (Matrix-Assisted Laser Desorption/Ionization)[6] (2.5 samples/s) or hybrid techniques[7] (0.5–1.3 samples/s for IR-MALDESI (Infrared
Matrix-Assisted Laser Desorption Electrospray Ionization)) have produced
encouraging results with respect to analytical speed. Although having
clear advantages over the routinely used label-based photometric readouts,
MS-based fast analysis applications are still not widely employed.However, ADE,[8] as well as LAP-MALDI
MS,[9] recently demonstrated new speeds of
up to 6 samples/s, reliably producing stable ion signals that are
well-separated from each other. The latter of these two approaches
is a new addition to the high-speed MS analysis tools with the inherent
speed advantage of laser-based techniques. Additional advantages of
LAP-MALDI are high ion signal stability,[10,11] which is crucial for fast sample scanning, and the production of
multiply charged analyte ions,[12] thus allowing
the employment of high-performing mass analyzers such as orbitraps
and modern Q-TOF instruments. In combination with its low matrix background,
LAP-MALDI facilitates the simultaneous detection of low-molecular
weight (metabolites, lipids) and high-molecular weight (peptides,
proteins) analytes,[13] outperforming conventional
solid MALDI on axial TOF instruments. LAP-MALDI MS and its associated
(offline) upfront sample preparation support large-scale analyses,
by using microtiter-plate format and multiple robotic preparation
platforms to feed one LAP-MALDI mass spectrometer.Importantly,
LAP-MALDI is inherently fast. On a commercial Q-TOF
instrument, recorded ion packets from individual desorption events
are <5 ms wide,[14] allowing an acquisition
rate of up to 200 desorption events per second. Thus, we further developed
LAP-MALDI with Q-TOF instrumentation by optimizing instrumental bottlenecks
such as spectral scan rates, laser repetition rate, sample plate movement,
and sample number per plate (see additional experimental details in
the Supporting Information), in order to
push the speed limits toward tens and ultimately hundreds of samples
per second.
Experimental Section
LAP-MALDI and MS Setup
The general
LAP-MALDI setup
can be found elsewhere.[9] For this work,
a diode-pumped solid-state (DPSS) laser was used with a wavelength
of 343 nm and a pulse repetition rate of 2000 Hz (Flare NX 343-0.2-2,
Coherent, Santa Clara, USA). MALDI sample plates were rastered as
described in the Supporting Information. The acquisition mode SONAR[15] (Waters)
was used with the quadrupole scanning being disabled and in RF-only
mode. Ion mobility gases were turned off. Each of the SONAR TOF ‘scans’
were stored in 200 consecutive spectra or ‘bins’, allowing
the acquisition and storage of up to 1000 spectra/s, while the temporal
resolution increased up to 0.93 ms per spectrum/bin.
Matrix Preparation
and Sample Spotting
CHCA was dissolved
in acetonitrile and water (1:1; v/v) to a concentration of 5 mg/mL.
After short sonication, propylene glycol (PG) was added at 60% by
volume. The matrix was mixed with a sample at a ratio of 1:1 (v/v),
and 1 or 0.3 μL of the mixture was spotted onto the stainless-steel
sample plate using a 384- or 1536-well format, respectively.
Peptide
Analysis
A total of 10 pmol of peptide was
used for each LAP-MALDI sample. The MALDI samples were analyzed in
each row by moving the sample plate at a constant speed of 50–200
mm/s. To ease postacquisition data processing, the start and end of
each sample row was marked with a sample using the analyte standard
Angiotensin I (Ang I) (40 pmol) and N-Hippuryl-His-Leu
hydrate (HHL) (10 pmol), respectively.
Enzyme Assay
Angiotensin-converting
enzyme (ACE) was
dissolved in 50 mM Tris buffer at pH 8.5 to yield 0.1 U/mL and mixed
with the substrate 1:1 (v/v, 320 pmol/μL Ang I or 100 pmol/μL
HHL). The mixture was incubated at 37.5 °C for several hours.Additional experimental details can be found in the Supporting Information.
Results and Discussion
For initial testing, a 384-well microtiter-format sample plate
was prepared by alternatingly spotting two peptides: bradykinin (Brdk)
and [Lys-des-Arg9]-bradykinin (Lys-Brdk). The sample plate
was automatically rastered with an analysis speed of >20 samples/s
per row. The overall speed for the sample plate was slightly reduced
to 17 samples/s, as additional time was needed to move from the last
sample in each row to the first sample in the next row due to restrictions
in the sample stage movement and data processing software (see additional
experimental details in Supporting Information). The total data acquisition time for an entire sample plate based
on the microtiter-plate format was less than 23 s. Figure S1 shows a diagrammatic scheme of the LAP-MALDI source
as used in this study.The data obtained from this initial analysis
clearly show well-separated
ion signals for all samples without any analyte carryover (see Figure a–d). As previously
reported, predominantly doubly charged peptide ions are observed.[12] Mass spectrometer scan rates were adjusted to
yield at least 10 data acquisitions across each sample, keeping the
detection deadtime (interscan delays; ISDs) to a minimum. This is
important as data acquisition gaps caused by ISDs lead to the loss
of some or all ion signals from a sample.
Figure 1
LAP-MALDI MS analysis
of a 384-well plate with alternating Brdk
and Lys-Brdk samples and different standards at the start and end
of each sample row at 20 samples/s for each row (average of 17 samples/s
for the entire plate). a) From top to bottom: extracted ion chromatogram
(XIC) of [HHL+H]+ (m/z 430.21, start marker), [Ang+2H]2+ (m/z 648.84, end marker), [Brdk+2H]2+ (m/z 530.78), [Lys-Brdk+2H]2+ (m/z 516.78), and total ion chromatogram
(TIC). Mass spectra of first (b) (Brdk) and last (c) (Lys-Brdk) peptide
sample of the plate. Singly (*) and doubly (**) charged analyte ions
are labeled. d) Enlargement of one row of analyte ion signals. From
top to bottom: XIC of [Brdk+2H]2+ (m/z 530.78), [Lys-Brdk+2H]2+ (m/z 516.78), and total ion chromatogram (TIC). Analyte
ion signal intensities for Brdk (e) and Lys-Brdk (f) using a 30% threshold.
LAP-MALDI MS analysis
of a 384-well plate with alternating Brdk
and Lys-Brdk samples and different standards at the start and end
of each sample row at 20 samples/s for each row (average of 17 samples/s
for the entire plate). a) From top to bottom: extracted ion chromatogram
(XIC) of [HHL+H]+ (m/z 430.21, start marker), [Ang+2H]2+ (m/z 648.84, end marker), [Brdk+2H]2+ (m/z 530.78), [Lys-Brdk+2H]2+ (m/z 516.78), and total ion chromatogram
(TIC). Mass spectra of first (b) (Brdk) and last (c) (Lys-Brdk) peptide
sample of the plate. Singly (*) and doubly (**) charged analyte ions
are labeled. d) Enlargement of one row of analyte ion signals. From
top to bottom: XIC of [Brdk+2H]2+ (m/z 530.78), [Lys-Brdk+2H]2+ (m/z 516.78), and total ion chromatogram (TIC). Analyte
ion signal intensities for Brdk (e) and Lys-Brdk (f) using a 30% threshold.Heatmaps for both peptides using simple ion signal
intensity thresholds
(see Figure e,f) reveal
a detection (classification) accuracy of >95% over multiple analyses
(n = 3, see Figure S2).
Greater accuracy is achieved at slightly lower throughput (∼99%
at 10 samples/s, n = 3, see Figure S3). Sample position effects, e.g., edge effects, are not evident
(see Figure S4). The few failures in the
correct sample assignment are currently mostly due to sample spotting,
which is envisaged to be improved by adequate robotic liquid handlers
capable of handling small volumes of liquid (<1 μL) for MALDI
sample preparation. To further automate the entire workflow, plate-changing
robotics can be used, which can achieve sample plate-swapping in around
5 s.[16] Overall high-throughput analysis
time could therefore be around 30 s per plate, allowing 120 plates
to be analyzed per hour with an adequate multistage robotic feeding
system. In 24 h, more than a million samples could be screened, in
principle.This level of sample throughput is desirable for
compound and assay
screening in pharmaceutics. Consequently, the platform’s applicability
to enzyme assays was also tested by monitoring the conversion of an
enzyme substrate and its product’s appearance simultaneously
(see Figure ). ACE
showed full transformation of its natural substrate Ang I and HHL,
commonly used in fluorescence experiments. The substrate ion signal
intensity substantially decreased after enzyme treatment, and new
ion signals at m/z 269.1589 and m/z 1046.5422 appeared, which can be attributed
to cleaved protonated His-Leu and converted Angiotensin II (Ang II)
with a mass measurement accuracy of 7 and 4 ppm, respectively. Although
some ion suppression due to the buffer can be observed (data not shown),
reproducible and visibly time-resolved peaks were observed at the
same acquisition speed as used for the peptide standards (see Figure ). With increasing
mass (mainly above 1000 Da), we observe the appearance of additional
(temporally delayed) low-intensity analyte ion peaks at these high-speed
acquisitions. These are most likely due to the complex ion path within
the employed Q-TOF/ion mobility instrument and are currently under
investigation.
Figure 2
LAP-MALDI MS analysis of an angiotensin-converting enzyme
assay
with alternating enzyme-treated and untreated samples for Angiotensin
I (Ang I, first row) and N-Hippuryl-His-Leu hydrate
(HHL, second row). a) From top to bottom: XIC of [HHL+H]+ (m/z 430.21, substrate), [Ang
I + 2H]2+ (m/z 648.84,
substrate), His-Leu (m/z 269.15,
product), and Angiotensin II [Ang II + H]+ (m/z 1046.54, product). b) Overlay of chromatograms
shown in a. Average data acquisition speed was 16 samples/s using
a stage speed of 100 mm/s.
LAP-MALDI MS analysis of an angiotensin-converting enzyme
assay
with alternating enzyme-treated and untreated samples for Angiotensin
I (Ang I, first row) and N-Hippuryl-His-Leu hydrate
(HHL, second row). a) From top to bottom: XIC of [HHL+H]+ (m/z 430.21, substrate), [Ang
I + 2H]2+ (m/z 648.84,
substrate), His-Leu (m/z 269.15,
product), and Angiotensin II [Ang II + H]+ (m/z 1046.54, product). b) Overlay of chromatograms
shown in a. Average data acquisition speed was 16 samples/s using
a stage speed of 100 mm/s.To further increase sample throughput, two other bottlenecks were
addressed. First, the speed of the translational stage for sample
plate movement was increased by a factor of up to 4 (from 50 to 200
mm/s). Second, the sample plate layout was changed to the 1536-well
format. Samples were spotted closer to each other and made smaller
as actual sample consumption per desorption event is minimal (less
than 1‰[12]).Tighter sample
spotting and a stage speed of 200 mm/s were first
tested with the ACE assay (see Figure ). These changes led to well-resolved peaks at >40
samples/s. With faster sample movement (≥200 mm/s), acceleration
and deceleration of the translational stage clearly show a broadening
effect for the first and last samples in each row. In addition, the
above-mentioned double-peaking for higher masses and the reduced number
of sampling points per sample currently limit the maximum analysis
speed to 40 samples/s for analytes with a mass >1000 Da. Interestingly,
the analysis of the lower-mass substrates (m/z 430.21) and products (m/z 269.15) clearly shows baseline-resolved analyte ion signals between
the alternating samples at a speed of 40 samples/s (Figure a).
Figure 3
LAP-MALDI MS analysis
of an angiotensin-converting enzyme assay
alternating enzyme-treated and untreated samples for Angiotensin I
using a 1536-well plate layout. a) 40 samples/s (stage movement speed
of 100 mm/s); b) 60 samples/s (stage movement speed of 200 mm/s);
c) overlay of XICs shown in a; d) overlay of XICs shown in b. TIC:
total ion chromatogram; XIC: extracted ion chromatogram.
LAP-MALDI MS analysis
of an angiotensin-converting enzyme assay
alternating enzyme-treated and untreated samples for Angiotensin I
using a 1536-well plate layout. a) 40 samples/s (stage movement speed
of 100 mm/s); b) 60 samples/s (stage movement speed of 200 mm/s);
c) overlay of XICs shown in a; d) overlay of XICs shown in b. TIC:
total ion chromatogram; XIC: extracted ion chromatogram.Next, extreme sample throughput levels were tested by analyzing
HHL (see Figure ).
While the first and last samples in each row are only slightly broadened
due to stage acceleration/deceleration at a speed around 40 samples/s,
peak broadening worsens and expands to other samples at even higher
speeds. Nonetheless, samples can still be separated in their TICs
(total ion chromatograms) and XICs (extracted ion chromatograms) at
60 samples/s (see Figure S5a,b). By analyzing
several samples of Brdk multiple times, coefficients of variation
(CVs) below 10% (<5% for [HHL+H]+) were achieved at
conditions corresponding to a speed of 9 samples/s (see Figure S6). Other MS-based techniques result
in similar variability but at significantly lower sample throughput.[17,18] Currently, higher speeds result in higher CVs, e.g., 15% at 20 samples/s.
Figure 4
LAP-MALDI
MS analysis of HHL using a 1536-well plate layout. a)
TIC and b) XIC of [HHL+H]+ at 41 samples/s. TIC: total
ion chromatogram; XIC: extracted ion chromatogram.
LAP-MALDI
MS analysis of HHL using a 1536-well plate layout. a)
TIC and b) XIC of [HHL+H]+ at 41 samples/s. TIC: total
ion chromatogram; XIC: extracted ion chromatogram.Finally, ultrahigh-throughput LAP-MALDI tandem mass spectrometry
(MS/MS) was demonstrated by analyzing Brdk and its collision-induced
dissociation (CID) fragment ions using a fixed window quadrupole.
Typical CID peptide fragment ions such as y- and b-type ions were
detected at a speed of 12 samples/s per row (see Figure S7). This initial data shows the potential of LAP-MALDI
for high-speed SRM/MRM applications.
Conclusion
In
summary, we have shown a step change in the speed for analyzing
individual samples by mass spectrometry. So far, up to 60 samples
per second can be well separated at the fwhm level by their analyte
ion signal. These significant speed increases by an order of magnitude
or more compared to previously published reports were achieved by
overcoming several bottlenecks such as data acquisition speed (adjusted
for ISD deadtimes), laser pulse repetition rate (increased to 2,000
Hz), and sample plate stage speed (increased to 200 mm/s) as well
as tighter sample spotting and the use of liquid matrices with their
high MALDI ion signal stability. Without the latter, the practical
analysis speed would be severely compromised due to the fluctuating
ion signal, typically observed with MALDI, thus impeding automated
data processing. As used in LAP-MALDI MS, liquid matrices also allow
for the generation of multiply charged ions, and together with an
atmospheric ion source, high-performing mass spectrometers and superior
MS/MS analyses can be employed; in addition, sample plates can be
exchanged significantly more quickly in AP-MALDI than in vacuum MALDI
sources. MS/MS analysis at these extreme speeds will take screening
of large sample sets to an even higher level, in particular with respect
to specificity and multiplexing using SRM/MRM approaches.The
method described here can be applied to a vast set of analytes,
and the demonstrated analysis speed is even greater than the speed
reported for MALDI imaging on commercial instrumentation,[19] which currently achieves around 20 pixels/s
(without gaps between samples). Future modifications of the mass spectrometry
acquisition software should allow data acquisition without data loss
between scans (due to ISD). Lower scan times and thus higher temporal
resolution will provide an additional boost to speed. In general,
the theoretical speed limit of the method is determined by the temporal
width of the ion plume formed during desorption. Further advancements
can be made by using adequate robotics for tighter sample positioning
(e.g., 6144-well microtiter-plate format) and employing sample stages
with even higher speed as well as acceleration/deceleration. In contrast
to other techniques like ADE,[18] sample
volumes can be <1 μL, and tighter sample layouts do not result
in increased analysis times but lead to a significant throughput improvement.Potential application areas for LAP-MALDI MS are in large-scale
(multiplex) population diagnostics and in screening of compound libraries
within the pharmaceutical industry, where a throughput of 1 million
samples per day or more is highly desirable.
Authors: Max A Müller; Mario Kompauer; Kerstin Strupat; Sven Heiles; Bernhard Spengler Journal: J Am Soc Mass Spectrom Date: 2020-12-28 Impact factor: 3.109
Authors: Ian Sinclair; Martin Bachman; Daniel Addison; Mattias Rohman; David C Murray; Gareth Davies; Elizabeth Mouchet; Michael E Tonge; Richard G Stearns; Lucien Ghislain; Sammy S Datwani; Lars Majlof; Eric Hall; Gareth R Jones; Emmy Hoyes; Joe Olechno; Richard N Ellson; Perdita E Barran; Steven D Pringle; Michael R Morris; Jonathan Wingfield Journal: Anal Chem Date: 2019-03-11 Impact factor: 6.986
Authors: Andrew Wagner; Jun Zhang; Chang Liu; Thomas R Covey; Timothy V Olah; Harold Bud N Weller; Wilson Z Shou Journal: Anal Chem Date: 2020-09-25 Impact factor: 6.986
Authors: Fan Pu; Andrew J Radosevich; James W Sawicki; David Chang-Yen; Nari N Talaty; Sujatha M Gopalakrishnan; Jon D Williams; Nathaniel L Elsen Journal: Anal Chem Date: 2021-04-22 Impact factor: 6.986
Authors: Roman P Simon; Tim T Häbe; Robert Ries; Martin Winter; Yuting Wang; Amaury Fernández-Montalván; Daniel Bischoff; Frank Runge; Wolfgang Reindl; Andreas H Luippold; Frank H Büttner Journal: SLAS Discov Date: 2021-07-26 Impact factor: 3.341
Authors: Michael Wleklinski; Bradley P Loren; Christina R Ferreira; Zinia Jaman; Larisa Avramova; Tiago J P Sobreira; David H Thompson; R Graham Cooks Journal: Chem Sci Date: 2018-01-04 Impact factor: 9.825
Authors: Kevan T Knizner; Michael C Bagley; Fan Pu; Nathaniel L Elsen; Jon D Williams; David C Muddiman Journal: J Mass Spectrom Date: 2022-06 Impact factor: 2.394